Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K233791
    Manufacturer
    Date Cleared
    2024-07-11

    (226 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K200374

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Drivewire 24 Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The Drivewire 24 Guidewire is intended to facilitate the selective placement of diagnostic catheters. This device is not intended for use in the coronary arteries.

    Device Description

    The Drivewire 24 Guidewire is a 0.024" diameter steerable guidewire with a deflectable tip to aid in accessing vasculature. The guidewire is supplied sterile (ETO sterilization) and is for single use only. The Drivewire 24 Guidewire is comprised of a stainless steel hypotube that is cut along its length to provide flexibility and tip deflection ability through control of the handle, an inner Nitinol braided flexible coil, an inner core wire, and a handle. The inner core wire runs inside the hypotube from the distal end to the handle. The distal end of the inner core wire is flattened, looped around and joined to the tip of the distal section of the hypotube, forming a deflectable tip. The hypotube is marked with fluoro-safe markers to provide visual clues to the user to initiate fluoroscopy guided insertion. In order to actuate the tip deflection in two directions, the Drivewire 24 Guidewire handle contains a tube assembly section. The handle is assembled to the proximal end of the core wire and controls the movement of the distal tip by pulling/pushing the inner moveable core wire, allowing the bending of the distal tip in two directions. The handle assembly has neutral landmarks to identify the location where the tip is straight. The Drivewire 24 Guidewire has a hydrophilic coating on its distal segment in order to reduce the friction of the guidewire while navigating. The Drivewire 24 Guidewire is provided with a torque accessory to facilitate use of the guidewire and is not intended to have patient contact.

    AI/ML Overview

    The information provided does not describe an AI/ML device but rather a medical guidewire. As such, the typical acceptance criteria and study designs for AI/ML devices (e.g., test sets, ground truth, expert adjudication, MRMC studies, standalone performance, training sets) are not applicable here.

    This document, K233791, is a 510(k) premarket notification for the Drivewire 24 Guidewire, a physical medical device. The submission demonstrates substantial equivalence to a predicate device (Aristotle 24 Guidewire) through non-clinical testing.

    Here's an analysis of the provided information, framed to address the prompt's questions where applicable, but explicitly noting the absence of AI/ML-specific details:

    1. A table of acceptance criteria and the reported device performance

    The document provides a table of "Non-Clinical Testing" which outlines various bench tests, their methods, and results indicating they met acceptance criteria. While specific numerical acceptance criteria are not always stated, the results confirm successful performance.

