Search Results
Found 1 results
510(k) Data Aggregation
(252 days)
The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic procedures, and sealing and transection of lymphatic vessels. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand control button.
The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments are sterile, single-patient-use instruments consisting of an ergonomic grip housing assembly with hand control buttons (MIN for minimum power level, MAX for maximum power level, and Advanced Hemostasis for large vessel sealing). An integrated audible and tactile mechanism in the grip housing indicates full trigger closure. The instruments have a clamp arm and coated curved blade that are designed to work through a 5 mm trocar, through a 5 mm reducer cap in a larger diameter trocar, or through an incision without the use of a trocar. The instrument shafts can be rotated 360° to facilitate visualization and access to targeted tissue. The three dashes on the instrument are intended to represent relative vessel size. The MAX button is typically used for smaller vessels where cutting speed is the fastest. The MIN button is typically used in slightly larger vessels and has reduced cutting speed. It is indicated for vessels up to 5 mm in size. The Advanced Hemostasis button is designed for larger vessels and is indicated for vessels up to 7 mm in size. In this mode, cutting speed is further reduced and hemostasis is maximized. This design is common to both subject and predicate devices and has been evaluated by FDA in K132612 (predicate device), K160752 (HARMONIC HD1000i), and K200841 (HARMONIC 1100). The instruments utilize Adaptive Tissue Technology. This provides the generator with the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate. The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are designed for use exclusively with the Generator G11 (GEN11) software version 2018-1 or later, last cleared under K200841 on 11 May 2020.
The provided text describes the HARMONIC 700 Shears and its substantial equivalence to a predicate device. Here's a breakdown of the acceptance criteria and supporting studies as presented in the document:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criterion (Category) | Specific Test/Metric | Reported Device Performance |
---|---|---|
Bench Testing | Tissue Pad Life | Evaluated, demonstrated substantial equivalence with the predicate device. |
Tissue Pad Removal Force | Evaluated, demonstrated substantial equivalence with the predicate device. | |
Instrument Grasping Force | Evaluated, demonstrated substantial equivalence with the predicate device. | |
Sealed Vessel Burst Pressure (vessels up to 7mm) | Evaluated on HARMONIC 700 Shears. Supplemental testing utilized porcine carotid arteries. The subject device's performance demonstrated substantial equivalence with the predicate device in vessels up to and including 7 mm. | |
Thermal Spread (MIN mode, MAX mode, Advanced Hemostasis) | Bench top testing evaluated vessels. The subject device performed as expected, and the results demonstrated substantial equivalence with the predicate device. | |
Preclinical Studies (Acute Animal) | Intra-operative tissue effects (vessels & pedicles) | Three acute porcine studies compared HARMONIC 700 Shears to the predicate (HARMONIC ACE+7 Shears). Arteries, veins, and pedicles of various sizes were evaluated. Results showed no statistical differences in intra-operative tissues effects between the three lengths of the subject device and the predicate. |
Thermal spread to vessel/vessel pedicles | Assessed in the acute porcine studies. | |
Preclinical Studies (Chronic Animal) | Post-operative performance (sealed blood vessels/pedicles) | A survival study compared HARMONIC 700 Shears to the predicate. Blood vessels and blood vessel pedicles were transected and sealed. The results demonstrated the subject device performed equivalent to the predicate. |
Biocompatibility | Patient-contacting materials | No new patient-contacting materials. All materials previously identified, reviewed, and cleared in the predicate device (K132612). Biocompatibility previously tested based on ISO 10993-1. |
Electrical Safety & EMC | Electrical Safety | Conducted in accordance with IEC 60601-1 and IEC 60601-2-2. |
Electromagnetic Compatibility | Conducted in accordance with IEC 60601-1-2. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
Sterilization/Shelf-Life | Sterility (SAL) | Sterilized to a 10^-6 sterility assurance level (SAL) through an EO sterilization process in accordance with ISO 11135. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
Shelf-life | Designated 5-years. |
2. Sample Size and Data Provenance for Test Set (Preclinical Studies)
- Acute Animal Testing: Performed in three acute porcine studies. The document does not specify the exact number of animals or vessels tested within these studies, but it mentions evaluating "arteries, veins, and pedicles of various sizes."
- Data Provenance: Porcine (animal) models. This is a form of prospective data collection for preclinical evaluation.
- Chronic Animal Testing: Performed in a survival study. The document does not specify the exact number of animals or vessels tested.
- Data Provenance: Animal (survival) model. This is a form of prospective data collection for preclinical evaluation.
3. Number of Experts and Qualifications for Ground Truth of Test Set
The document does not mention the use of human experts to establish ground truth for the test set in the context of the preclinical (animal) studies. The evaluation metrics (e.g., vessel burst pressure, thermal spread, intra-operative tissue effects, post-operative performance) are typically measured objectively by research staff or specific equipment.
4. Adjudication Method for Test Set
Not applicable, as ground truth for the preclinical studies was based on objective measurements rather than human reader interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence." The comparison is between the subject device and a predicate device through bench and animal testing.
6. Standalone (Algorithm Only) Performance
Not applicable. This device is a surgical instrument, not an AI or algorithm-based diagnostic/screening tool. The "Adaptive Tissue Technology" mentioned gives the generator "the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback," which is an integrated system feature, not a standalone algorithm being evaluated for performance against a ground truth.
7. Type of Ground Truth Used (for preclinical studies)
The ground truth for the preclinical studies was based on objective biological and mechanical outcomes measured in animal models (porcine).
- Acute Studies: Evaluation of "intra-operative tissues effects," "thermal spread," and the ability to "transected and sealed" vessels and pedicles.
- Chronic Studies: Evaluation of the long-term performance of "transected and sealed" blood vessels and blood vessel pedicles in a survival setting.
- Bench Testing: Direct physical measurements such as Tissue Pad Life, Tissue Pad Removal Force, Instrument Grasping Force, Sealed Vessel Burst Pressure, and objective measurements of thermal spread.
8. Sample Size for the Training Set
Not applicable. This device is a physical surgical instrument and does not involve AI/machine learning models that require a separate "training set" of data in the conventional sense for performance evaluation in this 510(k) submission. The "Adaptive Tissue Technology" is a functional design feature, not a trainable AI model in the context of this submission.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" as understood in AI/machine learning contexts for this device submission.
Ask a specific question about this device
Page 1 of 1