Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K081327
    Device Name
    RELIMESH
    Manufacturer
    Date Cleared
    2008-06-19

    (38 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K971745, K070625

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Relimesh prostheses are intended for use in the reconstruction of soft tissue deficiencies, such as but not limited to the repair of hernias and chest wall defects

    Device Description

    Relimesh prostheses are dual component meshes made of non-absorbable monofilament polypropylene mesh on one side and a layer of ePTFE on the other side.
    Relimesh is warp knitted in such a way that the mesh may be cut into preshaped designs without unraveling and will maintain excellent isotropic properties because of its knitted construction. The ePTFE layer is heat sealed to the polypropylene layer
    These meshes have the necessary strength, flexibility, and durability for the various stresses which may be encountered in the body.
    The devices are supplied as sterile, single-use surgical meshes.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Herniamesh Relimesh, a surgical mesh intended for the reconstruction of soft tissue deficiencies. The device's acceptance criteria and the study proving it meets these criteria are established through a comparison to legally marketed predicate devices, Composite Mesh (C.R. Bard Inc.) and Hermesh 7 (Herniamesh Srl).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Relimesh device are based on demonstrating substantial equivalence to predicate devices through functional and material characteristic comparisons. The specific acceptance criteria are that the Relimesh device's performance characteristics (suture retention, burst strength, tensile strength) are comparable to or better than the predicate devices, and that its material properties and intended use are similar. The reported device performance is shown in the table below:

    CharacteristicAcceptance Criteria (inferred by predicate)Relimesh PerformancePredicate 1: Hermesh 7 (K070625) PerformancePredicate 2: Composite Mesh (K971745) Performance
    Material/DesignMonofilament Polypropylene & ePTFEPolypropylene & ePTFE100% PolypropylenePolypropylene & ePTFE
    Mesh Weight (g/sq. m)Within historical range of predicates78 ± 6%19214
    Filament Diameter (μ)Within historical range of predicates12080160
    Thickness Mesh (mm)Within historical range of predicates0.5 ± 10%0.27 mm1.5 mm
    % PorosityWithin historical range of predicates86.891.983
    Suture Retention (N)Comparable to or better than predicates24.608.7526 ± 7.56
    Burst Strength (Kpa)Comparable to or better than predicates3783591316 ± 3.8
    Tensile Strength (N/cm)Comparable to or better than predicates24 ± 617.7N/A
    BiocompatibilityCompliant with ISO 10993Compliant--
    SterilitySAL 10^-6, EtO Residual
    Ask a Question

    Ask a specific question about this device

    K Number
    K031925
    Manufacturer
    Date Cleared
    2003-09-17

    (86 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K963619, K994328, K971745

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PROCEED Mesh may be used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

    Device Description

    PROCEED* Surgical Mesh is a sterile mesh designed for the repair of hernias and other fascial deficiencies. The mesh product has separate layers comprising of PROLENE* Soft Mesh, a nonabsorbable polypropylene mesh, a fabric of oxidized regenerated cellulose (ORC) and a polydioxanone polymer film.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for PROCEED* Trilaminate Surgical Mesh

    The provided document describes the predicate equivalence of the PROCEED* Trilaminate Surgical Mesh. The nature of this submission (a 510(k) for a medical device) means that the "acceptance criteria" are primarily related to demonstrating substantial equivalence to a legally marketed predicate device rather than meeting specific performance metrics with predefined thresholds like a diagnostic AI model.

