K Number
K213646
Date Cleared
2022-03-29

(145 days)

Product Code
Regulation Number
892.1680
Panel
RA
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Focus HD 35 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all generalpurpose diagnostic procedures. The device is not intended for mammography or dental applications.

Device Description

Focus HD 35 Detector is a kind of wireless digital flat panel detector. It supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 35cm×43cm.

The sensor plate of Focus HD 35 Detector is direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface. The major function of the Focus HD 35 Detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. Both kinds of detectors are the key component of DR system, enable to complete the digitalization of the medical X-ray imaging with the DR system software.

SDK(include iDetector) is intended to supply API interface for DR system manufacturers. DR system manufacturer control the detector by SDK interface. SDK is not intend to be used directly by other users beside DR system manufacturers.

AI/ML Overview

The provided text is a 510(k) summary for the iRay Technology Taicang Ltd. Focus HD 35 Detector. This document primarily focuses on establishing substantial equivalence to a predicate device (Mars1417X Wireless Digital Flat Panel Detector, K210316) rather than directly presenting a study proving that the device meets specific acceptance criteria related to a diagnostic task or clinical performance.

Medical device 510(k) clearances typically rely on demonstrating that the new device is as safe and effective as a legally marketed predicate device, often through engineering and performance testing. They do not usually involve clinical trials or studies like those required for novel AI/diagnostic devices that predict diagnoses or outcomes.

Therefore, many of the requested points regarding acceptance criteria, study details, human reader improvement with AI, standalone performance, ground truth, and training set information are not applicable or not explicitly detailed in this type of submission.

Here's a breakdown of what can be gleaned from the text, and where information is missing:


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria here refer to the technical specifications and safety standards required for substantial equivalence, rather than diagnostic performance metrics (e.g., sensitivity, specificity for disease detection). The comparison is primarily against the predicate device's specifications.

CharacteristicAcceptance Criteria (Predicate Device K210316)Reported Device Performance (Focus HD 35 Detector)Met/Not MetComments / Justification
Intended UseGeneral radiographic diagnosis for human anatomy (adult & pediatric), replacement for film/screen systems, not for mammography or dental.SameMetDirectly comparable Statement.
Indications for UseSame as Intended Use.SameMetDirectly comparable Statement.
Classification NameStationary X-ray systemSameMetDirectly comparable Statement.
Product CodeMQBSameMetDirectly comparable Statement.
Regulation Number21 CFR 892.1680SameMetDirectly comparable Statement.
PanelRadiologySameMetDirectly comparable Statement.
ClassificationClass IISameMetDirectly comparable Statement.
X-Ray Absorber (Scintillator)CsISameMetDirectly comparable Statement.
Installation TypeWireless, PortableSameMetDirectly comparable Statement.
Readout MechanismThin Film TransistorSameMetDirectly comparable Statement.
Image Matrix Size3500 x 4300 pixelsSameMetDirectly comparable Statement.
Pixel Size100μmSameMetDirectly comparable Statement.
ADC Digitization16 bitSameMetDirectly comparable Statement.
Effective Imaging Area350.0 mm x 430.0 mmSameMetDirectly comparable Statement.
Spatial ResolutionMin. 4.3 lp/mm5.0 lp/mmMet (Exceeded)The proposed device offers higher spatial resolution, indicating improved performance.
Modulation Transfer Function (MTF)0.65 at 1 lp/mmSameMetDirectly comparable Statement.
Detective Quantum Efficiency (DQE)0.54 at 1 lp/mm (RQA5, 2.5μGy)SameMetDirectly comparable Statement.
Power ConsumptionMax. 19WMax. 35.5WDifferentThis is a difference, but typically evaluated in terms of safety and functionality, not diagnostic performance. Addressed by electrical safety testing.
Communications (Wireless functionality)Wired (Gigabit Ethernet), Wireless (IEEE 802.11a/b/g/n/ac)SameMetDirectly comparable Statement.
Imaging Protect PlateCarbon Fiber PlateSameMetDirectly comparable Statement.
CoolingAir coolingSameMetDirectly comparable Statement.
Dimensions384 mm × 460 mm × 15 mmSameMetDirectly comparable Statement.
Detector IP gradeIP56SameMetDirectly comparable Statement.
Power input port4 pin port10 pin portDifferentMain modification from predicate device, but determined substantially equivalent after non-clinical studies.
Surface pressure (Uniform load)300 kg over whole area300 kg over whole areaMetDirectly comparable Statement.
Surface pressure (Local load)150 kg on 4 cm diameter100 kg on 4 cm diameterDifferentProposed device has a lower local load capacity. The document implies this is acceptable for substantial equivalence without further explanation of the impact.
Operation Temperature+10 ~ +35°C+5 ~ +35°CDifferentProposed device supports a wider temperature range, indicating improved or equivalent functionality.
Operation Humidity5 ~ 90% (Non-Condensing)SameMetDirectly comparable Statement.
Operation Atmospheric pressure70 ~ 106 kPaSameMetDirectly comparable Statement.
Operation AltitudeMax. 3000 metersSameMetDirectly comparable Statement.
Storage & Transportation Temperature-20 ~ +55°CSameMetDirectly comparable Statement.
Storage & Transportation Humidity5 ~ 95% (Non-Condensing)SameMetDirectly comparable Statement.
Storage & Transportation Atmospheric pressure60 ~ 106 kPa70 ~ 106 kPaDifferentProposed device has a slightly reduced range. Implied acceptable for substantial equivalence.
Storage & Transportation AltitudeMax. 3000 metersSameMetDirectly comparable Statement.
SoftwareSDK (include iDetector) supplies API interface for DR system manufacturers.SameMetDirectly comparable Statement.
Electrical SafetyMeet IEC/ES 60601-1All test results meet standard requirements.MetStated as part of non-clinical testing.
EMC TestingMeet IEC 60601-1-2All test results meet standard requirements.MetStated as part of non-clinical testing.
Biological EvaluationMeet ISO 10993-1Evaluation and test results assured safety.MetStated as part of non-clinical testing.