    TestAcceptance Criteria (Implied)Reported Device Performance
    Visual and Dimensional VerificationMeet specified overall length, diameter, bend diameter, coating length; no damage.Dimensional and visual inspection results meet acceptance criteria.
    Tip FlexibilityComparable to predicate device.Tip flexibility was shown to be comparable to the predicate device.
    Tip Deflection ForceLess than or equal to the predicate device's force.The force applied by the tip is equal to or less than the predicate device's.
    Simulated Use - Delivery & Retrieval ForceLess than the predicate device's force.The delivery and retrieval forces were less than the predicate device's.
    Simulated Use - Performance & CompatibilityDemonstrate intended use in challenging simulated conditions.All devices met acceptance criteria in demonstrating the intended use of the guidewire in challenging simulated use conditions.
    Usability EvaluationMeet acceptance criteria (as determined by physicians).The device met the acceptance criteria.
    TorqueabilityRequired initial rotation to tip response equivalent to or less than predicate device.All devices met acceptance criteria in having the required initial rotation to the point when the tip starts to respond be equivalent to or less than the predicate device.
    Kink ResistanceWithstand clinical use scenarios without kinking.All devices met acceptance criteria for kink resistance which represent clinical use scenarios.
    Fracture TestNo fractures, loosening, or failures when wound around a cylinder.All devices met the acceptance criteria of having no fractures, loosening, or failures.
    Flexing TestNo fractures, loosening, or failures after repeated bending.All devices met the acceptance criteria of having no fractures, loosening, or failures.
    Tensile ForceMeet peak tensile force criteria established by delivery and retrieval force testing.All joints met peak tensile force acceptance criteria established by delivery and retrieval force testing.
    Tip Mechanism DurabilityWithstand a predetermined number of handle actuations.All devices met acceptance criteria of withstanding a predetermined number of handle actuations.
    Torque StrengthNumber of turns to failure no less than a predetermined value.All devices met the acceptance criteria of having the number of turns to failure be no less than a predetermined value.
    Torquer PerformanceNo damage to handle or shaft; equivalent or higher tensile force and measured torque force at slipping to predicate device.All devices met acceptance criteria of having no damage to the handle or shaft, and equivalent or higher tensile force and measured torque force at slipping to predicate device.
    ParticulateParticulates equivalent to or less than the predicate device.All devices have particulates equivalent or less than the predicate device.
    LubricityMaximal and average force comparable to the predicate device.All devices met the acceptance criteria of having a maximal force and an average force comparable to the predicate device.
    Coating IntegrityNo coating separation after simulated use testing.All devices met the acceptance criteria of having no coating separation after simulated use testing.
    RadiopacityVisible under fluoroscopy.All devices were visible under fluoroscopy.
    CorrosionNo evidence of corrosion.The test sample did not have evidence of corrosion.
    Biocompatibility Tests (Cytotoxicity, Irritation, Sensitization, Hemocompatibility, Pyrogenicity, Acute Systemic Toxicity, Thrombogenicity)Meet ISO 10993 standards and specific test requirements.All tests confirmed that the Drivewire 24 Guidewire met biological safety requirements per the ISO 10993 standard, with specific positive results for each test (e.g., no toxicity, no irritation, not sensitizers, non-hemolytic, not complement activator, nonpyrogenic, no systemic toxicity, no thrombus formation).
    SterilizationAchieve SAL of 10^-6 or less by EtO.Demonstrate that the Ethylene Oxide (EtO) sterilization process will reliably sterilize the product loads to a sterility assurance level (SAL) of 10^-6 or less, according to the overkill half cycle approach and relevant standards.
    Shelf Life and Package IntegrityMeet acceptance criteria for 1 year shelf-life.The Drivewire 24 Guidewire device and package met all shelf life and package integrity test acceptance criteria supporting a shelf-life of one (1) year.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Bench Tests: The specific number of devices tested for each bench test is not explicitly stated, but the results refer to "All devices," implying a sufficient sample was used to statistically validate the findings against acceptance criteria. Data provenance for bench tests is typically from the manufacturer's internal labs.
    • Biocompatibility Tests: The description mentions "test article extract" or "test articles," indicating samples of the device or its materials were used according to ISO 10993 standards.
    • Animal Study: The study was conducted in "domestic swine." The number of animals or devices used per animal is not specified but is implicitly sufficient for a GLP study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable as this is a physical medical device, not an AI/ML diagnostic system requiring expert ground truth for interpretation. The "Usability Evaluation" mentioned that "Physicians evaluated the guidewire in clinically relevant simulated use models," implying expert input, but details on their number or qualifications are not provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is not an AI/ML diagnostic study requiring expert adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study was performed as this is a physical medical device. The efficacy is demonstrated through non-clinical testing and comparison to a predicate device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For a physical device like a guidewire, the "ground truth" is established through physical and material science measurements, functional performance testing against defined engineering specifications, and biological safety testing against established international standards (e.g., ISO 10993). In the animal study, the "ground truth" for thrombogenicity was the direct examination of devices, renal arteries, and downstream organs for thrombus absence.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. No training set for an AI/ML model was involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222690
    Manufacturer
    Date Cleared
    2023-04-24

    (230 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K200374

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SmartGUIDE guidewire is intended for general intravascular and peripheral vascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. SmartGUIDE is intended to be used only by physicians trained in percutaneous, intravascular techniques and procedures.

    Device Description

    SmartGUIDE is a deflectable quidewire for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. The device is available with a nominal outer diameter of 0.014" (0.36 mm) and overall length of 200 cm. The device is available sterile and is for single use only.

    AI/ML Overview

    The provided text describes the SmartGUIDE deflectable hydrophilic guidewire (K222690), a medical device, and its substantial equivalence to a predicate device. It details various tests conducted to demonstrate its safety and effectiveness, but it does not pertain to an Artificial Intelligence (AI) or machine learning device.

    Therefore, many of the requested categories related to AI/ML device studies (such as sample size for test sets and training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth for training) are not applicable to this document.

    However, I can extract and present the acceptance criteria and performance data for the physical medical device as described in the provided text.

    Here is the information based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    For the SmartGUIDE deflectable hydrophilic guidewire, the acceptance criteria are generally implied by conformity to established standards and guidance documents (e.g., FDA Guidewire Guidance, ISO 11070) and comparative equivalence to a predicate device. The "reported device performance" is consistently "Pass" or "Pass / Equivalent," indicating that the device met these criteria.

