Search Results
Found 30 results
510(k) Data Aggregation
(267 days)
) | Clinical Chemistry |
| 12 | 864.7675 | LJX
The UC-1800 Automatic Urine Analyzer is automated instrument which is intended for professional, in vitro diagnostic use only.
Depending on the reagent strips being used, the instruments perform semi-quantitative detection of the following analytes in urine: ascorbic acid, microalbumin, leukocytes, creatinine, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine hydrometer (optional) can determine the color and turbidity of urine. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.
The URIT 11FA urine reagent strips provide semi-quantitative tests for ascorbic acid, leukocytes, setone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 11FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.
The URIT 12FA urine reagent strips provide semi-quantitative tests for microalbumin, leukocytes, creatinine, ketone, urobilinogen, bilirubin, glucose, protein, specific gravity, blood and pH in urine and for qualitative determination of nitrite in urine. The URIT 12FA urine reagent strips are for use with the UC-1800 Automatic Urine Analyzer and are for professional, in vitro diagnostic use only. Test results may provide information regarding the status of carbohydrate metabolism, kidney and liver function, acid-base balance and bacteriuria.
UC-1800 Automatic Urine Analyzer is characterized by fully automated and simple operation. All you need to do is to set test strips and samples, press the START key, and the rest of operations are fully automated with UC-1800, which can measure samples continuously. For each measurement, the instrument automatically performs a series of operation: sample transmitting, sample aspirating, sample dropping, rinsing strip feeding and color identifying, etc. The instrument is used in conjunction with a serial of URIT urine test strips for measuring 15 parameters. Measure results are printed through either built-in printer or external printer.
Urine Reagent Strips is used to determine the components to be measured in urine by dry chemistry method together with urine analyzer. Various components to be tested in the urine can result in changes to the colors of corresponding reagent blocks on the Urine Reagent Strips. The depth of reaction color is proportional to the corresponding component to be tested in the urine. Qualitative and semi-quantitative detection can be conducted to the contents of the corresponding detected components. As a reagent for the determination of multiple components in human urine and the most basic test item for clinical urine routine test), it is suitable for the screening test or auxiliary diagnosis for clinical diagnosis, without the specificity for diseases or indications, and urine dry chemistry test is a screening test and cannot be used as a single diagnostic method.
The provided document describes the URIT UC-1800 Automatic Urine Analyzer and its associated reagent strips (URIT 11FA and 12FA Urine Reagent Strips). The information below summarizes the acceptance criteria and the studies performed to demonstrate the device meets these criteria.
1. Table of Acceptance Criteria and Reported Device Performance
The document presents the performance in terms of "Exact agreement" and "± 1 color block" agreement with expected values or between the proposed device and predicate devices. For most analytes, the criteria seem to be high exact agreement and 100% agreement within ±1 color block. Specific thresholds for acceptance were not explicitly stated as global criteria but are implied by the "Qualified" conclusions for individual tests. The comparison to predicates also uses agreement rates.
Below is a summary of the reported device performance from the "Precision / Reproducibility" section (Tables 8, 9, 10, 11) for repeatability and "Comparison Studies" section (Tables a.1, a.2, a.3) for agreement with predicate devices. Given the extensive number of analytes and concentration levels, key representative results are presented.
a. Repeatability (Within-Run Precision)
Reported as "Exact agreement" and "± 1 color block". All tests for all expected values show 100% agreement for "± 1 color block". Exact agreement varies slightly, but most are 100% or very close.
Test (Analyte) | Expected Value (Example) | Exact Agreement (Proposed Device: UC-1800 with 11FA/12FA strips) | Conclusion |
---|---|---|---|
Ascorbic acid (11FA) | -(0) mg/dL | 100% (60/60) | Qualified |
Nitrite (11FA/12FA) | - (Negative) | 100% (60/60) | Qualified |
Leukocyte (11FA/12FA) | -(0) leu/μL | 100% (60/60) | Qualified |
Ketone (11FA) | -(0) mg/dL | 100% (60/60) | Qualified |
Ketone (12FA) | +1(15) mg/dL | 96.7% (58/60) | Qualified |
Urobilinogen (11FA/12FA) | Normal EU/dL | 100% (60/60) | Qualified |
Bilirubin (11FA) | +2(2.0) mg/dL | 96.7% (58/60) | Qualified |
Bilirubin (12FA) | +2(2.0) mg/dL | 91.7% (55/60) | Qualified |
Glucose (11FA/12FA) | All | 100% (60/60) | Qualified |
Protein (11FA/12FA) | +1(30) mg/dL | 93.3% (56/60) | Qualified |
Specific Gravity (11FA) | 1.010 | 96.7% (58/60) | Qualified |
Specific Gravity (12FA) | All | 100% (60/60) | Qualified |
Blood (11FA) | +1(25) Cell/μL | 98.3% (59/60) | Qualified |
Blood (12FA) | All | 100% (60/60) | Qualified |
pH (11FA) | 7.0 | 96.7% (58/60) | Qualified |
pH (12FA) | All | 100% (60/60) | Qualified |
Microalbumin (12FA) | All | 100% (60/60) | Qualified |
Creatinine (12FA) | All | 100% (60/60) | Qualified |
Turbidity (Physical Module) | All | 100% (60/60) | Qualified |
Color (Physical Module) | All | 100% (60/60) | Qualified |
b. Reproducibility (Between-Run Precision)
Reported as "Exact agreement" and "± 1 color block". All tests for all expected values show 100% agreement for "± 1 color block". Exact agreement varies slightly, but most are 100% or very close.
Test (Analyte) | Expected Value (Example) | Exact Agreement (Proposed Device: UC-1800 with 11FA/12FA strips) | Conclusion |
---|---|---|---|
Ascorbic acid (11FA) | +2(50) mg/dL | 94.2% (113/120) | Qualified |
Protein (11FA) | +2(100) mg/dL | 99.2% (119/120) | Qualified |
Blood (11FA) | +2(80) Cell/μL | 95% (114/120) | Qualified |
pH (11FA) | 7.0 | 99.2% (119/120) | Qualified |
Nitrite (12FA) | All | 100% (120/120) | Qualified |
Blood (12FA) | +2(80) Cell/μL | 95.8% (115/120) | Qualified |
All other analytes (11FA/12FA) | Most levels | 100% (120/120) or very close | Qualified |
c. Comparison with Predicate Devices
Analyte (Strip) | Predicate Device | Agreement Type | Agreement Rate (%) (Reported value from the table) | Conclusion |
---|---|---|---|---|
Ascorbic acid (11FA) | Uritest-500B (K082811) | Complete agreement | 97.60% to 100% across all levels (e.g., 99.00% for +1(25) level) | Subst. Eq. |
Leukocyte (11FA/12FA) | Uritest-500B (K082811) | Complete agreement | 95.80% to 100% across all levels (e.g., 95.80% for +2(125)) | Subst. Eq. |
Ketone (11FA/12FA) | Uritest-500B (K082811) | Complete agreement | 97.40% to 100% across all levels (e.g., 97.40% for +1(15)) | Subst. Eq. |
Nitrite (11FA/12FA) | Uritest-500B (K082811) | Overall Agreement (OPA) | 99.40% (11FA); 98.70% (12FA) | Subst. Eq. |
Urobilinogen (11FA/12FA) | Uritest-500B (K082811) | Complete agreement | 95.50% to 100% across all levels (e.g., 95.50% for +1(2.0) on 12FA) | Subst. Eq. |
Bilirubin (11FA/12FA) | Uritest-500B (K082811) | Complete agreement | 95.20% to 100% across all levels (e.g., 95.20% for +2(2.0) on 12FA) | Subst. Eq. |
Glucose (11FA/12FA) | Uritest-500B (K082811) | Complete agreement | 96.30% to 100% across all levels (e.g., 96.30% for +1(100)) | Subst. Eq. |
Protein (11FA/12FA) | Uritest-500B (K082811) | Complete agreement | 95.50% to 100% across all levels (e.g., 95.50% for +1(30) on 12FA) | Subst. Eq. |
pH (11FA/12FA) | Uritest-500B (K082811) | Complete agreement | 97.80% to 100% across all levels (e.g., 97.80% for 6.5) | Subst. Eq. |
Specific Gravity (11FA/12FA) | Uritest-500B (K082811) | Complete agreement | 97.80% to 100% across all levels (e.g., 97.80% for 1.010 on 12FA) | Subst. Eq. |
Microalbumin (12FA) | Mission® U120 Ultra Urine Analyzer (K142391) | Complete agreement | 83.33% for 30mg/L, up to 98.21% for 150mg/L | Subst. Eq. |
Creatinine (12FA) | Mission® U120 Ultra Urine Analyzer (K142391) | Complete agreement | 90.91% for 300mg/dL, up to 97.74% for 50mg/dL | Subst. Eq. |
Color (Physical module) | AUTION MAX AX-4030 Urinalysis System (K093098) | Coincidence rate | 91.29% (Colorless) to 98.48% (Brown) | Subst. Eq. |
Turbidity (Physical module) | AUTION MAX AX-4030 Urinalysis System (K093098) | Coincidence rate | 96.00% (Turbid) to 100% (Micro turbid) | Subst. Eq. |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Repeatability (within-run) & Reproducibility (between-run):
- Sample Size:
- Repeatability: 60 measurements per concentration level per analyte (20 replicates x 3 instruments).
- Reproducibility: 120 measurements per concentration level per analyte (20 days x 2 runs/day x 1 time/run in 3 sites, with 1 instrument/site, 3 operators).
- Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective. It refers to "negative urines and spiked urines of known concentrations." This suggests controlled laboratory-prepared samples rather than direct patient samples.
- Sample Size:
-
Linearity/Assay Reportable Range:
- Sample Size: 63 measurements per concentration level per analyte (reference solutions tested 21 times on 3 UC-1800 machines with 3 lots of reagent strips).
- Data Provenance: Laboratory-prepared reference solutions, not clinical samples.
-
Analytical Sensitivity: No sample sizes mentioned, values are stated directly.
-
Critical Value: No sample sizes mentioned, values are stated directly.
-
Analytical Specificity (Interference, pH, Color, Specific Gravity):
- Sample Size: For interference studies, samples were tested 5 times on 2 UC-1800 machines with 2 batch numbers of 11FA and 12FA reagent strips. This totals 20 measurements per condition (5 tests x 2 machines x 2 strips).
- Data Provenance: Laboratory-prepared urine samples (negative samples prepared and spiked with interfering substances or adjusted for pH/color/SG).
-
Comparison Studies with Predicate Devices (Clinical Samples):
- Uritest-500B (K082811): 1000 clinical urine samples.
- Mission® U120 Ultra Urine Analyzer (K142391): 979 clinical urine samples.
- AUTION MAX AX-4030 Urinalysis System (K093098):
- Color: 1365 clinical urine samples.
- Turbidity: 1000 clinical urine samples.
