Search Results
Found 2 results
510(k) Data Aggregation
(200 days)
Zirconia Implants:
The Neodent Implant System is intended to be surgically placed in the bone of upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusion loading. Multiple teeth applications can be rigidly splinted. The implants with length of 5 mm (short implants) may be used only with two-stage surgical procedures. The recommended healing time before loading is between 10 to 12 weeks.
Zi Transmucosal Cover Screw and Healing:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted.
Zi Transmucosal Provisional Coping:
The Neodent Implant System is intended for surgical procedures in maxilla or mandible, providing support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single- or multi-unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Zi Transmucosal Abutment Replacement Screw:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Zi Transmucosal Universal Base:
The Universal Ceramic Base Zi Transmucosal 5.0 is an abutment placed over Neodent Zi Transmucosal 5.0 Ceramic Implant System in order to provide support for custom-made prosthetic restorations, such as copings or crowns. It may be used for cement or screw-retained single unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
Zirconia Base for Bridge:
The Zirconia Base for Bridge is an abutment placed over Neodent Zirconia Implants in order to provide support for custom-made prosthetic restorations. It may be used for cement or screw-retained multi-unit restorations. All digitally designed copings and/or crowns to be used with the Neodent Zirconia Base Abutment System are intended to be sent to Straumann for manufacture at a validated milling center.
Zirconia Base C:
The Zirconia Base C is an abutment placed over Neodent Zirconia Implants in order to provide support for customized prosthetic restorations, such as copings or crowns. It may be used for single-unit restorations that are screw- or cement-retained in esthetic areas over implants installed in the maxilla or mandible. All copings and/or crowns digitally designed for use with the Titanium Base C are to be designed using Sirona inLab software or Sirona CEREC Software and manufactured using a Sirona CEREC or inLab MC X or MC XL milling unit.
This premarket notification includes new ceramic devices into Neodent Implant System, which are compatible with Zirconia Implant System. The Zirconia Implants and Abutments proposed on this submission are similar to devices already cleared in previous submissions of Neodent Implant System – Zirconia Implant System, according to predicate devices described above. This submission intends to expand the portfolio with new solutions and diameter, in order to provide more treatment options to the customers.
The Zirconia Implants are manufactured in Zirconia Y-TZP and are available in Bone Level (BL) or Tissue Level (TL or Transmucosal) configurations. The Zirconia Implants (BL) are available in a diameter of 5.0 mm and lengths in a range of 8 to 13 mm. The Zi Transmucosal Implants (TL) are available in a diameter of 5.0 mm and lengths in a range of 5 to 11.5 mm.
The Zi Transmucosal Healing and Cover Screw are temporary abutments manufactured in PEEK and used during the healing phase. They are compatible with the Zi Transmucosal Implants Ø5.0. The Zi Transmucosal Healing Abutment is available in the heights of 2 and 3.5mm.
The Zi Transmucosal Provisional Coping is a temporary abutment made of polycarbonate (PC) and has a double function: used for molding procedures or production of provisional restoration.
The Zi Transmucosal Abutment Replacement Screw is a prosthetic component manufactured in titanium alloy and used to fix the fix the Zi Transmucosal Base to the Zi Transmucosal Implant.
The Zi Transmucosal Universal Base is a two-piece abutment of base and top-half prosthetic structure to provide support for customized single-unit restorations over Zi Transmucosal Implant (TL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 0.3, 1.0 and 1.5 mm. The top-half prosthetic structure to be used with Zi Transmucosal Universal Base must be designed and milled in a Straumann Validated Milling center, using the following restoration materials and dimensions:
Material: IPS e.max CAD HT, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm, Maximum angulation: 30°
Material: IPS e.max CAD LT, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm
Material: N!ce, Associated Material 510(k): K171773, Minimum wall thickness: 1.0 mm
Material: IVOCLAR Multilink cement, Associated Material 510(k): K130436, Minimum wall thickness: N/A
Material: Zirconia N!ce® LT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® HT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® XT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: PMMA N!ce, Associated Material 510(k): K071548, Minimum wall thickness: 0.7 mm
Material: Panavia—Kuraray Cement, Associated Material 510(k): K150704, Minimum wall thickness: N/A, Maximum angulation: N/A
The Zi Base for Bridge is a two-piece abutment of base and top-half prosthetic structure to provide support for customized multi-unit restorations over Zirconia Implants (BL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 1.5, 2.5 and 3.5 mm. The top-half prosthetic structure to be used with Zi Base for Bridge must be designed and milled in a Straumann Validated Milling center, using the following restoration materials and dimensions:
Material: Zirconia N!ce® LT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm, Maximum angulation: 30°
Material: Zirconia N!ce® HT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: Zirconia N!ce® XT, Associated Material 510(k): K222836, Minimum wall thickness: 0.4 mm
Material: PMMA N!ce, Associated Material 510(k): K071548, Minimum wall thickness: 0.7 mm
Material: Panavia—Kuraray Cement, Associated Material 510(k): K150704, Minimum wall thickness: N/A, Maximum angulation: N/A
