K Number
K973161

Validate with FDA (Live)

Date Cleared
1998-07-10

(322 days)

Product Code
Regulation Number
868.5905
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fisher & Paykel Healthcare HC200 CPAP Humidifier is a combined Continuous Positive Airway Pressure Blower and Heated Respiratory Humidifier (as per 73 BZD, 21 CFR §868.5905 and 73 BTT, 21 CFR §868.5450).

The HC200 is used to assist with patient breathing while sleeping, for the purpose of treating Obstructive Sleep Apnea (OSA). This is by the delivery of Continuous Positive Airway Pressure (CPAP) in order to prevent airway obstruction.

The addition of heated respiratory humidification to the device (normally included as an accessory to CPAP therapy) relieves the drying and irritating effects on the patient airways which may arise from use of a CPAP system.

The HC200 is for use on adult, spontaneously breathing (non ventilator-dependent) patients and is a prescription-use device.

Device Description

The Fisher & Paykel Healthcare HC200 CPAP Humidifier is a Continuous Positive Airway Pressure flow generator, as per 73 BZD, 21 CFR §868.5905. It includes a Heated Respiratory Humidifier, as per 73 BTT, 21 CFR §868.5450.

The HC200 is used to assist with patient breathing while sleeping, for the purpose of treating Obstructive Sleep Apnea (OSA). This is by the delivery of Continuous Positive Airway Pressure (CPAP) in order to prevent airway obstruction. The addition of heated respiratory humidification relieves the drying and irritating effects on patient airways which may arise from use of a CPAP system. The HC200 is for use on adult, spontaneously breathing (non ventilator-dependent) patients and is a prescription-use device.

A motorized fan assembly provides positive air pressure, which can be adjusted as prescribed by a physician. The fan speed is directly related to air pressure, and is controlled by software. An air filter is located at the back of the device leading to the input of the fan assembly. Warmth and moisture are added to the air as it passes over a heated body of water in a humidification chamber, which slides into place on a heaterplate at the front of the unit. The chamber connects directly to the CPAP generator section via a port at the back of the chamber. The temperature of the heaterplate can be adjusted by the user. Temperature is controlled by software, with the level of humidification delivered dependent on heaterplate temperature, rate of air flow, and ambient conditions of temperature and humidity.

Safety features for the device consist of software monitoring of CPAP generator. humidifier and power supply parameters, electronic monitoring of the microprocessor and a thermal cut-out on the heaterplate.

Controls include a pressure set potentiometer and mains power switch, on the right side of the device, with the mains power cord. A heaterplate adjustment potentiometer, mask test mode button and display LED are located on the front control panel.

The case enclosure components are a polycarbonate thermoplastic, and the heaterplate is cast aluminum. The humidification chamber is a clear thermoplastic top with an aluminum base. Accessories for the HC200 consist of a delivery tube and replacement humidification chambers.

AI/ML Overview

The provided document is a 510(k) summary for the Fisher & Paykel HC200 CPAP Humidifier and does not contain detailed acceptance criteria or a study with the specific information requested in your prompt regarding device performance metrics like AUC, sensitivity, specificity, or detailed experimental setups with sample sizes, expert involvement, and ground truth methodologies usually associated with AI/ML device evaluations.

This document describes a traditional medical device (CPAP Humidifier) clearance based on substantial equivalence to predicate devices, not an AI/ML powered device. Therefore, many of the requested fields are not applicable or cannot be extracted from this document.

However, I can extract the information related to the device's functional equivalence testing.

1. Table of Acceptance Criteria (Implied) and Reported Device Performance:

Since this is a substantial equivalence submission, the "acceptance criteria" are implied to be the performance characteristics of the predicate devices. The "reported device performance" indicates that the HC200 met these equivalent characteristics.

