K Number
K181024

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-05-23

(35 days)

Product Code
Regulation Number
862.1340
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended for the in vitro measurement of the following in urine: Leukocyte, Nitrite, Urobilinogen, Protein, pH, Blood, Specific gravity, Ketone, Bilirubin, Glucose. These strips are intended for prescription, in vitro diagnostic use only and they are visually read.

DUS 2GP reagent strips provide qualitative and semiquantitative measurements for protein, and glucose in urine specimens. Test results may provide information regarding the status of carbohydrate metabolism and kidney function.

DUS 5 reagent strips provide qualitative and semiquantitative measurements for leukocytes, nitrite, blood , protein, and glucose in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies and urinary tract infections.

DUS 10 reagent strips provide qualitative and semiquantitative measurements for specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin and blood in urine speciments are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections and liver function.

Device Description

The DUS Series are urine test strips with different reagent pads for the determination of specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin and blood affixed onto plastic strips. of which leukocyte, nitrite, urobilinogen, protein, pH, blood, specific gravity, ketone, bilirubin and glucose reagent pads are affixed onto the plastic strips. The reagent pads react with analytes in the urine giving a visible color. Results are confirmed by comparison of the test strip with the color chart on the container. For each color result for each analyte, a semiquantitative value is available on the box label (e.g. bilirubin results include negative, 1, 2, and 4 mg/dL) and the associated qualitative result (e.g. bilirubin results include negative. +. ++, +++).

AI/ML Overview

Here's an analysis of the provided document, outlining the acceptance criteria and study details for the DUS Reagent Strips:

Acceptance Criteria and Device Performance

The provided document does not explicitly state pre-defined acceptance criteria for the "Exact Agreement" or "Agreement within +/- one block (%)" in the method comparison study. However, the study results, which consistently show high percentages (generally in the high 90s and 100%) for both metrics across all analytes, implicitly represent the device meeting an expected high level of agreement with the predicate device.

For the linearity/assay reportable range, the acceptance criteria are implied by the reported "% Exact match." A high percentage (generally 97.7% to 100%) indicates acceptance.

For the detection limit study, the acceptance criteria are stated as "Sensitivity was defined as the cutoff for which ≥95% of the contrived pooled measurements were trace or the first positive result."

The interference study implicitly accepts the device if interference is either not observed at clinically relevant concentrations or if observed interferences are adequately addressed in the labeling.

Here's a table summarizing the reported device performance, where the values themselves act as the demonstration that the implied or stated acceptance criteria (high agreement, 95% detection, or non-interference/labeling for interference) were met.

