(145 days)
Focus HD 35 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all generalpurpose diagnostic procedures. The device is not intended for mammography or dental applications.
Focus HD 35 Detector is a kind of wireless digital flat panel detector. It supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 35cm×43cm.
The sensor plate of Focus HD 35 Detector is direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface. The major function of the Focus HD 35 Detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. Both kinds of detectors are the key component of DR system, enable to complete the digitalization of the medical X-ray imaging with the DR system software.
SDK(include iDetector) is intended to supply API interface for DR system manufacturers. DR system manufacturer control the detector by SDK interface. SDK is not intend to be used directly by other users beside DR system manufacturers.
The provided text is a 510(k) summary for the iRay Technology Taicang Ltd. Focus HD 35 Detector. This document primarily focuses on establishing substantial equivalence to a predicate device (Mars1417X Wireless Digital Flat Panel Detector, K210316) rather than directly presenting a study proving that the device meets specific acceptance criteria related to a diagnostic task or clinical performance.
Medical device 510(k) clearances typically rely on demonstrating that the new device is as safe and effective as a legally marketed predicate device, often through engineering and performance testing. They do not usually involve clinical trials or studies like those required for novel AI/diagnostic devices that predict diagnoses or outcomes.
Therefore, many of the requested points regarding acceptance criteria, study details, human reader improvement with AI, standalone performance, ground truth, and training set information are not applicable or not explicitly detailed in this type of submission.
Here's a breakdown of what can be gleaned from the text, and where information is missing:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria here refer to the technical specifications and safety standards required for substantial equivalence, rather than diagnostic performance metrics (e.g., sensitivity, specificity for disease detection). The comparison is primarily against the predicate device's specifications.
| Characteristic | Acceptance Criteria (Predicate Device K210316) | Reported Device Performance (Focus HD 35 Detector) | Met/Not Met | Comments / Justification |
|---|---|---|---|---|
| Intended Use | General radiographic diagnosis for human anatomy (adult & pediatric), replacement for film/screen systems, not for mammography or dental. | Same | Met | Directly comparable Statement. |
| Indications for Use | Same as Intended Use. | Same | Met | Directly comparable Statement. |
| Classification Name | Stationary X-ray system | Same | Met | Directly comparable Statement. |
| Product Code | MQB | Same | Met | Directly comparable Statement. |
| Regulation Number | 21 CFR 892.1680 | Same | Met | Directly comparable Statement. |
| Panel | Radiology | Same | Met | Directly comparable Statement. |
| Classification | Class II | Same | Met | Directly comparable Statement. |
| X-Ray Absorber (Scintillator) | CsI | Same | Met | Directly comparable Statement. |
| Installation Type | Wireless, Portable | Same | Met | Directly comparable Statement. |
| Readout Mechanism | Thin Film Transistor | Same | Met | Directly comparable Statement. |
| Image Matrix Size | 3500 x 4300 pixels | Same | Met | Directly comparable Statement. |
| Pixel Size | 100μm | Same | Met | Directly comparable Statement. |
| ADC Digitization | 16 bit | Same | Met | Directly comparable Statement. |
| Effective Imaging Area | 350.0 mm x 430.0 mm | Same | Met | Directly comparable Statement. |
| Spatial Resolution | Min. 4.3 lp/mm | 5.0 lp/mm | Met (Exceeded) | The proposed device offers higher spatial resolution, indicating improved performance. |
| Modulation Transfer Function (MTF) | 0.65 at 1 lp/mm | Same | Met | Directly comparable Statement. |
| Detective Quantum Efficiency (DQE) | 0.54 at 1 lp/mm (RQA5, 2.5μGy) | Same | Met | Directly comparable Statement. |
| Power Consumption | Max. 19W | Max. 35.5W | Different | This is a difference, but typically evaluated in terms of safety and functionality, not diagnostic performance. Addressed by electrical safety testing. |
| Communications (Wireless functionality) | Wired (Gigabit Ethernet), Wireless (IEEE 802.11a/b/g/n/ac) | Same | Met | Directly comparable Statement. |
| Imaging Protect Plate | Carbon Fiber Plate | Same | Met | Directly comparable Statement. |
| Cooling | Air cooling | Same | Met | Directly comparable Statement. |
| Dimensions | 384 mm × 460 mm × 15 mm | Same | Met | Directly comparable Statement. |
