Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112382
    Date Cleared
    2011-12-14

    (118 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K083784, K061958

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Terumo Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

    Device Description

    The Terumo Aspiration Catheter is a dual lumen rapid exchange catheter. The guidewire lumen is used to facilitate passage of a guide wire which must not exceed 0.014" (0.36 mm) in diameter. The larger extraction lumen allows the removal of thrombus (thrombi) by use of the included aspiration syringe through the extension line. The catheter has a proximal stiff region and a distal flexible region that is coated with hydrophilic polymer which generates lubricity when wet. On the distal tip a radiopaque marker band is incorporated. The proximal end of the catheter is equipped with a standard luer adapter to facilitate the attachment of the included extension line, stopcock and syringes. The provided stylet can be inserted in the catheter to assist in the delivery of the catheter to the vascular lesion. The included flushing tool is used to flush the guide wire lumen in preparation for use. A filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis or any thrombosis.

    AI/ML Overview

    The Terumo Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. The acceptance criteria and supporting studies are outlined below.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Bench Testing)Reported Device Performance
    ISO 10555-1 Conformance Tests
    Surface inspectionConformance to ISO 10555-1 standardsPassed
    Corrosion resistanceConformance to ISO 10555-1 standardsPassed
    Force at break (shaft, hub)Minimal force at break of 4N (more stringent than 3N requirement of ISO 10555-1)Passed
    Freedom from leakageConformance to ISO 10555-1 standardsPassed
    Shelf Life Performance Tests (t=0 vs. t=shelf life)
    Catheter force at breakNo significant changes in performanceMet all specifications; no significant changes observed
    Catheter lubricityNo significant changes in performanceMet all specifications; no significant changes observed
    Dimensional verificationNo significant changes in performanceMet all specifications; no significant changes observed
    Stylet coil bond tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
    Stylet to connector tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
    Extension line tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
    In-House Comparative Performance Tests
    Kink resistanceEquivalent or superior to predicate devices (PRONTO V3, Medtronic Export XT)Equivalent or superior to predicate devices
    Aspiration rateEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
    Thrombus aspiration capabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
    PushabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
    TrackabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
    Accessory Conformance Tests
    StyletConformance to ISO 10555-1 standardsPassed
    Stylet connectorConformance to ISO 594-1 & 2 standardsPassed
    Aspiration syringeConformance to ISO 7886-1 standardsPassed
    Aspiration syringe luerConformance to ISO 594-1 & 2 standardsPassed
    Extension LineConformance to ISO 10555-1 standardsPassed
    Extension Line luers and stopcockConformance to ISO 594-1 & 2 standardsPassed
    Flushing toolConformance to ISO 594-1 & 2 standardsPassed
    Biocompatibility Tests
    SensitizationMeets the requirements of ISO 10993-10Meets the requirements
    HemolysisNon-hemolytic (ASTM F756)Non-hemolytic
    Dog thrombo-resistanceNon-thrombogenic (ISO 10993-4)Non-thrombogenic
    Complement ActivationMeets the requirements of ISO 10993-4Meets the requirements
    Ames AssayMeets the requirements of ISO 10993-3Meets requirements
    Lymphoma Forward MutationNon-mutagenic (ISO 10993-3)Non-mutagenic
    Bone Marrow MicronucleusNon-clastogenic (ISO 10993-3)Non-clastogenic
    Rabbit PyrogenNon-pyrogenic (ISO 10993-11)Non-pyrogenic
    CytotoxicityNon-cytotoxic (ISO 10993-5)Non-cytotoxic
    Intracutaneous reactivityMeets the requirements of ISO 10993-10Meets requirements
    Acute Systemic toxicityNon-toxic (ISO 10993-11)Non-toxic
    Physicochemical profile (aged, sterile)Meets the requirements of USP 661Meets the requirements
    Cytotoxicity (aged, sterile)Non-cytotoxic (ISO 10993-5)Non-cytotoxic
    Hemolysis (aged, sterile)Non-hemolytic (ASTM F756)Non-hemolytic
    Sterilization and Residuals
    Sterility Assurance Level (SAL)10⁻⁶ (validated per ANSI/AAMI/ISO 11135)Achieved 10⁻⁶ SAL
    Ethylene Oxide (EO) residuals≤ 4 mg per device (per ANSI/AAMI/ISO 10993-7)Will meet requirements (prior to use)
    Ethylene Chlorohydrin (ECH) residuals≤ 9 mg per device (per ANSI/AAMI/ISO 10993-7)Will meet requirements (prior to use)
    Shelf Life36-months (based on performance testing and package integrity validation)36-months

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes bench testing and biocompatibility testing, not a clinical study with a "test set" in the context of patient data.

    • Bench Testing: The sample sizes for each specific bench test (e.g., force at break, aspiration rate) are not explicitly stated. However, the tests were conducted using both real-time aged and accelerated aged samples to assess performance over the specified shelf life.
    • Biocompatibility Testing: The number of samples for each biocompatibility test (e.g., Sensitization, Hemolysis, Pyrogen) is not explicitly stated. The tests were performed on non-aged (sterile) and aged (sterile) samples.
    • Data Provenance: The studies are described as "in-house bench testing" and "biocompatibility testing." This implies the data were generated in a laboratory setting, likely at the manufacturer's (Terumo Medical Corporation) or a contracted testing facility's site, and are not derived from human patients or a specific country of origin in the clinical sense. These are pre-market studies to demonstrate safety and performance characteristics.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The reported studies are bench and biocompatibility tests, which do not involve establishing ground truth by human experts in the context of diagnostic interpretation or clinical outcomes. The "ground truth" for these tests refers to established scientific and engineering standards (e.g., ISO, ASTM).

    4. Adjudication Method for the Test Set

    Not applicable. As these are bench and biocompatibility tests, there is no human adjudication process involved in the data collection or interpretation in the way there would be for a clinical trial or image analysis study. The results are compared against predefined pass/fail criteria from international and industry standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The provided document details pre-market bench and biocompatibility studies for a physical medical device (aspiration catheter). It does not describe an AI/software as a medical device or diagnostic imaging study that would typically involve an MRMC comparative effectiveness study to assess human reader performance with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. The Terumo Aspiration Catheter is a physical medical device, not an algorithm or software. Therefore, there is no "standalone" algorithm performance study described.

    7. Type of Ground Truth Used

    The "ground truth" for the various tests were defined by:

    • International Standards: ISO 10555-1, ISO 594-1 & 2, ISO 7886-1, ISO 10993 (various parts), ISO 11135, ISO 11607-1, ISO 11607-2.
    • National Standards: ASTM F756, ANSI/AAMI/ISO 11135, ANSI/AAMI/ISO 10993-7, USP 661.
    • Manufacturer Specifications: In-house bench testing compared the device's performance to predicate devices and likely internal design specifications that reflect expected performance characteristics (e.g., "equivalent or superior to predicate devices" for kink resistance, aspiration rate, etc.).

    8. Sample Size for the Training Set

    Not applicable. The studies described are for physical device testing and do not involve machine learning algorithms that require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1