    Acceptance Criteria (Implied by Test and Standard)Reported Device Performance Statement
    Bench Testing:
    Dimensional and Visual Verification (per FDA Guidewire Guidance & ISO 11070)Pass
    Simulated Use (per FDA Guidewire Guidance)Pass
    Tip Flexibility (per FDA Guidewire Guidance)Pass
    Torqueability (per FDA Guidewire Guidance)Pass
    Kink Resistance (per FDA Guidewire Guidance)Pass
    Fracture Test (per ISO 11070)Pass
    Flexing Test (per ISO 11070)Pass
    Tensile Strength and Tip Pull (per FDA Guidewire Guidance & ISO 11070)Pass
    Torque Strength (per FDA Guidewire Guidance)Pass
    Particulate Release (per FDA Guidewire Guidance)Pass
    Lubricity (per FDA Guidewire Guidance)Pass
    Coating Integrity (per FDA Guidewire Guidance)Pass
    Corrosion Resistance (per FDA Guidewire Guidance & ISO 11070)Pass
    Radiopacity (per FDA Guidewire Guidance)Pass
    Specific Bench Tests for In Situ Bending Mechanism:
    Tip actuation force (maximum tip deflection force within specified range)Pass
    Tip fatigue (durability of tip deflection mechanism to 30 full handle actuations)Pass
    Blood clot crossing (crossing of soft and hard clots)Pass
    Biocompatibility:
    Cytotoxicity (per ISO 10993)All tests confirmed biocompatibility
    Sensitization (per ISO 10993)All tests confirmed biocompatibility
    Intracutaneous reactivity (per ISO 10993)All tests confirmed biocompatibility
    Material mediated pyrogenicity (per ISO 10993)All tests confirmed biocompatibility
    Bacterial endotoxins (per ISO 10993)All tests confirmed biocompatibility
    Acute systemic toxicity (per ISO 10993)All tests confirmed biocompatibility
    Complement activation (per ISO 10993)All tests confirmed biocompatibility
    Hemolysis (per ISO 10993)All tests confirmed biocompatibility
    Thrombogenicity (per ISO 10993)All tests confirmed biocompatibility
    EtO sterilization residuals (per ISO 10993)All tests confirmed biocompatibility
    Sterilization and Shelf Life:
    Sterilization process validated (per EN ISO 11135)Device is non-pyrogenic; process validated
    Shelf-life (after 2 years accelerated aging)All acceptance criteria met
    Pre-Clinical Animal Testing (GLP study in domestic swine):
    Overall, in life health (Clinical health status of animals)Pass / Equivalent (Subject device equivalent to predicate device)
    Efficacy / functional evaluation (Maneuverability, navigability, etc.)Pass / Equivalent (Subject device equivalent to predicate device)
    Procedural, Subacute, and Chronic safety (Macro- and microscopic evaluation)Pass / Equivalent (Subject device equivalent to predicate device)
    Thrombogenicity (Thrombus formation, fibrin deposits, non-patent artery, etc.)Pass / Equivalent (Subject device equivalent to predicate device)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • This document describes a non-AI/ML medical device.
    • Sample Size (Pre-Clinical Animal Study): Not explicitly stated, but it was a "controlled good laboratory practice (GLP) study in domestic swine." The number of animals used is not specified.
    • Data Provenance: The pre-clinical animal study was conducted as a "controlled good laboratory practice (GLP) study in domestic swine." No country of origin is listed for the study data. It is a prospective study in animals.
    • For bench and biocompatibility testing, sample sizes are also not explicitly stated, but these are standard laboratory tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable as this is not an AI/ML device study requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for the device's performance comes from validated laboratory tests and the observation of physical and physiological responses in animal studies.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable as this is not an AI/ML device study involving human interpretation with adjudication. The assessment of device performance in bench and animal studies relies on established measurement protocols and scientific observation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable as this is not an AI/ML device. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For bench testing: Established physical and mechanical measurement standards and validated laboratory protocols (e.g., measuring forces, observing fractures, checking coating integrity).
    • For biocompatibility: Standardized in vitro and in vivo toxicology tests (e.g., cytotoxicity, sensitization, hemolysis) with predefined acceptance criteria as per ISO 10993.
    • For pre-clinical animal testing: Direct physiological observation, macroscopic and microscopic evaluation of tissues and organs (pathology), and assessment of device functionality and complications in vivo.

    8. The sample size for the training set

    • Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not applicable as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1