Search Results
Found 6 results
510(k) Data Aggregation
(121 days)
The Steripath® Micro Blood Collection System is a system to draw blood for in vitro diagnostic testing.
The Steripath® Micro Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood cultures drawn with standard procedure without manual diversion.
Additionally, components of the system may be used for infusion following sample collection after discomection of the Initial Specimen Diversion Device® (ISDD®).
The device, Steripath® Micro Blood Collection System, is a system to draw blood for in vitro diagnostic testing. The purpose of this submission is to expand the product line to include inlet components with the BD Venipuncture® UltraTouch™ Push Button Needle, outlet components with a transfer adapter, and universal models with no inlet component.
The Steripath® Micro Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion.
The device sequesters and diverts 0.5 - 1.0 mL of the initial specimen of blood (potentially contaminated) into a diversion chamber. Once diversion is complete, a subsequent blood sample flows through a second pathway within the device and is collected either directly into a culture bottle (not provided by Magnolia Medical Technologies), or into a syringe that is used to inoculate culture bottles. Additionally, components of the system may be used for infusion after disconnection of the Initial Specimen Diversion Device® (ISDD®).
The Steripath® Micro Blood Collection System, needle configurations, for the subject device are manufactured using non-sterile BD UltraTouch® needles, which are equivalent to the commercially available BD UltraTouch® Push Button™ Blood Collection Set (K212724), except for the sterilization step. The Steripath® Micro Blood Collection System and UltraTouch® needles have the same intended use. Incorporation of the needle with the ISDD® functions on the same technological characteristics and principles of operation with the difference that the subject device leverages the Initial Specimen Diversion Technology® (ISDT) and mechanically sequesters the initial specimen of blood. Inclusion of this technology does not raise any added questions of safety or effectiveness.
The different configurations of the Steripath® Micro Blood Collection System are made available by using different combinations of the inlet and outlet accessories. Proposed configurations are listed in Table 5.1.
The provided text describes a medical device, the Steripath® Micro Blood Collection System, and its 510(k) premarket notification for substantial equivalence to a predicate device. The information details the device's characteristics and various types of testing performed to demonstrate its safety and effectiveness.
Here's an analysis of the acceptance criteria and the study information as requested:
1. Table of Acceptance Criteria and Reported Device Performance
| Requirement | Description | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|---|
| Sequestration | The ISDD® shall sequester the diversion volume. | Effective sequestration of the diversion volume. | PASS |
| Minimum Vacuum performance. | The ISDD® shall meet its performance requirements under minimum vacuum conditions. | Performance within specifications under minimum vacuum. | PASS |
| Maximum Vacuum performance. | The ISDD® shall meet its performance requirements under maximum vacuum conditions. | Performance within specifications under maximum vacuum. | PASS |
| Diversion Volume. | The ISDD® shall meet the minimum and maximum diversion volume requirements. | Diversion volume between 0.5mL and 1.0mL. | PASS |
| Positive Pressure Maximum | The ISDD® shall remain functionally intact and safe under maximum positive pressure conditions. | Remaining functionally intact and safe under maximum positive pressure. | PASS |
| Sterilization | Compliance with gamma radiation process. | Sterility Assurance Level (SAL) of 10^-6 at 25kGy dose. | Complies |
| Aging/Shelf-Life (12 months) | Maintenance of functional properties over time. | Functional (needle insertion, blood collection, needle retraction, hub attachment, culture bottle septum puncture, blood collection, removal), diversion volume, and pressure rating maintained. | Validated to 12 months |
| Biocompatibility | Meets requirements for biological evaluation. | Compliance with ISO 10993-1 for indicated contact duration. | Meets requirements |
| Packaging Integrity / Shipping | Packaging maintain integrity during transport. | Compliance with ASTM D4169-16, Distribution Cycle 13, Assurance Level II; seal strength, package leaks, drop tests met. | Meets requirements |
| Functional (General) | Meets all functional and performance requirements for safe and effective performance. | Safe and effective performance for intended use. | Successfully meets all |
2. Sample Size Used for the Test Set and Data Provenance
The provided text does not explicitly state the sample sizes used for the individual performance tests (sequestration, vacuum, diversion volume, etc.). It describes the types of tests performed (e.g., Sterilization, Aging/Shelf-Life, Biocompatibility, Packaging Integrity, and Performance Testing) and states that "all design verification and validation activities were performed."