Performance CharacteristicAcceptance Criteria (Implied from Predicate Devices)Reported Device Performance (HC200)
CPAP Operating PressureSame range as predicate devicesOperates in the same CPAP operating pressure range
Heaterplate TemperatureSame range as predicate devicesOperates in the same heaterplate temperature range
Pressure Drop vs. FlowrateEquivalent to predicate CPAP systemsHas equivalent pressure drop vs. flowrate characteristics to predicate CPAP systems
Dynamic Pressure VariationEquivalent to predicate CPAP systemsHas equivalent dynamic pressure variation characteristics to predicate CPAP systems
Temperature DeliverySame levels as the predicate humidifierProvides the same levels of temperature over varying flowrates as the predicate humidifier, at the same heaterplate temperatures
Humidity DeliverySame levels as the predicate humidifierProvides the same levels of humidity over varying flowrates as the predicate humidifier, at the same heaterplate temperatures
Safety FeaturesEquivalent to predicate devices (software monitoring, electronic monitoring, thermal cut-out, compliance with IEC 601-1)Has similar safety features (software monitoring, electronic monitoring, thermal cut-out) and complies with IEC 601-1

2. Sample size used for the test set and the data provenance:

  • Sample size: Not specified. The document states "Testing carried out on the HC200 demonstrates that the device has equivalent performance..." but does not provide a sample size (e.g., number of units tested, duration of testing, specific conditions).
  • Data provenance: Not specified. This would typically be laboratory testing conducted by the manufacturer, Fisher & Paykel Healthcare, in New Zealand, as indicated by their address. It's a prospective test of newly manufactured devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable and not provided. The testing described is for a physical device's performance characteristics (pressure, temperature, humidity) against engineering specifications and predicate device performance, not for an AI/ML algorithm requiring expert interpretation of medical data.

4. Adjudication method for the test set:

Not applicable. There is no mention of expert adjudication for physical device performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is not an AI/ML device where human readers are involved in interpreting outputs.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a traditional medical device, not an AI/ML algorithm.

7. The type of ground truth used:

The ground truth for this type of device performance testing would be:

  • Engineering specifications/measurements: The established physical and electrical parameters for CPAP pressure, flow, temperature, and humidity.
  • Predicate device performance data: The documented performance characteristics of previously cleared CPAP systems and humidifiers (Respironics Inc, Solo™ CPAP System, K961626; ResCare Ltd, Sullivan® III Nasal CPAP System, K930656; Fisher & Paykel, HC100 Respiratory Humidifier, K915460). The HC200's performance was compared directly to these.

8. The sample size for the training set:

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

JUL 10 1998

Image /page/0/Picture/1 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are on the top line in a bold, sans-serif font. Underneath the top line is a horizontal line, and below that is the word "HEALTHCARE" in a thinner, sans-serif font.

Fisher & Paykel Electronic Box 14-348 Panmure. 4-9-574-0100

K973161

7 August, 1997

1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

510(k) Summary of Safety and Effectiveness Information

Model Number / Name: HC200 CPAP Humidifier

Ventilator, Non-Continuous (Respirator) - 73 BZD Classification Name: Anesthesiology Devices, 21 CFR §868.5905 (Class II) Respironics Inc, Solo™ CPAP System, K961626 Predicate Devices: ResCare Ltd, Sullivan® III Nasal CPAP System, K930656 Fisher & Paykel, HC100 Respiratory Humidifier, K915460

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The Fisher & Paykel Healthcare HC200 CPAP Humidifier is a Continuous Positive Airway Pressure flow generator, as per 73 BZD, 21 CFR §868.5905. It includes a Heated Respiratory Humidifier, as per 73 BTT, 21 CFR §868.5450.

The HC200 is used to assist with patient breathing while sleeping, for the purpose of treating Obstructive Sleep Apnea (OSA). This is by the delivery of Continuous Positive Airway Pressure (CPAP) in order to prevent airway obstruction. The addition of heated respiratory humidification relieves the drying and irritating effects on patient airways which may arise from use of a CPAP system. The HC200 is for use on adult, spontaneously breathing (non ventilator-dependent) patients and is a prescription-use device.

