Search Results
Found 3 results
510(k) Data Aggregation
(60 days)
The Flow 7000 Peripheral Vascular Coil is a receive-only quadrature-phased array RF coil, used for obtaining diagnostic images of the vasculature and soft tissue anatomy, extending from the abdomen to the foot region in Magnetic Resonance Imaging Systems. The indications for use are the same as for standard MR Imaging. The Flow 7000 Peripheral Vascular Coil is designed for use with the 1.5T Signa MRI scanner manufactured by GE Medical Systems.
The Flow 7000 Phased Array Peripheral Vascular Coil is designed to provide Magnetic Resonance Images of the abdominal aorta, peripheral vasculature and soft tissue anatomy. The Premier 7000 Phased Array Spine Coil is designed for use with the GE MR's Signa 1.5T scanner.
Anatomic Regions: Abdominal Vasculature, Peripheral Vasculature and soft tissue Nuclei Excited: Hydrogen
The indications for use are the same as for standard imaging:
The Signa 1.5T system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The Flow 7000 Peripheral Vascular Coil is a 12 element quadrature, phased array, receive-only coil. The coil consists of two sections: an upper and lower section, which are positioned above and below the patient's body (lower torso and legs). The coil sections which contain the coil elements and accessory coil electronics, is enclosed to prevent any exposure to the patient or environment.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Flow 7000 Peripheral Vascular Coil.
Important Note: The provided document is a 510(k) submission for a medical device (an MRI accessory coil). This type of submission relies on demonstrating substantial equivalence to existing legally marketed devices, rather than a clinical study proving its independent performance against pre-defined acceptance criteria in the way one would for a novel AI algorithm or drug. Therefore, many of the typical acceptance criteria and study elements requested will not be present in this context.
Acceptance Criteria and Device Performance for Flow 7000 Peripheral Vascular Coil (MRI Accessory)
As this is a 510(k) submission for an MRI accessory coil, the "acceptance criteria" are primarily based on demonstrating substantial equivalence to legally marketed predicate devices in terms of intended use, indications for use, safety, and technological characteristics. The "study" proving this involves a direct comparison to these predicates.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Demonstrated through Substantial Equivalence) | Reported Device Performance (Comparison to Predicate) |
---|---|
Intended Use | |
Imaging of vasculature and tissue extending from the torso to the feet. | Meets: Identical Intended Use as the Predicate Device: Peripheral Vascular Coil, Medical Advances Inc. (K964813) - "Imaging of vasculature and tissue extending from the torso to the feet." |
Indications for Use | |
Identical to routine MRI imaging (abdominal aorta, peripheral vasculature, soft tissue anatomy), compatible with GE MRI's Signa 1.5T scanner. | Meets: Identical Indications for Use as the Predicate Device: Peripheral Vascular Coil, Medical Advances Inc. (K964813) - "Identical to routine MRI imaging" (specifically for abdominal aorta, peripheral vasculature, and soft tissue anatomy, compatible with GE MR's Signa 1.5T scanner). The fundamental indications for MR imaging are also stated as "distribution of protons exhibiting NMR signal," "depend upon NMR parameters," and "display the soft tissue structure of the head and whole body," yielding diagnostic information when interpreted by a trained physician. |
Coil Material | |
Flame Retardant ABS/PVC Plastic alloy, Flame retardant Polyurethane Plastic, Flame retardant Vinyl fabric. | Meets (Similar to existing materials): Compared to: Profile 7000 Volume Neck Coil, USA Instruments (K964531); Leo 7000 Quadrature Knee Coil, USA Instruments (K971246); General Purpose Flex Coil, Picker International (K944469). The materials used are standard for MRI coils and have established safety profiles. |
Coil Design | |
Receive-only phased array design. | Meets (Similar to existing designs): Compared to: Insight 7000 Phased Array Torso Coil, USA Instruments (K972340); Phased Array C/T/L Spine Coil, USA Instruments (K980157). The receive-only phased array design is a well-established and safe design for MRI coils. |
Decoupling | |
RF Chokes with Switching Diodes. | Meets (Similar to existing methods): Compared to: Insight 7000 Phased Array Torso Coil, USA Instruments (K972340). This is a standard method for decoupling in MRI coils. |
Prevention of RF Burns | |
Does not transmit RF Power, Decoupling isolates coil elements from RF fields, Coil elements and circuitry enclosed in non-conductive housing. | Meets (Safety features are standard): Compared to: Insight 7000 Phased Array Torso Coil, USA Instruments (K972340). These are fundamental safety design principles for receive-only MRI coils to prevent RF-induced heating and burns. The device's receive-only nature is a primary safety feature. |