The Zi Base C is a two-piece abutment of base and top-half prosthetic structure to provide support for customized single-unit restorations over Zirconia Implants (BL). The base is manufactured in Zirconia Y-ZTP and used with a patient-specific top-half prosthetic structure. The two-piece abutment has a cementable portion of 4mm and is available with gingival heights of 1.5, 2.5, 3.5 and 4.5 mm. The top-half prosthetic structure to be used with Zi Base C must be designed and milled in a Sirona InLab Validated Workflow, using the following restoration materials and dimensions:
Material: IPS e.max CAD, Associated Material 510(k): K132209, Minimum wall thickness: 0.9 mm, Maximum angulation: 20°
Material: IVOCLAR Multilink cement, Associated Material 510(k): K130436, Minimum wall thickness: N/A, Maximum angulation: N/A
All these abutments have an internal connection with the implants (ZiLock) and the prosthetic platform is identical for all subject devices described in this submission. They are intended for single use and provided sterile via Ethylene Oxide method, along with undergoing moist heat sterilization after end-user customization.
The provided FDA 510(k) clearance letter and its associated summary for the Neodent Implant System - Zirconia Implant System contain extensive information about the device, its intended use, and comparisons to predicate devices. However, it does not include specific acceptance criteria with numerical thresholds directly stated within the tables, nor does it detail a study that directly proves the device meets such criteria in terms of performance metrics like sensitivity, specificity, or image quality assessments.
Instead, the submission focuses on demonstrating substantial equivalence to predicate devices through various tests, implying that if the new device performs similarly to or better than previously cleared devices, it meets the necessary standards. The performance testing section describes the types of tests conducted (e.g., dynamic fatigue, torsion, insertion, pull-out, and software validation), but it does not present clear quantitative acceptance criteria or the specific performance results in a comparative table format.
Therefore, many of the requested fields cannot be directly extracted from the provided text as they pertain more to the performance evaluation of AI/software in interpreting medical images, which is not the primary focus of this dental implant submission.
Here's an attempt to answer the questions based on the available information, noting where information is not explicitly provided in the document:
Acceptance Criteria and Device Performance Study for Neodent Implant System - Zirconia Implant System
The FDA 510(k) summary for the Neodent Implant System - Zirconia Implant System focuses on demonstrating substantial equivalence to predicate devices through a combination of bench testing, software validation, MRI compatibility, biocompatibility, and sterilization validation. It does not present specific quantitative acceptance criteria or performance metrics directly from a comparative study in the way one might expect for an AI/software-based medical device (e.g., sensitivity, specificity thresholds). Instead, the "acceptance criteria" are implicitly met by demonstrating that the proposed devices perform at a level substantially equivalent to legally marketed predicate devices under standardized testing conditions.
1. A table of acceptance criteria and the reported device performance
As mentioned, explicit numerical acceptance criteria and reported device performance in a comparative table (e.g., for diagnostic accuracy) are not provided in this 510(k) summary. The summary indicates that tests were conducted according to relevant ISO standards and FDA guidance, and the results demonstrated that the subject devices exhibit a level of performance substantial equivalent to the predicate and reference devices.
Below is a conceptual table based on the types of tests mentioned, noting that specific numerical acceptance criteria and performance data are not detailed in the provided text.
Acceptance Criteria Category (Implicit) | Standard/Guidance | Reported Device Performance (Summary) |
---|---|---|
Dynamic Fatigue Strength | ISO 14801, FDA Guidance (Class II Special Controls) | Demonstrated a level of performance substantial equivalent to predicate and reference devices in identical conditions. |
Torsion Strength | Not specified | Adequate torsion strength in accordance with recommended IFU installation torque. |
Insertion Torque | Not specified | Evaluated insertion torque in sawbones material (Bone type I, II, III, IV). (Specific values not given). |
Implant Surface Area & Pull-Out Strength | Not specified | Greater surface area compared to reference devices; higher resistance values in Pull Out Test. |
Software Validation (Sirona Digital Workflow) | Not specified | Accuracy requirement was met; critical design parameters (min wall thickness, max angulation) respected and monitored. |
MRI Compatibility | K182620, FDA Guidance | MR conditional labeling from K182620 is applicable; safe for scanning under previously established parameters. |
Biocompatibility | ISO 10993-1, ISO 10993-18, ISO 10993-5, FDA Guidance | Subject devices are equivalent in material and manufacturing processes to predicates; no new issues raised; no additional testing required. |
Sterilization Validation (Ethylene Oxide) | ISO 11135:2014 | Validated to a Sterility Assurance Level (SAL) of 1x10⁻⁶; residuals below max allowable limits per ISO 10993-7. |
Sterilization Validation (Moist Heat) | ISO 17665-1 | Validated using parameters described in IFU. |
Endotoxin Test | ANSI/AAMI ST72:2011, ISO 11737-3, US Pharmacopeia chapter 85 | Results |
Ask a specific question about this device
(574 days)
Indications for Use for iPhysio® Profile Designer:
iPhysio® System Profile Designers are indicated to be placed in the end of the implant placement to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. The healing abutments should be used only with comections. The healing abutments are intended for use up to 6 months.