- Data Provenance: The document explicitly states "clinical urine samples were collected." The country of origin is not specified but given the submitter's address (China), it is highly likely that these clinical samples were collected in China and were retrospective as the study compares against existing predicate devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not mention the use of experts to establish a "ground truth" for the test set in the traditional sense of medical image interpretation or clinical diagnosis. For chemical analyzers, the "ground truth" (or reference standard) is typically established by:
- Known concentrations: For repeatability, reproducibility, linearity, and analytical sensitivity, samples are prepared with known concentrations of the analytes.
- Reference methods: For the analytical specificity and comparison studies, the predicate devices themselves or established reference methods (e.g., 2,6-Dichlorophenolindophenol Titration Method for Ascorbic Acid, Lange method for Ketone, etc. as specified in Table 35 "Traceability") serve as the reference for comparison.
Therefore, the concept of "number of experts" and their "qualifications" for ground truth establishment, as typically applied in AI/ML performance evaluation (e.g., for image interpretation), is not directly applicable here. The ground truth is analytical and based on laboratory standards and established measurement techniques.
4. Adjudication Method for the Test Set
Not applicable in the context of this analytical device. As explained above, the "ground truth" refers to known concentrations or results from predicate/reference methods. There isn't a subjective interpretation by multiple experts that would require an adjudication method like 2+1 or 3+1.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an automated urine analyzer, not an AI/ML-driven diagnostic aid for human readers. It performs measurements to provide semi-quantitative results for various urine analytes. Therefore, there is no human-in-the-loop performance or comparison of human reader improvement with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the studies presented are all standalone performance evaluations of the device (UC-1800 Automatic Urine Analyzer with URIT 11FA/12FA Urine Reagent Strips) without human intervention in the measurement process. The device operates automatically to detect and report analyte levels. The comparison studies demonstrate its performance against existing predicate (standalone) analyzers.
7. The Type of Ground Truth Used
The ground truth used in the studies includes:
- Known Concentrations: For analytical performance studies such as precision (repeatability and reproducibility), linearity, analytical sensitivity, and analytical specificity (interference testing), samples were prepared with known, precisely measured concentrations of the target analytes or interfering substances.
- Predicate Device Measurements: For method comparison studies, the results obtained from the established predicate devices (Uritest-500B Urine Analyzer, Mission® U120 Ultra Urine Analyzer, and AUTION MAX AX-4030 Urinalysis System) served as the reference standard for comparison. These predicate devices also operate based on defined analytical principles.
- Reference Methods: Table 35 details the "Reference Method" used for traceability for each analyte (e.g., 2,6-Dichlorophenolindophenol Titration Method for Ascorbic Acid, Glucose Oxidase Method for Glucose, Acidometer measurement for pH, etc.). This indicates fundamental analytical standards are the basis for the stated detection ranges and performance.
8. The Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning. The UC-1800 Automatic Urine Analyzer uses reflectance photometry and other physical principles (refractometer for specific gravity, light-scattering for turbidity, light-transmission for color) to generate results, not an AI algorithm that learns from a dataset in the conventional sense. The "training" of such a system typically involves calibrating optical sensors and algorithms to known standards, which is part of the engineering design and quality control processes.
If interpretation of "training set" refers to the data used for the initial development and calibration of the device's measurement algorithms:
- No specific sample size for a "training set" for an AI model is reported because the device does not employ machine learning that requires a distinct "training set" and "test set" in the typical AI/ML development lifecycle.
- The system is calibrated using URIT urine control materials and calibration test strips (Table 9, "Calibration").
9. How the Ground Truth for the Training Set Was Established
As noted above, the device does not use an AI/ML model with a "training set" in the common understanding. The system's operational parameters and calibration are established using:
- Reference materials and calibrators: The device is calibrated using "URIT urine control materials and calibration test strips" (Table 9, "Calibration"). These control materials and calibration strips would have their values established using highly accurate reference methods or certified reference materials, ensuring traceability to scientific standards (as indicated in Table 35 "Traceability").
- Chemical principles: The underlying "ground truth" for the device's internal algorithms (i.e., how they convert optical signals to analyte concentrations) is based on established chemical reactions and physical measurement principles (e.g., reflectance photometry, refractometry, light scattering, dye-binding, enzymatic reactions). The "ground truth" for developing and fine-tuning these algorithms would be derived from rigorous scientific validation against these known chemical and physical properties.
Ask a specific question about this device
(65 days)
-----|-----------------------------------------------------|------------|
| LJX | 21 CFR § 864.7675
The Healgen URS Test Strips are for the qualitative detection of Leukocytes (white blood cells) and nitrite in urine as an aid in the screening of a Urinary Tract infection (UTI). It is intended for over-the-counter home use only.
Healgen URS Test Strips are in vitro diagnostic test devices for qualitative detection of leukocyte and nitrite in urine. The device is composed of two color pads aligned on a strip. One pad is employed for testing leukocyte and the other for nitrite by visually reading the color change of the pad and comparing with the corresponding blocks on a color chart.
The document describes the performance of the Healgen URS Test Strips for the qualitative detection of Leukocytes and Nitrite in urine to aid in UTI screening for over-the-counter home use.
Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" for each performance characteristic in a formal table. However, it presents various study results that implicitly serve as an indication of meeting performance expectations for a Class I device intended for OTC use. The implied acceptance is based on the presented results demonstrating sufficient sensitivity, precision, and user-friendability.
Here's a table summarizing the reported device performance for key characteristics:
Performance Characteristic | Analyte | Reported Device Performance |
---|---|---|
Limit of Detection (LOD) | Leukocyte | LOD verified at 15 cells/uL (100% detection at 30 and 15 cells/uL, 47% at 10 cells/uL, 7% at 5 cells/uL). |
Nitrite | LOD verified at 0.05 mg/dL (100% detection at 0.1 and 0.08 mg/dL, 83% at 0.06 mg/dL, 50% at 0.05 mg/dL). | |
Precision/Reproducibility | Leukocyte | 100% agreement with expected results across Negative, Trace (15 cells/uL), Small (70 cells/uL), Moderate (125 cells/uL), and Large (500 cells/uL) controls (N=45 for each control level). |
Nitrite | 100% agreement with positive results for 0.1 mg/dL and 0.3 mg/dL controls (N=45 for each control level). | |
Analytical Specificity | Leukocyte | No interference from various substances (Albumin, Ammonium Chloride, Bilirubin, Ciprofloxacin, Creatinine, Fructose, Galactose, Glycine, Hemoglobin, Lactose, Oxalic Acid, Phenazopyridine, Phenolphthalein, Potassium Chloride, Riboflavin, Sodium Nitrate, Sodium Nitrite, Sodium Phosphate, Sulfamethoxazole, Theophylline, Urea) at specified concentrations. |
Potential interference: High glucose (≥1000 mg/dL), high ascorbic acid (≥150 mg/dL) may decrease readings. pH >8.0 may cause false positive readings. Specific gravity >1.035 may cause false negative readings. | ||
Nitrite | No interference from various substances (same list as Leukocyte). | |
Potential interference: High ascorbic acid (≥150 mg/dL) may cause false negative readings. No effect of pH or specific gravity on nitrite testing. | ||
Stability | Both | Stable for 24 months at 2-30℃ based on real-time stability studies. |
Lay-user Study (Method Comparison) | Leukocyte | +% Agreement (Exact Match): Large (90.9%), Moderate (92%), Small (89.7%), Trace (88.5%), Negative (98.3%). |
+% Agreement (+/- Color Block): 100% for all levels. | ||
Nitrite | +% Agreement (Exact Match): 100% for both Positive and Negative, with 100% overall agreement. | |
Lay-user Instruction Survey | Instructions | All lay users indicated that the device instructions can be easily followed. Flesch-Kincaid reading analysis showed a reading grade level less than 7. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Sizes:
- Sensitivity Study (LOD): 30 samples/tests per concentration for each analyte (Leukocyte and Nitrite). Each sample was tested in duplicates.
- Precision/Reproducibility Study: 45 assays results for each of eight levels of control (total 3 lots x 3 sites x 2 operators x 5 days x 1 rep/day = 90 tests per control, though results are reported as N=45 for each control).
- Analytical Specificity: Not explicitly stated as a single test set size, but multiple urine samples (negative urine spiked with interfering substances) were tested with three device lots by three different operators for each substance.
- Lay-user Studies: 150 lay users with UTI symptoms.
- Data Provenance: The document does not explicitly state the country of origin for the data. However, for the precision study, it mentions "three clinical sites," implying clinical data from a real-world setting. The lay-user study involved recruitment of "150 lay users with UTI symptoms," suggesting prospective collection of real-world user data. The sensitivity and analytical specificity studies appear to be laboratory-based. It is not specified if any data was retrospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- LOD and Precision Studies: "Known concentrations" for spiked samples and "expected ranges/values" for controls are used as ground truth. The expertise involved in preparing these precisely calibrated samples and controls is implied but not explicitly detailed (e.g., analytical chemists, laboratory technicians). Operators are mentioned (5 for LOD, 2 at each of 3 sites for precision), but their role is to perform the tests, not establish ground truth.
- Analytical Specificity: Ground truth is established by "negative urine with different leukocyte and nitrite concentrations" which were likely precisely prepared laboratory samples.
- Lay-user Studies: The "comparison testing by healthcare professionals" served as the ground truth. The number and qualifications of these healthcare professionals are not specified beyond the general term "healthcare professionals."
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- The document does not describe a formal adjudication method (like 2+1 or 3+1) for resolving discrepancies in the test results.
- For the lay-user study, the comparison is made between the layperson's results and the results obtained by healthcare professionals, implying the healthcare professional's result is the definitive ground truth for comparison.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This device is a test strip for visual interpretation by lay-users, not an AI-assisted diagnostic device evaluated by professional readers. The study focuses on direct comparison of lay-user performance to healthcare professional results for the test strip itself, not on the impact of AI assistance on human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- No, this device is not an algorithm. It is a visual assay (colorimetric test strip) that requires human interpretation. Therefore, a standalone algorithm-only performance study is not applicable. The "standalone" performance here refers to the device's inherent chemical reaction and color development, which is then interpreted by a human.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- LOD, Precision, Analytical Specificity: Ground truth was established using known concentrations of spiked analytes in urine samples and control materials, presumably confirmed by established laboratory methods and validated reagents.
- Lay-user Study: Ground truth was established by results obtained by healthcare professionals using the same test strips. This implies the healthcare professionals' interpretation served as the 'expert' reference. It's not explicitly stated whether these healthcare professionals confirmed results with a gold standard like culture or microscopic analysis, or if their reading of the strip was considered the definitive truth for the purpose of the study.
8. The sample size for the training set
- The document describes performance studies for the Healgen URS Test Strips, which are a chemical assay (test strips) with visual interpretation. This is not a machine learning or AI-based device that typically requires a "training set." Therefore, the concept of a training set size is not applicable here.
9. How the ground truth for the training set was established
- As stated above, this is not an AI/ML device, so there is no training set or ground truth establishment for a training set in the context of machine learning. The studies described are for analytical and lay-user performance of a diagnostic test strip.