A motorized fan assembly provides positive air pressure, which can be adjusted as prescribed by a physician. The fan speed is directly related to air pressure, and is controlled by software. An air filter is located at the back of the device leading to the input of the fan assembly. Warmth and moisture are added to the air as it passes over a heated body of water in a humidification chamber, which slides into place on a heaterplate at the front of the unit. The chamber connects directly to the CPAP generator section via a port at the back of the chamber. The temperature of the heaterplate can be adjusted by the user. Temperature is controlled by software, with the level of humidification delivered dependent on heaterplate temperature, rate of air flow, and ambient conditions of temperature and humidity.

Safety features for the device consist of software monitoring of CPAP generator. humidifier and power supply parameters, electronic monitoring of the microprocessor and a thermal cut-out on the heaterplate.

{1}------------------------------------------------

510(k) Summary continued - Fisher & Paykel HC200 CPAP Humidifier

Controls include a pressure set potentiometer and mains power switch, on the right side of the device, with the mains power cord. A heaterplate adjustment potentiometer, mask test mode button and display LED are located on the front control panel.

The case enclosure components are a polycarbonate thermoplastic, and the heaterplate is cast aluminum. The humidification chamber is a clear thermoplastic top with an aluminum base. Accessories for the HC200 consist of a delivery tube and replacement humidification chambers.

The technological characteristics of the HC200 are equivalent to those of the predicate devices. The devices operate in the same CPAP operating pressure and heaterplate temperature ranges, and have similar features available to the user in both CPAP and humidification sections. The size, materials and technology used in the devices are equivalent, including the power supply, motor, control technology, enclosure components and air filter sections. The CPAP devices use microprocessor control with associated software safety features, and the humidifier devices use the same electronic and thermal safety features. They comply with international device safety standards for electromedical devices including IEC 601-1.

Testing carried out on the HC200 demonstrates that the device has equivalent performance in provision of both constant positive airway pressure and humidity to the two predicate device types.

The HC200 has equivalent pressure drop vs flowrate and dynamic pressure variation characteristics to the predicate CPAP systems. It provides the same levels of temperature and humidity over varying flowrates as the predicate humidifier, at the same heaterplate temperatures.

signed:

Chris Mander Fisher & Paykel Healthcare

te: 7 August 1997

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is simple and monochromatic.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 0 1998

Mr. Chris Mander Fisher & Paykel Electronics Limited 25 Carbine Road Panmure, Auckland NEW ZEALAND

K973161 Re: HC200 CPAP Humidifier Requlatory Class: II (two) Product Code: 73 BZD Dated: May 20, 1998 Received: May 22, 1998

Dear Mr. Mander:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Chris Mander

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Cillehan

Thomas J. Callahan, Ph. D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Fisher & Paykel Healthcare. The words "Fisher & Paykel" are in a large, bold font on the top line, with a line underneath. The word "HEALTHCARE" is in a smaller, non-bold font on the bottom line.

[510(k)] Number:

K97316 1

Fisher & Paykel Electronics Limited
25 Carbine Road, Panmure, Auckland, New Zealand
P.O. Box 14-348, Panmure, Auckland, New Zealand
Tel: +64-9-574-0100 Fax: +64-9-574-0158

6 August, 1997

Fisher & Paykel HC200 CPAP Humidifier

PREMARKET NOTIFICATION 510(k) INDICATIONS FOR USE STATEMENT

The Fisher & Paykel Healthcare HC200 CPAP Humidifier is a combined Continuous Positive Airway Pressure Blower and Heated Respiratory Humidifier (as per 73 BZD, 21 CFR §868.5905 and 73 BTT, 21 CFR §868.5450).

The HC200 is used to assist with patient breathing while sleeping, for the purpose of treating Obstructive Sleep Apnea (OSA). This is by the delivery of Continuous Positive Airway Pressure (CPAP) in order to prevent airway obstruction.

The addition of heated respiratory humidification to the device (normally included as an accessory to CPAP therapy) relieves the drying and irritating effects on the patient airways which may arise from use of a CPAP system.

The HC200 is for use on adult, spontaneously breathing (non ventilator-dependent) patients and is a prescription-use device.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark Kramer

Division of Cardiovascular, Respiratory, and Neurological Devic

510(k) Number K973161

Prescription Use
(Per 21 CFR 801.109)

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).