The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective data. The testing appears to be primarily laboratory-based engineering and performance verification testing rather than large-scale clinical studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the text. The type of testing described (e.g., sterilization, biocompatibility, mechanical performance) typically relies on established standards and laboratory protocols rather than human expert interpretation for "ground truth" in the way a diagnostic imaging study would.
4. Adjudication Method for the Test Set
This information is not applicable or not provided. Since the testing focuses on objective mechanical and biological properties against established standards, an adjudication method for conflicting expert opinions is not relevant in the context of the described tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done based on the provided text. The submission focuses on device characteristics and performance verification for substantial equivalence, not comparative effectiveness of human readers with or without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
No, this device is a physical medical device (blood collection system), not an AI algorithm. Therefore, "standalone" algorithm performance is not applicable.
7. The Type of Ground Truth Used
The ground truth used for the device's performance is based on established industry standards and engineering specifications for medical devices. For example:
- Sterilization: Standards like ISO 11137-1.
- Aging/Shelf-Life: Standards like ASTM F1980-16.
- Biocompatibility: Standards like ISO 10993-1.
- Packaging: Standards like ASTM D4169-16.
- Functional requirements (e.g., diversion volume, vacuum performance) are based on the device's design specifications.
8. The Sample Size for the Training Set
This device is not an AI/ML algorithm that requires a training set. Therefore, this information is not applicable.
9. How the Ground Truth for the Training Set Was Established
This device is not an AI/ML algorithm. Therefore, this information is not applicable.
Ask a specific question about this device
(210 days)
The Steripath® Micro Blood Collection System is a system to draw blood for in vitro diagnostic testing.
The Steripath® Micro Blood Collection System is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to reduce the frequency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion.
Additionally, components of the system may be used for infusion following sample collection after disconnection of the Initial Specimen Diversion Device® (ISDD®).
The Steripath® Micro Blood Collection System diverts and sequesters the initial portion of the blood specimen (potentially contaminated blood) in the diversion reservoir. When diversion is complete, a subsequent blood sample flows through a second pathway within the device. The subsequent blood sample is collected into a syringe that is used to inoculate culture bottles. Upon removal of the Initial Specimen Diversion Device®, ISDD®, components of the system can be used for infusion per the included manufacturer's instructions for use.
The subject device incorporates multiple syringe oulet accessory configurations that include and three sizes of syringes that are previously cleared as referenced below. The Steripath® Micro Blood Collection System are available:
| Steripath®Micro KitModelNumber | ISDD® | Inlet Accessory | Outlet Accessory |
|---|---|---|---|
| 4005-EN | P00353 | Luer Extension, 7"ICU Medical, Inc.Model B1754-NSK964435 | Syringe, 5mlBecton DickinsonModel 301027K980987 |
| 4010-EN | P00353 | Luer Extension, 7"ICU Medical, Inc.Model B1754-NSK964435 | Syringe, 10mlBecton DickinsonModel 301029K980987 |
| 4020-EN | P00353 | Luer Extension, 7"ICU Medical, Inc.Model B1754-NSK964435 | Syringe, 20mlBecton DickinsonModel 301031K980987 |
The provided document is a 510(k) summary for the Steripath® Micro Blood Collection System (K200661). It details the device's indications for use, technology, comparison to a predicate device, and performance testing. However, it does not contain specific acceptance criteria, reported device performance metrics (e.g., sensitivity, specificity, accuracy, or contamination rates), sample sizes for test sets, data provenance, information about expert ground truth, adjudication methods, MRMC study details, or standalone performance metrics for the aspects related to the "acceptance criteria and study that proves the device meets the acceptance criteria" as requested in the prompt.
The "Summary of Performance Testing" section lists some requirements and states "PASS" for each, but this is a high-level summary and doesn't provide the detailed numbers, methodologies, or study specifics that would typically characterize a study proving device performance against acceptance criteria. It primarily focuses on comparing the new device to its predicate and ensuring it meets safety and functionality standards.
Therefore, many aspects of your request cannot be fulfilled by the provided document.
Here's a breakdown of what can and cannot be extracted:
1. Table of Acceptance Criteria and Reported Device Performance:
The document mentions several requirements and functional tests:
- Sequestration
- Minimum Vacuum performance
- Maximum Vacuum performance
- Diversion Volume
- Positive Pressure Maximum
For each of these, the "Verification Test Result" is reported as "PASS". However, the actual acceptance criteria (e.g., minimum percentage of sequestration, specific vacuum pressure ranges, specific diversion volume ranges) are not detailed in this document. The "PASS" result indicates the device met unstated criteria.
