510(k) Data Aggregation
(264 days)
Total Hip Arthroplasty and Hemi Hip Arthroplasty
Total hip replacement or hip arthroplasty is indicated in the following conditions:
- A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
- Avascular necrosis of the femoral head.
- Acute traumatic fracture of the femoral head or neck.
- Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemi-arthroplasty, surface replacement arthroplasty, or total hip replacement.
- Certain cases of ankylosis.
Partial hip replacement or hip hemi-arthroplasty is indicated in the following conditions:
- Acute fracture of the femoral head or neck that cannot be appropriately reduced and treated with internal fixation.
- Fracture dislocation of the hip that cannot be appropriately reduced and treated with internal fixation.
- Avascular necrosis of the femoral head.
- Non-union of femoral neck fractures.
- Certain high subcapital and femoral neck fractures in the elderly.
- Degenerative arthritis involving only the femoral head in which the acetabulum does not require replacement.
- Pathology involving only the femoral head/neck and/or proximal femur that can be adequately treated by hip hemi-arthroplasty.
HA coated stems of the Corail Hip System are indicated for cementless use only.
The Corail AMT Hip is a tapered stem available both collarless and collared. The stems are manufactured from titanium alloy (Ti6Al4V) and plasma-sprayed with a biocompatible hydroxyapatite (HA) coating for bone fixation. The Corail AMT stems feature a 12/14 modular taper that accepts 12/14 heads with a wide range of diameters. The compatible components for the hip hemi-arthroplasty application are bipolar heads cleared via 510(k) K812672 and unipolar heads cleared via 510(k) K903084. Corail AMT stems are indicated for cementless use only.
The provided document describes the Corail AMTTM Hip Prosthesis, an orthopaedic device. The submission is a 510(k) to expand the cleared indication for use to include hip hemi-arthroplasty surgical application.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria Category | Specific Test/Characteristic | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Material | Composition | Ti-6Al-4V | Ti-6Al-4V (Same as predicate) |
Stem Surface | Coating for bone fixation | Plasma-sprayed with biocompatible hydroxyapatite (HA) coating | Plasma-sprayed with biocompatible hydroxyapatite (HA) coating (Same as predicate) |
Compatible Femoral Heads | Types of heads accepted | Bipolar and Unipolar | Bipolar and Unipolar (Same as predicate) |
Taper Design | Modular Taper | 12/14 taper | 12/14 taper (Same as predicate) |
Placement | Method of implantation | Press Fit: Cementless | Press Fit: Cementless (Same as predicate) |
Performance Data | Validation of taper/heads dimensional compatibility | Implied to perform acceptably for mechanical mating | Performed as "Validation of the taper/heads dimensional compatibility with Corail AMT Stems" |
Performance Data | Fatigue Resistance | Implied to withstand anticipated physiological loads and cycles | Performed as "Fatigue Resistance tests for the Corail AMT Stems" |
Clinical Equivalence (for expanded indication) | Safety and effectiveness for hip hemi-arthroplasty | Substantial equivalence to predicate devices and supported by existing clinical literature | Supported by a systematic literature review of Corail Hip System for femoral neck fractures, demonstrating no fundamental change in scientific technology and consistency with established safety and effectiveness. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Non-clinical (Performance) Tests: The document does not specify the exact sample sizes (number of devices tested) for the "Validation of the taper/heads dimensional compatibility" or the "Fatigue Resistance tests." It only states that these tests were performed. Given the nature of these tests, they would typically involve a defined number of test articles to meet industry standards (e.g., ISO or ASTM standards for implant testing).
- Clinical Evaluation: No new clinical tests were conducted. The clinical evaluation utilized a "systematic literature review of Corail Hip System for the treatment of femoral neck fractures." The provenance of the data in this systematic review would depend on the included studies, but the document does not provide details on country of origin or whether these studies were retrospective or prospective. It is a compilation of retrospective existing scientific literature.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This submission relies on non-clinical performance testing and a literature review for clinical evaluation, not on a test set requiring expert ground truth for interpretation of images or other data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There was no clinical test set requiring expert adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device for hip arthroplasty, not an AI-based diagnostic or imaging device that would typically involve MRMC studies or AI assistance for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a hip prosthesis, not an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Non-clinical (Performance) Tests: Ground truth for these tests is based on engineering specifications, material properties, and established test standards (e.g., demonstrating mechanical compatibility and fatigue life under simulated physiological conditions).
- Clinical Evaluation: The "ground truth" for the clinical evaluation is based on the outcomes data and findings reported in the systematic scientific literature review regarding the safety and effectiveness of the Corail Hip System for femoral neck fractures, by extension supporting the expanded indications for hemi-arthroplasty.
8. The sample size for the training set
Not applicable. This is a medical device for hip arthroplasty, not an AI or machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable. As above, there is no training set for this type of device submission.