Table of Reported Device Performance

AnalyteTest TypeReported Device Performance (Exact agreement / % Exact match / Positive Agreement / Notes)
Precision (Within-run, N=90)
UrobilinogenLevel 1 (Normal)Exact: 100%, +/-1 block: 100%
Level 2 (4mg/dL)Exact: 100%, +/-1 block: 100%
GlucoseLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (1000mg/dL)Exact: 100%, +/-1 block: 100%
BilirubinLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (4mg/dL)Exact: 100%, +/-1 block: 100%
KetonesLevel 1 (Positive)Exact: 100%, +/-1 block: 100%
Level 2 (Positive)Exact: 100%, +/-1 block: 100%
SGLevel 1 (1.020)Exact: 100%, +/-1 block: 100%
Level 2 (1.020)Exact: 98.9%, +/-1 block: 100%
BloodLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (200 RBC/uL)Exact: 100%, +/-1 block: 100%
pHLevel 1 (6)Exact: 100%, +/-1 block: 100%
Level 2 (7)Exact: 100%, +/-1 block: 100%
ProteinLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (100 mg/dL)Exact: 100%, +/-1 block: 100%
NitriteLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (Pos)Exact: 100%, +/-1 block: 100%
LeukocytesLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (70 WBC/uL)Exact: 100%, +/-1 block: 100%
Precision (Within-day, N=90)
UrobilinogenLevel 1 (Normal)Exact: 100%, +/-1 block: 100%
Level 2 (4mg/dL)Exact: 100%, +/-1 block: 100%
GlucoseLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (1000mg/dL)Exact: 100%, +/-1 block: 100%
BilirubinLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (4mg/dL)Exact: 100%, +/-1 block: 100%
KetonesLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (40mg/dL)Exact: 100%, +/-1 block: 100%
SGLevel 1 (1.020)Exact: 100%, +/-1 block: 100%
Level 2 (1.020)Exact: 100%, +/-1 block: 100%
BloodLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (200 RBC/uL)Exact: 100%, +/-1 block: 100%
pHLevel 1 (6)Exact: 100%, +/-1 block: 100%
Level 2 (7)Exact: 98.9%, +/-1 block: 100%
ProteinLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (100 mg/dL)Exact: 100%, +/-1 block: 100%
NitriteLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (Pos)Exact: 100%, +/-1 block: 100%
LeukocytesLevel 1 (Negative)Exact: 100%, +/-1 block: 100%
Level 2 (70 WBC/uL)Exact: 100%, +/-1 block: 100%
Linearity / Assay Reportable Range (N=90 per concentration)
UrobilinogenVarious Conc.97.7% - 100% Exact match
GlucoseVarious Conc.97.7% - 100% Exact match
BilirubinVarious Conc.96.6% - 100% Exact match
KetonesVarious Conc.97.7% - 100% Exact match
Specific GravityVarious Conc.94.4% - 100% Exact match
NitriteVarious Conc.100% Exact match
BloodVarious Conc.98.8% - 100% Exact match
pHVarious Conc.97.7% - 100% Exact match
ProteinVarious Conc.96.6% - 100% Exact match
LeukocytesVarious Conc.97.7% - 100% Exact match
Detection Limit (N=90 per concentration)
Urobilinogen2 mg/dL100% Positive Agreement
Glucose100 mg/dL100% Positive Agreement
Bilirubin1.0 mg/dL100% Positive Agreement
Ketones5 mg/dL100% Positive Agreement
Blood10 RBC/μL100% Positive Agreement
Protein15 mg/dL100% Positive Agreement
Nitrite0.05 mg/dL100% Positive Agreement
Leukocytes15 WBC/μL100% Positive Agreement
Analytical Specificity
(Various Analytes)(Various Interferents)No interference observed for a long list of compounds at specified concentrations. Observed interferences were noted and addressed in labeling limitations.
Method Comparison Study (Total N=867 samples; comparisons to Multistix 10SG)
UrobilinogenAll levelsExact: 92-100%, Within 1 block: 100%
GlucoseAll levelsExact: 92-100%, Within 1 block: 100%
BilirubinAll levelsExact: 92-100%, Within 1 block: 100%
KetonesAll levelsExact: 91-100%, Within 1 block: 100%
BloodAll levelsExact: 92-100%, Within 1 block: 100%
ProteinAll levelsExact: 91-100%, Within 1 block: 100%
NitriteAll levelsExact: 97-100%, Within 1 block: 100%
LeukocytesAll levelsExact: 93-100%, Within 1 block: 100%
pHAll levelsExact: 91-100%, Within 1 block: 100%
Specific GravityAll levelsExact: 91-100%, Within 1 block: 100%

Study Details

  1. Sample size used for the test set and the data provenance:

    • Method Comparison Study: Total of 867 samples.
      • Provenance: Fresh urine samples obtained at three medical facilities. The country of origin is not explicitly stated but implied to be South Korea, given the submitter's address. The data is prospective as samples were "processed within 4 hours."
    • Precision/Reproducibility: Two levels of commercially available urine-based control solutions.
      • Sample Size: 90 replicates for within-run (10 tests from 3 lots at 3 sites) and 90 replicates for within-day (1 test a day from 3 lots, at 3 sites for 10 days) for each level.
      • Provenance: Commercially available control solutions.
    • Linearity/Assay Reportable Range:
      • Sample Size: 90 replicates per concentration level (10 replicates with each of 3 lots of test strips).
      • Provenance: Samples created by spiking known concentrations of standard materials or by serial dilution of a high concentration urine sample with negative urine.
    • Detection Limit:
      • Sample Size: 90 replicates for each concentration (each sample concentration analyzed 30 times using 3 reagent strip lots).
      • Provenance: Negative urine spiked with standard materials.
    • Analytical Specificity:
      • Sample Size: 3 replicates using 3 lots of DUS 10 test strips for each concentration level of interfering substance.
      • Provenance: Urine sample pools prepared at 3 analyte concentrations (negative, low, high positive) spiked with potential interfering substances.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Precision, Linearity, Detection Limit, Analytical Specificity: Ground truth was established by the known concentrations of control solutions or spiked samples. Testing was performed by 3 medical technicians as reported in the linearity study and precision study, but their specific qualifications (e.g., years of experience) are not stated beyond being "medical technicians."
    • Method Comparison Study: The ground truth for the method comparison study was established by comparing the DUS 10 test strips results against a predicate device (Multistix 10SG). The testing was performed by three medical technicians at each of the clinical sites. Similar to above, their specific qualifications are not detailed beyond "medical technicians."
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • The document does not describe an explicit adjudication method (like 2+1 or 3+1) involving multiple human readers or experts resolving discrepancies for the test set in the traditional sense of image-based AI studies.
    • The method comparison study compares the new device's readings to the predicate device's readings, and the percentage agreement is calculated. The process for resolving discrepancies between the new device and the predicate device is not detailed, nor is there a mention of an expert panel reviewing cases.
    • For other studies (precision, linearity, detection limit), the ground truth is analytically determined by control concentrations.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader, multi-case (MRMC) comparative effectiveness study as typically understood for AI-assisted diagnostic tools (i.e., human readers with and without AI assistance) was not explicitly performed or described in this document.
    • The studies focus on the performance characteristics of the device itself (the reagent strips) and its comparison to a predicate device, which is also a reagent strip for visual reading. The device's use is "visually read" by operators, but the study design is not one of AI assistance to human readers.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This device is a reagent strip that is visually read. It is not an AI algorithm. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" does not directly apply in the context of an AI device. The performance data presented (precision, linearity, detection limits, analytical specificity, method comparison) are essentially standalone performance characteristics of the physical reagent strip, with human visual interpretation being the intended mode of operation for reading the results. The method comparison specifically assesses this human-read performance against a predicate device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Method Comparison Study: The ground truth for comparative analysis was the results obtained from the predicate device (Multistix 10SG).
    • Precision, Linearity, Detection Limit, Analytical Specificity: The ground truth was analytically determined concentrations derived from commercial control solutions or spiked urine samples with known analyte levels.
  7. The sample size for the training set:

    • This document is for a traditional in-vitro diagnostic (IVD) reagent strip that is visually read, not an AI/ML device. Therefore, there is no "training set" in the context of machine learning model development. The assays are based on chemical reactions, and the performance is inherent to the chemical formulation and strip manufacturing.
  8. How the ground truth for the training set was established:

    • As there is no AI/ML component or "training set" in the context of an algorithm, this question is not applicable. The chemical reactions on the strips are designed to react to specific analytes, and their performance is validated through the studies described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 23, 2018

DFI Co., Ltd. % Ho Dong Yang, CEO Onbix Corporation #821 Samil Plaza, 14, Dogok-ro 1-Gil, Gangnam-gu Seoul, 06523 Korea

Re: K181024

Trade/Device Name: DUS 2GP Reagent Strips, DUS 5 Reagent Strips, DUS 10 Reagent Strips Regulation Number: 21 CFR 862.1340 Regulation Name: Urinary glucose (nonguantitative) test system Regulatory Class: Class II Product Code: JIL, JIO, JIR, LJX, JMT, CDM, CEN, JRE, JIN, JJB Dated: April 11, 2018 Received: April 18, 2018

Dear Ho Dong Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181024

Device Name

DUS 2GP Reagent Strips, DUS 5 Reagent Strips, DUS 10 Reagent Strips

Indications for Use (Describe)

DUS 2GP Reagent Strips, DUS 5 Reagent Strips, DUS 10 Reagent Strips

This device is intended for the in vitro measurement of the following in urine: Leukocyte, Nitrite, Urobilinogen, Protein, pH, Blood, Specific gravity, Ketone, Bilirubin, Glucose. These strips are intended for prescription, in vitro diagnostic use only and they are visually read.

DUS 2GP reagent strips provide qualitative and semiquantitative measurements for protein, and glucose in urine specimens. Test results may provide information regarding the status of carbohydrate metabolism and kidney function.

DUS 5 reagent strips provide qualitative and semiquantitative measurements for leukocytes, nitrite, blood , protein, and glucose in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies and urinary tract infections.

DUS 10 reagent strips provide qualitative and semiquantitative measurements for specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin and blood in urine speciments are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections and liver function.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY [K181024]

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR §807.92.

Submitter Information:DFI Co., Ltd.388-25, Gomo-ro, Jillye-myeon, Gimhae-si,Gyeongsangnam-do, 621-881, KoreaTel. +82-55-346-1882 Fax: +82-55-346-1883
Contact Person:Ho Dong, YangOnbix Corporation#821 Samil Plaza, 14, Dogokro 1-gil Gangnam-guSeoul, Korea (06253)Tel: *82-2-566-3360 / Fax: *82-2-6280-3360Email: onbix@naver.com

Date Summary Prepared: May 17, 2018

Device Information:
Trade Name(s):DUS 2GP Reagent Strips, DUS 5 Reagent Strips, DUS 10 Reagent Strips
Common Name:urinary test system
Panel:clinical chemistry

Regulatory information

NameRegulationProductcodeDeviceclass
Urinary Glucose(non-quantitative) test system21 CFR 862.1340JILII
Occult Blood test21 CFR 864.6550JIOII
Urinary Protein or Albumin(nonquantitative) test system21 CFR 862.1645JIRI
Leukocyte peroxidase test21 CFR 864.7675LJXI
Nitrite (nonquantitative) test system21 CFR 862.1510JMTI
Urinary Urobilinogen(nonquantitative) test system21 CFR 862.1785CDMI
Urinary pH (nonquantitative) test21 CFR 862.1550CENI
Specific Gravity21CFR 862.2800JREI
Ketones (nonquantitative) testsystem21 CFR 862.1435JINI
Urinary Bilirubin and its conjugates(nonquantitative) test system21 CFR 862.1115JJBI

{4}------------------------------------------------

Predicate Device Information:

K992257 / MULTISTIX 10 SG

Device Description:

DUS 2GP Reagent Strips, DUS 5 Reagent Strips, DUS 10 Reagent Strips

The DUS Series are urine test strips with different reagent pads for the determination of specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin and blood affixed onto plastic strips. of which leukocyte, nitrite, urobilinogen, protein, pH, blood, specific gravity, ketone, bilirubin and glucose reagent pads are affixed onto the plastic strips. The reagent pads react with analytes in the urine giving a visible color. Results are confirmed by comparison of the test strip with the color chart on the container. For each color result for each analyte, a semiquantitative value is available on the box label (e.g. bilirubin results include negative, 1, 2, and 4 mg/dL) and the associated qualitative result (e.g. bilirubin results include negative. +. ++, +++).