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified for the listed performance tests.
- Data Provenance: Not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not provided. The performance tests described (e.g., sequestration, vacuum performance) are technical/engineering tests, not clinical evaluations requiring expert interpretation of ground truth in the context of diagnostic accuracy.
4. Adjudication method for the test set:
- Not applicable/Not provided. These are technical tests, not diagnostic interpretations.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a blood collection system, not an AI diagnostic device that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a mechanical blood collection system, not an algorithm.
7. The type of ground truth used:
- For the listed performance tests (Sequestration, Vacuum performance, Diversion Volume, Positive Pressure Maximum), the "ground truth" would be established by the physical and functional specifications of the device, measured by engineering testing methods. It is not expert consensus, pathology, or outcomes data in a clinical diagnostic sense. The document states "all design verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met."
8. The sample size for the training set:
- Not applicable. This is a mechanical device, not an AI/machine learning model that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable.
In summary, the provided document focuses on substantiating equivalency to a predicate device through general functional and safety testing, rather than presenting detailed clinical performance study results against specific, quantitatively defined acceptance criteria for diagnostic accuracy or contamination rates with detailed study methodologies. The "PASS" results for functional requirements confirm the device meets its design specifications but lacks the granular data requested in your prompt.
Ask a specific question about this device
(193 days)
The Steripath® Gen2 Blood Collection System is a system to draw blood for in vitro diagnostic testing.
The Steripath® Gen2 Blood Collection is indicated for use as a blood collection system that diverts and sequesters the initial specimen prior to collection of a subsequent test sample to requency of blood culture contamination when contaminants are present in the initial blood sample compared to blood cultures drawn with standard procedure without manual diversion.
Additionally, components of the system may be used for infusion following sample collection after disconnection of the Initial Specimen Diversion Device® (ISDD®). Venipuncture needles are indicated for short term infusion (less than 2 hours).
The Steripath® Gen2 Blood Collection System diverts and sequesters the initial portion of the blood specimen (potentially contaminated blood) in the diversion reservoir. When diversion is complete, a subsequent blood sample flows through a second pathway within the device. The subsequent blood sample is collected either directly into a culture bottle (not provided by Magnolia Medical Technologies), or into a syringe that is used to inoculate culture bottles. Upon removal of the ISDD®, components of the system can be used for infusion per the included manufacturer's instructions for use (note: infusion with butterfly needles is limited to less than 2hrs). The subject device incorporates multiple configurations that include various inlet and outlet accessories that are previously cleared as referenced below.
The Steripath® Gen2 Blood Collection System is a single use, sterile, mechanical device that diverts and sequesters the initial 1.5mL to 2.0mL of blood from the patient. The system consists of an Initial Specimen Diversion Device® (ISDD®) made of injection molded, medical grade plastics. Off-the-Shelf (OTS) components provide the interface to the patient vasculature, and to the culture bottle or syringe for subsequent sample collection. Upon removal of the ISDD®, components of the system can be used for infusion per the included manufacturer's instructions for use (note: infusion with butterfly needles is limited to less than 2hrs).
Here's an analysis of the provided text to extract information about the acceptance criteria and the study proving the device's performance, structured as requested.
The provided document describes a 510(k) premarket notification for the "Steripath® Gen2 Blood Collection System." This type of submission primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving enhanced effectiveness through a primary clinical endpoint study as might be required for a PMA. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are geared towards showing functional equivalence, safety, and a reasonable expectation of effectiveness, rather than a direct clinical performance study against a defined clinical endpoint for de novo approval.
The primary claim of effectiveness for the Steripath® Gen2 is its ability to reduce the frequency of blood culture contamination. While clinical studies are summarized, the core of the FDA's acceptance for a 510(k) clearance hinges on functional and performance testing, and the demonstration that the new feature (initial specimen diversion) does not raise new questions of safety or effectiveness.
Acceptance Criteria and Reported Device Performance
The "acceptance criteria" for the device are largely derived from its functional requirements and safety standards. The reported device performance indicates that the device passed all these verification tests.