Radio Frequency Absorption | |
Coil is a receive-only coil and does not transmit RF power. | Meets (Inherently safe design): Compared to: Insight 7000 Phased Array Torso Coil, USA Instruments (K972340). As a receive-only coil, it does not contribute to SAR (Specific Absorption Rate) heating of the patient. This characteristic is inherently safe regarding RF absorption. |
Formation of Resonant Loops | |
Decoupling isolates coil elements from RF fields, length of cable and stiffness does not permit looping. | Meets (Safety features are standard): Compared to: Insight 7000 Phased Array Torso Coil, USA Instruments (K972340). Design considerations (decoupling, cable length/stiffness) are implemented to prevent the formation of resonant loops which could lead to unsafe heating. |
Study Details (Relevant to 510(k) Substantial Equivalence)
Given the nature of a 510(k) for an MRI accessory, a "study" in the traditional sense of a clinical trial with specific endpoints and statistical power calculations is not typically performed or required in the document. Instead, the "study" is a comparison and justification of equivalence to predicate devices based on design and safety characteristics.
Here's how the requested information maps to this type of submission:
-
Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
- Not applicable in the provided text. There is no "test set" of patient data or images described for evaluating the coil's performance in a clinical study. The evaluation is based on engineering design, material specifications, and comparison to already approved predicate devices.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not applicable. No expert review of a test set is documented. The "ground truth" for a 510(k) submission for a device like this is primarily regulatory acceptance of the safety and performance features as equivalent to existing devices.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. No test set or adjudication process is mentioned.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an MRI coil, not an AI algorithm. Therefore, an MRMC study related to AI assistance is not relevant.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is an MRI coil, not a standalone algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For this 510(k), the "ground truth" for demonstrating safety and effectiveness is the established safety and efficacy profile of the identified predicate devices. The new device's design, materials, and functional principles are compared against these predicates to show that it raises no new questions of safety or effectiveness.
-
The sample size for the training set:
- Not applicable. This device does not involve a "training set" in the context of machine learning or data-driven model development. Its development is based on engineering principles and existing MRI technology.
-
How the ground truth for the training set was established:
- Not applicable. As there is no training set, this question is not relevant.
Summary of the 510(k) Process for this device:
The "study" for the Flow 7000 Peripheral Vascular Coil is a comparative analysis demonstrating that it is substantially equivalent to legally marketed predicate devices. This is a regulatory pathway that allows new devices to be marketed without requiring entirely new clinical trials if they can be shown to be as safe and effective as existing products.
The key aspects of this demonstration are:
- Identical Intended Use and Indications for Use to the primary predicate (Peripheral Vascular Coil, Medical Advances Inc., K964813).
- Similar (or equivalent safe) Technological Characteristics in terms of coil material, design, decoupling mechanisms, and fundamental safety principles (e.g., receive-only, prevention of RF burns, prevention of resonant loops) as compared to various other predicate MRI coils.
The FDA's letter states: "We have determined the device is substantially equivalent... to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified..." This confirms that the manufacturer successfully demonstrated equivalence, which is the "acceptance criteria" for a 510(k) submission.
Ask a specific question about this device
(87 days)
The Premier 7000 Phased Array CTL Spine Coil is a receiveonly phased array RF coil, used for obtaining diagnostic images of the spine region (cervical, thoracic and lumbar anatomy) in Magnetic Resonance Imaging Systems. The indications for use are the same as for standard MR Imaging. The Premier 7000 Phased Array CTL Spine Coil is designed for use with the 1.5T Signa MRI scanner manufactured by GE Medical Systems.
The Premier 7000 Phased Array Spine Coil is designed to provide Magnetic Resonance Images of the spine anatomy. The Premier 7000 Phased Array Spine Coil is designed for use with the GE MR's Signa 1.5T scanner.