Indications for Use:

This device is intended for the in vitro measurement of the following in urine: Leukocyte, Nitrite, Urobilinogen, Protein, pH, Blood, Specific gravity, Ketone, Bilirubin, Glucose. These strips are intended for prescription, in vitro diagnostic use only and they are visually read.

DUS 2GP reagent strips provide qualitative and semiquantitative measurements for protein and glucose in urine specimens. Test results may provide information regarding the status of carbohydrate metabolism and kidney function.

DUS 5 reagent strips provide qualitative and semiquantitative measurements for leukocytes, nitrite, blood, protein, and glucose in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders and kidney function anomalies and urinary tract infections.

DUS 10 reagent strips provide qualitative and semiquantitative measurements for specific gravity, pH. leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin a n d blood in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections and liver function.

Comparison to Predicate Device(s):

This device is equivalent to the predicate devices in its intended use and technological characteristics, including:

indications for use technological characteristics performance properties

Summary of testing:

The performance characteristics of the candidate strips were verified by method comparison, precision, detection limit, interference and specificity studies.

Testing is summarized below.

{5}------------------------------------------------

1. Performance Characteristics

DUS 10 reagent strips were used for all analytical studies, because DUS 2GP and DUS 5 reagent strips contain subsets of the analytes in the DUS 10 reagent strips.

a) Precision/Reproducibility

Precision testing was done in accordance with CLSI EP5-A3.. The repeatability/ reproducibility were determined at three clinical sites, using commercially available control solutions. Studies were performed by medical technicians, using 2 levels of commercially available urine based control solutions. For within-run precision studies, 10 test strips from 3 different lots at 3 sites were tested (30 tests at each site = 90 replicates per level). For within-day precision studies, 1 test strip a day from 3 different lots was tested at 3 sites for 10 days (30 tests at each site = 90 replicates per level). The results are summarized below:

Level 1 control
Within-run (n=90)Within-day (n=90)
ItemTest ResultsExactagreement(%)Agreementwithin+/-oneblock(%)Exactagreement(%)Agreementwithin+/-oneblock(%)
UrobilinogenNormal100100100100
GlucoseNegative100100100100
BilirubinNegative100100100100
KetonesNegative100100100100
SG1.020100100100100
BloodNegative100100100100
pH6100100100100
ProteinNegative100100100100
NitriteNegative100100100100
LeukocytesNegative100100100100
Level 2 control
Within-run (n=90)Within-day (n=90)
ItemTest ResultsExactagreement(%)Agreementwithin+/-oneblock(%)Exactagreement(%)Agreementwithin+/-oneblock(%)
Urobilinogen4mg/dL100100100100
Glucose1000mg/dL100100100100
Bilirubin4mg/dL100100100100
Ketones40mg/dL100100100100
SG1.02098.9100100100
Blood200 RBC/uL100100100100
pH710010098.9100
Protein100 mg/dL100100100100
NitritePos100100100100
Leukocytes70 WBC/uL100100100100

{6}------------------------------------------------

b) Linearity /assay reportable range.

Linearity / assay reportable range was evaluated in-house. Samples were created by spiking known concentration of each standard material or by serial dilution of a high concentration urine sample with negative urine. Testing was performed by 3 medical technicians using 3 lots of test strips. Each concentration level was tested in 10 replicates with each lot of test strips. The results are summarized below:

AnalyteUrine sampleconcentrationColor blockoutput% Exact match
UrobilinogenNegative/Norm0.1100 % (90/90)
1 mg/dL1 mg/dL100 % (90/90)
2 mg/dL2 mg/dL97.7 % (88/90)
4 mg/dL4 mg/dL98.8 % (89/90)
8 mg/dL8 mg/dL98.8 % (89/90)
GlucoseNegativeNeg.100 % (90/90)
100 mg/dL±(100 mg/dL)100 % (90/90)
250 mg/dL+(250 mg/dL)97.7 % (88/90)
500 mg/dL++(500 mg/dL)98.8 % (89/90)
1000 mg/dL+++(1000 mg/dL)97.7 % (88/90)
2000 mg/dL++++(2000mg/dL)100 % (90/90)
BilirubinNegativeNeg.100 % (90/90)
1 mg/dL+98.8 % (89/90)
2 mg/dL++96.6 % (87/90)
4 mg/dL+++100 % (90/90)
KetonesNegativeNeg.100 % (90/90)
5 mg/dL±(5 mg/dL)100 % (90/90)
15 mg/dL+(15 mg/dL)97.7 % (88/90)
40 mg/dL++(40 mg/dL)98.8 % (89/90)
80 mg/dL+++(80 mg/dL)98.8 % (89/90)
160 mg/dL++++(160 mg/dL)100 % (90/90)
AnalyteUrine sampleconcentrationColor blockoutput% Exact match
Specific gravity1.0001.000100 % (90/90)
1.0051.00594.4 % (85/90)
1.0101.010100 % (90/90)
1.0151.01596.6 % (87/90)
1.0201.02097.7 % (88/90)
1.0251.02596.6 % (87/90)
1.0301.030100 % (90/90)
NitriteNegativeNeg.100% (90/90)
0.05 mg/dLTrace100 % (90/90)
10 mg/dLPos100 % (90/90)
BloodNegativeNeg.100 % (90/90)
10 RBC/μLTrace100 % (90/90)
25 RBC/μL+(25 RBC/μL)98.8 % (89/90)
80 RBC/μL++(80 RBC/μL)100 % (90/90)
200 RBC/μL+++(200RBC/μL)100 % (90/90)
pH55100 % (90/90)
6698.8 % (89/90)
6.56.598.8 % (89/90)
7797.7 % (88/90)
7.57.597.7 % (88/90)
8898.8 % (89/90)
8.58.5100% (90/90)
ProteinNegativeNeg.100 % (90/90)
15 mg/dLTrace98.8 % (89/90)
30 mg/dL+(30 mg/dL)98.8 % (89/90)
100 mg/dL++(100 mg/dL)96.6 % (87/90)
300 mg/dL+++(300 mg/dL)97.7 % (88/90)
100 mg/dL+++++(1000 mg/dL)100 % (90/90)
LeukocytesNegativeNeg.100 % (90/90)
15 WBC/μLTrace100 % (90/90)
70 WBC/μL+(70 WBC/μL)97.7 % (88/90)
125 WBC/μL++(125 WBC/μL)97.7 % (88/90)
500 WBC/μL+++(500 WBC/μL)100 % (90/90)

{7}------------------------------------------------

{8}------------------------------------------------

c) Detection limit

A sensitivity study was performed to evaluate the limits of detection for each of the analytes. Negative urine was spiked with standard materials to obtain 3 levels across the measuring range for each analyte concentration. 90 replicates were obtained for each concentration (each sample concentration was analyzed 30 times using 3 reagent strip lots). Sensitivity was defined as the cutoff for which ≥95% of the contrived pooled measurements were trace or the first positive result.

Urobilinogen
Conc.NegativePositive(+ 2mg/dL)Positive Agreement(%)
1.8 mg/dL88291.1%
2 mg/dL090100%
2.3 mg/dL090100%
Result cut off2 mg/dL
Glucose
Conc.NegativePositivePositive Agreement
(± 100mg/dL)(%)
80 mg/dL58594.4%
100 mg/dL090100%
120 mg/dL090100%
Result cut off100mg/dL
Bilirubin
Conc.NegativePositive(+ 1.0mg/dL)Positive Agreement(%)
0.8 mg/dL68493.3%
1.0 mg/dL090100%
1.2 mg/dL090100%
Result cut off1.0 mg/dL

{9}------------------------------------------------

Ketones
Conc.NegativePositivePositive Agreement
(± 5mg/dL)(%)
4 mg/dL58594.4%
5 mg/dL090100%
6 mg/dL090100%
Result cut off5 mg/dL
Blood
PositivePositive Agreement
Conc.Negative(±5 RBC/μL)(%)
8 RBC/μL78392.2%
10 RBC/μL090100%
12 RBC/μL090100%
Result cut off10 RBC/μL
Protein
NegativePositive(±15 mg/dL)Positive Agreement(%)
Conc.
12mg/dL68493.3%
15mg/dL090100%
18mg/dL090100%
Result cut off15mg/dL
Nitrite
PositivePositive Agreement
Conc.Negative(0.05 mg/dL)(%)
0.04mg/dL68493.3%
0.05mg/dL090100%
0.06mg/dL090100%
Result cut off0.05mg/dL

{10}------------------------------------------------

Leukocytes
Conc.NegativePositive(Trace15WBC/μL)Positive Agreement(%)
12 WBC/μL58594.4%
15 WBC/μL090100%
18WBC/μL090100%
Result cut off15 WBC/μL

The reportable ranges for each analyte are summarized below:

TestReportable Range
UrobilinogenQualitative: normal to 3+
Semi-quantitative: 0.1 to 8 mg/dL
GlucoseQualitative: Negative to 4+
Semi-quantitative : Negative to 2000 mg/dL
BilirubinQualitative: Negative to 3+
Semi-quantitative : Negative to 4 mg/dL
KetonesQualitative: Negative to 4+
Semi-quantitative : Negative to 160 mg/dL
Specific GravitySemi-quantitative : 1.000 to 1.030
BloodQualitative: Negative to 3+
Semi-quantitative : Negative to 200 RBC/ μL
pHSemi-quantitative : 5 to 8.5
ProteinQualitative: Negative to 4+
Semi-quantitative : Negative to 1000 mg/dL
NitriteQualitative: Negative to Pos
Semi-quantitative : Negative to 10 mg/dL
LeukocytesQualitative: Negative to 3+
Semi-quantitative : Negative to 500 WBC/μL

{11}------------------------------------------------

d) Analytical specificity

Potential interferents and drugs were evaluated to assess their interfering effect on the performance of DUS 10 test strips. Urine sample pools prepared at 3 analyte concentrations (negative, low positive and high positive) were spiked with potential interfering substances at various interference concentrations and analyzed in 3 replicates using 3 lots of DUS 10 test strips. Interference was defined as change in output of ±1 color block between the spiked and unspiked control sample. No interference was observed for the following compounds at the concentrations evaluated below:

Potential Interfering SubstanceHighest concentration at whichno interference was observed
Albumin (protein)750 mg/dL
Ascorbic acid15 mg/dL
Acetaminophen25 mg/dL
Azo gantrinsin (Sulfamethoxazol)140 mg/dL
Bilirubin3 mg/dL
Captropril100 mg/dL
Sodium hypochlorite0.1%
Creatinine800 mg/dL
Nitrofurantoin10 mg/dL
Oxalic acid30 mg/dL
p-amino Salicylic acid375 mg/dL
β-D-Glucose (Glucose)1000 mg/dL
Phenazopyridine25 mg/dL
Riboflavin12 mg/dL
Selenium150 mg/dL
Forrmaldehyde100 mg/dL
Hemoglobin3 mg/dL
Lithium acetoacetate (ketones)20 mg/dL
Tetracycline20 mg/dL
Urobilinogen10 mg/dL
pH7
Specific gravity1.030

The following table shows the substances that did interfere with one more of the analytes tested with the DUS 10 test strips. Results are expressed as the lowest concentration of interfering substances that exhibit interference and the resulting change in output of color block:

{12}------------------------------------------------

AnalytePotential InterferingSubstanceConcentration at whichInterference wasobservedChange incolor blockoutput
UrobilinogenAzo gantrinsin(Sulfamethoxazol)≥105mg/dL+1
Bilirubin≥4mg/dL+1
Formaldehyde≥150mg/dL-1
Nitrofurantoin≥40mg/dL+1
p-Amino salicylic acid≥750mg/dL+1
Phenazopyridine≥37mg/dL+1
Riboflavin≥25mg/dL+1
Selenium≥220mg/dL+1
GlucoseAscorbic Acid≥40mg/dL-1
Bilirubin≥6mg/dL+1
Chlorine Bleach≥1%+1
Formaldehyde≥150mg/dL+1
Lithium≥40mg/dL-1
NaCl (Specific gravity)≤1.000+1
BilirubinAscorbic Acid≥40mg/dL-1
p-Amino salicylic acid≥1500mg/dL+1
Phenazopyridine≥37mg/dL+1
Selenium≥220mg/dL+1
KetonesBilirubin≥4mg/dL+1
Captopril≥150mg/dL+1
NaCl (Specific gravity)≥1.040+1
BloodAlbumin≥1000mg/dL-1
Ascorbic Acid≥30mg/dL-1
Bilirubin≥6mg/dL+1
Captopril≥150mg/dL-1
Chlorine Bleach≥0.5%+1
Formaldehyde≥150mg/dL+1
NaCl (Specific gravity)≥1.040-1
Nitrofurantoin≥30mg/dL+1
Riboflavin≥50mg/dL+1
Urobilinogen≥20mg/dL-1
pH---
Specific GravitypH>8+1
ProteinAcetaminophen50mg/dL+1
Bilirubin≥6mg/dL+1
Hemoglobin≥5mg/dL+1
Nitrofurantoin≥20mg/dL+1
Riboflavin≥25mg/dL-1
Urobilinogen≥20mg/dL-1
pH≥9+1
NitriteAscorbic Acid≥40mg/dL-1
Bilirubin≥4mg/dL+1
Nitrofurantoin≥20mg/dL+1
Phenazopyridine≥37mg/dL+1
Riboflavin≥50mg/dL+1
Selenium≥300mg/dL+1
Urobilinogen≥15mg/dL+1
LeukocytesAlbumin≥1000mg/dL-1
Bilirubin≥4mg/dL+1
Captopril≥150mg/dL-1
Chlorine Bleach≥0.5%+1
Formaldehyde≥150mg/dL+1
NaCl (Specific gravity)≥1.040-1
Nitrofurantoin≥30mg/dL+1
Oxalic Acid≥45mg/dL-1
Glucose≥1500mg/dL-1
Phenazopyridine≥50mg/dL+1
Riboflavin≥37mg/dL+1
Selenium≥300mg/dL+1
Tetracycline≥40mg/dL-1
Urobilinogen>15mg/dL+1

{13}------------------------------------------------

{14}------------------------------------------------

The labeling includes the following limitations to address the interferences observed:

As with all laboratory tests, definitive diagnostic or therapeutic decisions should not be based on any single result of method. Substances that cause abnormal urine color may affect the readability of test pads in urinalysis reagent strips.