    The study presented focuses entirely on non-clinical laboratory and in-vivo animal testing to demonstrate comparability and intended performance. It does not involve human readers, AI algorithms, or clinical trials with human subjects.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Demonstrated Substantial Equivalence to Predicate)Reported Device Performance (Summary of Findings)
    Technological Characteristics: The device should have similar technological characteristics to the predicate devices.PROCEED* has similar technological characteristics as the predicate devices. The addition of an Oxidized Regenerated Cellulose (ORC) layer to the standard base surgical mesh material is common to other predicate devices (W.L. GORE & Associates, Inc. DualMesh** Biomaterial (K963619), the Genzyme Corp. Sepramesh** Biosurgical Composite (K994328) and the Bard Composix** Mesh (510K K971745), with a performance characteristic of minimizing tissue attachment to the base mesh material. With the addition of the ORC layer, the PROCEED* Mesh continues to meet the functional requirements of a surgical mesh.
    Performance: The device should perform comparably to standard surgical mesh devices indicated for hernia repair and other fascial deficiencies.Non-clinical laboratory testing was performed, demonstrating that the device is comparable to standard surgical mesh devices that are indicated for hernia repair and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result. Additionally, in-vivo testing was provided, showing that the device performed as intended and as claimed.
    Intended Use/Indications: The device's intended use and indications for use should be consistent with predicate devices.The intended use for PROCEED* Mesh is for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result, which is consistent with the predicate device (PROLENE Soft Polypropylene Mesh) and other similar mesh products.

    2. Sample size used for the test set and the data provenance

    The document specifies non-clinical laboratory testing and in-vivo testing.

    • Sample size for test set: Not explicitly stated. The nature of the non-clinical lab and in-vivo testing (which implies animal studies in this context) would typically involve specific numbers of samples or animals, but these are not provided in this summary.
    • Data provenance: Not explicitly stated, but based on the context of a 510(k) submission, this data would primarily be generated by the manufacturer (Ethicon, Inc.) through their own testing facilities or contracted labs. The "in-vivo testing" suggests animal studies were conducted. It is retrospective in the sense that the testing was completed before the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The study does not involve expert readers establishing ground truth for image interpretation or similar diagnostic tasks. The "ground truth" for a surgical mesh device would be derived from physical property measurements and biological responses in animal models, not expert interpretations of data.

    4. Adjudication method for the test set

    This question is not applicable. There is no adjudication method described as the study does not involve human interpretation or consensus for a test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study is for a surgical mesh and does not involve AI assistance, human readers, or diagnostic performance evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone (algorithm only) performance study was not done. This study is for a surgical mesh and does not involve any algorithms or AI.

    7. The type of ground truth used

    The ground truth used for this type of device would be:

    • Benchtop Data: Physical and mechanical properties of the mesh (e.g., tensile strength, pore size, burst strength, pliability) compared to predicate devices and engineering specifications.
    • In-vivo (Animal) Data: Biological response and performance in animal models (e.g., tissue integration, adhesion formation, inflammatory response, degradation profile, functional outcome related to hernia repair).

    8. The sample size for the training set

    This question is not applicable. There is no "training set" in the context of this 510(k) submission for a non-AI surgical device. All testing described falls under verification and validation activities.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set mentioned or implied by the type of device and study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K994328
    Manufacturer
    Date Cleared
    2000-03-02

    (71 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K971745

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sepramesh™ Biosurgical Composite is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias.

    Device Description

    Sepramesh™ Biosurgical Composite (Sepramesh™) is a dual-component (absorbable and non-absorbable), sterile prosthesis designed for the reconstruction of soft tissue deficiencies. Sepramesh™ is constructed of a polypropylene mesh that is coated on one side with a bioresorbable coating composed of sodium hyaluronate (HA) and carboxymethylcellulose (CMC).

    The uncoated side of the mesh allows a prompt fibroblastic response through the interstices of the mesh, encouraging tissue ingrowth similar to polypropylene mesh alone and providing support for soft tissue repair. The HA/CMC side of the mesh provides a hydrophilic bioresorbable coating separating the mesh from underlying tissue and organ surfaces during the critical wound-healing period to minimize tissue attachment to the mesh. Shortly after placement, the HA/CMC coating becomes a hydrated gel that is slowly resorbed from the site of placement within 5-7 days and excreted from the body within 30 days.

    AI/ML Overview

    This 510(k) summary describes a medical device called Sepramesh™ Biosurgical Composite and compares it to legally marketed predicate devices to establish substantial equivalence.