Anatomic Regions: Spine Nuclei Excited: Hydrogen
The indications for use are the same as for standard imaging:
The Signa 1.5T system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The Premier 7000 Phased Array CTL Spine Coil is a multi-element phased array receive-only coil. The coil is shaped to conform to the contours of the spine and has an removable anterior section for imaging the anterior cervical region. The elements and associated circuitry are enclosed in a housing made of plastic materials which are fire rate and have high impact and tensile strength.
The provided documentation for the K980157 Premier 7000 Phased Array CTL Spine Coil is a 510(k) premarket notification. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing extensive clinical study data to prove novel performance criteria.
Therefore, the submission does not include a study that proves the device meets specific acceptance criteria in the way a traditional clinical trial or performance study would. Instead, the "acceptance criteria" here are aligned with demonstrating substantial equivalence in terms of intended use, indications for use, materials, design, and safety mechanisms compared to predicate devices.
Here's an analysis based on the provided text, addressing your questions where information is available:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the characteristics of the predicate devices. The "Reported Device Performance" for the Premier 7000 Phased Array Spine Coil is its ability to match or be similar in these characteristics.
Parameter | Acceptance Criteria (Based on Predicate Devices) | Premier 7000 Phased Array Spine Coil Performance (as reported) |
---|---|---|
Intended Use | Imaging of the spine (as in K932693, K911806) | Imaging of the spine. |
Indications for Use | Identical to routine MRI imaging (as in K932693, K911806) | Identical to routine MRI imaging: produces images corresponding to proton distribution, dependent on NMR parameters, displaying soft tissue structure, useful for diagnosis by a trained physician. Anatomic Regions: Spine, Nuclei Excited: Hydrogen. |
Coil Material | Use of materials comparable to those in legally marketed devices (e.g., K964531, K971246, K962677, K972826, K944469) | ABS/PVC Plastic alloy, Polyurethane Plastic, Glass fiber reinforced polyester (Flame retardant Fiberglass), Naughahyde (fabric material) |
Coil Design | Receive-only phased array design (as in K972340, K932693) | Receive-only phased array design |
Decoupling | RF chokes with switching diodes (as in K972340) | RF Chokes with Switching Diodes |
Prevention of RF Burms | Mechanisms to prevent RF burns (as in K972340) | Does not transmit RF Power, Decoupling isolates coil elements from RF fields during RF transmission, Coil elements and circuitry enclosed in non-conductive housing. |
Radio Frequency Absorption | Limited or no RF power transmission for receive-only coils (as in K972340) | Coil is a receive only coil and does not transmit RF power |
Formation of Resonant Loops | Mechanisms to prevent formation of resonant loops (as in K972340) | Decoupling isolates coil elements from RF fields during RF transmission, Length of cable and stiffness does not permit looping. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not applicable in the context of this 510(k) submission. A clinical "test set" with patient data to evaluate performance metrics directly is not described. The evaluation is based on comparing device specifications and technical characteristics to predicates.
- Data Provenance: Not applicable. The submission is a comparison of technical specifications and intended use against existing, cleared devices.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
- Number of Experts: Not applicable. Ground truth, in the sense of expert consensus on image quality or diagnostic accuracy, is not part of this 510(k) submission.
- Qualifications of Experts: Not applicable.
4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set
- Adjudication Method: Not applicable. No test set involving human interpretation and adjudication is described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No. This device is an MRI accessory (a coil), not an AI-powered diagnostic system.
- Effect Size: Not applicable.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
- Standalone Performance Study: No. This is a hardware component with its performance intrinsically linked to the MRI scanner it integrates with, not a standalone algorithm.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: The "ground truth" for this 510(k) submission is the established safety and effectiveness profile of the predicate devices. The applicant asserts that their new device shares the same fundamental scientific technology and intended use, and does not raise new questions of safety or effectiveness compared to these predicates. This is a regulatory "ground truth" rather than a clinical one.
8. The Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This is not an AI/machine learning device requiring a training set.
9. How the Ground Truth for the Training Set was Established
- How Ground Truth for Training Set was Established: Not applicable.
Ask a specific question about this device
(83 days)
The Picker International EDGE and VISTA systems are indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
The EDGE/VISTA system enhancement package includes changes in the computer subsystem, the gradient subsystem, the RF subsystem, the magnet subsystem, the patient handling subsystem, and the magnet enclosure. The main features provided with this enhancement are a 61 cm body coil, three gradient performance levels, water cooled gradients, an increase in the SAR first controlled operating mode limit, and optional increased reconstruction rates.