1. Table of Acceptance Criteria and the Reported Device Performance:
| Acceptance Criteria (Requirement) | Description | Reported Device Performance (Verification Test Result) |
|---|---|---|
| Unidirectional movement | Operation of the ISDD® actuator shall result in unidirectional movement. | PASS |
| Diversion state negative pressure | In the diversion state, the ISDD® shall generate negative pressure in the diversion chamber and inlet flow path. | PASS |
| Minimum diversion volume | The ISDD® shall meet the minimum diversion volume requirement. | PASS |
| Diversion compliance | The ISDD shall sequester the diversion volume prior to opening the second sample path. | PASS |
| Fully actuated blood collection | When fully actuated the ISDD shall allow flow through the second sample path. | PASS |
| Actuation Lock | When fully actuated, the ISDD® shall lock-out travel of the actuator. | PASS |
| Actuation force, maximum | The ISDD® shall require less than the maximum force to actuate. | PASS |
| Actuation, blocked inlet | With the inlet blocked, the ISDD® shall remain safe during operation. | PASS |
| Winged needle accessory | The Steripath® Gen2 Blood Collection System shall be supplied with commercially available, sharps-safe, winged, hypodermic needle assembly. | PASS |
| Backflow prevention | The ISDD® shall not be operable in a manner that allows blood towards patient. | PASS |
| Sterilization | The system is sterilized using validated Ethylene Oxide (EO) processes in conformance with ANSI/AAMI/ISO 11135:2014. | Conforms (validated) |
| Aging/Shelf Life Test | The system is validated to achieve a real-time 1-year shelf-life, with Accelerated Aging performed in conformity with ASTM F1980-16. | Conforms (validated, 1-year shelf-life) |
| Biological Safety (Biocompatibility Tests) | The system meets the requirements of ANSI/AAMI/ISO 10993-1:2009/(R)2013 for a short duration (<24hrs), blood path indirect, contacting device. Testing included Cytotoxicity, Sensitization, Irritation (intracutaneous reactivity), Acute System Toxicity, and Hemocompatibility. | Conforms (passed tests) |
| Packaging Integrity Testing / Shipping Tests | The system meets the requirements of ASTM D4169-16, Distribution Cycle 13, Assurance Level II. | Conforms (passed tests) |
Study Proving the Device Meets Acceptance Criteria:
The document describes "Functional and Performance Testing" which includes the "Key Functional and Performance Requirements" listed in the table above. It states: "The system meets its functional requirements for safe and effective performance as noted below." All listed verification test results are "PASS."
Additionally, the submission references clinical testing to support the device's efficacy claim regarding contamination reduction, although the document does not explicitly state these clinical results as "acceptance criteria" for the 510(k) clearance itself, but rather as supportive evidence for the indication for use related to contamination reduction. These are presented as summaries of human studies:
| Study | Institution | Total Samples Collected | Samples Collected Using Steripath® | Reduction in Contamination % Using Steripath® |
|---|---|---|---|---|
| Rupp | U. of Nebraska Medical Center (UNMC) | 1,808 | 904 | 87.6% |
| Bell | Lee Health | 41,685 | 6,293 | 82.8% |
No details are given on specific statistical acceptance criteria for these clinical studies within this summary document (e.g., minimum percentage reduction or hypothesis testing results), as the primary focus for a 510(k) is often substantial equivalence based on safety and functional performance.
Additional Information for the Study
The request includes several points typically associated with AI/ML model validation studies. The provided document concerns a mechanical blood collection device and not an AI/ML system. Therefore, many of these points are not applicable or directly addressed in the provided FDA 510(k) summary for a non-AI device. I will address them to the extent possible, noting when a point is not applicable.
2. Sample size used for the test set and the data provenance:
- Functional/Performance Testing (Test Set): The document states "The Steripath® Gen2 Blood Collection System has been found to conform to its System, Labeling, Controls, Interfaces, Accessory, Functional, Physical, Biological Safety and Packaging requirements." and "functional requirements for safe and effective performance." The specific number of devices tested for each functional requirement (e.g., for "Unidirectional movement" or "Actuation force") is not specified in this summary document.
- Clinical Testing (Test Set):
- Rupp Study: 1,808 total samples, 904 collected using Steripath®.
- Bell Study: 41,685 total samples, 6,293 collected using Steripath®.
- Data Provenance:
- Functional/Performance: Implied to be internal testing by Magnolia Medical Technologies, Inc.
