K Number
K221790
Date Cleared
2023-02-28

(252 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic procedures, and sealing and transection of lymphatic vessels. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand control button.

Device Description

The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments are sterile, single-patient-use instruments consisting of an ergonomic grip housing assembly with hand control buttons (MIN for minimum power level, MAX for maximum power level, and Advanced Hemostasis for large vessel sealing). An integrated audible and tactile mechanism in the grip housing indicates full trigger closure. The instruments have a clamp arm and coated curved blade that are designed to work through a 5 mm trocar, through a 5 mm reducer cap in a larger diameter trocar, or through an incision without the use of a trocar. The instrument shafts can be rotated 360° to facilitate visualization and access to targeted tissue. The three dashes on the instrument are intended to represent relative vessel size. The MAX button is typically used for smaller vessels where cutting speed is the fastest. The MIN button is typically used in slightly larger vessels and has reduced cutting speed. It is indicated for vessels up to 5 mm in size. The Advanced Hemostasis button is designed for larger vessels and is indicated for vessels up to 7 mm in size. In this mode, cutting speed is further reduced and hemostasis is maximized. This design is common to both subject and predicate devices and has been evaluated by FDA in K132612 (predicate device), K160752 (HARMONIC HD1000i), and K200841 (HARMONIC 1100). The instruments utilize Adaptive Tissue Technology. This provides the generator with the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate. The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are designed for use exclusively with the Generator G11 (GEN11) software version 2018-1 or later, last cleared under K200841 on 11 May 2020.

AI/ML Overview

The provided text describes the HARMONIC 700 Shears and its substantial equivalence to a predicate device. Here's a breakdown of the acceptance criteria and supporting studies as presented in the document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Category)Specific Test/MetricReported Device Performance
Bench TestingTissue Pad LifeEvaluated, demonstrated substantial equivalence with the predicate device.
Tissue Pad Removal ForceEvaluated, demonstrated substantial equivalence with the predicate device.
Instrument Grasping ForceEvaluated, demonstrated substantial equivalence with the predicate device.
Sealed Vessel Burst Pressure (vessels up to 7mm)Evaluated on HARMONIC 700 Shears. Supplemental testing utilized porcine carotid arteries. The subject device's performance demonstrated substantial equivalence with the predicate device in vessels up to and including 7 mm.
Thermal Spread (MIN mode, MAX mode, Advanced Hemostasis)Bench top testing evaluated vessels. The subject device performed as expected, and the results demonstrated substantial equivalence with the predicate device.
Preclinical Studies (Acute Animal)Intra-operative tissue effects (vessels & pedicles)Three acute porcine studies compared HARMONIC 700 Shears to the predicate (HARMONIC ACE+7 Shears). Arteries, veins, and pedicles of various sizes were evaluated. Results showed no statistical differences in intra-operative tissues effects between the three lengths of the subject device and the predicate.
Thermal spread to vessel/vessel pediclesAssessed in the acute porcine studies.
Preclinical Studies (Chronic Animal)Post-operative performance (sealed blood vessels/pedicles)A survival study compared HARMONIC 700 Shears to the predicate. Blood vessels and blood vessel pedicles were transected and sealed. The results demonstrated the subject device performed equivalent to the predicate.
BiocompatibilityPatient-contacting materialsNo new patient-contacting materials. All materials previously identified, reviewed, and cleared in the predicate device (K132612). Biocompatibility previously tested based on ISO 10993-1.
Electrical Safety & EMCElectrical SafetyConducted in accordance with IEC 60601-1 and IEC 60601-2-2.
Electromagnetic CompatibilityConducted in accordance with IEC 60601-1-2. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Sterilization/Shelf-LifeSterility (SAL)Sterilized to a 10^-6 sterility assurance level (SAL) through an EO sterilization process in accordance with ISO 11135. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Shelf-lifeDesignated 5-years.

2. Sample Size and Data Provenance for Test Set (Preclinical Studies)

  • Acute Animal Testing: Performed in three acute porcine studies. The document does not specify the exact number of animals or vessels tested within these studies, but it mentions evaluating "arteries, veins, and pedicles of various sizes."
    • Data Provenance: Porcine (animal) models. This is a form of prospective data collection for preclinical evaluation.
  • Chronic Animal Testing: Performed in a survival study. The document does not specify the exact number of animals or vessels tested.
    • Data Provenance: Animal (survival) model. This is a form of prospective data collection for preclinical evaluation.

3. Number of Experts and Qualifications for Ground Truth of Test Set

The document does not mention the use of human experts to establish ground truth for the test set in the context of the preclinical (animal) studies. The evaluation metrics (e.g., vessel burst pressure, thermal spread, intra-operative tissue effects, post-operative performance) are typically measured objectively by research staff or specific equipment.

