Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K192241

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Halyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (Vidaza) (25 mg/mL) Bendamustine HCl (5 mg/mL) Bleomycin Sulfate (Blenoxane) (15 mg/mL) Bortezomib (Velcade) (1 mg/mL) Busulfan (6 mg/mL) Capecitabine (26 mg/mL) Carboplatin ( 10 mg/mL) Carfilzomib (2 mg/mL) Cetuximab (Erbitux) (2 mg/mL) Cisplatin (1 mg/mL) Cladribine (1 mg/mL) Cyclophosphamide (20 mg/mL) Cytarabine HCl (Cytosine) (100 mg/mL) Dacarbazine (DTIC) (10 mg/mL) Dactinomycin (0.5 mg/mL) Daunorubicin HCl (5 mg/mL) Decitabine ( 5 mg/mL) Docetaxel HCl (20 mg/mL) Doxorubicin HCl (2 mg/mL) Epirubicin HCl (Ellence) (2 mg/mL) Etoposide (20 mg/mL) Floxuridine (1 00mg/mL) 5-Fluorouracil (50 mg/mL) Gemcitabine HCl (38 mg/mL) Idarubicin HCl (1 mg/mL) lfosfamide (IFEX) (50 mg/mL) Irinotecan HCl (20 mg/mL) Lenvatinib (20mg/mL) Leuprolide Acetate Salt (5 mg/mL) Mechlorethamine HCl (1 mg/mL) Melphalan HCl (5 mg/mL) Methotrexate (25 mg/mL) Mitomycin C (0.5 mg/mL) Mitoxantrone HCl (2 mg/mL) Nelarabine (5 mg/mL) Oxaliplatin (5 mg/mL) Paclitaxel (Taxol) (6 mg/mL (6,000 ppm) Pemetrexed (25 mg/mL) Raltitrexed (0.5 mg/mL) Sorafenib Tosylate (200 mg/mL) Streptozocin (100mg/mL) Tamoxifen (2 mg/mL) Teniposide (10 mg/mL) Topotecan HCl (1 mg/mL) Trisenox (Arsenic Trioxide) (1 mg/mL) Vinblastine Sulfate (1 mg/mL) Vincristine Sulfate (Oncovin) (1 mg/mL) Vinorelbine (10 mg/mL)

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: Carmustine (3.3 mg/mL), breakthrough detected at 22.9 minutes

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 40 minutes: ThioTEPA (10 mg/mL), breakthrough detected at 37.1 minutes

    Warning: Do not use with Carmustine or ThioTEPA

    The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 mL) Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution

    The Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (Vidaza) (25 mg/mL) Bendamustine HCl (5 mg/mL) Bleomycin Sulfate (Blenoxane) (15 mg/mL) Bortezomib (Velcade) (1 mg/mL) Busulfan (6 mg/mL) Capecitabine (26 mg/mL) Carboplatin ( 10 mg/mL) Carfilzomib (2 mg/mL) Cetuximab (Erbitux) (2 mg/mL) Cisplatin (1 mg/mL) Cladribine (1 mg/mL) Cyclophosphamide (20 mg/mL) Cytarabine HCl (Cytosine) (100 mg/mL) Dacarbazine (DTIC) (10 mg/mL) Dactinomycin (0.5 mg/mL) Daunorubicin HCl (5 mg/mL) Decitabine ( 5 mg/mL) Docetaxel HCl (20 mg/mL) Doxorubicin HCl (2 mg/mL) Epirubicin HCl (Ellence) (2 mg/mL) Etoposide (20 mg/mL) Eribulin Mesylate (0.5 mg/mL) Fludarabine (25 mg/mL) 5-Fluorouracil (50 mg/mL) Gemcitabine HCl (38 mg/mL) Idarubicin HCl (1 mg/mL) lfosfamide (IFEX) (50 mg/mL) Irinotecan HCl (20 mg/mL) Lenvatinib (20 mg/mL) Leuprolide Acetate Salt (5 mg/mL) Mechlorethamine HCl (1 mg/mL) Melphalan HCl (5 mg/mL) Methotrexate (25 mg/mL) Mitomycin C (0.5 mg/mL) Mitoxantrone HCl (2 mg/mL) Oxaliplatin (5 mg/mL) Paclitaxel (Taxol) (6 mg/mL (6,000 ppm) Pemetrexed (25 mg/mL) Raltitrexed (0.5 mg/mL) Ritux1mab ( 10 mg/mL) Sorafenib Tosylate (200 mg/mL) Tamoxifen (2 mg/mL) Topotecan HCl (1 mg/mL) Trisenox (Arsenic Trioxide) (1 mg/mL) Vinblastine Sulfate (1 mg/mL) Vincristine Sulfate (Oncovin) (1 mg/mL) Vinorelbine (10 mg/mL)

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 20 minutes: Carmustine (BCNU) (3.3 mg/ml), breakthrough detected at 14.8 minutes