Urinary ascorbic acid concentrations as low as 40mg/dl can cause interference in specimens with low concentrations of glucose, blood, nitrite and bilirubin

Urobilinogen: The absence of urobilinogen in the specimen cannot be determined. The test area will react with interfering substances known to react with Ehrlich's reagent, such as paminosalicylic acid. Drugs containing azo gantrisin (sulfamethoxazol) or high bilirubin may give a masking golden color. Preservative formaldehyde may cause false negatives. The test is not reliable method for the detection of porphobilinogen.

Glucose: Ascorbic acid (>40 mg/dL) may cause false negative results at low levels (100 mg/dL) of glucose. Ketones reduce the sensitivity of the test. Moderately high ketone levels (≥40 mg/dL) may cause false negatives for specimens containing a small amount of glucose (100 mg/dL). Chlorine Bleach (≥1%), low SG (≤1.000) and formaldehyde (≥150 mg/dL) may cause false positive results.

Bilirubin: Metabolites of drugs, such as selenium(≥220 mg/dL) or phenazopyridine(≥37 mg/dL) may cause false positives. p-Amino salicylic acid(≥1500 mg/dL) can produce a yellow-orange to red color response, which may lead to false positive bilirubin readings. Ascorbic acid (≥40 mg/dL) may cause false negatives.

Ketones: Low level false positive reactions may be seen in highly concentrated urine specimens (SG>1.004) or in specimens containing large amounts of levodopa drug metabolites such as captopril. The uroprotective drug mesna (sodium 2-mercaptoethane sulfonate) and other free-sulfhydryl compounds produce false-positive results in ketone methods that are based on the Legal reaction (alkaline sodium nitroprusside).

pH: If excessive urine remains on the strip because of improper test procedure, it is possible that the acidic buffer in the protein pad comes out and affects the neighboring pH pad, decreasing the measured pH. This phenomenon is called "run-over effect."

Blood: Elevated specific gravity (≥1.040) or protein in urine (≥1000 mg/dL) may reduce the sensitivity of the blood test pad leading to lower results. Microbial peroxidase associated with urinary tract infections may cause false positive results. Ascorbic acid concentrations (>30 mg/dl) may cause false negatives at trace levels of blood. Substances that cause abnormal urine color, such as drugs containing azo dyes, nitrofurantoin and riboflavin may cause false positive result. Strong oxidizing cleaning agents such as chlorine bleach cause false positive results.

Specific Gravity (SG): Alkaline urine (pH≥8) may cause false positive results.

Protein: False positive results may be found in strongly basic urine (pH>9). Protein testing should not be performed using turbid urine specimens. Metabolites of drugs such as acetaminophen may cause false positives. Hemoglobin may also cause false positives.

{15}------------------------------------------------

Pigments such as bilirubin and azo-containing compounds cause false positive results.

Nitrite: Ascorbic acid (≥40mg/dL) may cause false negative results with urine containing low levels of nitrite (<0.05mg/dL). The negative result does not always mean that the patient is free from bacteriuria. Pink spots or pink edges should not be interpreted as positive results. Medications such as phenazopyridine or other azo-containing compounds or other dyes cause false positive results.

Leukocvte: The test results may not always be consistent with the leukocyte cell number by the microscopic examination. Strong oxidizing agents such as chlorine bleach (≥0.5%) and formaldehyde (≥150 mg/dL) may cause false positive results. Falsely low results may be seen with elevated glucose (≥1500 mg/dL), high SG (≥1.040), oxalic acid (≥45 mg/dL) and high level of albumin (≥1000 mg/dL). Drugs causing decreased or false negative results include tetracycline, captopril. Substance that color urine, such as a nitrofurantoin. riboflavin, selenium and bilirubin, may cause false positive results.