Ask a specific question about this device
(1211 days)
Leukocyte, peroxidase
test | LJX | Class I | 21 CFR 864.7675
The AUTION MAX AX-4060 Urinalysis System (AUTION MAX) is comprised of the AUTION MAX AX-4060 automated urine analyzer and AUTION Sticks 9EB multi-parameter test strips.
The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB test strips is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories.
AUTION Sticks 9EB test strips are test strips for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders.
Special conditions for use statements: Prescription use only. AUTION Sticks 9EB test strips are not to be read visually.
The AUTION MAX AX-4060 Urinalysis System is a fully automated urine analyzer that provides a semi-quantitative or qualitative measurement for glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, specific gravity, turbidity, and color tone. The AUTION MAX AX-4060 Urinalysis System consists of AUTION MAX AX-4060 urine analyzer and AUTION Sticks 9EB test strips.
The AUTION Sticks 9EB test strips consist of a plastic strip containing 9 dry chemistry reagent pads impregnated with chemical substances for the determination of the above analytes in urine. These substances if present in urine leads to a chemical reaction that results in a color change, which is measured by the AUTION MAX AX-4060 urine analyzer based on spectrophotometry. In addition, three additional parameters, specific gravity, turbidity, and color tone are directly measured based on reflectance refractometry, transmitted and scattered light measurements.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary:
Device: AUTION MAX AX-4060 Urinalysis System
Predicate Device: AUTION MAX AX-4030 Fully Automated Urinalysis System
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as distinct numerical targets in the document. Instead, the study aims to demonstrate substantial equivalence to the predicate device. Therefore, the "acceptance criteria" are implicitly understood as matching or performing comparably to the predicate device's established performance, particularly in terms of agreement percentage. The reported device performance is presented as the agreement between the proposed device (AUTION MAX AX-4060) and the predicate device (AUTION MAX AX-4030).
Key Performance Metric: Percentage of Exact Match and Percentage within +/- 1 Color Block Match compared to the predicate device.
Analyte | Acceptance Criteria (Implicit: Substantial Equivalence to Predicate) | Reported Device Performance (Exact Match %) | Reported Device Performance (+/- 1 CB Match %) |
---|---|---|---|
GLU | High agreement (Exact and +/- 1 CB match) | 98.1% | 100.0% |
PRO | High agreement (Exact and +/- 1 CB match) | 94.5% | 100.0% |
BIL | High agreement (Exact and +/- 1 CB match) | 99.5% | 100.0% |
URO | High agreement (Exact and +/- 1 CB match) | 99.2% | 100.0% |
pH | High agreement (Exact and +/- 1 CB match) | 92.9% | 100.0% |
BLD | High agreement (Exact and +/- 1 CB match) | 95.8% | 100.0% |
KET | High agreement (Exact and +/- 1 CB match) | 99.0% | 100.0% |
NIT | High agreement (Exact and +/- 1 CB match) | 99.7% | 100.0% |
LEU | High agreement (Exact and +/- 1 CB match) | 98.6% | 100.0% |
Turbidity | High agreement (Exact and +/- 1 CB match) | 98.6% | 100.0% |
Color tone | High agreement (Exact and +/- 1 CB match) | 82.7% | 100.0% |
(Note: The +/- 1 CB Match
indicates results falling within one color block difference from the predicate, which is often considered acceptable for semi-quantitative tests.)
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set (Method Comparison): A total of 1374 samples were used. This included:
- 1333 natural patient samples
- 41 spiked samples
- Data Provenance: The study was conducted at two (2) clinical laboratories. The document does not specify the country of origin, but given the FDA submission, it's likely conducted in or in accordance with standards for the U.S. and/or Japan (where the manufacturer is located). It is a prospective comparison study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This study does not involve human experts establishing a "ground truth" for each sample in the typical sense of diagnostic imaging or pathology. Instead, the "ground truth" for the method comparison study is the result obtained from the legally marketed predicate device (AUTION MAX AX-4030 Automated Urinalysis System). The study demonstrates the correlation and agreement of the new device's readings with the predicate device's readings. No information is provided about experts interpreting results aside from the automated systems.
4. Adjudication Method for the Test Set
Since the "ground truth" is established by the predicate device's results and the comparison is between two automated systems, there was no human adjudication process (e.g., 2+1, 3+1 consensus) described for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This study evaluates the performance of an automated urinalysis system against a predicate automated system, not the improvement of human readers with AI assistance.
6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)
Yes. The primary study detailed, especially the Method Comparison (Section 14), assesses the standalone performance of the AUTION MAX AX-4060 Urinalysis System. It compares the results generated by the new automated system directly against those generated by the predicate automated system, without human intervention in the result interpretation or decision-making process during the comparison phase. The device itself is an automated system intended to be used by trained operators in clinical laboratories, but the performance data presented is for the automated analyzer's output.
7. Type of Ground Truth Used
The ground truth for the method comparison study was the results obtained from the legally marketed predicate device (AUTION MAX AX-4030 Automated Urinalysis System). For the initial precision study (Tables 3-5), the "Expected Result" for quality controls served as the reference.
8. Sample Size for the Training Set
The document is a 510(k) summary for a medical device (urinalysis system), not a typically "AI-driven" software device that undergoes explicit training. The device determines analytes based on chemical reactions and spectrophotometry (reflectance refractometry, transmitted and scattered light measurements). Therefore, there is no explicit "training set" in the context of machine learning algorithms usually discussed. The system is likely calibrated and validated during its development, but this is a different process than training an AI model.
9. How the Ground Truth for the Training Set Was Established
As noted above, there is no explicit "training set" in the machine learning sense for this device. The chemical principles and measurement methods are well-established for urinalysis. Any calibration or internal development would have relied on known concentrations and reference methods for each analyte to ensure accurate colorimetric or other measurements. The basis of the technology relates to established chemical reactions on the test pads (e.g., Glucose oxidase reaction for glucose, Protein-error reaction for protein, etc.).
Ask a specific question about this device
(147 days)
Leukocyte, peroxidase
test | LJX | Class I | 21 CFR 864.7675
The cobas u 601 urinalysis test system is comprised of the cobas u 601 urine analyzer and the cobas u pack.
The cobas u 601 urine analyzer when used with the cobas u pack is a fully automated urinalysis system intended for the in vitro qualitative or semi- quantitative determination of urine analytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, color and erythrocytes, as well as clarity. These measurements are useful in the evaluation of renal, urinary, hepatic and metabolic disorders. This system is intended to be used by trained operators in clinical laboratories.
The cobas u pack is a cassette loaded with cobas u 601 test strips for the in vitro qualitative or semi-quantitative determination of pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, color and erythrocytes in urine with the cobas u 601 urine analyzer. These measurements are useful in the evaluation of renal, urinary, hepatic and metabolic disorders.
The cobas u 601 Urinalysis Test System consists of the following components:
- . cobas u 601 urine analyzer
- . cobas u pack
The cobas u 601 urine analyzer is a fully automated urine analysis system. It is optimized for high throughput workloads in the professional environment. The cobas u 601 urine analyzer performs a maximum theoretical throughput of up to 240 samples per hour.
The cobas u 601 analyzer consists of several major components:
- Rack transport system
- Liquid handling system
- Test strip cassette compartment
- Automated test strip processing area
- Photometer which is a 4 wavelength reflectance measuring unit based on a Complementary Metal Oxide Semiconductor chip used in digital cameras (CMOS sensor)
- Physical Measurement Cell (PMC): flow cell connected to an optical detector ●
- Touch Screen
- Inbuilt Computer
The functions of the cobas u 601 urine analyzer include:
- Sample loading and transport ●
- . Sample identification
- Robotic pipetting of samples onto test pads on test strips
- Robotic aspiration of samples into the PMC
- . Controlled incubation
- . Photometric measurement of test strips
- Optical determination in the PMC
- Automatic disposal of used test strips ●
- . Result readout
- Result memory
- Optional formats for data output including electronic result communication
The operating system will be a Microsoft Windows for embedded devices. The system will use a Postgres/SQL database.
The cobas u 601 urine analyzer is designed to be inter-connected mechanically and electronically with another urine sediment analyzer (cobas u 701) in order to create a urine work area (cobas® 6500).
The cobas u pack is a cassette containing 400 tests strips. The cobas u 601 analyzer will use the cobas u pack to dispense single test strips for each sample.
Each test strip has ten individual test pads that are used to test for different substances or characteristics. The test strips are analyzed automatically through the analyzer. One test strip is used per sample. When a strip is dispensed for use by the cobas u 601, an aliquot of the urine sample is pipetted onto each of the test pads. The resulting color changes are measured photometrically.
The test strip in the cobas u pack cassette ("cassette test strip") is a multi-parameter urine analysis test strip, with test pads for blood (Erythrocytes), Leukocytes, Nitrite, Proteins, Glucose, Ketones, Bilirubin, Urobilinogen, Color and pH.
Here's an analysis of the provided text, focusing on the acceptance criteria and the study proving the device meets them:
Device: cobas u 601 urinalysis test system
Predicate Devices: cobas u 411 (for pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and erythrocytes), Urisys 2400 (for specific gravity, color, and clarity).
Based on the provided 510(k) Summary
, the term "acceptance criteria" is not explicitly defined as a single, overarching set of numerical thresholds for all performance metrics. Instead, the document describes the studies performed and their results, implying that demonstrating acceptable performance within clinical ranges and in comparison to predicate devices constitutes meeting the "acceptance criteria" for substantial equivalence.
For each study, the "acceptance criteria" are implied by the reported results meeting the necessary performance for a diagnostic device, particularly demonstrating "exact agreement" or "agreement ± 1 block" within clinically relevant ranges and comparable to the predicate devices.
The information is extracted from the "NON-CLINICAL PERFORMANCE EVALUATION" and "CLINICAL PERFORMANCE DATA" sections.
1. Table of Acceptance Criteria (Implied) and Reported Device Performance
Since explicit numerical acceptance criteria for each test in a pass/fail format are not directly stated as "acceptance criteria," the table below presents the implied acceptance criteria (what the study aimed to demonstrate as acceptable performance) and the reported device performance as found in the document. The primary method for showing acceptance is often "exact agreement" or "agreement ± 1 block" with the predicate device/reference.
