Search Results
Found 3 results
510(k) Data Aggregation
(144 days)
The INFINITY Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to an R50 Bedside recorder, either directly or via the INFINITY Network.
The INFINITY Modular monitors are capable of monitoring:
- Heart rate .
- Respiration rate .
- Invasive pressure .
- Non-invasive pressure .
- Arrhythmia .
- Temperature .
- Cardiac output .
- Arterial oxygen saturation .
- Pulse rate .
- Apnea .
- ST Segment Analysis ●
- 12-Lead ST Segment Analysis
- tcn02/tcp(02 .
- EEG signals .
- FiQ2 .
The SCIO and MultiGas/MultiGas+ modules sample breathing gases from adults and pediatrics. The gas modules continuously measure and produce real time and derived gas information of CO2, NO, O2 and one of the anesthetic agents, halothane, isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the INFINITY monitors.
With etCO2 the monitors can measure end tidal carbon dioxide, inspired carbon dioxide, and respiration with etCO2 and instream or side-stream measurement mode; and with etCO2+Respiratory Mechanics, spirometry and carbon dioxide can be monitored.
The monitors can interface with specific third party devices via an MIB protocol converter.
The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
The devices are intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult population only, and tcpO2 which is to be used in the neonatal population only when the patient is not under gas anesthesia.
The INFINITY Modular monitor VF4 software release includes a modification in support of the Scio gas module for dual-anesthetic agent display.
The provided document describes a Special 510(k) Notification for modifications to the INFINITY Modular MonitorsVF4 software release, specifically in support of the Scio gas module for dual-anesthetic agent display. This is a modification to an existing, legally marketed device, and the submission primarily focuses on demonstrating that the modified device is as safe and effective as the previous versions.
Therefore, the document does not contain information about the development and testing of a new AI/ML device that would typically involve acceptance criteria, performance metrics, ground truth establishment, or clinical studies involving human readers. Instead, it describes a verification and validation study to ensure that the modifications to the existing software and hardware do not introduce new safety or efficacy issues.
Here's an analysis based on the information provided, recognizing that the scope is for device modification rather than a new AI/ML product:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantitative acceptance criteria or detailed performance metrics in the format of a table. Instead, it refers to:
Acceptance Criteria Category | Reported Device Performance |
---|---|
Safety and Effectiveness | "Testing in accordance with internal design control procedures has verified that the INFINITY Modular Monitors VF4 modifications with Scio are as safe and effective as the previous released versions of the monitors." |
New Issues Introduced | "Verification and validation testing of VF4 software, as well as testing applicable to the hardware modifications for the new Draeger look indicate no new issues relative to safety and efficacy." |
The implicit acceptance criterion is that the modified device performs similarly to or better than the predicate device without introducing any new safety or efficacy concerns. The "reported device performance" is a qualitative statement of conformity.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample size for any test set or the data provenance. It generally states that "verification and validation testing of VF4 software" was performed. Given this is a software modification for an existing monitor, the testing would likely involve simulation, bench testing, and potentially some form of in-house functional testing using synthetic or recorded physiological data, but no details are provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable to the type of submission described. "Ground truth" in the context of device performance, especially for AI/ML, refers to a definitive correct answer for a given input. For this type of physiological monitoring device modification, the "ground truth" would be the known physiological parameters (e.g., gas concentrations) used during testing, established by calibrated reference equipment, not by human experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies where human readers interpret data, and there's a need to resolve discrepancies. This submission pertains to software and hardware modifications to a physiological monitor, not diagnostic image interpretation.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic aids where human interpretation is involved. The INFINITY Modular Monitors are physiological monitors that directly display parameters. The modifications relate to displaying dual-anesthetic agents.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document implies that the device's performance (i.e., its ability to monitor and display physiological parameters) was assessed independently through "verification and validation testing." However, it does not explicitly use the term "standalone performance" in the context of an algorithm. The device itself, as a monitor, operates in a "standalone" fashion by measuring and displaying data, and the software modifications were tested to ensure this function remains effective.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The type of "ground truth" for a physiological monitor would typically be measurements from calibrated reference instruments or known physiological states (e.g., a gas mixture with a known concentration of anesthetic agents). The document does not explicitly state how ground truth was established, but it would be part of standard verification and validation procedures for such devices. For example, to test the accuracy of gas measurements, the Scio module would be exposed to precisely mixed gas samples with known concentrations, and the device's readings would be compared to these known values.
8. The sample size for the training set
Not applicable. This is a modification to an existing physiological monitor, not an AI/ML product that undergoes a training phase with a dataset.
9. How the ground truth for the training set was established
Not applicable. As above, there is no training set mentioned or implied for this type of device modification.
