K Number
K970855

Validate with FDA (Live)

Device Name
CLAVE CONNECTOR
Manufacturer
Date Cleared
1997-06-24

(106 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration set for the administration of fluids to a patient through a cannula placed in the vein or artery.

Device Description

The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to Intravascular administration set.

AI/ML Overview

This submission for the ICU Medical Inc. CLAVE CONNECTOR is a 510(k) premarket notification, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific performance criteria. Therefore, the provided text does not contain the detailed information necessary to complete most of the requested sections about acceptance criteria and a study.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

This information is not provided in the 510(k) summary. 510(k) submissions primarily focus on demonstrating substantial equivalence to a legally marketed predicate device, rather than detailed performance against acceptance criteria in the way a new efficacy study might.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided. The document outlines a comparison to a predicate device, not a performance study with a test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. Ground truth establishment, if any, would be relevant for a de novo or PMA submission, not a 510(k) in this format.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not provided. The device is an "Intravascular Administration Set" and not an AI-powered diagnostic device, so an MRMC study is not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not provided. As mentioned, this is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not provided.

8. The sample size for the training set

This information is not provided. There is no mention of a "training set" as this is not an AI/ML device.

9. How the ground truth for the training set was established

This information is not provided.


Summary of what is present:

The document is a 510(k) Summary of Safety and Effectiveness for the ICU Medical Inc. CLAVE CONNECTOR. It demonstrates substantial equivalence to the IVAC SmartSite™.

  • Device Description: A single-use, sterile, non-pyrogenic device intended as an accessory to intravascular administration sets for fluid administration.
  • Intended Use: As an accessory to intravascular administration set for the administration of fluids to a patient through a cannula placed in the vein or artery.
  • Comparison to Predicate Device (IVAC SmartSite™):
    • Product Labeling: Both are sterile, non-pyrogenic (for CLAVE), and include directions for use.
    • Intended Use: CLAVE - Needleless injection port to access any vein or artery. IVAC - Needleless valve (luer lock). These appear equivalent in function.
    • Design: Both are one-piece designs activated by luer connection to allow fluid flow.
    • Materials: Specific materials are listed for the CLAVE (Polycarbonate, Polyester, Silicone Rubber, Polypropylene, Flurosilicone), while the predicate's materials are "Unknown" in this comparison table. This implies that the materials chosen for the CLAVE were deemed acceptable.

The FDA's letter (K970855) indicates that the device was found substantially equivalent to devices marketed prior to May 28, 1976, or to reclassified devices. This decision assumes compliance with Good Manufacturing Practice (GMP) regulations.

Conclusion:

The provided text is a 510(k) premarket notification. It focuses on comparison to a predicate device to establish substantial equivalence for market clearance, rather than detailing a study that directly proves the device meets specific performance acceptance criteria using quantitative metrics from a test set. Therefore, most of the requested fields regarding acceptance criteria, study details, and AI-specific information cannot be filled from the given text.

{0}------------------------------------------------

ICU Medical Inc. CLAVE CONNECTOR

JUN 24 1997

510(k) Notification Page 21 of 24

K970855

Appendix H. 510(k) Summary of Safety and Effectiveness 17.

The following section is included as required by the Safe Medical Device Act (SMDA) of 1990.

Name:ICU Medical, Inc.
Address:951 Calle AmanecerSan Clemente, CA 92673

Contact Person: Arlene Dutchik Phone Number: (714)366-2183 Fax Number: (714)366-8368

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K970855

Applicant Information:

Date Prepared: May 14, 199 Name: ICU Medical Inc. Address: 951 Calle Amanecer San Clemente, CA 92673

Contact Person: Arlene Dutchik Phone Number: (714)366-2183 (714)366-8368 Fax Number:

Device Information:

Trade Name: Clave Connector Common Name: Intravascular Administration Set Classification Name: Intravascular Administration Set

Equivalent Device:

IVAC SmartSite™

Device Description:

The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to Intravascular administration set.

Intended Use:

As an accessory to intravascular administration set for the administration of fluids to a patient through a cannula placed in the vein or artery.

{1}------------------------------------------------

ICU Medical Inc. CLAVE CONNECTOR

i.

510(k) Notification Page 22 of 24

Comparison To Predicate Device:

Characteristic ComparedCLAVE® NEEDLELESS CONNECTORIVAC NEEDLE FREE VALVE ADMINISTRATION SETS (SMARTSITE™)
Product LabelingSterile Use, Non-Pyrogenic, Sterile fluid pathway in unopened, undamaged package. Directions for use on labeling.Sterile Use, Directions for use on labeling
Intended UseNeedleless injection port to access any vein, or artery.Needleless valve (luer lock)
DesignOne piece design activated by luer connection to allow fluid flow.One piece design activated by luer connection to allow fluid flow.
MaterialsInternal Conduit-PolycarbonateHousing-PolyesterSilicone Seal-Silicone RubberRing-PolypropyleneLubricant-FlurosiliconeBreather Cap-PolypropylenePackaging-Medical packaging grade fiber-free peelable paper lidding and pouching material.Unknown

.. ...

:

Aline Antchik

Arlene Dutchik Regulatory Affairs Specialist ICU Medical, Inc.

نر

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The seal is black and white and appears to be a scanned image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Arlene Dutchik Requlatory Affairs Specialist ICU Medical, Incorporated 951 Calle Amanecer San Clemente, California 92673

JUN 24 1997

Re : K970855 Trade Name: Clave Connector Regulatory Class: II Product Code: FPA Dated: May 14, 1997 Received: May 19, 1997

Dear Ms. Dutchik:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਬੋ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Paqe 2 - Ms. Dutchik

This letter will allow you to begin marketing your device as described in your 510 k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.

Sincerely yours,
Patricia Cicerite for

Timothy A. Ulatowski Director

Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

ICU Medical Inc. CLAVE CONNECTOR 510(k) Notification Page 16 of 24

12. Appendix C. Indications for Use

The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration set for the administration of fluids to a patient through a cannula placed in the vein or artery.

(Division Sign-Off) Patricia Crucenta
Division of Dental, Infection Control,
and General Hospital Devices

10 General Hospital Devices
510(k) Number K970855

Prescription Use
(Per 21 CFR 801.109)

05/14/97510KCLAVE-1.doc

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.