Search Results
Found 4 results
510(k) Data Aggregation
(176 days)
The Venture 0.014" catheters are indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the coronary and peripheral vasculature. The OTW version (Model 5821) may also be used for manual delivery of saline solution or diagnostic contrast agents.
The Venture 0.014" catheters are single lumen catheters with a radiopaque deflectable distal tip designed to provide guidewire support during interventional procedures in the coronary and peripheral vasculature. The Venture 0.014" catheters are compatible with 0.014" guidewires and are available in two configurations. The rapid exchange configuration consists of a hypotube shaft with a rapid exchange (RX) lumen and deflectable tip on the distal segment. The over the wire configuration consists of an over-the-wire (OTW) lumen that runs the entire length of the catheter with a deflectable tip. The Venture 0.014" catheter tips are deflectable 90° by rotating a deflection knob on the control handle of the proximal end and a hydrophilic coating on the distal end of the catheter enhances deliverability to the target vasculature. The Venture 0.014" catheters are sterilized with ethylene oxide.
This is a 510(k) summary for a medical device (Venture 0.014" Catheter), not an AI/ML device study. Therefore, many of the requested categories for AI/ML device studies are not applicable or cannot be extracted from this document.
However, I can extract the acceptance criteria and performance data for the device's functional integrity as described in the document.
1. Table of Acceptance Criteria and Reported Device Performance
Performance Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Deliver, Deflect, and Retract | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Guidewire Steerability | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Tip Deflection Angle | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Tip Rebound | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Tip Rotation | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Tip Tensile | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Surface Defects | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Guidewire Movement | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Torque Robustness | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Luer to Extension Tensile | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Extension to Back Handle Tensile | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Liquid Leakage from Fitting Assembly under Pressure | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Static Pressure | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Coating Lubricity/Durability | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Throw Length | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Particulates | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
HPC Fluid Drops | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Therapeutic Agent Conditioning | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Dynamic Pressure | Met specified criteria (implicit from conclusion) | Results met specified acceptance criteria. |
Biocompatibility Testing (ISO 10993-1): | ||
Cytotoxicity | Non-cytotoxic | Device is non-cytotoxic. |
Sensitization | Non-sensitizing | Device is non-sensitizing. |
Irritation | Non-irritating | Device is non-irritating. |
Material Mediated Pyrogenicity | Non-pyrogenic (implied by "Material Mediated Pyrogenicity") | Not explicitly stated but implied by "Passing results...demonstrate...". |
Hemolysis | Non-hemolytic | Device is non-hemolytic. |
Complement Activation | Not an activator of the complement system | Device is not an activator of the complement system. |
Acute Systemic Toxicity | Non-systemically toxic | Device is non-systemically toxic. |
Thrombogenicity | Thromboresistant | Device is thromboresistant. |
2. Sample size used for the test set and the data provenance:
- The document does not specify exact sample sizes for each bench test. It generically states "The results of the verification tests met the specified acceptance criteria."
- Data provenance is not explicitly stated, but these are bench tests performed by the manufacturer, Vascular Solutions, LLC. This is not a study on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This is not an AI/ML study involving expert ground truth for medical images or clinical outcomes. These are engineering bench tests.
4. Adjudication method for the test set:
- Not applicable. This is not an AI/ML study involving expert interpretation or adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the bench tests, the "ground truth" is established by engineering specifications and standards for device performance (e.g., specific force, angle, integrity measurements).
- For biocompatibility, the ground truth is established by the results of standardized biological tests according to ISO 10993-1.
8. The sample size for the training set:
- Not applicable. This is not an AI/ML device, so there is no training set in the AI sense.
9. How the ground truth for the training set was established:
- Not applicable.
Ask a specific question about this device
(30 days)
The Venture 038 catheter is indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the peripheral vasculature. It may also be used for manual delivery of saline solution or diagnostic contrast agents.
The Venture 038 catheter is an over-the-wire single lumen catheter with a radiopaque deflectable distal tip. The deflectable distal tip is actuated by rotating a knob on the control handle of the proximal end of the catheter. The Venture 038 catheter has a working length of 120 cm or 70 cm and has a hydrophilic coating on the distal 45 cm of the catheter. The Venture 038 catheter provides support to 0.035" guidewires and is compatible with 6F introducer sheaths and 8F guide catheters.
The provided document is a 510(k) summary for the Venture 038 catheter, a medical device. It does not describe an AI/ML powered device or a study involving human readers or AI assistance. Therefore, many of the requested fields are not applicable.