4. Adjudication Method for Test Set

Not applicable, as ground truth for the preclinical studies was based on objective measurements rather than human reader interpretation requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence." The comparison is between the subject device and a predicate device through bench and animal testing.

6. Standalone (Algorithm Only) Performance

Not applicable. This device is a surgical instrument, not an AI or algorithm-based diagnostic/screening tool. The "Adaptive Tissue Technology" mentioned gives the generator "the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback," which is an integrated system feature, not a standalone algorithm being evaluated for performance against a ground truth.

7. Type of Ground Truth Used (for preclinical studies)

The ground truth for the preclinical studies was based on objective biological and mechanical outcomes measured in animal models (porcine).

  • Acute Studies: Evaluation of "intra-operative tissues effects," "thermal spread," and the ability to "transected and sealed" vessels and pedicles.
  • Chronic Studies: Evaluation of the long-term performance of "transected and sealed" blood vessels and blood vessel pedicles in a survival setting.
  • Bench Testing: Direct physical measurements such as Tissue Pad Life, Tissue Pad Removal Force, Instrument Grasping Force, Sealed Vessel Burst Pressure, and objective measurements of thermal spread.

8. Sample Size for the Training Set

Not applicable. This device is a physical surgical instrument and does not involve AI/machine learning models that require a separate "training set" of data in the conventional sense for performance evaluation in this 510(k) submission. The "Adaptive Tissue Technology" is a functional design feature, not a trainable AI model in the context of this submission.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" as understood in AI/machine learning contexts for this device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA logo, which includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 28, 2023

Ethicon Endo-Surgery, LLC Hillary Lewis Regulatory Affairs Specialist 4545 Creek Road Cincinnati. Ohio 45242

Re: K221790

Trade/Device Name: HARMONIC 700 Shears Regulatory Class: Unclassified Product Code: LFL Dated: January 26, 2023 Received: January 27, 2023

Dear Hillary Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting

{1}------------------------------------------------

combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MarkDigitally signed by
MarkMark Trumbore -S
Trumbore -S Date: 2023.02.2815:44:07 -05'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221790

Device Name HARMONIC™ 700 Shears

Indications for Use (Describe)

The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in genecal, pediatric, gynecologic, thoracic procedures, and sealing and transection of lymphatic vessels. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand control button.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Company Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969

Contact Hillary Lewis Regulatory Affairs Specialist, Regulatory Affairs Ethicon Endo-Surgery, LLC. Telephone: (224) 345-1077 Email: hlewis6(@its.jnj.com

Date Prepared

28 February 2023

Trade Name:HARMONIC® 700 Shears
Common Name:Instrument, Ultrasonic Surgical
Classification Regulation:Unclassified
Classification Name:Instrument, Ultrasonic Surgical
Product Code:LFL
Device Class:Class II
Panel:79, General and Plastic Surgery

Device and Classification Information - HARMONIC 700 Shears

Device Description - HARMONIC 700 Shears

The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are intended for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments are sterile, single-patient-use instruments consisting of an ergonomic grip housing assembly with hand control buttons (MIN for minimum power level, MAX for maximum power level, and Advanced Hemostasis for large vessel sealing).

An integrated audible and tactile mechanism in the grip housing indicates full trigger closure. The instruments have a clamp arm and coated curved blade that are designed to work through a 5 mm trocar, through a 5 mm reducer cap in a larger diameter trocar, or through an incision without the use of a trocar. The instrument shafts can be rotated 360° to facilitate visualization and access to targeted tissue. The three dashes on the instrument are intended to represent

{4}------------------------------------------------

relative vessel size. The MAX button is typically used for smaller vessels where cutting speed is the fastest. The MIN button is typically used in slightly larger vessels and has reduced cutting speed. It is indicated for vessels up to 5 mm in size. The Advanced Hemostasis button is designed for larger vessels and is indicated for vessels up to 7 mm in size. In this mode, cutting speed is further reduced and hemostasis is maximized. This design is common to both subject and predicate devices and has been evaluated by FDA in K132612 (predicate device), K160752 (HARMONIC HD1000i), and K200841 (HARMONIC 1100).

The instruments utilize Adaptive Tissue Technology. This provides the generator with the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate.

The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are designed for use exclusively with the Generator G11 (GEN11) software version 2018-1 or later, last cleared under K200841 on 11 May 2020.

Indications for Use - HARMONIC 700 Shears

The HARMONIC 700, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic procedures, and sealing and transection of lymphatic vessels. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand control button.

