Search Results
Found 2 results
510(k) Data Aggregation
(191 days)
The Micro-Stick® Set is indicated for percutaneous introduction of up to a 0.038 inch guidewire or catheter into the vascular system following a small 21 gauge needle stick. The Micro-Stick® Set is not intended for use in the coronary or cerebral vasculature.
The Micro-Stick® Set is designed to provide dilation of the initial entry point and provide a channel for attaining vascular access. The coaxial introducer consists of an inner dilator within a slightly shorter outer sheath.
The kit contains a 21 gauge introducer needle, a .018" diameter guidewire, and a 4F or 5F coaxial introducer. The coaxial introducer is offered in a standard and stiff version.
The Micro-Stick® Set is packaged sterile.
Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and study proving the device meets them:
No AI/ML device is described in the provided text. The document describes a medical device called the "Micro-Stick® Set," which is an introducer set used for vascular access. The testing performed is in-vitro bench testing to ensure reliable design and performance, not clinical studies involving human readers or AI algorithms.
Therefore, many of the requested sections related to AI/ML device studies (sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable to this document.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document describes in-vitro bench testing for the Micro-Stick® Set. It states that the device was tested "to assure reliable design and performance in accordance with ISO standards and/or internal procedures." However, specific numerical acceptance criteria or detailed performance results (e.g., exact force at break in Newtons, precise leakage rates) are not provided in this summary. It only lists the types of tests performed and a general conclusion of conformance.
Acceptance Criterion (Type of Test) | Reported Device Performance |
---|---|
Guidewire Passage | Met requirements (implied) |
Air Leakage | Met requirements (implied) |
Liquid Leakage | Met requirements (implied) |
Force at Break | Met requirements (implied) |
Simulated Use | Met requirements (implied) |
Biocompatibility (ISO 10993) | Met requirements |
Note: The specific numerical or descriptive acceptance levels for these tests are not detailed in the summary. The summary only confirms that the "devices meet the performance criteria of design verification as specified by ISO standards and test protocols."
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified in the document. The testing was in-vitro bench testing, so there isn't a "test set" in the context of patient data.
- Data Provenance: The testing was "in-vitro" and conducted as "bench performance data." This implies laboratory testing of the physical device components, not involving patient data from a specific country or retrospective/prospective collection.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not an AI/ML device, and the testing was in-vitro bench testing. No human experts were used to establish ground truth for this type of performance evaluation.
4. Adjudication method for the test set
Not applicable. This is not an AI/ML device, and no human readers/experts were involved in assessing the 'ground truth' of the in-vitro performance.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document does not describe an AI/ML device, nor does it conduct MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This document does not describe an AI/ML algorithm.
7. The type of ground truth used
The "ground truth" for the in-vitro tests would be the established ISO standards and internal procedures/specifications for physical device performance (e.g., a specific leakage rate is acceptable, a certain force is required for breakage, a guidewire must pass smoothly). These are engineering and material science metrics, not clinical ground truth like pathology or expert consensus.
8. The sample size for the training set
Not applicable. This document does not describe an AI/ML device, and therefore there is no training set mentioned.
9. How the ground truth for the training set was established
Not applicable. There is no training set for an AI/ML device described here.
Ask a specific question about this device
(128 days)
K990705, P970004
The Percutaneous Introducer Set is intended for percutaneous introduction, manipulation, and removal of stimulation leads.
The Percutaneous Introducer Set is a single use device allowing for introduction, manipulation, and removal of stimulation leads after percutaneous entry is gained with a needle. The sheath is made of HDPE (High Density Polyethylene) with a HDPE hub. It is slightly tapered at the distal end to make a smooth transition to the dilator. The dilator is made of 304 stainless steel tubing tapered at the distal end for smooth insertion. It has a molded HDPE hub. The guidewire directional guide is made of 304 stainless steel wire with radiuses at each end. It has depth markings etched into the surface along the distal end.
This document, K013120, is a 510(k) summary for a Percutaneous Introducer Set. The device is intended for the introduction, manipulation, and removal of stimulation leads.