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: ThioTEPA (10 mg/ml), breakthrough detected at 23.9 minutes

    Warning: Do not use with Carmustine or ThioTEPA

    The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 mL) Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution

    The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes: Cyclosporin A (100 mg/mL) Cytovene (Ganciclovir) (10 mg/mL) Retrovir (Zidovudine) (10 mg/mL)

    The Halyard STERLING NITRILE-XTRA* Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs. Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (Vidaza) (25 mg/mL) Bendamustine HCl (5 mg/mL) Bleomycin Sulfate (Blenoxane) (15 mg/mL) Bortezomib (Velcade) (1 mg/mL) Busulfan (6 mg/mL) Capecitabine (26 mg/mL) Carboplatin ( 10 mg/mL) Carfilzomib (2 mg/mL) Cetuximab (Erbitux) (2 mg/mL) Cisplatin (1 mg/mL) Cladribine (1 mg/mL) Cyclophosphamide (20 mg/mL) Cytarabine HCl (Cytosine) (100 mg/mL) Dacarbazine (DTIC) (10 mg/mL) Dactinomycin (0.5 mg/mL) Daunorubicin HCl (5 mg/mL) Decitabine ( 5 mg/mL) Docetaxel HCl (20 mg/mL) Doxorubicin HCl (2 mg/mL) Epirubicin HCl (Ellence) (2 mg/mL) Etoposide (20 mg/mL) Floxuridine (100 mg/mL) Fludarabine (25 mg/mL) 5-Fluorouracil (50 mg/mL) Gemcitabine HCl (38 mg/mL) Idarubicin HCl (1 mg/mL) lfosfamide (IFEX) (50 mg/mL) Irinotecan HCl (20 mg/mL) Lenvatinib (20 mg/mL) Leuprolide Acetate Salt (5 mg/mL) Mechlorethamine HCl (1 mg/mL) Melphalan HCl (5 mg/mL) Methotrexate (25 mg/mL) Mitomycin C (0.5 mg/mL) Mitoxantrone HCl (2 mg/mL) Nelarabine (5 mg/mL) Oxaliplatin (5 mg/mL) Paclitaxel (Taxol) (6 mg/mL (6,000 ppm) Pemetrexed (25 mg/mL) Ralt1trexed (0.5 mg/mL) Rituximab ( 10 mg/mL) Sorafenib Tosylate (200 mg/mL) Streptozocin (100 mg/mL) Tamoxifen (2 mg/mL) Teniposide (10 mg/mL) Topotecan HCl (1 mg/mL) Trisenox (Arsenic Trioxide) (1 mg/mL) Vinblastine Sulfate (1 mg/mL) Vincristine Sulfate (Oncovin) (1 mg/mL) Vinorelbine (10 mg/mL)

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: Carmustine (3.3 mg/ml), breakthrough detected at 25.2 minutes

    The following chemotherapy drugs and concentration showed breakthrough detected in less than 40 minutes: ThioTEPA (10 mg/ml), breakthrough detected at 35.5 minutes

    Warning: Do not use with Carmustine or ThioTEPA

    The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 mL) Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution

    Device Description

    The subject devices are disposable, grey-colored, chlorinated, nitrile, powder-free, textured fingertip, ambidextrous, non-sterile patient examination gloves that are packed in a cardboard dispenser box.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the U.S. FDA for Halyard STERLING Nitrile Powder-Free Exam Gloves and related products. It outlines the regulatory classification, intended use, and a comparison of technological characteristics and performance data against predicate and reference devices to demonstrate substantial equivalence.

    Based on the content, this document describes physical and chemical testing of medical gloves, not an AI/ML-enabled medical device. Therefore, many of the requested
    details, such as sample size for test/training sets, data provenance, expert
    ground truth establishment, adjudication methods, MRMC studies, standalone
    algorithm performance, and specific effect sizes for human readers with/without
    AI assistance, are not applicable to this document.

    However, I can extract the acceptance criteria and reported device performance for these physical products based on the provided tables.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for these medical gloves are based on various ASTM and ISO standards for physical properties, residual powder, and biological safety. The "Reported device performance" directly aligns with the "Subject Devices" column in the "PERFORMANCE CHARACTERISTICS OF THE SUBJECT DEVICE" table.

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard & Requirement)Reported Device Performance (Subject Devices)
    ASTM D5151-06 Standard Test Method for Detection of Holes in Medical Gloves:
    • Meets 2.5% AQL (Acceptable Quality Level) requirement for leakage.Halyard STERLING Nitrile Powder-Free Exam Gloves, Halyard STERLING SG SENSI-GUARD Powder-Free Nitrile Exam Gloves, Halyard STERLING NITRILE-XTRA Powder-Free Exam Gloves:
    • Meets the 2.5% AQL requirement for leakage.
    ASTM D6124-06 Standard Test Method for Residual Powder on Medical Gloves:
    • Residual powder
    Ask a Question