Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K170783
    Device Name
    CX50N(CX50YQS)
    Manufacturer
    Date Cleared
    2017-03-31

    (16 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K160351

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CX50N(CX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifically designed for digital mammography applications and digital breast tomosynthesis applications.

    Device Description

    CX50N(CX50YQS) is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for color image display. The WIDE controller board is installed into the PACS workstation computer system to display PACS medical images.

    AI/ML Overview

    The provided text describes the WIDE Corporation CX50N(CX50YQS) LCD Monitor System, which displays medical images for review and analysis, specifically for digital mammography and digital breast tomosynthesis. The document is a 510(k) summary indicating substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study conducted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    ParameterAcceptance Criteria (Predicate Device K160326)Reported Device Performance (CX50N(CX50YQS))
    Intended UseDisplaying and viewing medical images for diagnosis in digital mammography PACS, FFDM, and breast tomosynthesis.Displaying and viewing digital medical images for review and analysis by trained medical practitioners, specifically designed for digital mammography and digital breast tomosynthesis.
    Response Time25ms (On/Off)25ms (On/Off)
    LCD Panel Size21.3"21.3"
    Resolution2560 x 20482560 x 2048
    Pixel Pitch0.165mm x 0.165mm0.165mm x 0.165mm
    Brightness1000cd/m²800cd/m²
    Contrast Ratio1300 : 1800 : 1
    Input SignalDVI-I, DisplayPortDVI-I, DisplayPort
    Power Supply100~240 VAC, 50/60Hz100~240 VAC, 50/60Hz
    Color/MonochromeColorColor
    SafetyCompliance with IEC 60601-1 and IEC 60601-1-2 (Electromagnetic Compatibility) standards.Compliance achieved with IEC 60601-1 and IEC 60601-1-2.
    EffectivenessMeeting acceptance criteria for resolution, luminance, contrast, and noise as specified in standards.Tests conducted for resolution, luminance, contrast, and noise met the acceptance criteria specified in the standards.

    2. Sample size used for the test set and the data provenance:

    The document does not specify a "test set" in the context of medical images or patient data. The testing described focuses on the monitor's technical specifications and compliance with international standards for electrical safety and electromagnetic compatibility. Therefore, there is no mention of data provenance (e.g., country of origin, retrospective/prospective) related to imaging data. The device itself (the monitor) underwent non-clinical testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Since this is a medical display device, not an AI-powered diagnostic tool, the concept of "ground truth" derived from expert interpretation of medical images does not apply in the context of the described testing. The 'ground truth' for the acceptance criteria would be the established technical standards and performance specifications for medical displays.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable, as there was no test set involving human interpretation of medical images.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a medical monitor, not an AI-assisted diagnostic tool. No MRMC comparative effectiveness study was reported.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is an LCD monitor; it does not perform algorithmic analysis or have a standalone diagnostic function. Its performance is evaluated based on its ability to accurately display images.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the safety and effectiveness testing of the monitor, the "ground truth" was established by international and FDA-recognized consensus standards (IEC 60601-1, IEC 60601-1-2) which define the technical performance requirements for medical electrical equipment and displays. The tests for resolution, luminance, contrast, and noise are objective measurements against these predefined technical specifications.

    8. The sample size for the training set:

    Not applicable. This device is a hardware monitor, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1