The device is indicated for use in obtaining diagnostic images to aid the physician with diagnosis. The system can be used to perform radiographic imaging of various portions of the human body, including the skull, spinal column, extremities, chest, abdomen and other body parts. The device is not indicated for use in mammography
The DRX-Compass System is a general purpose x-ray system used for acquiring radiographic images of various portions of the human body. The system consists of a combination of components including various models of high voltage x-ray generators, control panels or workstation computers, various models of patient support tables, wall-mounted image receptors/detectors for upright imaging, various models of tube support devices, x-ray tube, and collimator (beam-limiting device). The DRX-Compass can be used with digital radiography (DR) and computed radiography (CR) receptors. Smart Features are added to the DRX-Compass system to provide remote capabilities for existing functions of the DRX-Compass system. These remote capabilities simplify exam set up and improve workflow for the operator while preparing for the patient exposure. The "smart features", described below, are designed to reduce the technologist's manual tasks and to speed up workflow for existing features of the system. These improvements are referred to as "smart features" in the product documentation. Implementation of these "smart features" does not change the intended use of the system.
The provided text does not contain detailed information about specific acceptance criteria and a study that comprehensively proves the device meets those criteria for the DRX-Compass system. The document is a 510(k) summary for the FDA, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive efficacy study for new features.
However, based on the information provided, I can extract the relevant details that are present and explain why some requested information is not available in this document.
Here's a breakdown of what can be inferred and what is missing:
1. Table of Acceptance Criteria and Reported Device Performance
The document mentions that "Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device." However, the specific acceptance criteria themselves (e.g., specific thresholds for DQE/MTF, or performance metrics for the "smart features") are not explicitly detailed in this 510(k) summary. Similarly, the reported device performance values against those specific criteria are also not provided.
The closest information related to performance is:
Acceptance Criteria (Inferred/General) | Reported Device Performance (Inferred/General) |
---|---|
Image quality of additional detectors equivalent to predicate. | Flat panel detector DQE/MTF data shows the additional detectors (DRX Plus 2530, Focus HD 35, Focus HD 43, Lux 35) are equivalent in image quality to DRX Plus detectors cleared with the predicate. |
Compliance with electrical safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-54). | Device complies with listed electrical safety standards. |
Compliance with usability standards (IEC 60601-1-6, IEC 62366). | Device complies with listed usability standards. |
No new risks identified that raise additional questions of safety and performance (ISO 14971). | All product risks have been mitigated; no changes to risk control measures; testing indicates substantial equivalence. |
"Smart Features" (Real-time Video, LLI, Collimation, Patient Picture) simplify exam setup and improve workflow without changing intended use. | These features are designed to reduce manual tasks and speed up workflow. (No specific quantitative performance metrics provided in this document). |
2. Sample Size Used for the Test Set and Data Provenance
This information is not provided in the 510(k) summary. The document states "Non-clinical testing such as standards testing are the same as that of the predicate. The verification and validation testing of the modified device demonstrates that the modified device performs as well as the predicate and is substantially equivalent." without detailing the specific sample sizes or data provenance for these tests. For imaging performance, it mentions DQE/MTF data for detectors, but not the sample size of images or patients used for performance evaluation of the overall system or its new "smart features."
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not provided in the 510(k) summary. The document focuses on technical verification and validation, and comparison to a predicate device, rather than a clinical study requiring expert consensus on ground truth.
4. Adjudication Method for the Test Set
This information is not provided in the 510(k) summary. Given the absence of specific clinical study details or expert ground truth establishment, no adjudication method would be mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A MRMC comparative effectiveness study is not mentioned in this document. The submission's focus is on demonstrating substantial equivalence through technical testing and compliance with recognized standards, particularly for the "smart features" which are described as workflow enhancements rather than diagnostic AI tools requiring reader performance studies. There is no mention of AI assistance for human readers or associated effect sizes.
6. Standalone (Algorithm Only) Performance Study
A standalone performance study of an algorithm without human-in-the-loop is not explicitly mentioned in this document. The "smart features" are described as functionalities to assist the operator, implying human-in-the-loop operation, rather than a standalone diagnostic algorithm. The document mentions "Flat panel detector DQE/MTF data shows that the additional detectors supported by the modified device (DRX-Compass) are equivalent in image quality to that of the DRX Plus detectors cleared with the predicate," which is a technical performance metric for the detector component, not an algorithm's diagnostic performance.
7. Type of Ground Truth Used
The type of ground truth used for any performance evaluation is not explicitly stated. For the detector performance, DQE/MTF data refers to physical image quality metrics rather than a diagnostic ground truth (like pathology or clinical outcomes). For the "smart features," their evaluation appears to be based on functional verification and validation of their workflow enhancement capabilities, rather than comparison to a ground truth for diagnostic accuracy.
8. Sample Size for the Training Set
This information is not provided in the 510(k) summary. The document does not describe the use of machine learning algorithms that would typically require a training set. The "smart features" appear to be rule-based or real-time processing functionalities rather than learning algorithms.
9. How Ground Truth for the Training Set Was Established
Since there is no mention of a training set or machine learning, details on establishing its ground truth are not provided.
In summary, the 510(k) submission for the DRX-Compass focuses on demonstrating substantial equivalence to a predicate device by:
- Ensuring the modified device's indications for use are identical.
- Confirming compliance with recognized electrical safety and performance standards (AAMI ES60601-1, IEC 60601-1-6, IEC 60601-1-3, IEC 60601-2-54, IEC 62366).
- Applying risk management (ISO 14971) to ensure no new risks are introduced.
- Showing that new components (e.g., additional detectors) maintain equivalent image quality (e.g., DQE/MTF data).
- Asserting that new "smart features" improve workflow without changing the device's intended use or safety profile.
The document does not provide the kind of detailed clinical study data often found for AI/ML-based diagnostic devices, including specific acceptance criteria values, sample sizes for test or training sets, expert qualifications, or adjudication methods, as these may not be typically required for modifications to a stationary X-ray system primarily focused on workflow enhancements and component upgrades.