Parameter / Study Type | Implied Acceptance Criteria (Goal) | Reported Device Performance (Achieved) |
---|---|---|
Analytical Sensitivity (Lowest concentration for ≥90% detection) | To detect analytes at specified low concentrations with high confidence (≥90% detection). | LEU: 10 Leu/μL (meets criteria of ≥90% detection) |
NIT: 0.045 mg/dL (meets criteria of ≥90% detection) | ||
PRO: 9 mg/dL albumin (meets criteria of ≥90% detection) | ||
GLU: 25 mg/dL (meets criteria of ≥90% detection) | ||
KET: 4 mg/dL (meets criteria of ≥90% detection) | ||
BIL: 0.6 mg/dL (meets criteria of ≥90% detection) | ||
UBG: 1.15 mg/dL (meets criteria of ≥90% detection) | ||
ERY: 7 Ery/μL (meets criteria of ≥90% detection) | ||
Drug & Endogenous Interferences | No significant interference from tested therapeutic drugs and endogenous substances at specified concentrations, that would impact clinical interpretation. Reported interferences are acknowledged and included in labeling claims. | No Interference: Acetaminophen, Amoxicillin, Biotin, Cefoxitin, Furosemide, Gabapentin, Gentamycin Sulfate, Ibuprofen, Levodopa, Lisinopril, Metformin, Methyldopa, Methenamine + Methylene blue, N-Acetyl-Cysteine, Ofloxacin, Phenazopyridine, Salicyluric acid, Tetracycline, ß-3-Hydroxybutyrate, Human IgG, Uric acid. |
Interference (listed in method sheet): Specific interferences with various analytes (ERY, LEU, NIT, PRO, GLU, KET, UBG, BIL) by therapeutic drugs and endogenous substances were identified and reported in tables showing "No Interference up to" a certain concentration, and the "Effect above stated concentration." These are noted in the product labeling. | ||
Color Interference | The system's color compensation functionality should ensure accurate analyte measurements despite urine color variations. For negative parameters, 100% negative results; for positive, 100% exact agreement during color compensation. | Results: For all tested parameters (Nitrite, Ketone, Glucose, Bilirubin, Urobilinogen, Erythrocytes, Leukocytes) at both negative/normal and positive concentrations, both with and without added color interferents (Bilirubin for orange, Erythrocyte for red, Urobilinogen for brown; Hemoglobin, Sunset Yellow, Lignin for specific parameters), the exact agreement was consistently 100% when color compensation was active, with minor exceptions (e.g., Leu: 90% and 20% exact agreement for Sunset Yellow at 40 Leu/µL, though the table notes 100% for negative). The summary implies the system performs as expected. |
Shelf-life Stability | cobas u pack stable for 15 months at room temperature. | The cobas u pack is stable at room temperature for 15 months. |
On-board Stability | cobas u pack stable for 14 days during operation on the system. | The cobas u pack is stable up to 14 days during operation on the system. |
Repeatability (Within-run precision) | 100% exact agreement for controls (negative/normal and positive analyte concentrations). | Achieved: 100% exact agreement for all analytes (pH, ERY, LEU, PRO, GLU, KET, UBG, BIL, COL) at both Level 1 (Neg/Norm) and Level 2 (high positive) controls. |
Intermediate Precision | High percentage of exact agreement for controls. | Achieved: 100% exact agreement for most analytes (pH, ERY, PRO, GLU, KET, UBG, BIL, COL). LEU Level 2 had 95.2% exact agreement. |
Method Comparison (vs. cobas u 411) | High exact agreement and overall agreement with predicate, along with acceptable sensitivity and specificity. Thresholds varied by parameter. | ERY: Exact Agreement (100% fit): 85-100% (6/6 ranges passed); Overall: 99%; Specificity: 99%; Sensitivity: 99%. |
LEU: Exact Agreement: 88-99% (4/4 ranges passed); Overall: 99%; Specificity: 99%; Sensitivity: 97%. | ||
NIT: Exact Agreement: 99-100% (2/2 ranges passed); Overall: 100%; Specificity: 99%; Sensitivity: 100%. | ||
KET: Exact Agreement: 88-99% (5/5 ranges passed); Overall: 99%; Specificity: 99%; Sensitivity: 97%. | ||
GLUC: Exact Agreement: 86-100% (5/5 ranges passed); Overall: 99%; Specificity: 99%; Sensitivity: 100%. | ||
PRO: Exact Agreement: 87-98% (5/5 ranges passed); Overall: 99%; Specificity: 98%; Sensitivity: 100%. | ||
BIL: Exact Agreement: 91-100% (4/4 ranges passed); Overall: 99%; Specificity: 100%; Sensitivity: 98%. | ||
UBG: Exact Agreement: 87-99% (5/5 ranges passed); Overall: 99%; Specificity: 99%; Sensitivity: 98%. | ||
pH: Exact Agreement: 70-97% (6/6 ranges passed); Overall: 95%; Specificity: 98%. | ||
Method Comparison (vs. Urisys 2400 for Color) | High agreement rates for color classification. | Agreement rates: Pale yellow (81%), Yellow (70%), Amber (65%), Brown (88%), Orange (68%), Red (91%). Overall agreement implied by diagonal matches. |
Method Comparison (vs. Urisys 2400 for Clarity) | High exact agreement and agreement ± 1 color block. | Exact agreement (%): Clear (89%), Light Turbid (80%), Turbidity (84%). |
Agreement ± 1 color block: 100% for all clarity categories. | ||
Sample Carryover | No risk to patient safety due to carryover. | Results met pre-defined acceptance criteria for BIL, GLU, KET, LEU, ERY, NIT, PRO, UBG, pH, COL and SG. Deviations were observed for Clarity but considered to pose no risk due to low medical relevance. |
2. Sample Sizes and Data Provenance
-
Test Set Sample Sizes:
- Analytical Sensitivity: Multiple samples tested for each analyte, each sample measured 20 times on each of 3 instruments using 3 reagent test strip lots. (e.g., 20 measurements x 3 instruments x 3 lots = 180 total per sample condition per analyte if all combinations were tested). Specific number of "samples" (unique spiked concentrations) not explicitly stated for each analyte.
- Drug and Endogenous Interferences: Urine pools (negative/normal and first positive range) tested at 2 concentrations of interferents. Multiple replicates measured. Number of replicates not specified for each condition.
- Color Interference: Final test solutions (for each parameter) tested in a 10-fold determination. (e.g., 10 measurements per condition).
- Stability:
- Real-time: Defined set of samples (native urine, artificial urine, low/high spiked urine) measured with n=10 determinations at each time point (0, 3, 13, 16 months).
- On-board: 400 tests over 15 days from a single cassette (using native and artificial urine samples).
- Precision (Repeatability): Controls measured in 2 runs, 21 determinations each, producing n=42 results per control.
- Precision (Intermediate Precision): Controls measured in 21 days with 2 runs per day and duplicate measurements per control, producing n=84 results per control.
- Method Comparison (cobas u 411): "fresh samples" used to cover claimed ranges. Specific total number of samples for comparison is not explicitly stated.
- Method Comparison (Urisys 2400):
- Color: 478 total samples.
- Clarity: 1364 total samples.
- Sample Carryover: Not specified, but involved testing low/negative and high concentration samples.
-
Data Provenance: The document does not explicitly state the country of origin for the data or whether the studies were retrospective or prospective. Given it's a 510(k) submission for an in vitro diagnostic device, these are typically prospective laboratory studies conducted by the manufacturer, often at their R&D facilities or contracted clinical sites.
3. Number of Experts and Qualifications for Ground Truth
- The document does not specify the number of experts used to establish ground truth for the test sets.
- It also does not specify the qualifications of these experts.
- For urinalysis strips, ground truth is typically established by reference methods such as quantitative chemical assays or microscopic examination, rather than relying solely on human expert consensus on visual interpretation of the strips themselves. The comparison is made against a "reference system" which implies an objective and validated method.
4. Adjudication Method for the Test Set
- The document does not mention any adjudication method (e.g., 2+1, 3+1) for the test sets. For objective chemical measurements like those performed by this device, human adjudication of "ground truth" is typically less relevant than the use of quantitative reference methods.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- A MRMC comparative effectiveness study was not conducted. This device is an automated urinalysis system, and its performance is evaluated against established analytical methods and predicate devices, not by comparing human reader performance with and without AI assistance. The "AI" component is implicit in the automated analysis of reflectance, which is a core function of the machine's software, but it's not presented as an AI-assistive tool for human readers.
6. Standalone (Algorithm Only) Performance
- Yes, the performance data presented (e.g., Analytical Sensitivity, Accuracy/Method Comparison studies) represents the standalone performance of the cobas u 601 urinalysis test system. It is an automated device designed to perform urinalysis without human interpretation of the test strip results; the human role is in operating the system and interpreting the numerical/qualitative results provided by the machine.
7. Type of Ground Truth Used
- The ground truth used for these studies generally aligns with:
- Reference System/Predicate Device Comparison: For the method comparison studies, the cobas u 411 and Urisys 2400 systems served as "reference systems" against which the new device's qualitative and semi-quantitative results were compared.
- Spiked Samples/Defined Concentrations: For sensitivity, interference, stability, and precision studies, the ground truth was established by preparing urine samples with precisely known concentrations of analytes or interfering substances ("spiking the negative urine pool with the appropriate agent," "known concentrations").
- Control Materials: For precision studies, standardized control materials with known values were used.
8. The Sample Size for the Training Set
- The document does not provide information regarding the sample size for a "training set." This type of device, based on reflectance photometry and chemical reactions, typically relies on predetermined algorithms derived from extensive analytical characterization of the strip chemistry and optical properties, rather than "training" an AI model in the conventional machine learning sense using a large, distinct "training set" of patient data. The development process would involve calibration and algorithm refinement using controlled samples, but not necessarily a "training set" as defined in AI/ML contexts with expert-labeled ground truth for each case.
9. How the Ground Truth for the Training Set Was Established
- Since there's no explicit mention of a "training set" in the context of an AI/ML model for this device, the question of how ground truth was established for it is not applicable in the provided document. The "training" in this context would refer to the calibration and algorithm development process, which relies on the principles of analytical chemistry and physics inherent to reflectance photometry to accurately read the color changes on the test strips. This would involve precise chemical and optical characterization using known standards and samples.
Ask a specific question about this device
(106 days)
Leukocyte peroxidase test | LJX | Class I | 21 CFR §864.7675
The inui In-Home Urine Analysis Test System consists of the inui In-Home Urine Analysis Device and the inui Urine Analysis Mobile Application. The inui In-Home Urine Analysis Test System is intended for detecting the following parameters in urine: Protein, Glucose, Leukocytes, Nitrites, and Ketones. The test results provide information regarding the status of Urinary Tract Infections (UT), proteinuria, glucosuria, and ketonuria. These results can be used as an aid for monitoring kidney functions and metabolic disorders (e.g. diabetes mellitus), and can be used in the screening for Urinary Tract Infections (UTI).
The inui In-Home Urine Analysis Device is intended for use as a prescription home use device.
The inui In-Home Urine Analysis Test System consists of inui In-Home Urine Analysis Device and the inui Urine Analysis Mobile Application. It is an in vitro diagnostic device comprised of a disposable test paddle, a urine collection cup, a gray background sheet, and a mobile application ("inui App"). The plastic "paddle" contains multiple chemistry test pads (CTPs) and 1 quick response (QR) barcode. CTPs contain pre-dried chemicals that react to specific substances in a urine sample. A color reaction occurs on the CTP based on the amount of substance in the urine sample. The color reaction is captured as an image using the mobile device camera and should only be read using the inui App. The results are reported as specified by the measurement range for each test. The QR Code is used for quality control purposes to track lot number, expiration date, and paddle use.
The inui In-Home Urine Analysis System reports semi-quantitative or qualitative results for each test parameters.
Here's a breakdown of the acceptance criteria and the study details for the inui In-Home Urine Analysis Test System, based on the provided FDA 510(k) summary:
Acceptance Criteria and Reported Device Performance
General Acceptance Criteria from Analytical Performance:
- Precision and Reproducibility: Proportion of agreement for each level and analyte tested to be above 95% to the expected values.