- Clinical: Retrospective or prospective status is not explicitly stated in this summary, but the studies were conducted at U. of Nebraska Medical Center (UNMC) and Lee Health (USA, implicitly, as it's an FDA submission). The Rupp study is identified as a "company sponsored study."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not Applicable (N/A) for this device. This pertains typically to AI/ML image analysis, where ground truth often requires expert annotation. For a mechanical blood collection device, "ground truth" for contamination would typically be laboratory culture results, not expert consensus on images.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- N/A. Again, this is typically for AI/ML studies involving human reader review. The clinical studies here involve laboratory results (blood culture contamination rates), which inherently establish their 'truth' through a standardized, objective process.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This is a mechanical device, not an AI assistance tool for human readers. The clinical studies compared the device's performance against standard procedure without manual diversion, effectively comparing a new collection method to an older one, not human reader performance with or without AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- N/A. This device does not have an "algorithm only" mode; it is a physical medical device used for blood collection.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the functional/performance requirements, the "ground truth" is adherence to engineering specifications and relevant standards.
- For the clinical studies, the "ground truth" for contamination reduction would be laboratory blood culture results (positive for contaminants vs. negative), which constitutes clinical outcome data relevant to the device's intended use.
8. The sample size for the training set:
- N/A. As a physical device, there is no "training set" in the context of machine learning. The device design and manufacturing process are developed through engineering and design controls, not data training.
9. How the ground truth for the training set was established:
- N/A. Same reason as above.
Ask a specific question about this device
(100 days)
To administer IV fluids/medication to a patient's vascular system.
The Victus I.V. Administration and Extension Sets are single use, sterile, non-pyrogenic devices used to administer filtered intravenous solutions and/or nutritional solution to a patient's vascular system via a catheter venous site under gravity-controlled flow. When used, the purpose of the filter is for filtering solution during intravascular administration. The device may include a needle-free valve injection site/connector which eliminates the use of needle to access the set during IV administration and aids in the prevention of needlestick injuries.
This document describes a 510(k) premarket notification for Victus Intravascular (I.V.) Administration Sets and Victus I.V. Extension Sets. The submission details changes and additions to the product line, including substituting a Flow Regulator for a Roller Clamp, replacing PVC tubing, substituting a Needle-Free connector, and adding a filter and check valve to some configurations. The document asserts that these changes do not affect the safety and effectiveness of the devices.
The information provided is primarily focused on the declaration of substantial equivalence to predicate devices and the scope of the device's intended use. It does not include detailed acceptance criteria or a study proving the device meets specific performance criteria in the format requested. The only mention of testing is "The VICTUS IV Extension Set and IV Administration Set have undergone Mechanical testing to verify performance," without detailing the tests, criteria, or results.
Therefore, most of the requested information cannot be extracted from this document.
Here's what can be extracted based on the provided text, with significant gaps for the unavailable information:
1. A table of acceptance criteria and the reported device performance
Unfortunately, the provided text does not contain a table of acceptance criteria or specific reported device performance metrics beyond the statement "The VICTUS IV Extension Set and IV Administration Set have undergone Mechanical testing to verify performance." No quantitative results or predefined thresholds are mentioned.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample size used for any mechanical testing or the data provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable and not present as this is a mechanical medical device, not an AI or diagnostic imaging device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable and not present. No adjudication method is mentioned for any testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable as this is a mechanical medical device and not an AI-assisted diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable as this is a mechanical medical device and does not involve an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For mechanical devices, "ground truth" typically refers to established engineering standards or regulatory requirements that the device must meet (e.g., flow rate, tensile strength, sterility). The document states that the devices underwent "Mechanical testing to verify performance" and that the "Technological characteristics of the Victus Sets are substantially equivalent to the referenced predicates." This implies that the performance was compared to established standards for similar predicate devices, though the specific 'ground truth' metrics are not detailed.
8. The sample size for the training set
This information is not applicable and not present as this is a mechanical medical device, not an AI or machine learning device.
9. How the ground truth for the training set was established
This information is not applicable and not present as this is a mechanical medical device, not an AI or machine learning device.
Ask a specific question about this device
(152 days)
Universal Spikes are used to build Intravascular Administration Sets which are Single Use, sterile, non- pyrogenic device and provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter inserted into a vein.