1. Table of acceptance criteria and the reported device performance
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Simulated Use | Not specified (implied to meet functional requirements) | Passed |
Kink Radius | Not specified (implied to meet functional requirements) | Passed |
Visual Inspection | Not specified (implied to meet functional requirements) | Passed |
Tip Deflection | Not specified (implied to meet functional requirements) | Passed |
Tip Rotation | Not specified (implied to meet functional requirements) | Passed |
Tensile Strength | Not specified (implied to meet functional requirements) | Passed |
Liquid Leak | Not specified (implied to meet functional requirements) | Passed |
Liquid Leak Under Pressure | Not specified (implied to meet functional requirements) | Passed |
Injection Pressure | Not specified (implied to meet functional requirements) | Passed |
Aspiration | Not specified (implied to meet functional requirements) | Passed |
Corrosion | Not specified (implied to meet functional requirements) | Passed |
Particulate Count | Not specified (implied to meet functional requirements) | Passed |
Biocompatibility (Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Pyrogenicity, Hemolysis, Complement Activation, In Vitro Hemocompatibility, Coagulation, Thrombogenicity) | Per ISO 10993-1 | Passed |
Note: The document states "The results of the verification tests met the specified acceptance criteria," but it does not explicitly list the specific quantitative or qualitative acceptance criteria for each test. Instead, it generally indicates that the device passed these tests.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document describes bench tests and biocompatibility tests, not studies with test sets in the context of AI/ML or clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable as the document describes a physical medical device and its bench testing, not an AI/ML algorithm requiring expert ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the document does not describe a study involving adjudication of a test set by human experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable as the document describes a physical medical device and its bench testing, not an AI/ML powered device or a MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable as the document describes a physical medical device and its bench testing, not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This information is not applicable as the document describes a physical medical device and its bench testing, not an AI/ML algorithm requiring a ground truth in that context. The "ground truth" for the device's performance would be the physical properties and functional results observed during the bench tests, compared against design specifications and relevant standards.
8. The sample size for the training set
This information is not applicable as the document describes a physical medical device and its bench testing, not an AI/ML algorithm requiring a training set.
9. How the ground truth for the training set was established
This information is not applicable as the document describes a physical medical device and its bench testing, not an AI/ML algorithm.
Ask a specific question about this device
(152 days)
The Twin-Pass catheter is intended to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and to subselectively infuse/deliver diagnostic agents.
The Twin-Pass Torque is a dual lumen catheter consisting of an over-the-wire (OTW) lumen that runs the full length of the catheter, exiting just proximal to the distal tip, and a rapid exchange (RX) delivery lumen on the distal segment. The two exit ports are each marked by a radiopaque platinum-iridium marker band and the device shaft has two positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, which are not visible under fluoroscopy.
This document is a 510(k) premarket notification for the "Twin-Pass Torque" catheter. It focuses on demonstrating substantial equivalence to a predicate device rather than detailing a clinical study with acceptance criteria for device performance based on patient outcomes or diagnostic accuracy. Therefore, information regarding reader studies, ground truth establishment, or multi-reader multi-case analyses is not applicable.
Here's an breakdown of the available information:
1. A table of acceptance criteria and the reported device performance
The document lists performance tests conducted to establish substantial equivalence. For each test, the document states that "The results of the verification tests met the specified acceptance criteria." However, the specific numerical acceptance criteria or detailed numerical performance results are not provided in this summary.
Acceptance Criteria (General) | Reported Device Performance (General) |
---|---|
Met specified acceptance criteria | Passed Kink resistance |
Met specified acceptance criteria | Passed Guidewire insertion |
Met specified acceptance criteria | Passed Tip flexibility |
Met specified acceptance criteria | Passed Distal shaft flexibility |
Met specified acceptance criteria | Passed Proximal shaft support |
Met specified acceptance criteria | Passed Coating lubricity/Durability |
Met specified acceptance criteria | Passed Radiopacity |
Met specified acceptance criteria | Passed Guidewire deflection |
Met specified acceptance criteria | Passed Distal tip length |
Met specified acceptance criteria | Passed Torque control |
Met specified acceptance criteria | Passed Torque transmission |
Met specified acceptance criteria | Passed Tensile strength |
Met specified acceptance criteria | Passed Torque strength |
Met specified acceptance criteria | Passed Torque robustness |
Met specified acceptance criteria | Passed Hub luer tests (air leak, burst, compatibility) |
Met specified acceptance criteria | Passed Hydrophilic coating particulate |
Met specified acceptance criteria | Passed Package integrity |
Met specified acceptance criteria (ISO 10993-1) | Passed Cytotoxicity |
Met specified acceptance criteria (ISO 10993-1) | Passed Sensitization |
Met specified acceptance criteria (ISO 10993-1) | Passed Irritation |
Met specified acceptance criteria (ISO 10993-1) | Passed Systemic toxicity |
Met specified acceptance criteria (ISO 10993-1) | Passed Pyrogenicity |
Met specified acceptance criteria (ISO 10993-1) | Passed Hemocompatibility |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document describes performance tests (bench testing), not a clinical trial with a "test set" in the context of patient data. The sample sizes for these engineering and biocompatibility tests are not specified in this summary. The data provenance is also not specified, but given it's bench testing, it would originate from the manufacturer's testing facility.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is not a study assessing diagnostic accuracy or human interpretation. It's about mechanical and biological performance of a medical device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document does not describe a MRMC study or an AI-powered device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This document is for a physical medical device (catheter), not an algorithm or AI.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
For the performance tests, the "ground truth" would be the engineering and physical standards and specifications established for the device's functional characteristics (e.g., kink resistance, tensile strength) and the biological response standards for biocompatibility (e.g., ISO 10993-1).