- Limit of Detection (LOD) and Linearity: Established for each level and analyte.
- Specificity and Interference Testing: Identification of interfering substances and their minimum concentration levels causing false positives/negatives. The device should produce an error message for substances introducing colors outside the analyte's normal range.
Clinical Performance Acceptance Criteria:
- Clinical Accuracy (Method Comparison): Lay-user clinical test results should be comparable to those of professional users using the predicate device. Specific agreement percentages are implicitly set by the reported data in Table 5.
- Usability: Lay users should be able to follow procedural steps to obtain successful test results with a specified performance percentage.
- Lay-user Reproducibility: Lay-users should obtain the same results after repeated testing with a high reproducibility percentage.
1. Table of Acceptance Criteria and the Reported Device Performance
Parameter / Indication for Use | Acceptance Criteria (Implicit from Study Design) | Reported Device Performance |
---|---|---|
Analytical Performance | ||
Precision & Reproducibility | >95% agreement to expected values for each level/analyte. | Achieved: "The proportion of agreement for each level and analyte tested met the established acceptance criteria and was found to be above 95% agreement to the expected values for both precision and reproducibility." (Page 7) |
Limit of Detection (LOD) | Established for each level/analyte. | Achieved: LODs were determined, and a "Summary of the sensitivities are provided in Table 3." (Page 7) |
Specificity / Interference | Identification of interfering substances, their levels, and a mechanism for handling abnormal urine colors (e.g., error message). | Achieved: "Thirty-nine potentially interfering exogenous and endogenous substances were evaluated." The device "produces an error message with no result (invalid) for any substance that introduces colors outside the range for that analyte." (Page 7-8) Acetoacetic Acid consistently produced an error message. Table 4 lists interfering substances and their levels. Other identified interferences include: |
- Protein: Albumin (>3000 mg/dL), Bilirubin (>10 mg/dL), Hemoglobin (>100 mg/dL), Chlorhexidine (>40 mg/dL), Riboflavin (>5 mg/dL), Hypochlorite (>375 mg/dL), Specific Gravity (>1.025).
- Leukocyte: Human Leukocyte Esterase (>0.025 U/mL), Hemoglobin (>150 mg/dL), Albumin HSA (>3000 mg/dL), Bilirubin (>10 mg/dL), Sodium Chloride (>324 mg/dL), Hypochlorite (>375 mg/dL), Chlorhexidine (>60 mg/dL), Microbial Peroxidase (>0.65%), Riboflavin (>5 mg/dL), Sodium Acetate (>600 mg/dL), Sodium Bicarbonate (>630 mg/dL), Specific Gravity (>1.020), Urobilinogen (>4 mg/dL).
- Nitrite: Sodium Nitrite (>5 mg/dL), Human Hemoglobin (>100 mg/dL), Hypochlorite (>375 mg/dL), Human Leukocyte (>0.0375 U/mL), Urobilinogen (>4 mg/dl), Sodium Bicarbonate (>945mg/dl), Sodium Acetate (>900 mg/dL), Hypochlorite (>375 mg/dL), Sodium Nitrite (>7.5 mg/dL).
- Ketone: Lithium Acetoacetate (>40 mg/dL), Hypochlorite (>375 mg/dL), Sodium Nitrite (>7.5 mg/dL).
- Glucose: D-(+)-Glucose (>500 mg/dL), Hypochlorite (>375 mg/dL), Bilirubin (invalid results at all conc), Lithium Acetoacetate (>80 mg/dL), Sodium Chloride (>486 mg/dL). (Pages 8-9) |
| Clinical Performance | | |
| Method Comparison (Lay User vs. Predicate) | Lay user results comparable to professional users with predicate device; high agreement percentages. An explicit threshold for each analyte/level is implied by the successful submission. | Achieved; "clinical test results that are comparable to those of a professional users".
Protein: Negative (83% exact, 100% within 1 block), Trace (75% exact, 100% within 1 block), Moderate (90% exact, 100% within 1 block), Large (100% exact, 100% within 1 block).
Glucose: Negative (100% exact, 100% within 1 block), Low (74% exact, 100% within 1 block), Moderate to Large (100% exact, 100% within 1 block).
Leukocyte: Negative (100% exact), Positive (97% exact).
Nitrite: Negative (96% exact), Positive (100% exact).
Ketone: Negative/Trace (99% exact, 100% within 1 block), Small (67% exact, 100% within 1 block), Moderate to Large (100% exact, 100% within 1 block). (Table 5, Page 10) |
| Usability | Lay users can follow all procedural steps to obtain a successful test result, with a high overall performance percentage. | Achieved: Overall performance of 88.9% (1st paddle), 86.7% (2nd paddle), and 91.1% (3rd paddle). (Page 10) |
| Lay-user Reproducibility | High reproducibility (e.g., 100%) for lay-user repeated testing. | Achieved: "reproducibility of the inui Device in the hands of the lay user was 100%." (Page 10) |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size:
- Clinical Accuracy (Method Comparison): Two studies were conducted using 190 and 91 lay users, respectively.
- Usability Study: 45 lay users.
- Lay-User Reproducibility Study: 10 lay users.
- Data Provenance: The document does not explicitly state the country of origin. It indicates that the clinical evaluation followed an "Investigational Review Board approved clinical protocol," suggesting prospective data collection.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- The ground truth for the clinical accuracy study was established by comparing lay-user results from the inui device to results obtained from the predicate device, Siemens Multistix® 10SG, "by professional users."
- The document does not specify the exact number of professional users or their specific qualifications (e.g., "radiologist with 10 years of experience"). It just refers to them as "professional users."
4. Adjudication Method for the Test Set
- The document describes a direct comparison method where lay-user results from the inui device are compared against results from the predicate device read by professional users.
- It does not mention an explicit adjudication method like "2+1" or "3+1" to resolve discrepancies between multiple professional readers or to establish a consensus ground truth beyond the professional reading of the predicate device. The professional reading of the predicate device serves as the reference.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study involving AI assistance for human readers was not explicitly described.
- The study focuses on the standalone performance of the inui system (which incorporates an app for image capture and analysis) when used by lay-users, against a predicate device read by professional users.
- The system itself is the "AI assistance" in the sense that the mobile app performs the color analysis traditionally done visually. The comparison is between the inui system's performance by lay users and the predicate device's performance by professional users, not human readers with vs. without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, in spirit, the core of the inui device's performance is essentially standalone algorithm performance within the context of a "lay-user-in-the-loop" setting.
- The "inui App" performs the electronic color analysis via the mobile phone camera. The "lay user reproducibility study" and the "clinical accuracy study" demonstrate the device's (including the algorithm's) performance when operated by lay users. The comparison in the clinical accuracy study is effectively between the inui device's algorithm output (managed by lay users) and a predicate device's visual interpretation by professionals.
7. The Type of Ground Truth Used
- The ground truth for the clinical accuracy (method comparison) study was established by comparison to a legally marketed predicate device (Siemens Multistix® 10SG device) assessed by professional users. This is a form of "reference method" ground truth.
- For the analytical performance studies (Precision, LOD, Specificity), the ground truth was based on expected values and established protocols in accordance with CLSI guidelines.
8. The Sample Size for the Training Set
- The document is a 510(k) summary, which focuses on validation data. It does not provide information on the sample size for the training set used to develop the inui device's mobile application algorithm.
9. How the Ground Truth for the Training Set Was Established
- The document does not provide details on how the ground truth for the training set was established, as it pertains to the development process of the device which is typically not included in a 510(k) summary.
Ask a specific question about this device
(35 days)
I |
| Leukocyte peroxidase test | 21 CFR 864.7675
This device is intended for the in vitro measurement of the following in urine: Leukocyte, Nitrite, Urobilinogen, Protein, pH, Blood, Specific gravity, Ketone, Bilirubin, Glucose. These strips are intended for prescription, in vitro diagnostic use only and they are visually read.
DUS 2GP reagent strips provide qualitative and semiquantitative measurements for protein, and glucose in urine specimens. Test results may provide information regarding the status of carbohydrate metabolism and kidney function.
DUS 5 reagent strips provide qualitative and semiquantitative measurements for leukocytes, nitrite, blood , protein, and glucose in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies and urinary tract infections.
DUS 10 reagent strips provide qualitative and semiquantitative measurements for specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin and blood in urine speciments are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections and liver function.
The DUS Series are urine test strips with different reagent pads for the determination of specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin and blood affixed onto plastic strips. of which leukocyte, nitrite, urobilinogen, protein, pH, blood, specific gravity, ketone, bilirubin and glucose reagent pads are affixed onto the plastic strips. The reagent pads react with analytes in the urine giving a visible color. Results are confirmed by comparison of the test strip with the color chart on the container. For each color result for each analyte, a semiquantitative value is available on the box label (e.g. bilirubin results include negative, 1, 2, and 4 mg/dL) and the associated qualitative result (e.g. bilirubin results include negative. +. ++, +++).
Here's an analysis of the provided document, outlining the acceptance criteria and study details for the DUS Reagent Strips:
Acceptance Criteria and Device Performance
The provided document does not explicitly state pre-defined acceptance criteria for the "Exact Agreement" or "Agreement within +/- one block (%)" in the method comparison study. However, the study results, which consistently show high percentages (generally in the high 90s and 100%) for both metrics across all analytes, implicitly represent the device meeting an expected high level of agreement with the predicate device.
For the linearity/assay reportable range, the acceptance criteria are implied by the reported "% Exact match." A high percentage (generally 97.7% to 100%) indicates acceptance.
For the detection limit study, the acceptance criteria are stated as "Sensitivity was defined as the cutoff for which ≥95% of the contrived pooled measurements were trace or the first positive result."
The interference study implicitly accepts the device if interference is either not observed at clinically relevant concentrations or if observed interferences are adequately addressed in the labeling.
Here's a table summarizing the reported device performance, where the values themselves act as the demonstration that the implied or stated acceptance criteria (high agreement, 95% detection, or non-interference/labeling for interference) were met.