| Detector IP grade | IP56 | Same | Met | Directly comparable Statement. |
| Power input port | 4 pin port | 10 pin port | Different | Main modification from predicate device, but determined substantially equivalent after non-clinical studies. |
| Surface pressure (Uniform load) | 300 kg over whole area | 300 kg over whole area | Met | Directly comparable Statement. |
| Surface pressure (Local load) | 150 kg on 4 cm diameter | 100 kg on 4 cm diameter | Different | Proposed device has a lower local load capacity. The document implies this is acceptable for substantial equivalence without further explanation of the impact. |
| Operation Temperature | +10 ~ +35°C | +5 ~ +35°C | Different | Proposed device supports a wider temperature range, indicating improved or equivalent functionality. |
| Operation Humidity | 5 ~ 90% (Non-Condensing) | Same | Met | Directly comparable Statement. |
| Operation Atmospheric pressure | 70 ~ 106 kPa | Same | Met | Directly comparable Statement. |
| Operation Altitude | Max. 3000 meters | Same | Met | Directly comparable Statement. |
| Storage & Transportation Temperature | -20 ~ +55°C | Same | Met | Directly comparable Statement. |
| Storage & Transportation Humidity | 5 ~ 95% (Non-Condensing) | Same | Met | Directly comparable Statement. |
| Storage & Transportation Atmospheric pressure | 60 ~ 106 kPa | 70 ~ 106 kPa | Different | Proposed device has a slightly reduced range. Implied acceptable for substantial equivalence. |
| Storage & Transportation Altitude | Max. 3000 meters | Same | Met | Directly comparable Statement. |
| Software | SDK (include iDetector) supplies API interface for DR system manufacturers. | Same | Met | Directly comparable Statement. |
| Electrical Safety | Meet IEC/ES 60601-1 | All test results meet standard requirements. | Met | Stated as part of non-clinical testing. |
| EMC Testing | Meet IEC 60601-1-2 | All test results meet standard requirements. | Met | Stated as part of non-clinical testing. |
| Biological Evaluation | Meet ISO 10993-1 | Evaluation and test results assured safety. | Met | Stated as part of non-clinical testing. |
2. Sample size used for the test set and the data provenance
- Sample size used for the test set: Not applicable for a typical 510(k) submission based on technical equivalence. The "test set" here refers to the actual physical device and its components undergoing engineering and safety tests, not a dataset for diagnostic performance evaluation.
- Data provenance: Not applicable. The data is from engineering and safety tests conducted on the physical device and from a comparison of specifications with the predicate. There is no mention of country of origin for clinical or image data, as such data is not the primary basis for this type of submission. The tests were likely conducted by the manufacturer or accredited labs.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. This device is an X-ray detector, a hardware component. Its clearance is based on technical specifications, safety, and performance as an imaging capture device, not on its ability to produce a specific diagnosis from images or generate ground truth. There is no "ground truth" in the diagnostic sense being established for a test set of images within this submission.
4. Adjudication method for the test set
- Not Applicable. As no clinical diagnostic performance study with images and ground truth is described, there's no adjudication method.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is not an AI-powered diagnostic device. It's a digital X-ray detector. Therefore, no MRMC study comparing human readers with and without AI assistance was performed or reported here.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This is a hardware device (X-ray detector), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not Applicable. For this hardware device submission, "ground truth" relates to physical measurements and compliance with engineering standards (e.g., spatial resolution, DQE, electrical safety standards). There is no clinical ground truth (pathology, expert consensus on images, etc.) associated with this submission.
8. The sample size for the training set
- Not Applicable. This is a hardware device; no "training set" in the context of machine learning or AI is mentioned or relevant to its clearance.
9. How the ground truth for the training set was established
- Not Applicable. As no training set is mentioned, no ground truth for it was established.
Summary of the Study (as described in the 510(k) submission):
The "study" referenced in the document is primarily a technical assessment and comparison to establish substantial equivalence with a legally marketed predicate device (K210316, Mars1417X Wireless Digital Flat Panel Detector).