2. Method Comparison study

The method comparison study was conducted at three clinical sites with a total of867 samples. Three medical technicians performed the testing at each site. Fresh urine samples were obtained at each medical facility and the samples were processed within 4 hours. The results from the DUS 10 test strips were compared to the predicate device (Multistix 10SG). The results of the method comparison study for the combines sites are shown in the tables below:

URO (Total)Predicate device (mg/mL)
New device(mg/mL)Norm1248
8539
44832
241122
1801
Norm535
Total535841179041
Exact agreement100%95%96%92%95%
Within 1 block100%100%100%100%100%

Results for urinary urobilinogen (URO):

{16}------------------------------------------------

Results for urinary glucose (GLU):

GLU (Total)Predicate device (mg/mL)
New device(mg/mL)Neg10025050010002000
2000123
10002381
5002582
2502703
100701
Neg594
Total5947273634124
Exact agreement100%97%96%92%93%96%
Within 1 block100%100%100%100%100%100%

Results for urinary Bilirubin (BIL):

BIL (Total)Predicate device (mg/mL)
New deviceNeg12
(mg/mL)
344
2833
182
Neg644
Total644869047
Exact agreement100%તે તે જ જીરી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્ય92%94%
Within 1 block100%100%100%100%

Results for urinary Ketones (KET):

KET (Total)Predicate device (mg/mL)
New device(mg/mL)Neg5154080160
1609
80437
40592
15285
5588
Neg602
Total602609363399
Exact agreement100%97%91%94%95%100%
Within 1 block100%100%100%100%100%100%

{17}------------------------------------------------

Results for urinary Blood (BLD):

BLD (Total)Predicate device (mg/mL)
New device (mg/mL)Neg102580200
2008153
48081467
22591415
101575
Neg228
Total228166154159160
Exact agreement100%95%92%92%96%
Within 1 block100%100%100%100%100%

Results for urinary Protein (PRO):

PRO (Total)Predicate device (mg/mL)
New device (mg/mL)Neg15301003001000
1000216
3005491
10061011
3021253
151252
Neg419
Total4191371331095217
Exact agreement100%91%94%93%94%94%
Within 1 block100%100%100%100%100%100%

Results for urinary Nitrite (NIT):

NIT (Total)Predicate device (mg/mL)
New device (mg/mL)NegTracePos
Pos4309
Trace1216
Neg427
Total427125315
Exact agreement100%97%98%
Within 1 block100%100%100%

{18}------------------------------------------------

LEU (Total)Predicate device (mg/mL)
New device(mg/mL)Neg1570125500
5004106
12551318
7061196
151003
Neg379
Total379106127141114
Exact agreement100%94%94%93%93%
Within 1 block100%100%100%100%100%

Results for urinary Leukocytes (LEU):

Results for urinary pH (pH):

pH (Total)Predicate device (mg/mL)
New device(mg/mL)566.577.588.5
8.5539
851023
7.591064
761485
6.581763
6314010
5896
Total9215419216011611142
Exactagreement97%91%92%93%91%92%93%
Within 1 block100%100%100%100%100%100%100%

{19}------------------------------------------------

SG (Total)Predicate device (mg/mL)
New device(mg/mL)1.0001.0051.0101.0151.0201.0251.030
1.030677
1.02561055
1.02091663
1.01571617
1.01091447
1.00511005
1.000490
Total5010915617717911482
Exact agreement98%92%92%91%93%92%94%
Within 1 block100%100%100%100%100%100%100%

Results for urinary Specific Gravity (SG):

Expected values/Reference range:

Urobilinogen: The normal urobilinogen range is 0.1 to 1.0 Ehrlich unit /dl. If results exceed the concentration of 2.0 mg/d1, the patient and the urine specimen should be evaluated further.

Glucose: The kidney normally excretes small amounts of glucose. Concentrations of 100mg/dl may be considered as abnormal if found consistently.

Bilirubin: Normally no bilirubin is detectable in urine by even the most sensitive methods. Even trace amounts of bilirubin are sufficiently abnormal to require further investigation.

Ketones: Ketone bodies should not be detected in normal urine specimens with this reagent.

pH: Urine values generally range from pH 4.5 - 8.

Blood: Normally, no hemoglobin is detectable in urine. When hemoglobin appears in urine it may indicate kidney disease or a urinary tract disorder. Blood may often be found in the urine of menstruating females.

Specific Gravity (SG): The normal SG of urine ranges from 1.001 to 1.035.

Protein: Normal urine specimens ordinarily contain some protein (<20mg/dL) therefore only

{20}------------------------------------------------

persistent elevated levels of urine protein indicate kidney or urinary tract disease. The persistent results of trace level or greater indicate significance proteinuria and thus further clinical testing is needed to evaluate the significance of results.

Nitrite: Normally no nitrite is detectable in urine.

Leukocyte: Normally no leukocytes are detectable in urine.

Levey AS, Coresh J, Balk E, et al. National Kidney Foundation practice guidelines for chronic kidney disease: evaluation, classification, stratification. Ann Intern Med. 139:137-147; 2003

Conclusion

The new device (DUS 2GP Reagent Strips, DUS 5 Reagent Strips, DUS 10 Reagent Strips) has the same device characteristics as the predicate device. The results of the testing show that the new device is substantially equivalent to the predicate device.

§ 862.1340 Urinary glucose (nonquantitative) test system.

(a)
Identification. A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.(b)
Classification. Class II.