Intended Use - HARMONIC 700 Shears

The subject device and the predicate device have the same intended use. Both devices are ultrasonic surgical instruments intended to cut and control bleeding from soft tissues and vessels during open and endoscopic procedures. Both are intended for use as an adjunct to or substitute for electrosurgery, lasers or manual scalpels in several different clinical specialties, for sealing and transecting vessels, and for dissection. Both devices coagulate vessels up to and including 7 mm in diameter. Both devices are a surgical tool used by surgeons during open and endoscopic procedures that share the same function to simultaneously cut and cauterize tissue and vessels using ultrasonic vibration. Both devices have the same tissue/body contact and limited duration of use from a biocompatibility perspective.

Predicate Device and Substantial Equivalence Comparison - HARMONIC 700 Shears

For purposes of the substantial equivalence comparisons, the following predicate device was selected:

· HARMONIC ACE+7 Shears with Advanced Hemostasis cleared under K132612 on 17 October 2013

{5}------------------------------------------------

The purpose of this 510(k) submission is to notify the Agency of the intent to commercialize a new device, HARMONIC 700 Shears.

Technological Characteristics: The fundamental technological characteristics (i.e., design, material, chemical composition, energy source) of the HARMONIC 700 Shears are equivalent to the predicate. The subject device uses a different version of GEN11 software (2018-1) than the predicate (2013 1).

Comparison of Technological Characteristics with the Predicate Device:

Performance Data: Verification testing for the design modifications was provided and raises no new issues of safety or effectiveness.

Biocompatibility Testing: There were no new patient-contacting materials for the subject device, and all patient contacting materials for the subject device were previously identified, provided, reviewed, and cleared in the predicate device under K132612. The biocompatibility of the patient contact materials was previously tested based on ISO 10993-1: Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing.

Electromagnetic Compatibility and Electrical Safety: Electrical safety testing and electromagnetic compatibility was conducted for the subject device in accordance with IEC 60601-1-2 for electromagnetic compatibility and IEC 60601-1 and IEC 60601-2-2 for electrical safety.

Sterilization/Shelf-Life: No new changes have been made to the sterilization process or parameters; HARMONIC 700 Shears are sterilized to a 10th sterility assurance level (SAL) through an EO sterilization process in accordance with ISO 11135. There have been no new materials, either patient contacting or non-patient contacting, added to the subject device that would impact the previously validated sterilization process or parameters reported from the predicate device, K132612. The designated shelf-life for HARMONIC 700 Shears is 5-years.

Bench Testing: Tissue Pad Life, Tissue Pad Removal Force, Instrument Grasping Force, and Sealed Vessel Burst Pressure were evaluated on HARMONIC 700 Shears to determine substantial equivalence to the predicate device. The mechanical design between the subject and predicate device is identical. Supplemental Sealed Vessel Burst Pressure testing evaluated the device performance and ability to divide and seal vessels. Porcine carotid arteries were utilized in this testing. Thermal Spread bench top testing evaluated vessels utilizing the MIN mode, MAX mode, and Advanced Hemostasis. The subject device performed as expected, the results of the bench testing demonstrated substantial equivalence with the predicate device in vessels up to and including 7 mm.

{6}------------------------------------------------

Preclinical Studies:

Acute Animal Testing

Acute Animal Testing was performed in three acute porcine studies to evaluate the three lengths of the subject device, HARMONIC 700 Shears as compared to the predicate (control) device, HARMONIC ACE+7 Shears. Arteries, veins, and pedicles of various sizes were evaluated to determine equivalency of the subject device to the predicate. Additionally, the thermal spread to the vessel/vessel pedicles was assessed. Vessels were transected and sealed using the MIN/MAX Modes and the Advanced Hemostasis Mode. The results of the acute studies demonstrated no statistical differences in the intra-operative tissues effects between the three devices lengths. Results of the testing between the three lengths supported the equivalency of the HARMONIC 700 Shears to the predicate device.

Chronic Animal Testing

Chronic Animal Testing was performed in a survival study with the subject device, HARMONIC 700 Shears as compared to the predicate device. Blood vessels and blood vessel pedicles were The results of the survival study demonstrated the subject device transected and sealed. performed equivalent to the predicate.

Clinical Testing

This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence

Conclusion: In demonstrating substantial equivalence of the subject HARMONIC 700 Shears to the predicate device, Ethicon Endo-Surgery, LLC. evaluated the indication for use, intended use and technological characteristics. The HARMONIC 700 Shears, as indicated for vessel/vessel pedicles up to and including 7 mm, are substantially equivalent to the predicate device; they share the same intended use and equivalent technological characteristics. The subject device is as safe and effective as the predicate device.

N/A