Ask a specific question about this device
(88 days)
Use of the INFINITY MultiView WorkStation Telemetry System is indicated for adult and pediatric patient populations in an environment where patient care is provided by Healthcare Professionals (Physicians, Nurses, Technicians) when the professional determines that a device is required to measure and produce visual and audible alarms for any one or more of the following parameters:
- . Heart rate
- ECG Arrhythmia Analysis ●
- Arterial oxygen saturation ●
- Pulse rate
- ST segment analysis .
The INFINITY MultiView WorkStation (MVWS) Telemetry System with TruST is indicated for use when 12-Lead ECG monitoring with a reduced set of electrodes is desired. Reconstructed leads are intended for real-time assessment of ST segment changes.
Siemens' INFINITY MultiView WorkStation Telemetry with TruST enables continuous 12-lead ECG monitoring with a reduced number of electrodes. With TruST only 6 lead wires are required to acquire 12 leads of ECG. TruST relies on conventional lead placement and the INFINITY TruST telemetry transmitters. The TruST matrix uses general reconstruction coefficients to compute 4 precordial leads.
Here's a breakdown of the acceptance criteria and study information for the Siemens Infinity MultiView WorkStation Telemetry System with TruST, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) summary does not explicitly state numerical performance acceptance criteria (e.g., specific sensitivity, specificity, accuracy thresholds) or precise reported device performance metrics for the TruST feature. Instead, it relies on substantial equivalence to predicate devices and describes the intended functionality. However, we can infer the acceptance criteria are met if the device demonstrates equivalent performance to the predicate devices in terms of its ability to monitor 12-lead ECG with a reduced set of electrodes and reconstruct leads for real-time ST segment assessment.
Feature / Performance Aspect | Acceptance Criteria (Inferred from Substantial Equivalence) | Reported Device Performance (as described in the document) |
---|---|---|
ST Segment Analysis (12-Lead ECG with reduced electrodes) | The device should effectively reconstruct 12-lead ECG from a reduced set of electrodes and provide real-time assessment of ST segment changes, similar to the predicate devices. This implies accurate ST segment measurement and trending. | "The INFINITY MultiView WorkStation Telemetry with TruST enables continuous 12-lead ECG monitoring with a reduced number of electrodes. With TruST only 6 lead wires are required to acquire 12 leads of ECG. TruST relies on conventional lead placement and the INFINITY TruST telemetry transmitters. The TruST matrix uses general reconstruction coefficients to compute 4 precordial leads." |
"Reconstructed leads are intended for real-time assessment of ST segment changes." | ||
Intended Use compatibility with reduced electrode set | The device should be usable for its stated intended use (12-lead ECG monitoring and ST segment assessment) with the reduced electrode configuration. | "The INFINITY MultiView WorkStation (MVWS) Telemetry System with TruST is intended for 12-Lead ECG monitoring with a reduced set of electrodes." |
Physiological Parameter Monitoring (general) | The system should accurately measure and produce visual/audible alarms for specified physiological parameters, in line with predicate devices. | "...intended to measure and produce visual and audible alarms for one or more physiological parameters" (Heart rate, ECG Arrhythmia Analysis, Arterial oxygen saturation, Pulse rate, ST segment analysis). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size used for any test set or the data provenance (e.g., country of origin, retrospective/prospective). It mentions "Assessment of clinical performance data for equivalence: Section J," implying that clinical data was submitted in that section, but the summary document itself does not contain those details.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not provide information on the number of experts used or their qualifications for establishing ground truth.
4. Adjudication Method for the Test Set
The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor does it provide an effect size for human readers with vs. without AI assistance. The focus of this 510(k) is on the device's technical capability to reconstruct leads and monitor ST segments, implying a standalone performance assessment or equivalence claim rather than a human-in-the-loop improvement study.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
The description of the TruST system, specifically that "The TruST matrix uses general reconstruction coefficients to compute 4 precordial leads" and that "Reconstructed leads are intended for real-time assessment of ST segment changes," strongly suggests that a standalone (algorithm-only) performance assessment would have been done to validate the accuracy of these reconstructed leads against a conventionally acquired 12-lead ECG. However, the summary document does not explicitly describe this study or its results in detail. The focus is on establishing substantial equivalence.
7. Type of Ground Truth Used (Test Set)
The document does not explicitly state the type of ground truth used for the test set. Given the nature of ECG monitoring and ST segment analysis, it's highly probable that the ground truth would have been established by:
- Comparison to conventionally acquired 12-lead ECGs (e.g., a full 12-lead ECG from the same patient at the same time, reviewed by experts).