The document indicates that the device's substantial equivalence is based on its similarity to a predicate device, the MedAmicus Epidural Introducer (K994097), in terms of intended use, basic design, manufacturing methods, and materials.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document
does NOT list specific acceptance criteria with quantifiable metrics, nor does it provide a detailed table of reported device performance against such criteria.
Instead, it offers a general statement: "Performance data indicates that the Percutaneous Introducer Set meets the functional requirements and specifications for this device."
To infer potential "acceptance criteria" based on the provided information, we can consider the aspects of similarity to the predicate device and relevant material properties:
Acceptance Criterion (Inferred from document) | Reported Device Performance (as stated in document) |
---|---|
Intended Use Equivalence: Percutaneous introduction, manipulation, and removal of stimulation leads. | "The Percutaneous Introducer Set is intended for percutaneous introduction, manipulation, and removal of stimulation leads." "The intended use... are similar to the predicate device..." |
Basic Design Equivalence: Physical structure and components (sheath, dilator, guidewire). | "The basic design... are similar to the predicate device..." Detailed description of components: HDPE sheath, HDPE hub, 304 stainless steel dilator, 304 SS guidewire, platinum iridium marker band. |
Manufacturing Methods Equivalence: Processes used to produce the device. | "The methods of manufacturing... are similar to the predicate device..." "The sheath materials and manufacturing processes for the Percutaneous Introducer Set are equivalent to those used on the predicate MedAmicus device..." |
Material Equivalence: Composition of parts that come into contact with the patient or leads. | "The materials are similar to the predicate device..." "Materials used in the Percutaneous Introducer Set are identical to materials previously approved/cleared for market release." (Specific examples: HDPE identical to K990705, 304 SS identical to P970004, Platinum Iridium identical to P970004). |
Biocompatibility: Safety for contact with tissue. | "Applications referred to below are for Externally Communicating - Blood Path Indirect - Limited Exposure which includes all the testing required above." "Since the materials are identical to those previously approved/cleared, there are no new issues involving safety and effectiveness." |
Sterility: Assurance of being free from viable microorganisms. | "Present sterilization cycle assures a Sterility Assurance Level (SAL) of >10-6." |
Packaging Equivalence: Maintenance of sterility and physical integrity. | "The packaging materials for the Percutaneous Introducer Set are equivalent to the predicate MedAmicus Epidural Introducer. These materials consist of a double sterile barrier using Tyvek/Mylar pouches." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not provide information on a specific "test set" with a defined sample size or data provenance (e.g., country of origin, retrospective or prospective). The submission relies on establishing substantial equivalence to a predicate device, and the "performance data" mentioned is a general statement, not a detailed study report.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not provided in the document. The submission focuses on device equivalence rather than a study requiring expert-established ground truth for a test set.
4. Adjudication Method
This information is not provided in the document. As no specific test set or expert ground truth establishment is described, neither is an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A MRMC comparative effectiveness study was not performed or reported in this 510(k) submission. The document aims for substantial equivalence based on physical and functional characteristics, not on the impact of the device on human readers' performance with or without AI assistance.
6. Standalone Performance Study
A standalone (algorithm-only without human-in-the-loop performance) study was not performed or reported in this 510(k) submission. This is a medical device, not an AI/algorithm-based diagnostic tool. The "performance data" mentioned is a general statement about meeting functional requirements.
7. Type of Ground Truth Used
The concept of "ground truth" as typically used in the context of diagnostic performance studies (e.g., expert consensus, pathology, outcomes data) does not apply directly to this 510(k) submission. The "truth" for this submission is based on the demonstration of substantial equivalence to a legally marketed predicate device and the "identicality" of materials to those previously approved.
8. Sample Size for the Training Set
This information is not applicable and therefore not provided. This is not an AI/machine learning device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable and therefore not provided. As there is no training set for an AI/machine learning model, there is no ground truth establishment for such a set.
Ask a specific question about this device
Page 1 of 1