-
Methodology:
- Side-by-side comparison of technical specifications: The document provides tables comparing parameters like intended use, indications for use, physical dimensions, imaging characteristics (spatial resolution, MTF, DQE), environmental operating conditions, and interfaces between the proposed device and the predicate device.
- Non-clinical testing: This included:
- Electrical safety and EMC testing: Performed according to IEC/ES 60601-1 and IEC 60601-1-2 standards.
- Biological Evaluation: Materials contacting operators' or patient's skin evaluated with ISO 10993-1.
- Functional performance testing: The document implies that areas with minor differences (e.g., power consumption, local load capacity, operating temperature range, atmospheric pressure range) were still considered substantially equivalent following non-clinical studies. The improved spatial resolution (5.0 lp/mm vs. Min. 4.3 lp/mm) was presented as a positive difference.
-
Conclusion: The manufacturer concluded that the Focus HD 35 Detector is substantially equivalent to the predicate device regarding safety and effectiveness, based on the non-clinical studies and comparison of technical characteristics. The main identified differences (DC input port, power consumption, local load, operating temperature range, storage/transportation atmospheric pressure) were implicitly determined not to raise new questions of safety or effectiveness.
{0}------------------------------------------------
March 29, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a seal on the left and the FDA acronym with the agency's name on the right. The seal features an eagle-like figure, while the text reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "ADMINISTRATION" appearing in a smaller font size below the rest of the name.
iRay Technology Taicang Ltd. % Junjie Qian Registration & Regulation Affairs Engineer No.33 Xinggang Rd., Taicang Port Economic Technological Development Zone Taicang, Jiangsu 215434 CHINA
Re: K213646
Trade/Device Name: Focus HD 35 Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: October 20, 2021 Received: November 4, 2021
Dear Junjie Qian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213646
Device Name Focus HD 35 Detector
Indications for Use (Describe)
Focus HD 35 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all generalpurpose diagnostic procedures. The device is not intended for mammography or dental applications.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart G) |
|---|---|
| ----------------------------------------------------------------------------------------- | ---------------------------------------------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
iRay Technology Taicang Ltd.
510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
(As Required by 21 CFR 807.92)
1. Date Prepared [21 CFR 807.92(a)(1)]
October 13, 2021
2. Submitter's Information [21 CFR 807.92(a)(1)]
| Company Name: | iRay Technology Taicang Ltd. |
|---|---|
| Company Address: | No.33 Xinggang Road, Taicang Port Economic andTechnological Development Zone, Jiangsu, China 21543- |
| Contact Person: | Junjie.Qian |
| Phone: | 0512-53690872 |
| Fax: | 0512-53690872 |
| Email: | junjie.qian@iraygroup.com |
3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | Focus HD 35 Detector |
|---|---|
| Common Name: | Solid State X-Ray Imager |
| Model Name: | Focus HD 35 |
| Classification Name: | Stationary X-Ray System |
| Product Code: | MQB |
| Regulation Number: | 21 CFR 892.1680 |
| Device Class: | Class II |
{4}------------------------------------------------
4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[
The identification predicates within this submission are as follows:
| Manufacturer: | iRay Technology Taicang Ltd. |
|---|---|
| Trade Name: | Wireless Digital Flat Panel Detector |
| Model Name: | Mars1417X |
| Product Code: | MQB |
| Classification Name: | Stationary X-Ray System |
| Regulation Number: | 21 CFR 892.1680 |
| Device Class: | Class II |
| FDA 510 (k) #: | K210316 |
5. Description of the Device [21 CFR 807.92(a)(4)]
Focus HD 35 Detector is a kind of wireless digital flat panel detector. It supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 35cm×43cm.
The sensor plate of Focus HD 35 Detector is direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface. The major function of the Focus HD 35 Detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. Both kinds of detectors are the key component of DR system, enable to complete the digitalization of the medical X-ray imaging with the DR system software.
SDK(include iDetector) is intended to supply API interface for DR system manufacturers. DR system manufacturer control the detector by SDK interface. SDK is not intend to be used directly by other users beside DR system manufacturers.