8. The sample size for the training set
Not applicable. There is no software or algorithm involved that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(99 days)
The CenterCross ULTRA Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices.
The CenterCross Ultra Catheter is a sterile, single lumen over-the-wire support catheter to be used in conjunction with steerable guidewires to access discrete regions of the coronary and peripheral vasculature to facilitate exchange of guidewires and other interventional devices.
The CenterCross Ultra Catheter consists of an outer shaft and a removable handle/inner shaft assembly that allows for manual device manipulation. A key element of the device is an expandable and retractable nitinol structure, which, when deployed by the physician, expands to the artery wall to aid interventionalists in establishing greater support near the treatment site.
Subsequent to conventional guidewire placement, atherectomy devices, PTCA catheters, and/or stents may be used to provide therapeutic benefit. The CenterCross Ultra Catheter in and of itself does not provide therapeutic beyond simple facilitation of guidewire and device support. The CenterCross Ultra Catheter is similar in its design and achieves its intended use by means of the same mechanisms as the predicate device.
This document is a 510(k) premarket notification for the Roxwood Medical CenterCross Ultra Catheter. As such, it is focusing on demonstrating substantial equivalence to a predicate device rather than presenting a study where acceptance criteria for a novel device performance were established and met.
Therefore, many of the requested categories (such as sample size for test set, data provenance, number and qualifications of experts, adjudication method, MRMC study, training set information, and direct effect size of AI improvement) are not applicable to the information provided in this document. This document describes bench testing for performance and characterization, not clinical study results evaluating diagnostic accuracy or reader performance with an AI device.
Here's an extraction of the relevant information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of "acceptance criteria" paired with "reported device performance" in a manner typical for a clinical study with specific metrics like sensitivity, specificity, or reader improvement. Instead, it states that "All testing was performed in accordance with recognized standards" and the "collective results of the non-clinical testing demonstrate that the CenterCross Ultra Catheter meets the established specifications necessary for consistent performance for its intended use and is substantially equivalent to the predicate device."
The types of testing performed and confirmed to meet specifications are:
Acceptance Criteria Category | Reported Device Performance |
---|---|
Dimensional Inspection | Met established specifications |
Expansion Force | Met established specifications |
Simulated Use | Met established specifications |
Hydrophilic Coating | Met established specifications |
Leak | Met established specifications |
Contrast Flush Injection | Met established specifications |
Tip Flexibility & Kink Resistance | Met established specifications |
Torque Strength | Met established specifications |
Tensile Strength | Met established specifications |
Particulate | Met established specifications |
Radiopacity | Met established specifications |
Corrosion | Met established specifications |
LAL (Limulus Amebocyte Lysate) | Met established specifications |
EO Residuals (Ethylene Oxide) | Met established specifications |
Packaging and Shelf-Life | Met established specifications |
Biocompatibility | Met established specifications |
2. Sample Size Used for the Test Set and Data Provenance
Not applicable. This document refers to bench testing of a medical device, not a study involving human subjects or data collected from patients. Therefore, there is no "test set" in the context of patient data, nor data provenance (country of origin, retrospective/prospective).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Ground truth for a device's physical and mechanical performance is established through engineering specifications and recognized testing standards, rather than expert consensus on patient data.
4. Adjudication Method for the Test Set
Not applicable. This concept applies to human reader studies where disagreement needs to be resolved.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This is a conventional medical device (catheter), not an AI-powered diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the loop performance) Was Done
Not applicable. This is not an algorithm or AI device.
7. The Type of Ground Truth Used
The "ground truth" for the device performance is based on established engineering specifications, recognized industry standards, and relevant regulatory requirements for the physical, mechanical, and biological properties of percutaneous catheters. This includes objective measurements of dimensions, forces, material integrity, and biological safety as detailed in the testing categories.
8. The Sample Size for the Training Set
Not applicable. This is a medical device, not an AI model requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable.
Ask a specific question about this device
Page 1 of 1