    Here's an analysis of the provided information, framed by your requested criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by demonstrating "substantially equivalent" performance to predicate devices across key characteristics. The performance results are reported as meeting or exceeding these criteria for most aspects.

    Feature / Acceptance CriteriaSepramesh™ Biosurgical Composite PerformanceEquivalence Claim
    IndicationReconstruction of soft tissue deficiencies, such as for the repair of herniasSubstantially Equivalent to predicate devices (Bard® Mesh, Bard® Composix™ Mesh, Mersilene™ Mesh)
    Labeling Claims (Adhesion)HA/CMC surface minimizes tissue and visceral adhesions to deviceSubstantially Equivalent (compared to Bard® Composix™ Mesh's ePTFE claim; Bard® Mesh & Mersilene™ had none)
    Mesh Design (Knit)Single bar knit from 6 mil monofilament polypropylene fiberSubstantially Equivalent to Bard® Mesh
    Mesh Pore SizeSubstantially Equivalent to predicate devicesSubstantially Equivalent
    Burst StrengthSubstantially Equivalent to predicate devicesSubstantially Equivalent
    Suture RetentionSubstantially Equivalent to predicate devicesSubstantially Equivalent
    Tissue IngrowthComplete tissue incorporation of implantSubstantially Equivalent to Bard® Mesh and Bard® Composix™ Mesh
    Tissue Attachment (Adhesions) to MeshDecreased compared to Bard® Mesh and Composix™Substantially Equivalent (or better than predicate where applicable)
    SterilizationGammaSubstantially Equivalent (compared to EtO, Gamma, Steam of predicates)
    Sizes3"x6" to 8"x12"Substantially Equivalent (compared to ranges of predicates)
    Biocompatibility & SafetyNon-toxic, non-mutagenic, non-sensitizing, biocompatible and safeMeets requirements (based on ISO 10993 and GLP)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document mentions "in vivo in a rabbit hernia repair model" for effectiveness testing and comparisons. However, the specific number of rabbits or individual test samples used in this study is not provided.
    • Data Provenance: The in vivo study used a "rabbit hernia repair model." The country of origin for the data is not specified, but the submission is to the US FDA. The study is described as an in vivo comparison, suggesting it was prospective for the purpose of this submission, although the precise nature of the study (e.g., dedicated study for this submission vs. previously conducted data) is not detailed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable or not provided in this 510(k) summary. The "ground truth" for material performance and biological response (like adhesion and tissue ingrowth) in in vivo studies is typically established through direct observation, histology, and quantitative measurements by scientific and medical professionals involved in the study. There's no mention of a panel of experts formally adjudicating the in vivo results in the way it might occur for, e.g., diagnostic imaging.

    4. Adjudication Method for the Test Set

    Not explicitly described. For in vivo studies, data is typically collected, analyzed, and interpreted by researchers and veterinarians. While there are typically standardized protocols and objective measurements, a formal "adjudication method" involving multiple reviewers (like 2+1 or 3+1) for the in vivo results as is common in clinical imaging trials is not mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    No MRMC comparative effectiveness study was done, and AI assistance is not relevant to this device. This device is a surgical mesh, a physical implant, not an AI-powered diagnostic or assistive tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. As stated above, this is a physical surgical mesh, not an algorithm.

    7. The Type of Ground Truth Used

    For the in vivo studies, the ground truth was based on:

    • Direct observation and measurement of biological responses:
      • Adhesion formation (qualitative and quantitative assessment of extent and tenacity).
      • Tissue ingrowth (histological assessment of cellular response and incorporation into the mesh interstices).
      • Biocompatibility assays (e.g., toxicity, mutagenicity, sensitization studies).
    • Mechanical testing results: Burst strength, suture retention.

    These are established through recognized scientific and histological methods for material science and in vivo testing in animal models.

    8. The Sample Size for the Training Set

    Not applicable. This summary describes a physical medical device (surgical mesh), not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1