Compatible Implant Systems:
Manufacturer | Implant System | Implant Diameter (mm) | Platform Diameter (mm) | Platform Name |
---|---|---|---|---|
Nobel Biocare | NobelActive® | 3.5 | 3.5 | NP |
Nobel Biocare | NobelActive® | 4.3, 5.0 | 4.3, 5.0 | RP |
Nobel Biocare | NobelReplace® | 3.5 | 3.5 | NP |
Nobel Biocare | NobelReplace® | 4.3, 5.0 | 4.3, 5.0 | RP |
Nobel Biocare | NobelParallel™ | 3.75 | 3.5 | NP |
Nobel Biocare | NobelParallel™ | 4.3, 5.0 | 4.3, 5.0 | RP |
Straumann | Bone Level | 3.3 | 3.3 | NC |
Straumann | Bone Level | 4.1, 4.8 | 4.1, 4.8 | RC |
Straumann | Bone Level Tapered | 3.3 | 3.3 | NC |
Straumann | Bone Level Tapered | 4.1, 4.8 | 4.1, 4.8 | RC |
Zimmer Biomet | Trabecular Metal | 3.7, 4.1 | 3.5 | 3.5 |
Zimmer Biomet | Trabecular Metal | 4.7 | 4.5 | 4.5 |
Zimmer Biomet | Tapered Screw Vent | 3.7, 4.1 | 3.5 | 3.5 |
Zimmer Biomet | Tapered Screw Vent | 4.7 | 4.5 | 4.5 |
Indications for Use for iPhysio® PEEK Temporary Abutment:
The iPhysio® System PEEK Temporary Abutment is an abutment placed on the iPhysio® System Profile Designer to provide support for prosthetic structures for up to 6 months. It can be used in single- or two-stage procedures and is intended to be placed out of occlusion.
The iPhysio® System is a set of two-piece titanium healing abutments (iPhysio® Profile Designers) be screwed into the implant using compatible anodized titanium screws during the 1st or the 2nd surgical intervention. The iPhysio® Profile Designer abutment is manufactured from titanium alloy (Ti-6Al-4V ELI) per ISO 5832-3 and includes a zirconium nitride (ZrN) coating. iPhysio® Profile Designers are available in multiples anatomical shapes and gingival heights, and are compatible with Nobel BioCare (NobelActive®, NobelReplace®, NobelParallel™), Straumann (Bone Level / Bone Level Tapered), and Zimmer Biomet (Trabecular Metal, Tapered Screw-Vent®) implants. All iPhysio® Profile Designers are straight (0°). Supporting screws are manufactured from titanium alloy (Ti-6Al-4V ELI) per ISO 5832-3 and are anodized to a color that aligns with its corresponding height. The iPhysio® Profile Designer healing abutment screw is for placing the iPhysio® Profile Designer in the implant in the patient's mouth.
The iPhysio® Profile Designer remains in place throughout the bone and gingival healing process as well as during the taking of a digital or conventional impression, preserving the emergence profile during the impression. The abutment is only removed once to place and finally screw in the final abutment. Anatomical healing, implant impression, and placement of the temporary crown may all be done without removing the iPhysio® Profile Designer.
A compatible PEEK temporary abutment may be clipped to the iPhysio® Profile Designer to which a non-load bearing temporary crown may be then placed for improved aesthetics during healing. To use the PEEK temporary abutment, the iPhysio® Profile Designer healing abutment screw is replaced with a compatible temporary abutment screw. The temporary crown is placed over the PEEK temporary abutment following modification of the abutment to the appropriate shape. The PEEK temporary abutment is connected onto the iPhysio® Profile Designer by a combination of friction and cement. All iPhysio® temporary abutments are manufactured from PEEK per ASTM F2026.
This document is part of a 510(k) summary for a medical device called "iPhysio® System." It details the comparison of the subject device to predicate devices to establish substantial equivalence, which is a key requirement for FDA clearance. The document focuses on non-clinical performance testing rather than studies involving human data or AI.
Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, and ground truth establishment cannot be fully provided directly from this document as it pertains to clinical performance and AI algorithm validation, which are not the primary focus of this 510(k) submission for this specific type of dental device.
However, I can extract the information related to acceptance criteria for the non-clinical performance testing and the reported results:
1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical Testing)
Test | Test Method Summary | Acceptance Criteria (Implied by Result) | Reported Device Performance |
---|---|---|---|
Reverse Engineering Analysis | Reverse engineering analysis of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility. Relevant dimensions and tolerances from subject device screw and abutment connections were analyzed for compatibility with FDA cleared implants as specified in the device labeling. | "Device met all predetermined acceptance criteria" | PASS (compatibility confirmed) |
ZrN Coating Characterization | Per FDA Guidance "Root-form Endosseous Dental Implants and Endosseous Dental Abutments - Class II Special Controls Guidance Document for Industry and FDA Staff": | ||
• ZrN Coating Cross Section SEM Imaging Test Report | |||
• ZrN Coating Elemental Analysis and SEM Imaging Test Report | |||
• ZrN Coating Hardness Abrasion Scratch Test | |||
• ZrN Coating Tensile Shear and Abrasion Test | "Coating met all predetermined acceptance criteria" | PASS (coating properties confirmed) | |
MR Compatibility Testing | Per ASTM F2503-20 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment and FDA Guidance Document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" | Test results should lead to "MR Conditional" labeling | PASS (Device is MR Conditional) |
Fatigue Testing (Temporary Abutments) | Fatigue testing was conducted to evaluate the temporary abutments in a worst-case test configuration. The test setup was based on a modified version of ISO 14801:2016. | "Device met all predetermined acceptance criteria" | PASS (withstood worst-case loading) |
Biocompatibility | Evaluated per ISO 10993-1 and FDA Guidance for: Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material Mediate Pyrogenicity, Implantation, Subacute/subchronic toxicity, chronic toxicity, and carcinogenicity via chemical characterization and toxicological risk assessment. | "Found to be biocompatible for their intended contact level and duration" | PASS (Biocompatible) |
Sterilization Validation | Gamma sterilization validation per ISO 11137-2 for SAL 10^-6 for worst-case packaging. Steam sterilization per ISO 17665-1 for SAL 10^-6 for non-sterile components. Sterile packed components tested per ISO 11607-1 for shelf life. | SAL specified (10^-6) for sterility, validation of labeled shelf life | PASS (Sterility and shelf life validated) |
Note: The document explicitly states "Device met all predetermined acceptance criteria" or "Coating met all predetermined acceptance criteria" for several tests, which implies the acceptance criteria were established prior to testing and the device successfully met them. The specific numerical or qualitative thresholds for these criteria are not detailed in this summary for each test but are generally understood to align with relevant international standards (e.g., ISO 14801 for fatigue).
The following points cannot be answered from the provided text, as this is a 510(k) summary for a dental implant accessory, primarily relying on non-clinical (bench) testing and substantial equivalence arguments, rather than clinical trials or AI/image analysis studies.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Not applicable / Not provided. The "tests" referred to are non-clinical bench tests (e.g., mechanical fatigue, material characterization), not clinical studies with human patient data. Sample sizes would refer to the number of physical device units tested for each non-clinical performance test. Data provenance in this context would be the lab where the bench tests were performed, which is not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable / Not provided. Ground truth, in the context of diagnostic AI or clinical studies, refers to disease labels or interpretations. For non-clinical bench testing, the "ground truth" is adherence to established engineering standards and material properties, validated by testing methodologies. No human experts are establishing "ground truth" in the clinical sense for these types of tests.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable / Not provided. Adjudication methods are relevant for resolving discrepancies in expert interpretations in clinical or imaging studies. This document describes bench testing, where outcomes are determined by physical measurements and adherence to engineering standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This document explicitly states: "No clinical data was included in this submission." Therefore, no MRMC study or AI assistance evaluation was performed as part of this submission. The device (dental healing abutment) does not involve AI.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- No. As noted above, this device is a physical dental implant component, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not applicable / Not provided. For this submission, which relies on non-clinical testing, the "ground truth" is adherence to established engineering specifications, material standards (e.g., ISO 5832-3 for titanium alloy, ASTM F2026 for PEEK), and performance test methods (e.g., ISO 14801 for fatigue testing).
8. The sample size for the training set
- Not applicable. This device does not involve machine learning or a "training set."
9. How the ground truth for the training set was established
- Not applicable. This device does not involve machine learning or a "training set."
Ask a specific question about this device
Page 1 of 1