2 of 10
{5}------------------------------------------------
6. Intended Use [21 CFR 807.92(a)(5)]
6.1. Indications for use
Focus HD 35 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all generalpurpose diagnostic procedures. The device is not intended for mammography or dental applications.
6.2. Suitable patient
It is suitable for providing digital X-ray imaging for DR system to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients, but not intended for mammography or dental applications. The remaining notes depend on the DR system.
6.3. Processing of input and output
When Focus HD 35 Detector works continuously, it can automatically distinguish X-ray and output an imaging for diagnosis of disease, injury, or of any applicable health problem.
| Item | Predicate Device:Mars1417X Wireless DigitalFlat Panel Detector | Proposed Device:Focus HD 35 Detector |
|---|---|---|
| 510(K) Number | K210316 | |
| Intended Use | Mars1417X Wireless DigitalFlat Panel Detector isindicated for digital imagingsolution designed forproviding generalradiographic system in allgeneral-purpose diagnosticprocedures. | same |
| Item | Predicate Device:Mars1417X Wireless DigitalFlat Panel Detector | Proposed Device:Focus HD 35 Detector |
| Indications for Use | Mars1417X wireless digitalflat panel detector is indicatedfor digital imaging solutionsdesigned to provide generalradiographic diagnosis forhuman anatomy includingboth adult and pediatricpatients. It is intended toreplace film/screen systems inall general-purposediagnostic procedures. Thedevice is not intended formammography or dentalapplications. | Same |
| ClassificationName | Stationary X-ray system | Same |
| Product Code | MQB | Same |
| Regulation Number | 21 CFR 892.1680 | Same |
| Panel: | Radiology | Same |
| Classification: | II | Same |
| X-Ray Absorber(Scintillator): | CsI | Same |
| Installation Type: | Wireless, Portable | Same |
| ReadoutMechanism: | Thin Film Transistor | Same |
| Image Matrix Size: | $3500 \times 4300$ pixels | Same |
| Item | Predicate Device:Mars1417X Wireless DigitalFlat Panel Detector | Proposed Device:Focus HD 35 Detector |
| Pixel Size: | 100μm | Same |
| ADC Digitization | 16 bit | Same |
| EffectiveImagingArea: | 350.0 mm × 430.0 mm | Same |
| Spatial Resolution: | Min. 4.3lp/mm | 5.0 lp/mm |
| ModulationTransferFunction(MTF) | 0.65 at 1 lp/mm | Same |
| DetectiveQuantumEfficiency(DQE) | 0.54 at 1 lp/mm (RQA5,2.5μGy) | Same |
| PowerConsumption: | Max. 19W | Max.35.5W |
| Communications:(Wirelessfunctionality) | a) Wired (only forservice) : GigabitEthernet (1000BASE-T)b) Wireless: IEEE802.11a/b/g/n/ac (2.4GHz / 5 GHz) | Same |
| ImagingprotectPlate: | Carbon Fiber Plate | Same |
| Cooling: | Air cooling | Same |
| Dimensions: | 384 mm × 460 mm × 15 mm | Same |
| Item | Predicate Device:Mars1417X Wireless DigitalFlat Panel Detector | Proposed Device:Focus HD 35 Detector |
| Detector IP grade | IP56 | Same |
| Power input port | 4 pin port | 10 pin port |
| Surface pressure | Uniform load: 300 kg overthe whole area of the surface; | Uniform load: 300 kg over thewhole area of the surface; |
| Local load: 150 kg on anarea 4 cm diameter of center | Local load: 100 kg on an area 4cm diameter of center | |
| Operation: | Temperature: +10 ~ +35°C | Temperature: +5 ~ +35°C |
| Humidity: 5 ~ 90%(Non-Condensing) | Humidity: 5 ~ 90%(Non-Condensing) | |
| Atmospheric pressure: 70 ~ 106 kPa | Atmospheric pressure: 70 ~ 106 kPa | |
| Altitude: Max. 3000 meters | Altitude: Max. 3000 meters | |
| Storage andTransportation:( detector ) | Temperature: -20 ~ +55°C | Temperature: -20 ~ +55°C |
| Humidity: 5 ~ 95%(Non-Condensing) | Humidity: 5 ~ 95%(Non-Condensing) | |
| Atmospheric pressure: 60 ~ 106 kPa | Atmospheric pressure: 70 ~ 106 kPa | |
| Altitude: Max. 3000 meters | Altitude: Max. 3000 meters | |
| Software | SDK(include iDetector) isintend to supply API interfacefor DR system manufacturers.DR system manufacturercontrol the detector by SDKinterface. SDK is not intendto use directly by other users | Same |