The provided document is a 510(k) summary for a GE Healthcare product, specifically an enhancement package for their EDGE/VISTA MR systems. This type of submission is for demonstrating substantial equivalence to a predicate device, not for proving that a device meets specific clinical acceptance criteria through a study.
Therefore, the document does not contain the information requested for acceptance criteria and a study proving the device meets those criteria in the way a clinical trial or performance study would. It focuses on comparing the technical specifications and safety parameters of the new device to existing, legally marketed predicate devices.
Here's why the requested information isn't present and what the document does provide:
- No Acceptance Criteria in the context of clinical performance: The document implicitly demonstrates "acceptance" by showing the device's technical specifications and safety profiles are comparable to, or improved upon, predicate devices, indicating it meets general regulatory safety and effectiveness expectations for MR systems. There are no explicit clinical performance metrics (e.g., sensitivity, specificity for a diagnostic task) being tested.
- No "Study" as a traditional clinical trial: The document describes modifications and enhancements to an existing MR system. The "proof" is through a comparative table highlighting substantial equivalence in engineering, safety, and operational features. It's not a study in the sense of a clinical investigation with a test set, ground truth established by experts, etc.
- No details on sample size, data provenance, expert ground truth, adjudication, MRMC, or standalone performance: These concepts are relevant to clinical performance studies, which are not detailed in this 510(k) summary for a system enhancement.
However, I can extract information related to the technical comparisons made for demonstrating "substantial equivalence," which serves as the "proof" in this context:
1. Table of "Acceptance Criteria" (Implicit Safety/Performance Parameters) and Reported Device Performance
Since there are no explicit "acceptance criteria" for clinical performance, I will present the key technical and safety specifications that are being compared to demonstrate substantial equivalence, considering these as the "criteria" for regulatory acceptance in this context.
Item/Criterion (Implicit) | Predicate Devices (EDGE/VISTA K931544, K954646) Performance | EDGE/VISTA with System Enhancement Package Performance | Comparison Outcome (Proving "Acceptance") |
---|---|---|---|
Computer Subsystem | Dedicated Ethernet, single monitor, second Ethernet for external network. | Similar display/database computer, monitor, Ethernet. DICOM v3.0 compliant. Optional increased reconstruction rates. | Substantially Equivalent; added modern communication (DICOM) and optional performance improvements. |
Image Storage (Short Term/Archival) | Magnetic Disk/Optical Disk | Same | Substantially Equivalent. |
Gradient Coils | Air-cooled (Standard EDGE) or Water-cooled (High Performance Gradients) | Water-cooled self-shielded for all performance levels. | Improved/Equivalent; standardizes to the higher performance, water-cooled model. |
Max. Gradient Strength | 16 or 27 mT/m | 16, 20 or 27 mT/m | Improved/Equivalent; adds an intermediate gradient strength option. |
Max. Slew Rate | 25 or 72 mT/m/msec | 25, 40, 72 mT/m/msec | Improved/Equivalent; adds an intermediate slew rate option. |
Gradient Amps | 135 A RMS, 220 A peak, 350 V OR 270 A RMS, 440 A peak, 600 V | 150 A RMS, 300 A peak, 400 V OR 270 A RMS, 440 A peak, 600 V | Improved/Equivalent; slight increase in the lower-tier amp values. |
RF Amplifiers | Single frequency RF amplifiers. | Single frequency RF amplifiers with increased output power (for larger body coil). | Improved/Equivalent; adjusted for new 61cm body coil maintain equivalent B1 field. |
Body Coil | 58 cm quadrature multi-conductor transmit/receive coil. | 61 cm quadrature multi-conductor transmit/receive coil. | Improved; larger body coil for potentially greater patient comfort/versatility. |
Patient Handling | Computer controlled patient transport. | Computer controlled patient transport with increased weight capacity. New patient illumination. | Improved; increased capacity and comfort features. |
Time Varying Magnetic Field (dB/dt) | Standard EDGE: ≤ 20 T/s. High Performance: Normal ≤ 40 T/s, Controlled 40-60 T/s. | Normal Operating Mode: ≤ 40 T/s. First Controlled Operating Mode: **40 T/s |
Ask a specific question about this device
Page 1 of 1