2. Sample size used for the test set and the data provenance

  • Sample size used for the test set: Not applicable for a typical 510(k) submission based on technical equivalence. The "test set" here refers to the actual physical device and its components undergoing engineering and safety tests, not a dataset for diagnostic performance evaluation.
  • Data provenance: Not applicable. The data is from engineering and safety tests conducted on the physical device and from a comparison of specifications with the predicate. There is no mention of country of origin for clinical or image data, as such data is not the primary basis for this type of submission. The tests were likely conducted by the manufacturer or accredited labs.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This device is an X-ray detector, a hardware component. Its clearance is based on technical specifications, safety, and performance as an imaging capture device, not on its ability to produce a specific diagnosis from images or generate ground truth. There is no "ground truth" in the diagnostic sense being established for a test set of images within this submission.

4. Adjudication method for the test set

  • Not Applicable. As no clinical diagnostic performance study with images and ground truth is described, there's no adjudication method.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is not an AI-powered diagnostic device. It's a digital X-ray detector. Therefore, no MRMC study comparing human readers with and without AI assistance was performed or reported here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This is a hardware device (X-ray detector), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not Applicable. For this hardware device submission, "ground truth" relates to physical measurements and compliance with engineering standards (e.g., spatial resolution, DQE, electrical safety standards). There is no clinical ground truth (pathology, expert consensus on images, etc.) associated with this submission.

8. The sample size for the training set

  • Not Applicable. This is a hardware device; no "training set" in the context of machine learning or AI is mentioned or relevant to its clearance.

9. How the ground truth for the training set was established

  • Not Applicable. As no training set is mentioned, no ground truth for it was established.

Summary of the Study (as described in the 510(k) submission):

The "study" referenced in the document is primarily a technical assessment and comparison to establish substantial equivalence with a legally marketed predicate device (K210316, Mars1417X Wireless Digital Flat Panel Detector).

  • Methodology:

    • Side-by-side comparison of technical specifications: The document provides tables comparing parameters like intended use, indications for use, physical dimensions, imaging characteristics (spatial resolution, MTF, DQE), environmental operating conditions, and interfaces between the proposed device and the predicate device.
    • Non-clinical testing: This included:
      • Electrical safety and EMC testing: Performed according to IEC/ES 60601-1 and IEC 60601-1-2 standards.
      • Biological Evaluation: Materials contacting operators' or patient's skin evaluated with ISO 10993-1.
      • Functional performance testing: The document implies that areas with minor differences (e.g., power consumption, local load capacity, operating temperature range, atmospheric pressure range) were still considered substantially equivalent following non-clinical studies. The improved spatial resolution (5.0 lp/mm vs. Min. 4.3 lp/mm) was presented as a positive difference.
  • Conclusion: The manufacturer concluded that the Focus HD 35 Detector is substantially equivalent to the predicate device regarding safety and effectiveness, based on the non-clinical studies and comparison of technical characteristics. The main identified differences (DC input port, power consumption, local load, operating temperature range, storage/transportation atmospheric pressure) were implicitly determined not to raise new questions of safety or effectiveness.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.