510(k) Data Aggregation

    K Number
    K041035
    Manufacturer
    Date Cleared
    2004-06-09

    (49 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K014074,K973161,K003561

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TREND 110 device is for treatment of obstructive sleep apnea (OSA) in adult patients weighing at least 30 kg. These devices are not intended for use with ventilator-dependent patients.

    The TREND 110 device provides continuous positive airway pressure.

    The AquaTREND Humidifier accessory provides humidification of the air delivered to the patient.

    Device Description

    Hoffrichter TREND 110 devices provide mono-level CPAP treatment levels over the clinician settable range of 4 to 18 cmH₂O.

    The TREND 110 CPAP device is provided in a compact lightweight enclosure for easy bedside use. A 2 x 16 character LCD display with keys on the operator panel provides the user with displayed information about the treatment level as well as access to menus for soft-start ramp, clinical prescription (set by doctor, PIN code protected) and hours of operation. TREND 110's power supply automatically supply adapts to regional mains voltage (115 to 230 VAC).

    Heated humidification is available by installing the AquaTREND /// heated humidifier accessory. A key with LED indicator permits the user to select the humidification level. Moisture contacting materials in the humidifier and therapy tube meet biocompatibility requirements.

    Treatment pressure is measured as close as possible to the patient mask Freathern of a measuring tube integrated into the therapy tubing. This maximizes accuracy and control of pressure and minimizes losses due to the therapy delivery circuit.

    TREND 110 has a user selectable ramp function, which starts with a reduced pressure and gradually increases to the treatment pressure to make it more comfortable to fall asleep. The user can adjust the ramp time in 10 minute increments to 0 (no ramp), 10, 20, or 30 minutes.

    Standard accessories include treatment tubing with integrated measurement tube and headgear. Commercially available masks and swivels are specified.

    Hoffrichter CPAP devices are prescription use devices, and are not intended for life supporting or life sustaining applications.

    AI/ML Overview

    This 510(k) summary describes a CPAP device and humidifier accessory. The provided text does not include the typical structure of acceptance criteria and a detailed study report that would fully answer your request.

    Medical device approval (especially for 510(k) submissions) often relies on demonstrating substantial equivalence to a predicate device rather than strictly meeting pre-defined performance acceptance criteria in the same way a new drug might. Therefore, the "acceptance criteria" here are generally implied to be "comparable performance to the predicate devices" across various technical and safety benchmarks.

    Based on the provided document, here's the information that can be extracted and inferred:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Electrical SafetyPassed applicable standards
    Electromagnetic Compatibility (EMC)Passed applicable standards
    Shock and Vibration ResistancePassed applicable standards
    Environmental Temperature and Humidity TolerancePassed applicable standards
    Static Pressure Performance (comparable to predicate)Met specified requirements; comparable to predicate devices
    Dynamic Pressure Performance (comparable to predicate)Met specified requirements; comparable to predicate devices
    Humidification Performance (comparable to predicate)Met specified requirements; comparable to predicate devices
    Software Verification to RequirementsVerified to requirements
    Software Validation to Intended UseValidated to meet intended use by parameter and event testing
    Biocompatibility of Moisture-Contacting MaterialsMet biocompatibility requirements
    Equivalence to Predicate DevicesDemonstrated to be as safe and effective as, and perform in a manner equivalent to, the predicate devices: Hoffrichter SCALAR/VECTOR Plus, Fisher and Paykel HC200, and Invacare Poseidon.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated for any of the non-clinical tests. The document refers to testing "the TREND 110 and AquaTREND III devices" and comparisons "with predicate devices," implying a laboratory or engineering test environment rather than a patient-based test set.
    • Data Provenance: Not explicitly stated, but the tests were conducted by the manufacturer, HOFFRICHTER GmbH, in Germany. These are non-clinical engineering and bench tests, not clinical data from patients.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Not applicable. This document describes non-clinical engineering and bench tests for device functionality and safety. The "ground truth" for these tests would be established by engineering specifications, regulatory standards, and performance characteristics of predicate devices, not by expert medical consensus on patient data.