Table of Reported Device Performance
Analyte | Test Type | Reported Device Performance (Exact agreement / % Exact match / Positive Agreement / Notes) |
---|---|---|
Precision (Within-run, N=90) | ||
Urobilinogen | Level 1 (Normal) | Exact: 100%, +/-1 block: 100% |
Level 2 (4mg/dL) | Exact: 100%, +/-1 block: 100% | |
Glucose | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (1000mg/dL) | Exact: 100%, +/-1 block: 100% | |
Bilirubin | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (4mg/dL) | Exact: 100%, +/-1 block: 100% | |
Ketones | Level 1 (Positive) | Exact: 100%, +/-1 block: 100% |
Level 2 (Positive) | Exact: 100%, +/-1 block: 100% | |
SG | Level 1 (1.020) | Exact: 100%, +/-1 block: 100% |
Level 2 (1.020) | Exact: 98.9%, +/-1 block: 100% | |
Blood | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (200 RBC/uL) | Exact: 100%, +/-1 block: 100% | |
pH | Level 1 (6) | Exact: 100%, +/-1 block: 100% |
Level 2 (7) | Exact: 100%, +/-1 block: 100% | |
Protein | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (100 mg/dL) | Exact: 100%, +/-1 block: 100% | |
Nitrite | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (Pos) | Exact: 100%, +/-1 block: 100% | |
Leukocytes | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (70 WBC/uL) | Exact: 100%, +/-1 block: 100% | |
Precision (Within-day, N=90) | ||
Urobilinogen | Level 1 (Normal) | Exact: 100%, +/-1 block: 100% |
Level 2 (4mg/dL) | Exact: 100%, +/-1 block: 100% | |
Glucose | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (1000mg/dL) | Exact: 100%, +/-1 block: 100% | |
Bilirubin | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (4mg/dL) | Exact: 100%, +/-1 block: 100% | |
Ketones | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (40mg/dL) | Exact: 100%, +/-1 block: 100% | |
SG | Level 1 (1.020) | Exact: 100%, +/-1 block: 100% |
Level 2 (1.020) | Exact: 100%, +/-1 block: 100% | |
Blood | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (200 RBC/uL) | Exact: 100%, +/-1 block: 100% | |
pH | Level 1 (6) | Exact: 100%, +/-1 block: 100% |
Level 2 (7) | Exact: 98.9%, +/-1 block: 100% | |
Protein | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (100 mg/dL) | Exact: 100%, +/-1 block: 100% | |
Nitrite | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (Pos) | Exact: 100%, +/-1 block: 100% | |
Leukocytes | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (70 WBC/uL) | Exact: 100%, +/-1 block: 100% | |
Linearity / Assay Reportable Range (N=90 per concentration) | ||
Urobilinogen | Various Conc. | 97.7% - 100% Exact match |
Glucose | Various Conc. | 97.7% - 100% Exact match |
Bilirubin | Various Conc. | 96.6% - 100% Exact match |
Ketones | Various Conc. | 97.7% - 100% Exact match |
Specific Gravity | Various Conc. | 94.4% - 100% Exact match |
Nitrite | Various Conc. | 100% Exact match |
Blood | Various Conc. | 98.8% - 100% Exact match |
pH | Various Conc. | 97.7% - 100% Exact match |
Protein | Various Conc. | 96.6% - 100% Exact match |
Leukocytes | Various Conc. | 97.7% - 100% Exact match |
Detection Limit (N=90 per concentration) | ||
Urobilinogen | 2 mg/dL | 100% Positive Agreement |
Glucose | 100 mg/dL | 100% Positive Agreement |
Bilirubin | 1.0 mg/dL | 100% Positive Agreement |
Ketones | 5 mg/dL | 100% Positive Agreement |
Blood | 10 RBC/μL | 100% Positive Agreement |
Protein | 15 mg/dL | 100% Positive Agreement |
Nitrite | 0.05 mg/dL | 100% Positive Agreement |
Leukocytes | 15 WBC/μL | 100% Positive Agreement |
Analytical Specificity | ||
(Various Analytes) | (Various Interferents) | No interference observed for a long list of compounds at specified concentrations. Observed interferences were noted and addressed in labeling limitations. |
Method Comparison Study (Total N=867 samples; comparisons to Multistix 10SG) | ||
Urobilinogen | All levels | Exact: 92-100%, Within 1 block: 100% |
Glucose | All levels | Exact: 92-100%, Within 1 block: 100% |
Bilirubin | All levels | Exact: 92-100%, Within 1 block: 100% |
Ketones | All levels | Exact: 91-100%, Within 1 block: 100% |
Blood | All levels | Exact: 92-100%, Within 1 block: 100% |
Protein | All levels | Exact: 91-100%, Within 1 block: 100% |
Nitrite | All levels | Exact: 97-100%, Within 1 block: 100% |
Leukocytes | All levels | Exact: 93-100%, Within 1 block: 100% |
pH | All levels | Exact: 91-100%, Within 1 block: 100% |
Specific Gravity | All levels | Exact: 91-100%, Within 1 block: 100% |
Study Details
-
Sample size used for the test set and the data provenance:
- Method Comparison Study: Total of 867 samples.
- Provenance: Fresh urine samples obtained at three medical facilities. The country of origin is not explicitly stated but implied to be South Korea, given the submitter's address. The data is prospective as samples were "processed within 4 hours."
- Precision/Reproducibility: Two levels of commercially available urine-based control solutions.
- Sample Size: 90 replicates for within-run (10 tests from 3 lots at 3 sites) and 90 replicates for within-day (1 test a day from 3 lots, at 3 sites for 10 days) for each level.
- Provenance: Commercially available control solutions.
- Linearity/Assay Reportable Range:
- Sample Size: 90 replicates per concentration level (10 replicates with each of 3 lots of test strips).
- Provenance: Samples created by spiking known concentrations of standard materials or by serial dilution of a high concentration urine sample with negative urine.
- Detection Limit:
- Sample Size: 90 replicates for each concentration (each sample concentration analyzed 30 times using 3 reagent strip lots).
- Provenance: Negative urine spiked with standard materials.
- Analytical Specificity:
- Sample Size: 3 replicates using 3 lots of DUS 10 test strips for each concentration level of interfering substance.
- Provenance: Urine sample pools prepared at 3 analyte concentrations (negative, low, high positive) spiked with potential interfering substances.
- Method Comparison Study: Total of 867 samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Precision, Linearity, Detection Limit, Analytical Specificity: Ground truth was established by the known concentrations of control solutions or spiked samples. Testing was performed by 3 medical technicians as reported in the linearity study and precision study, but their specific qualifications (e.g., years of experience) are not stated beyond being "medical technicians."
- Method Comparison Study: The ground truth for the method comparison study was established by comparing the DUS 10 test strips results against a predicate device (Multistix 10SG). The testing was performed by three medical technicians at each of the clinical sites. Similar to above, their specific qualifications are not detailed beyond "medical technicians."
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not describe an explicit adjudication method (like 2+1 or 3+1) involving multiple human readers or experts resolving discrepancies for the test set in the traditional sense of image-based AI studies.
- The method comparison study compares the new device's readings to the predicate device's readings, and the percentage agreement is calculated. The process for resolving discrepancies between the new device and the predicate device is not detailed, nor is there a mention of an expert panel reviewing cases.
- For other studies (precision, linearity, detection limit), the ground truth is analytically determined by control concentrations.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader, multi-case (MRMC) comparative effectiveness study as typically understood for AI-assisted diagnostic tools (i.e., human readers with and without AI assistance) was not explicitly performed or described in this document.
- The studies focus on the performance characteristics of the device itself (the reagent strips) and its comparison to a predicate device, which is also a reagent strip for visual reading. The device's use is "visually read" by operators, but the study design is not one of AI assistance to human readers.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- This device is a reagent strip that is visually read. It is not an AI algorithm. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" does not directly apply in the context of an AI device. The performance data presented (precision, linearity, detection limits, analytical specificity, method comparison) are essentially standalone performance characteristics of the physical reagent strip, with human visual interpretation being the intended mode of operation for reading the results. The method comparison specifically assesses this human-read performance against a predicate device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Method Comparison Study: The ground truth for comparative analysis was the results obtained from the predicate device (Multistix 10SG).
- Precision, Linearity, Detection Limit, Analytical Specificity: The ground truth was analytically determined concentrations derived from commercial control solutions or spiked urine samples with known analyte levels.
-
The sample size for the training set:
- This document is for a traditional in-vitro diagnostic (IVD) reagent strip that is visually read, not an AI/ML device. Therefore, there is no "training set" in the context of machine learning model development. The assays are based on chemical reactions, and the performance is inherent to the chemical formulation and strip manufacturing.
-
How the ground truth for the training set was established:
- As there is no AI/ML component or "training set" in the context of an algorithm, this question is not applicable. The chemical reactions on the strips are designed to react to specific analytes, and their performance is validated through the studies described.
Ask a specific question about this device
(268 days)
| JIR | I |
| Leukocyte peroxidase | 21 CFR 864.7675
The DUS R-50S System provides a qualitative measurements for specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin, blood, microalbumin and creatinine in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections and liver function. The system is intended for prescription, in vitro diagnostic use only.
The DUS R-50S (Urine Chemistry system) is a portable analyzer. It is designed to read only DUS Series for urinalysis. This analyzer reports semi-quantitatively assays for 12 urine analytes [Leukocyte, Nitrite, Urobilinogen, Protein, pH, Blood, Specific gravity, Ketone, Bilirubin, Glucose, Microalbumin, Creatinine]. Reagent strip results are automatically displayed on the screen. The DUS R-50S is intended for in vitro diagnostic use only
The DUS R-50S (Urine Chemistry system) device is intended for qualitative and semi-quantitative measurements of various analytes in urine specimens to aid in diagnosing metabolic disorders, kidney function anomalies, urinary tract infections, and liver function.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally implied by the requirement for "very high concordance of between 90-100%" with the predicate device and the "majority of analysis display 90-100 % concordance over all blocks." The detailed "Exact agreement (%)" and "Within one block (%)" values for each analyte serve as the reported device performance against these implicit acceptance criteria.
Analyte | Acceptance Criteria (Implied) | Reported Device Performance (Exact agreement %) | Reported Device Performance (Within one block %) |
---|---|---|---|
Urobilinogen | 90-100% concordance | 99.3% | 100% |
Glucose | 90-100% concordance | 99.2% | 100% |
Bilirubin | 90-100% concordance | 99% | 100% |
Ketones | 90-100% concordance | 98.7% | 100% |
Blood | 90-100% concordance | 97% | 100% |
Protein | 90-100% concordance | 98% | 100% |
Nitrite | 90-100% concordance | 99.5% | 100% |
Leukocytes | 90-100% concordance | 98% | 100% |
pH | 90-100% concordance | 94.3% | 100% |
Specific Gravity | 90-100% concordance | 94.9% | 100% |
Creatinine | 90-100% concordance | 94.3% | 100% |
Microalbumin | 90-100% concordance | 98.4% | 100% |
2. Sample Size Used for the Test Set and Data Provenance
The method comparison study was conducted with a total of 867 samples. The data provenance is described as being collected at three clinical sites. The document does not specify the country of origin of the data, but the submitter information lists DFI Co., Ltd. in Korea. The study appears to be retrospective, as it compares the new device results against a legally marketed predicate device using collected samples.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and their Qualifications
The document does not explicitly mention the use of experts to establish ground truth for the test set. Instead, the performance is evaluated by "method comparison" against a predicate device (Siemens Clinitek Status + urine chemistry instrument). The predicate device's results are considered the reference for comparison.
4. Adjudication Method for the Test Set
There is no mention of an adjudication method involving human experts for the test set. The comparison is directly between the new device and the predicate device.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted as per the provided information. The study focuses on comparing the new device's analytical performance against a predicate device, not on human reader performance with or without AI assistance.
