Search Results
Found 2 results
510(k) Data Aggregation
(232 days)
The Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the Newport HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs).
The Newport HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
The Newport HT70 is a state of the art ventilator that combines ruggedness, exceptional mobility, and ease of use with clinical proficiency to provide ventilatory support for infant, pediatric and adult patients. It has a durable exterior and robust overall design that stands up to the harsh environments found in emergency response, transport and homecare environments. The HT70's unique design provides maximum mobility and safety for short or long distance transport of critically ill patients and also for patients who are going about their normal activities of daily life.
The HT70 is user friendly with all controls easily found on the front touchscreen and panel; there are no complicated menus or difficult sequences to follow in order to make necessary adjustments for common operations. It may be operated from AC or DC external power sources or from the "hot-swappable" Integrated Battery System. Any time the ventilator is connected to external power, the Integrated Battery System is charged, including while the ventilator is in use.
The Newport HT70™ Family of Ventilators is a device intended to provide continuous or intermittent positive pressure mechanical ventilatory support.
Here's an analysis based on the provided 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Design, Performance, and Safety Specifications (General) | "All test results confirm the device to meet its design, performance, and safety specifications." |
Electromagnetic Compatibility | Confirmed to meet specifications. |
Electrical Safety | Confirmed to meet specifications. |
Software and Hardware Verification and Validation | Confirmed to meet specifications. |
Mechanical Properties | "All testing demonstrates that the Newport HT70 performs as intended and has acceptable mechanical properties when used in accordance with its labeling." |
Substantial Equivalence to Predicate Devices | "The device's intended use is comparable to the referenced predicate devices, and its operating principles, ventilation modes and performance parameters are comparable to the predicate devices." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated for specific tests. The document refers to "a series of tests" but does not quantify the number of units or test iterations.
- Data Provenance: The tests are non-clinical and would have been conducted by the manufacturer, Newport Medical Instruments, Inc., likely at their facilities in Costa Mesa, CA, USA. The data is retrospective in that it's prior to the 510(k) submission, but it's new data generated for this specific device.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
- This information is not applicable to this 510(k) submission. The Newport HT70 Ventilator is a medical device designed to perform a physical function (mechanical ventilation) and its performance is evaluated against engineering, electrical, and software specifications, rather than against an expert-derived ground truth based on interpretation (e.g., image analysis for disease detection).
4. Adjudication Method for the Test Set
- Not applicable. As the performance assessment is based on objective measurements against predefined specifications, an adjudication method for reconciling expert opinions is not relevant.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of human readers improving with AI vs. without AI assistance
- No. An MRMC study is not applicable here. This device is a ventilator, not an AI-assisted diagnostic or decision-support tool where human readers interact with AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable in the context of AI algorithms. While the ventilator's software and hardware perform functions independently of continuous human-in-the-loop operation, this is not an "algorithm-only" performance as understood in AI/ML contexts. The device's performance is tested as a complete system of hardware and software.
7. The Type of Ground Truth Used
- The "ground truth" for the non-clinical tests were the engineering design specifications, performance requirements, and safety standards (e.g., electromagnetic compatibility standards, electrical safety standards, internal performance metrics for ventilation parameters). For substantial equivalence, the "ground truth" was the performance and operating principles of the legally marketed predicate devices.
8. The Sample Size for the Training Set
- Not applicable. This submission is for a conventional medical device (ventilator) and does not involve AI/Machine Learning, therefore there is no "training set."
9. How the Ground Truth for the Training Set was Established
- Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.
Ask a specific question about this device
(29 days)
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via . mask).
- Assist/Control, SIMV, or CPAP modes of ventilation. ●
- Breath types including Volume, Pressure Control and Pressure Support. ●
The ventilator is suitable for use in institutional, home, or transport settings.
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is suitable for use in institutional, home and transport settings, and is applicable for adult and pediatric patients weighing at least 5 kg (11 lbs.), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via . mask).
- . Assist/Control, SIMV, or CPAP modes of ventilation.
- Breath types including Volume, Pressure Control and Pressure Support. .