    4. Adjudication Method for the Test Set

    • Not applicable. As the tests described are non-clinical engineering tests, there is no mention of an adjudication method in the way it would apply to clinical expert reviews.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No. The document explicitly states "Clinical Tests Submitted: (None)". Therefore, no MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    • Not applicable. This device is a CPAP machine, a physical device that delivers therapy, not an algorithm or AI system for diagnosis or interpretation. The "software" mentioned refers to the embedded control software for the physical device, not a diagnostic algorithm.

    7. The type of ground truth used

    • For the non-clinical tests (e.g., pressure, humidification, safety), the "ground truth" was based on:
      • Applicable industry standards (e.g., for electrical safety, EMC).
      • Manufacturer's internal specified requirements.
      • Performance characteristics of the predicate devices for comparison.
    • No patient-derived ground truth (e.g., pathology, outcomes data) was used, as no clinical studies were submitted.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a hardware product with embedded control software, not an AI or machine learning model that requires a "training set" of data in the typical sense. The software was verified and validated against requirements and intended use, which is a different process than machine learning model training.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As there is no "training set" for an AI or machine learning model, this question does not apply. The "ground truth" for the device's control software would be its functional specifications and intended performance parameters.
    Ask a Question

    Ask a specific question about this device

    K Number
    K014074
    Manufacturer
    Date Cleared
    2003-07-23

    (590 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K973161

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Scalar, Scalar Plus, Vector, and Vector Plus devices are for treatment of obstructive sleep apnea (OSA) in adult patients weighing at least 30 kg. These devices are not intended for use with ventilator-dependent patients.

    The Scalar, Scalar Plus, Vector Plus devices provide continuous positive airway pressure. The Scalar Plus and Vector Plus devices provide humidification of the air delivered to the patient.

    Device Description

    Hoffrichter SCALAR and VECTOR CPAP devices provide mono-level CPAP treatment levels over the clinician settable range of 3 to 18 cmH2O.

    All devices are constructed on a common platform, including case, mechanical and pneumatic components, operator panel, and electronic control system. A 2 x 16 character display with keys on the operator panel provides the user with displayed information about the treatment level as well as access to menus for time/date, ramp and other functions.

    VECTOR's power supply automatically supply adapts to regional mains voltage (115 to 230 VAC), whereas SCALAR is set to the region (115 or 230 VAC) by rear panel switch.

    Models with a Plus designation include integrated heated humidification. An adjustment dial permits the user to select the humidification level. Moisture contacting materials in the breathing circuit meet biocompatibility requirements.

    Treatment pressure is measured as close as possible to the patient mask by means of a measuring tube integrated into the therapy tubing. This maximizes accuracy and control of pressure and minimizes losses due to the therapy delivery circuit.

    Hoffrichter CPAP devices have a user selectable ramp function, which starts with a reduced pressure and slowly increases to the treatment pressure to make it more comfortable to fall asleep. The user can adjust the ramp time from 0 to 60 minutes. A mask test may be initiated at the full treatment pressure before the ramp begins to check for mask leaks.

    Accessories include treatment tubing with integrated measurement tube and headgear. Commercially available masks and swivels are specified.

    A serially connected remote control device, the MULTI 20, supplied for clinician use during patient tests, emulates operator panel functions.

    Hoffrichter CPAP devices are prescription use devices, and are not intended for life supporting or life sustaining applications.

    AI/ML Overview

    The provided document describes the Hoffrichter Vector, Vector Plus, Scalar, and Scalar Plus CPAP devices. Based on the submission, here's a breakdown of the acceptance criteria and the study information:

    Description of the Study and Acceptance Criteria

    The submission is a 510(k) premarket notification, which seeks to demonstrate substantial equivalence to a previously legally marketed device (the predicate device). This means the study's primary goal is to show that the new device is as safe and effective as the predicate device. For CPAP devices, this typically involves demonstrating that the device can deliver the intended therapy (continuous positive airway pressure) within specified ranges and that it meets relevant safety standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a formal table with specific numerical acceptance criteria for pressure accuracy or other performance metrics, nor does it report precise measured values for the Hoffrichter CPAP devices. Instead, the performance is described in comparison to the predicate device.