Universal Spikes are used to build Intravascular Administration Sets which are Single Use, sterile, non- pyrogenic device and provides access for the administration of fluids from a controliner to a patient's vascular system through the administration set's needle or catheter inserted into a ven. Each set will be manufactured to unique specifications using dimensions, components, and configurations specified by the customer. Standard sets will also be offered. Components used in the IV sets may be either manufactured by ICU medical or purchased from other manufacturers. Components will be assembled into configurations specified by the customer and packaged.
The provided text is a 510(k) summary for the ICU Medical Universal Single-Use Spikes. It states that the device is substantially equivalent to a predicate device (K964435 - IV Sets- ICU Medical, Inc.) based on having the same intended use and materials.
However, the document does not contain any information regarding specific acceptance criteria, performance studies, sample sizes, ground truth establishment, expert qualifications, or comparative effectiveness studies (MRMC or standalone).
The document focuses on establishing substantial equivalence for regulatory purposes rather than describing a detailed performance study with acceptance criteria. It mentions "Safety and Performance" and states that "ICU Medical Universal Single-Use Spikes conform to the requirements of published international standards as well as those FDA recognized standards and/or published guidelines prior to marketing the device." This implies adherence to general standards relevant to such medical devices, but it does not specify quantitative acceptance criteria or the studies used to demonstrate compliance to them in this document.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and the reported device performance
- Sample size used for the test set and the data provenance
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Adjudication method for the test set
- MRMC comparative effectiveness study results
- Standalone performance study results
- Type of ground truth used
- Sample size for the training set
- How the ground truth for the training set was established
This information is typically found in detailed testing reports, design validation documentation, or other sections of a 510(k) submission, but it is not present in the provided "Summary of Safety and Effectiveness."
Ask a specific question about this device
(33 days)
To administer IV fluids/medication to the patient's vascular system through a needle-free system that aids in the elimination of needle-stick injury.
The Victus I.V. Administration Sets are single use, sterile, non-pyrogenic devices used to administer I.V. fluids/medication to a patient's vascular system via gravity control.
The provided text describes a 510(k) premarket notification for the Victus IV Administration Sets. However, it does not contain the detailed acceptance criteria, study design, or performance data that you've requested regarding device performance. The document explicitly states:
"The Victus I.V. Administration Sets have undergone performance and safety testing to verify mechanical properties and biocompatibility using FDA recognised standards."
This indicates that testing was performed, but the results of that testing (i.e., acceptance criteria and reported performance) are not included in this document. Instead, this document is a summary for a 510(k) submission, confirming that the device is substantially equivalent to predicate devices. Substantial equivalence means it has the same intended use and similar technological characteristics, and any differences don't raise new questions of safety or effectiveness.
Therefore, I cannot populate the table or answer the specific questions about "the study that proves the device meets the acceptance criteria" using the provided text. The document focuses on regulatory approval based on substantial equivalence, not on a detailed clinical or performance study report.
Here's what I can extract based on the limited information:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not provided in the document | Not provided in the document |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not provided. The document states "performance and safety testing" was done, but gives no details about the sample size, type of test set, or data provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable/Not provided. This type of information is relevant for studies involving human interpretation (e.g., medical imaging AI). For an IV administration set, "ground truth" would typically be established through engineering specifications, material science testing, and biological assays, not expert consensus on interpretations. No details on specific experts or their qualifications are mentioned.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not provided. Adjudication methods are typically used when multiple human experts provide opinions that need to be reconciled, such as in clinical studies evaluating diagnostic accuracy. This is not mentioned or relevant for the type of device and testing described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI device. It is a traditional medical device (IV administration set).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Inferred: For this type of device, ground truth would be established by engineering specifications, material properties, biocompatibility standards, and functional performance benchmarks (e.g., flow rates, leak integrity, particulate matter, pyrogenicity). The document mentions "FDA recognised standards" were used, implying these types of criteria formed the "ground truth." Specific details are not provided.
8. The sample size for the training set:
- Not applicable/Not provided. This is not an AI device that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable. This is not an AI device.
In summary: The provided 510(k) summary is a regulatory document focused on demonstrating substantial equivalence to pre-existing devices, rather than a detailed scientific study report outlining specific performance criteria and test results. It confirms that "performance and safety testing" was conducted using "FDA recognized standards," but the specifics of those tests and their outcomes are not included in this document.
Ask a specific question about this device
Page 1 of 1