6. Standalone Performance Study
Yes, a standalone performance study was done in the form of a method comparison study where the DUS R-50S (DUS R-50S instrument, DUS10 and DUS2AC reagent strips) was compared to the predicate device (Siemens Clinitek Status + urine chemistry instrument using Multistix 10SG and CLINITEK Microalbumin 2 test strips). This evaluates the algorithm and device's performance directly.
7. Type of Ground Truth Used
The "ground truth" for the test set was the results obtained from the predicate device, specifically the Siemens Clinitek Status + urine chemistry instrument using Multistix 10SG and CLINITEK Microalbumin 2 test strips.
8. Sample Size for the Training Set
The document does not explicitly specify a separate sample size for a "training set." The performance evaluation focuses on the method comparison study using 867 samples. For a 510(k) submission for an in vitro diagnostic device, the manufacturer typically performs internal validation and verification studies during development, but the detailed breakdown of training data vs. test data for algorithm development is not typically part of the regulatory submission summary provided.
9. How the Ground Truth for the Training Set Was Established
Since a "training set" is not explicitly mentioned or detailed, the method for establishing its ground truth is not provided. If an algorithm was developed using machine learning, the ground truth for training would generally be established by laboratory reference methods or expert interpretation of the results from traditional methods, similar to how the predicate device results are used for the test set.
Ask a specific question about this device
(251 days)
Regulation Classification Section:
Class I: 21 CFR 864.7675 (meets the limitations of exemptions per
The Scanostics UTI Check Application test system consists of the Scanostics UTI Check Application and the URS-2L (UTI) Urine Test Strips. The test system is intended for the qualitative detection of nitrite and leukocytes in urine as an aid in the screening of urinary tract infections (UTI). It is intended for over-the-counter home use only.
The Scanostics UTI Check Application test system consists of the Scanostics UTI Check Application and the URS-2L (UTI) Urine Test Strips. The Scanostics UTI Check Application measures the colour developed in two reaction zones (leukocytes and nitrite pads) on the UTI test strip following application of a urine sample. The developed colors are then compared to calibration colours located on the Scanostics backing material and the result for each pad is determined based on the minimum colour distance between the developed colours and calibration colours.
The URS-2L UTI Test Strip comprises of two reagent pads areas, which are absorbent material saturated with chemically active substances, then dried and affixed to the plastic strip with double-sided adhesive.
The backing card comprises of thirty-two (32) calibration colour blocks and three (3) black and white alignment squares printed onto a rigid card stock and die cut to provide a groove for the strip (preventing strip misalignment). The use of the backing card is primarily to compensate for different lighting environments as it allows the calibration colours and the test strip to be subjected to the same conditions as the reagent pads for comparison - this would not be possible if the calibration colours were stored within the application itself.
The representative platform for the test system is the iPhone 6 with iOS 9. The application has been proven to be compatible on the iPhone 6S (iOS 9) and iPod touch (8th gen with iOS 8 and iOS 9). The application's core technology is based on four (4) patents held by Teco Diagnostics (US 8655009, 8877140, 8911679 and 8506901).
Here's an analysis of the provided text to extract the acceptance criteria and study details for the Scanostics UTI Check Application Test System.
Please note: The provided text is a 510(k) summary, which is a regulatory document. It summarizes the findings rather than presenting the full study report. Therefore, some information like specific statistical metrics for acceptance criteria and the exact number of participants in all study phases might not be explicitly detailed.
Acceptance Criteria and Reported Device Performance
The document does not explicitly state numerical acceptance criteria (e.g., sensitivity > X%, specificity > Y%). Instead, it indicates that the device should perform "comparably" to the predicate device in the hands of lay users and "satisfactorily" in non-clinical studies. The ultimate acceptance criterion implicitly relies on the demonstration of substantial equivalence to the legally marketed predicate device, Clinistrip (URS-10).
Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|
Obtain comparable testing data to a trained HCP using predicate device | "Results indicate that the intended users were able to obtain comparable testing data when using the Scanostics UTI Check Application as a trained HCP using the Clinistrip (URS-10)." This implies that the performance in the hands of lay users matches that of a healthcare professional using the established predicate. The specific metrics (e.g., positive agreement, negative agreement) are not provided in this summary. |
Satisfactory performance in non-clinical studies | "Testing results indicate that the Scanostics UTI Check application test system performs satisfactorily when used appropriately." This covers method comparison, precision, detection limit, interference, specificity, shelf life and stress studies, and several flex studies, but specific quantitative results or thresholds are not detailed. |
Substantial equivalency to predicate device | "The study results demonstrate a substantial equivalency on performance between the Scanostics UTI Check application test system and the predicate device, Clinistrip (URS-10) test strips." This is the overarching conclusion of the submission. |
Study Details
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated for specific metrics like sensitivity/specificity studies. The "lay user studies were performed at three sites," suggesting a multicenter approach, but the number of participants or urine samples is not provided in detail.
- Data Provenance: The document does not specify the country of origin of the data. It seems to be a prospective study, as it describes "lay user studies were performed," implying new data collection rather than retrospective analysis.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document implies a comparison against "a trained HCP using the Clinistrip (URS-10)" as a reference for the lay user study. However, it does not specify the number of HCPs, their qualifications, or how a definitive "ground truth" was established for each sample (e.g., if a definitive culture result was also used). Given the nature of a 510(k) submission focusing on substantial equivalence, the "ground truth" for the test set often stems from the established predicate device's performance.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not described in the provided text.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study in the traditional sense was not performed. This device is an AI-powered reader for a diagnostic strip, intended for over-the-counter home use only by lay users. The comparison was between the lay user with the Scanostics app and a "trained HCP using the Clinistrip (URS-10)" (the predicate device). The focus is on enabling accurate interpretation by lay users, not improving expert reader performance.
- Therefore, an "effect size of how much human readers improve with AI vs without AI assistance" is not applicable in the context of this study design as presented.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- The "Scanostics UTI Check Application test system" inherently involves the app (algorithm) reading the test strip. The "lay user studies" describe the performance when the human (lay user) operates the system. However, the non-clinical studies ("method comparison, precision, detection limit, interference, specificity, shelf life and stress studies as well as several flex studies") would largely assess the standalone performance of the algorithm's ability to read the strips under various controlled conditions, separate from a human's interpretation error. The document states these studies found the device performers "satisfactorily," implying the algorithm itself functions as intended.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The primary ground truth for comparison in the lay user study appears to be the results obtained by a "trained HCP using the Clinistrip (URS-10)" (the predicate device). For the non-clinical studies, the "ground truth" would likely be established through controlled laboratory testing with known concentrations for precision, detection limit, specificity, etc., but this is not explicitly detailed. It does not mention pathology or outcomes data as direct ground truth for this device's performance.
-
The sample size for the training set:
- The document does not provide information about the sample size used for the training set of the Scanostics UTI Check Application's algorithm.
-
How the ground truth for the training set was established:
- This information is not provided in the 510(k) summary. Training set ground truth would typically be established through expertly-read images of a large number of test strips, potentially correlated with laboratory-confirmed analyte concentrations, but these details are absent.
Ask a specific question about this device
(31 days)
peroxidase test | LJX | I | 21 CFR§864.7675
The iChem® VELOCITY™ automated urine chemistry system is an in vitro diagnostic device used to automate the urine chemistry analysis profile using iChem® VELOCITY ™ Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in an iQ®200 Series system, a configuration given the proprietary name iRICELL™ as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity; semi-quantitative results for glucose, blood, leukocyte esterase, bilirubin, urobilingen, pH, protein, ketones and ascorbic acid; and qualitative results for nitrites, color and clarity.
iChemVELOCITY strips are intended for use only with the iChem Velocity analyzer. In particular, they are not intended for visual reading. The iChemVELOCITY is not intended to be used as a Point of Care (POC) analyzer.
These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver function. Tests performed using the iChemVELOCITY are intended for clinical laboratory and in vitro diagnostic use only.
The iChemVELOCITY is an automated urine chemistry system performing measurements of defined physical and chemical constituents in urine. The system utilizes iChemVELOCITY urine chemistry test strips which are read in the Strip Reader Module (SRM) by measuring light reflectance, in addition the iChemVELOCITY measures color and clarity using light transmittance through Color/Clarity Module (CCM) and specific gravity using refractive index through Specific Gravity Module (SGM).
The subject of this submission is a design change to the Color/Clarity Module (CCM) of the Color/Clarity/Specific Gravity Module (CGM) subassembly of the iChemVELOCITY. The CCM measures the color and clarity of a urine sample using transmitted and scattered light, incident on a color sensor. An algorithm then converts the output of the different channels of the sensor into semi-quantitative colors. The CCM is being changed to replace end-of-life components (color sensor and scatter light source) and update the color detection algorithm.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Device: iChem®VELOCITY™ Automated Urine Chemistry System with redesigned Color/Clarity Module (CCM)
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" with numerical targets for the redesigned CCM. Instead, it aims to demonstrate substantial equivalence to the predicate device (iChemVELOCITY with the original CCM). Therefore, the reported performance is compared to the predicate's expected performance, with the implicit criterion being that the redesigned CCM performs at least as well as, or is in agreement with, the predicate.
Here's an interpretation of the performance data in the context of implied acceptance:
Feature/Metric | Acceptance Criteria (Implied for Substantial Equivalence to Predicate) | Reported Device Performance (Redesigned CCM vs. Predicate) |
---|---|---|
Color Agreement | High percentage of exact match and ±1 block agreement with predicate device's color readings. | Refer to "Method Comparison Results Summary" and "Precision Results Summary" below for detailed numbers. All ±1 block agreement is 100% or 99.6%. Exact match ranges from 36% (Straw) to 100%. |
Clarity Agreement | High percentage of exact match and ±1 block agreement with predicate device's clarity readings. | Refer to "Method Comparison Results Summary" and "Precision Results Summary" below for detailed numbers. All ±1 block agreement is 100% or 97.2%. Exact match ranges from 29.2% (Slightly Cloudy) to 100%. |
Fit | The redesigned CCM must remain unchanged in fit within the iChemVELOCITY system. | "All testing successfully passed the acceptance criteria, and the test results demonstrated equivalence to the predicate." "The testing results confirmed that the redesigned CCM was fit..." |
Function | The redesigned CCM must produce equivalent results to the predicate design. | "All testing successfully passed the acceptance criteria, and the test results demonstrated equivalence to the predicate." "The testing results confirmed that the redesigned CCM was...functionally compatible and equivalent to the original CCM." |
New Hazards | No new hazards introduced by the design change. | "no new hazards were introduced with the implementation of this change and therefore safety associated with the operation of this device remains unchanged." |
User Experience | The design change should not negatively impact the user experience. | "did not negatively impact the user experience." |
Detailed Performance Tables from the Document:
Method Comparison Results Summary (Page 8):
Redesigned CCM / Comparator | Colorless | Straw | Yellow | Amber | Red | Blue |
---|---|---|---|---|---|---|
Color | ||||||
Blue | 18 | |||||
Red | 3 | 18 | ||||
Amber | 5 | 18 | ||||
Yellow | 11 | 61 | 9 | |||
Straw | 13 | |||||
Colorless | 18 | 12 | ||||
Exact match: | 100.0% | 36% | 92% | 60.0% | 100.0% | 100.0% |
± 1 Block: | 100.0% | 100.0% | 100.0% | 100.0% | 100.0% | 100.0% |
Redesigned CCM / Comparator | Clear | Slightly Cloudy | Cloudy | Turbid |
---|---|---|---|---|
Clarity | ||||
Turbid | 32 | |||
Cloudy | 1 | 22 | 1 | |
Slightly Cloudy | 7 | 13 | ||
Clear | 87 | 16 | 1 | |
Exact match: | 100.0% | 29.2% | 61.1% | 97.0% |
± 1 Block: | 100.0% | 100.0% | 97.2% | 100.0% |
Precision Results Summary (Page 9):
Color/Clarity | Test level | Number (Total) | Exact agree | ±1 block agree | % Exact agreement | % ±1 block agreement |
---|---|---|---|---|---|---|
Color | Colorless | 240 | 240 | 240 | 100% | 100% |
Straw | 240 | 238 | 240 | 99.2% | 100% | |
Yellow | 240 | 240 | 240 | 100% | 100% | |
Amber | 240 | 240 | 240 | 100% | 100% | |
Red | 240 | 237 | 239 | 98.8% | 99.6% | |
Blue | 240 | 240 | 240 | 100% | 100% | |
Clarity | Clear | 240 | 240 | 240 | 100% | 100% |
Slightly Cloudy | 240 | 228 | 240 | 95.0% | 100% | |
Cloudy | 240 | 240 | 240 | 100% | 100% | |
Turbid | 240 | 240 | 240 | 100% | 100% |
2. Sample size used for the test set and the data provenance
- Sample Size: The "Precision Results Summary" table on page 9 indicates that for each color and clarity level category, 240 samples were tested.
- Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. It only states that "Performance testing of the iChemVELOCITY with the redesigned CCM was conducted."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not specify the number of experts used or their qualifications for establishing ground truth. The "Method Comparison" results compare the "Redesigned CCM" to a "Comparator." Given the context of a design change to an existing device, the "Comparator" likely refers to the predicate iChemVELOCITY device with the original CCM. The ground truth, in this case, would be the results generated by the predicate device, not necessarily external human experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The document does not mention any adjudication method, as the comparison is primarily machine-to-machine (redesigned CCM vs. predicate CCM).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an automated urine chemistry system, and the study focuses on the performance of the redesigned CCM against its predicate, not on human reader performance or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance study was done. The entire study focuses on the performance of the automated iChemVELOCITY system with the redesigned CCM. There is no mention of human-in-the-loop performance evaluation for the color and clarity measurements. The device is intended "for clinical laboratory and in vitro diagnostic use only" and "is not intended for visual reading."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The primary ground truth for the performance study is the predicate device's performance. The study aims to show that the redesigned CCM produces results equivalent to the iChemVELOCITY with the original CCM, which was previously cleared.
8. The sample size for the training set
The document does not specify a separate training set size. The study describes a design change to existing hardware/software components, including an updated algorithm. It's possible that the "new firmware has been written and installed on the CCM's microprocessor... to update the CCM algorithm" implies some form of development/training, but the specifics and size are not provided in this regulatory summary.
9. How the ground truth for the training set was established
Since no specific training set and its size are explicitly mentioned, the method for establishing its ground truth is also not described. If algorithm development involved training, the ground truth would likely have been established using reference methods during the design phase of the original or updated algorithm, but this is not detailed in the provided text. The document focuses on the verification that the redesigned component performs equivalently to the existing, cleared device.
Ask a specific question about this device
(269 days)
) |
| 864.7675 | LJX | Class I
The AUTION ELEVEN Semi-Automated Urinalysis System provides a qualitative and semi-quantitative measurements for glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, specific gravity and color tone in urine specimens. The system is intended for in vitro diagnostic use in screening patient populations found in clinical laboratories.
The AUTION ELEVEN Semi-Automated Urinalysis System consists of the following:
- · AUTION ELEVEN model AE-4022 Urine Analyzer (device component)
- · AUTION Sticks 10EA Test Strips (reagent component)
The AUTION ELEVEN Semi-Automated Urinalysis System provides a qualitative and semi-quantitative measurement for glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, specific gravity and color tone. The system is intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The AUTION ELEVEN Semi-Automated Urinalysis System consists of AUTION ELEVEN model AE-4022 urine analyzer and AUTION Sticks 10EA test strips.
The AUTION sticks 10EA consist of a plastic strip containing 10 pads impregnated with chemicals specific for the determination of a particular analyte. The chemical reaction with the urine results in a color change which is measured by the AUTION ELEVEN AE-4022 device, resulting in a display and print out indicating analyte concentration. The AUTION ELEVEN technology provides fast results that can be used along with other diagnostic information to rule out certain disease states and to determine if microscopic analysis is needed.
The semi-automated nature of the device requires the user to dip an AUTION Stick 10EA test strip into a patient urine specimen and place it on the instrument. The instrument processes the test strip, allowing 60 seconds for the chemical reactions to occur on the test strip reagent pads. After 60 seconds, the device measures the amount of reflectance generated from each reagent pad and converts the reflectance measurements to qualitative and semi-quantitative results for physician use.
Here's a breakdown of the acceptance criteria and study information for the AUTION ELEVEN Semi-Automated Urinalysis System, extracted from the provided text:
Acceptance Criteria and Device Performance
Precision Results
Analyte | Acceptance Criteria (Exact match%) | Reported Device Performance (Exact match%) | Acceptance Criteria (+/- 1 Color Block%) | Reported Device Performance (+/- 1 Color Block%) |
---|---|---|---|---|
Glucose | Not explicitly stated | 98% (Repeatability), 99% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Protein | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Bilirubin | Not explicitly stated | 99% (Repeatability), 99% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Urobilinogen | Not explicitly stated | 99% (Repeatability), 100% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
pH | Not explicitly stated | 98% (Repeatability), 99% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Specific Gravity | Not explicitly stated | 96% (Repeatability), 97% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Blood | Not explicitly stated | 99% (Repeatability), 100% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Ketones | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Nitrite | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Leukocytes | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) | Not explicitly stated | 100% (Repeatability), 100% (Reproducibility) |
Note: The document states that "results from bench testing met pre-determined acceptance criteria and support a determination of substantial equivalence." However, the specific numerical acceptance criteria for "Exact match %" and "+/- 1 Color Block %" were not explicitly listed in the tables provided for precision. The reported device performance values are the results obtained from the study.
Method Comparison Results (Accuracy)
Analyte | Acceptance Criteria (Exact Agreement with Predicate) | Reported Device Performance (Exact Agreement with Predicate) | Acceptance Criteria (Agreement within ± 1 color block) | Reported Device Performance (Agreement within ± 1 color block) |
---|---|---|---|---|
Glucose | Not explicitly stated | 98% | Not explicitly stated | 100% |
Protein | Not explicitly stated | 86% | Not explicitly stated | 100% |
Bilirubin | Not explicitly stated | 100% | Not explicitly stated | 100% |
Urobilinogen | Not explicitly stated | 98% | Not explicitly stated | 100% |
pH | Not explicitly stated | 89% | Not explicitly stated | 100% |
Specific Gravity | Not explicitly stated | 81% | Not explicitly stated | 99% |
Blood | Not explicitly stated | 92% | Not explicitly stated | 100% |
Ketones | Not explicitly stated | 96% | Not explicitly stated | 100% |
Nitrite | Not explicitly stated | 99% | Not explicitly stated | 100% |
Leukocyte | Not explicitly stated | 92% | Not explicitly stated | 100% |
Note: Similar to precision, the document states "The overall conclusion from the clinical evaluation is that the results are acceptable and support a determination of substantial equivalence," implying the reported accuracy rates met their internal acceptance criteria. However, the specific numerical acceptance thresholds were not provided in these tables.
Detection Limits/Sensitivity Results
The acceptance criteria here are that the "Actual Concentration" for a given semi-quantitative rank should result in the device reporting that rank (or a higher rank) at a certain "Percent in Rank." These are implicitly the acceptance criteria for sensitivity. The table shows the "Actual Concentration" and the "Percent in Rank" achieved. For instance, for Glucose, the acceptance criterion for the 4+ rank might be that at 750 mg/dL, it should be in the 4+ rank at least 50% of the time (reported 87%).
Interfering Substances
The acceptance criteria are implicitly that the identified interferents cause a specific shift in color blocks as described (e.g., False negative (-2 to -3 color block change) for Glucose with Ascorbic acid). The study identifies what interference occurs and its magnitude, implying these findings were deemed acceptable within the context of device labeling.
Study Details
-
Sample size used for the test set and the data provenance:
- Precision Study:
- Sample Size: Not a direct patient sample size. Three (3) levels of quality controls were used for each analyte.
- Data Provenance: Not specified, but likely controlled laboratory conditions given the use of commercial controls and specified dilutions/spiking.
- Method Comparison Study (Accuracy):
- Sample Size: 8395 tests of individual analytes. The number of unique patient samples is not explicitly stated, but the note mentions "Each site collected urine patient samples from their clinical laboratory or obtained them from nearby hospitals."
- Data Provenance: Clinical patient samples, collected and refrigerated within 2 hours of collection for up to 24 hours prior to testing. Country of origin is not specified, but the clinical sites suggest real-world data collection. The study is prospective in the sense that samples were collected for the purpose of this study.
- Precision Study:
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Precision Study: Not applicable. Ground truth was established by the known concentrations of the quality control materials.
- Method Comparison Study: Ground truth was established by comparison against "2 commercially available urinalysis predicates, 1 semi-automated urine analyzer was used for specific gravity comparison and one fully-automated urine analyzer was used to compare all the rest of the analytes." No human experts were explicitly mentioned for ground truth establishment for this comparison. The predicate devices served as the reference standard.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable for either the precision or method comparison studies as ground truth was established either by known control concentrations or by predicate device performance, not by human expert consensus requiring adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is an automated and semi-automated urinalysis system, and the studies focused on its analytical performance against predicate devices and known controls, not on human reader performance or AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the performance characteristics studies (Precision, Method Comparison, Detection Limits, Interfering Substances) represent standalone (algorithm only) performance of the AUTION ELEVEN Semi-Automated Urinalysis System. While a human dips the strip, the measurement and interpretation of the color change are performed by the automated device's optical unit and algorithms. The system then provides qualitative and semi-quantitative results automatically.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Precision Study: Known concentrations of quality control materials.
- Method Comparison Study: Comparison against established predicate urinalysis devices for each analyte.
-
The sample size for the training set:
- No information about a "training set" is provided in the document. This type of device (urinalysis analyzer) is typically developed and validated using analytical samples (spiked, diluted, known concentrations) and clinical samples, rather than a machine learning "training set" in the conventional sense. The "Performance Characteristics" section details the validation of the device's measurement accuracy and precision.
-
How the ground truth for the training set was established:
- Not applicable, as no training set was explicitly mentioned or described for this type of device validation.
Ask a specific question about this device
Page 1 of 3