    Acceptance Criteria (Implied)Reported Device Performance
    Safety: Compliance with applicable medical device standardsPassed all tests for Electrical Safety, Electromagnetic Compatibility, Heated Humidifiers, Shock and Vibration, and Environmental Temperature and Humidity.
    Pressure Delivery: Provide mono-level CPAP treatment.Delivers CPAP treatment levels over a clinician settable range of 3 to 18 cmH2O. Improved pressure control and accuracy compared to the predicate device due to pressure measurement at the mask via an integrated measurement hose.
    Humidification: (For Plus models)Integrated heated humidification with an adjustment dial for the user to select the humidification level. Motion contacting materials meet biocompatibility requirements.
    Ramp Function: User-selectable ramp.User can adjust ramp time from 0 to 60 minutes.
    Dynamic and Static Pressure Performance:Met or exceeded the performance specifications of the predicate device (Fischer and Paykel HC200 Nasal CPAP with Heated Humidifier). The Hoffrichter CPAP devices employ a more advanced pressure measurement system (at the mask) compared to the predicate device, which sets pressure by a manual blower speed adjustment, implying better performance in this aspect.
    Software Validation:Software across all platforms was verified to requirements and validated to meet intended use by pressure testing and event testing.
    Biocompatibility:Moisture contacting materials in the breathing circuit meet biocompatibility requirements.
    Pneumatic Components:Constructed on a common platform, including case, mechanical and pneumatic components, similar to the predicate device.
    Intended Use Equivalence:Indications for use are the same as the predicate device: treatment of obstructive sleep apnea (OSA) in adult patients weighing at least 30 kg, not intended for use with ventilator-dependent patients (with available humidification for Plus models).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document refers to "the Hoffrichter CPAP devices" being tested, implying a representative number of units, but no specific count is given for the test set of devices or patients.
    • Data Provenance: The document does not provide details on data provenance (e.g., country of origin, retrospective/prospective). Given the nature of a 510(k) for device performance testing, it's highly likely these were bench tests (non-clinical) and not conducted on human subjects as part of this submission. The address of the submitting company (Hoffrichter GmbH) is in Germany, suggesting the tests were conducted in Germany or Europe.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable to this submission. The tests performed were non-clinical (bench testing) and compared the device's technical performance against established standards and the predicate device's specifications. There was no "ground truth" derived from expert consensus on medical conditions or interpretations.

    4. Adjudication Method for the Test Set

    This section is not applicable. Since no human interpretation or ground truth establishment based on expert consensus was involved, there was no adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission explicitly states "Clinical Tests Submitted: (2) (None)". The study focused on non-clinical (bench) testing of the device's technical specifications and safety.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This is partially applicable in the sense that the device's performance was evaluated purely on its own technical capabilities (standalone) during bench testing. However, it's important to note that this is not an "algorithm only" in the context of AI/software performance. It's the performance of a physical medical device. The "software" component was verified and validated to requirements, which implies its performance characteristics were evaluated in a standalone manner.

    7. The Type of Ground Truth Used

    The "ground truth" or reference standards used for the non-clinical tests were:

    • Applicable standards for medical devices: e.g., for Electrical Safety, Electromagnetic Compatibility, Heated Humidifiers, Shock and Vibration, Environmental Temperature and Humidity.
    • Performance specifications of the predicate device: The Fischer and Paykel HC200 Nasal CPAP with Heated Humidifier (K973161).

    8. The Sample Size for the Training Set

    This section is not applicable. The device is a hardware medical device with embedded control software, not an AI or machine learning algorithm that requires a "training set" in the conventional sense. The software was verified and validated, meaning it met its specified requirements and functions as intended, but this doesn't involve a machine learning training phase.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable for the reasons stated in point 8.