| Item | Predicate Device: | Proposed Device: |
| Mars1417X Wireless DigitalFlat Panel Detectorbeside DR systemmanufacturers. | Focus HD 35 Detector | |
| Utilized FDAguidancedocuments | 1. Guidance for theSubmission of 510(k)'sfor Solid State X-rayImaging Devices;2. The 510(k) Program:Evaluating SubstantialEquivalence in PremarketNotifications[510(k)];3. Content of PremarketSubmissions forManagement ofCybersecurity in MedicalDevices;4. Radio FrequencyWireless Technology inMedical Devices;5. Guidance for “PremarketAssessment of PediatricMedical Devices”;6. Guidance for “PediatricInformation for X-rayImaging DevicePremarket Notifications”. | Same |
| Item | Predicate Device:Mars1417X Wireless DigitalFlat Panel Detector | Proposed Device:Focus HD 35 Detector |
| 7. The Special 510(k)Program8. Design Control GuidanceFor Medical DeviceManufacturers9. Appropriate Use ofVoluntary ConsensusStandards in PremarketSubmissions for MedicalDevices |
7. Technological Characteristic [21 CFR 807.92(a)(6)]
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
8. System requirements to operate with other radiographic system components
-
- Recommended Generator Specification:
Energy range: 40~150kVp
- Recommended Generator Specification:
mA range: 10~1000mA (depending on the generator power)
ms range: 106300ms to produce 0.11000mAs (depending on the generator power) Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have any questions regarding the compatibility issue for other generators, please contact the distributor or manufacturer's service office.
-
- Application Program Interface (API) for system integration manufacturer Minimum configuration: Focus HD 35 Detector connected via wireless communication.
| Operating System: | Windows embedded |
|---|---|
| CPU: | Intel Core i3- 8100 3.6GHz 4C 65W |
| Memory: | 16GB (2x8GB) DDR4 2666 DIMM |
8 of 10
{11}------------------------------------------------
Hard Disk: 1TB
X-ray exposure mode 3)
The AED trigger module is a unit can connect X-ray signal in the Focus HD 35 Detector. Once there is X-ray generator exposure exist, the AED trigger module will detect the X-ray radiation and output signal to the detector. Until the exposure finished, the detector will receive a signal which represent the end of exposure from the inner trigger module and begin to acquire the image.
Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)] 9.
- Electrical Safety and EMC testing: 1) Electrical, mechanical, environmental safety according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC 60601-1-2. All test results are meet the standard requirements.
-
- Biological Evaluation:
The materials of the detector which contact operators' or patient's skin have been evaluated with the ISO 10993-1. And the evaluation results and test result assured the safety the same as the predicate device.
-
- Non-clinical Considerations:
Main modification from the predicate device is DC input port change(from 4 pin port to 10 pin port ) . This DC input port is only designed and applicable for DR system manufacturer.
- Non-clinical Considerations:
The non-clinical studies have been performed and the results have shown that sections of the non-clinical consideration mentioned in the 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices' are substantially equivalent to the non-clinical consideration of predicate device on the Market (Mars1417X, K210316).
-
- Clinical Consideration:
Intended use, fundamental scientific technology, regulatory requirement, non-clinical performance, labeling, quality-assurance program and software keep the same with
- Clinical Consideration:
9 of 10
{12}------------------------------------------------
those of predicate device. There is no any change about clinical performance from predicate device.
10. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, iRay Technology Taicang Ltd. concludes that Focus HD 35 Detector is substantially equivalent to predicate device with regards to safety and effectiveness.
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.