The modification intended to be cleared by this submission is: - The addition of Spontaneous Breathing Trial (SBT) function allowing the clinician to . more easily determine a patient's ability to be weaned from ventilation.
The provided 510(k) summary for the Pulmonetic Systems LTV 1000 Ventilator (K051767) primarily focuses on demonstrating substantial equivalence to predicate devices, particularly regarding a new "Spontaneous Breathing Trial (SBT)" function. This type of submission, for a ventilator modification, does not typically include detailed performance studies with acceptance criteria in the same way an AI/ML device would.
Based on the provided document, here's an analysis concerning acceptance criteria and study information:
1. A table of acceptance criteria and the reported device performance:
The document doesn't present a table of quantitative acceptance criteria and reported 'performance' in the sense of a diagnostic or predictive device. Instead, it compares characteristics of the modified LTV 1000 Ventilator with its predicate devices to demonstrate equivalence.
Characteristic (LTV Modification) | Acceptance Criteria (Implied) | Reported Device Performance (LTV 1000 with SBT Function) |
---|---|---|
Spontaneous Breathing (SBT) Function | Functionality is comparable to predicate, with beneficial enhancements and no new patient risk or technology. | Software Application: The software application to allow input presets for CPAP mode/parameters and alarm parameters is an addition to the existing LTV 1000 preset menu options. |
SBT Time Duration | Time duration meets clinician requests (potentially longer). | Offers SBT time duration via menu selection from 15-120 minutes. (Predicate LTV 1000 required manual CPAP mode/parameter setting; Engstrom Carestation offered 5-60 minutes). This longer duration is provided at the request of clinicians. |
Safety and Risk | No introduction of additional or new patient risk. | The software application "does not introduce any additional or new patient risk or involve the application of new technology." (Comparison to predicate LTV 1000, which required manual CPAP settings and had an alarm for trial end). The predicate Engstrom Carestation also required manual CPAP settings. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
This submission is for a modification to a medical device (ventilator software feature), not a diagnostic or AI/ML device that would typically involve a "test set" of patient data for performance evaluation in the way a clinical study would. Therefore, there is no specific sample size, test set, or data provenance mentioned in relation to proving the SBT function's performance. The evaluation is based on engineering design, comparison to predicate devices, and potentially internal validation/verification testing which is not detailed in this public summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. As this is not a diagnostic device relying on expert interpretation for ground truth, there's no mention of experts establishing ground truth for a test set. The statement about "Longer time duration is provided at the request of clinicians" suggests input from medical professionals informed the feature design, but not for establishing ground truth in a performance study.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. There is no test set or adjudication method described in the context of this submission.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI-assisted diagnostic device, and therefore, no MRMC study or AI assistance effect size is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. The device is a ventilator with a software function. The "performance" is inherent to the function's design and operation, not an algorithm providing a diagnostic output independently.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable in the traditional sense of a diagnostic performance study. The "ground truth" for the SBT function would be its adherence to engineering specifications, correct implementation of the defined parameters (e.g., timing, mode switching), and safe operation in a simulated or actual clinical environment. This type of validation is typically conducted through internal engineering testing, which is not detailed here but implied by regulatory compliance.
8. The sample size for the training set:
Not applicable. This is not an AI/ML device that would require a 'training set'.
9. How the ground truth for the training set was established:
Not applicable. This is not an AI/ML device.
Summary:
The 510(k) summary for K051767 focuses on demonstrating "substantial equivalence" of a modified ventilator with a new Spontaneous Breathing Trial (SBT) function to previously cleared predicate devices. The "acceptance criteria" here are implicitly tied to ensuring the new function is safe, effective, and performs comparably or superiorly without introducing new risks compared to existing methods or devices. The "study" proving this largely relies on design comparison, engineering assessment of the software modification, and the assertion that it does not introduce new technology or patient risk. It does not involve a clinical performance study with patient data, ground truth establishment, or human reader evaluations as would be expected for a diagnostic or AI/ML device.
Ask a specific question about this device
Page 1 of 1