- Expert consensus based on interpretation of conventional 12-lead ECGs.
- Potentially, correlation with clinical outcomes or pathology if the clinical study was designed to assess diagnostic accuracy for specific conditions (though this is less likely for a 510(k) summary focused on technical equivalence for monitoring).
8. Sample Size for the Training Set
The document does not specify the sample size used for the training set.
9. How the Ground Truth for the Training Set Was Established
The document does not describe how the ground truth for any potential training set was established. Since the TruST system involves "general reconstruction coefficients," it implies a model or algorithm was developed. For such a model, a training set would typically have ground truth established by:
- Conventionally acquired 12-lead ECGs that are considered the gold standard for full 12-lead information.
- Expert cardiologists or electrophysiologists who interpret these full 12-lead ECGs and potentially correlate them with patient data or outcomes to develop the "ground truth" for ST segment morphology and deviation.
Ask a specific question about this device
(22 days)
The SC 8000 Bedside monitor is capable of monitoring:
- . Heart rate
- Respiration rate .
- Invasive pressure .
- Non-invasive pressure .
- Arrhythmia ●
- Temperature .
- Cardiac output .
- Arterial oxygen saturation .
- . Pulse rate
- (central) apnea ●
- ST segment analysis .
- 12-Lead ST Segment Analysis ●
- tcpO2/CO2 .
The device is intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
The device is intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output and ST Segment Analysis which are intended for use in the adult and pediatric populations only; Arrhythmia which is intended for use in the adult population only; and tcpO2/CO2 which is to be used in the neonatal population only.
The SC 8000 bedside monitor is a modified version of the SC 7000/SC 9000XL INFINITY Modular Bedside Monitors (K982730). The SC 8000 supports all SC 7000 software based options and offers all of the standard I/O connections of the SC 7000 plus interface plate. The hardware has been reengineered as a unit without a display that allows customers the option of providing their own medical-grade display.
The Siemens SC 8000 Bedside Monitor is a modified version of the SC 7000/SC 9000XL INFINITY Modular Bedside Monitors. This device is intended to monitor various physiological parameters and produce alarms if these parameters vary beyond preset limits.
Here's an analysis of the provided text based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) summary does not explicitly state specific acceptance criteria or provide a table of performance data for the SC 8000 Bedside Monitor. The document focuses on establishing substantial equivalence to a predicate device (Siemens SC 7000/SC9000 XL INFINITY Modular Bedside Monitors, K982730) rather than presenting a detailed performance study with defined acceptance criteria for the new device.
The "Intended Use" section lists the parameters the device is capable of monitoring, but it does not specify performance metrics (e.g., accuracy, sensitivity, specificity, or error margins) for these parameters.
Therefore, I cannot generate the requested table of acceptance criteria and reported device performance from the provided text.
2. Sample Size Used for the Test Set and Data Provenance
The document states, "Assessment of non-clinical performance data for equivalence: See Section L" and "Assessment of clinical performance data for equivalence: Not applicable." This indicates that clinical performance data was not considered applicable for demonstrating equivalence.
Since no clinical performance study for the SC 8000 is described, there is no information on sample size, test set, or data provenance (country of origin, retrospective/prospective).
3. Number of Experts Used to Establish Ground Truth and Their Qualifications
As no clinical performance study is described as applicable for the SC 8000, there is no information on the number of experts used to establish ground truth or their qualifications.
4. Adjudication Method
Given that no clinical performance study or ground truth establishment for the SC 8000 is described, there is no information on an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document states, "Assessment of clinical performance data for equivalence: Not applicable." This clearly indicates that a multi-reader multi-case (MRMC) comparative effectiveness study was not performed or considered necessary for this submission. Therefore, there is no information on the effect size of human readers with vs. without AI assistance.
6. Standalone (Algorithm Only) Performance Study
The document does not describe any standalone algorithm performance studies for the SC 8000. The focus is on the device's functional capabilities and its equivalence to a predicate device, not on the performance of a specific embedded algorithm in isolation. Therefore, no standalone performance study is explicitly mentioned.
7. Type of Ground Truth Used
As clinical performance data was deemed "Not applicable" for this submission, the document does not specify the type of ground truth used for evaluating the SC 8000's performance. The equivalence relies on the predicate device's established performance and the changes being non-substantive from a clinical perspective.
8. Sample Size for the Training Set
The document does not describe the development or training of any algorithms for the SC 8000. It refers to the device as a "modified version" of existing monitors. Therefore, there is no information regarding the sample size for a training set.
9. How the Ground Truth for the Training Set Was Established
Since no training set is described (as per point 8), there is no information on how the ground truth for a training set was established.
Ask a specific question about this device
Page 1 of 1