Search Results
Found 2 results
510(k) Data Aggregation
(129 days)
The ANT-X System is indicated for use in conjunction with fluoroscopy imaging in percutaneous nephrolithotomy (PCNL) procedures to aid in needle positioning and alignment process.
The ANT-X System is an automated needle positioning medical device that is used in conjunction with fluoroscopy to support percutaneous interventions (PCNL). There are 4 main components to the system - the ANT-X Device. ANT-X Controller Box. ANT- X Needle Holder, and ANT-X Software. It is intended to be used in an operating theatre equipped with standard surgical equipment which are generally found in operating rooms.
The ANT-X Device holds a sterile, disposable needle-guide kit, the ANT-X Needle Holder. The main bulk of ANT-X Device is made of biocompatible polyetheretherketone (PEEK) material. PEEK is preferred to build the robot because it is radiolucent and lightweight while being able to provide relatively high material strength. The Device has a fixed base made of three pairs of high stiffness parallel linkages allow millimetric planar needle adjustment. The targeted insertion point will act as a pivot point for the needle. The device will be mounted onto the operating table with a 6-DOF instrument holder (FISSO Articulated arm) such that it can be precisely oriented to the area of interest on the patient.
The ANT-X Controller Box powers the ANT-X Device and is the integrating point for the whole ANT-X System. Images from the C-arm fluoroscopy machine will be passed into the ANT-X Controller Box and then to the ANT-X Software, in real-time. The user will then use the ANT-X Software to command the ANT-X Device to commence needle alignment between the chosen insertion point on the skin surface, and the desired target point within the body. Validation of the puncture trajectory and the final needle insertion will be performed by the surgeon.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally implied by the performance data presented, aiming to demonstrate safety and effectiveness comparable to predicate and reference devices, and adherence to relevant standards. Specific numerical acceptance criteria are not explicitly listed in a "criteria" column, but rather the performance values obtained are reported and deemed to meet "specifications" or be "within acceptable limit."
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Biocompatibility: Device components (ANT-X Needle Holder) contacting the patient are biocompatible for indirect contact (≤ 24h). | Passed: Tested in accordance with ISO 10993-1:2018 (Cytotoxicity, Sensitization, Irritation, Material-mediated pyrogenicity, Acute systemic toxicity). Considered biologically safe and biocompatible. |
Sterility: ANT-X Needle Holder achieves a Sterility Assurance Level (SAL) of 10⁻⁶. | Passed: Validated in accordance with ISO 11135, achieving SAL of 10⁻⁶. ETO and ECH residual levels within acceptable limits. |
Shelf Life: ANT-X Needle Holder maintains sterility and quality for one year. | Passed: Real-time studies confirmed sterility and quality maintained at 1 year. |
Electrical Safety: System complies with relevant electrical safety standards. | Passed: Complies with IEC 60601-1. |
Electromagnetic Compatibility (EMC): System complies with relevant EMC standards. | Passed: Complies with IEC 60601-1-2. |
Software Verification & Validation: Software meets design specifications and FDA guidance. | Passed: Verification and validation testing conducted; deemed "moderate" level of concern. |
Software Tracking Accuracy: ANT-X System software provides accurate tracking of individual points. | Passed: Tracking accuracy met design acceptance criteria. |
Positional Repeatability: Robot achieves consistent positional accuracy. | Passed: System met specifications, repeatability accuracy within acceptable limit. |
Bullseye Alignment Accuracy: ANT-X System provides accurate alignment between user-selected points. | Passed: ANT-X System met specifications. |
Puncture Accuracy: ANT-X System achieves successful puncture of target size. | Passed: ANT-X System met specifications. |
Animal Study Performance: Functionality and performance demonstrated in animal model. | Successful: Performance concluded to be consistent and safe for fluoroscopy-guided procedures. |
Clinical Study - Safety: Comparable complication rates to standard of care (US-guided PCNL). | Comparable Safety: No significant differences in overall complication rates (p=0.396) or 1-month complication rates (p=0.88) between RAFG and USG groups. Complications not related to ANT-X device use. |
Clinical Study - Efficacy (Non-inferiority for SFR): Stone-Free Rate (SFR) non-inferior to US-guided PCNL. | Comparable Efficacy: SFR at 3 months: USG 70.6%, RAFG 83.3% (p=0.26). Overall results comparable. |
Clinical Study - Efficacy (Secondary outcomes): Improved puncture efficiency (e.g., fewer attempts, shorter time). | Improved Efficiency: Significantly fewer needle punctures (RAFG: 1.83 vs. USG: 2.51, p=0.025). Significantly shorter median needle puncture duration (RAFG: 5.5 min vs. USG: 8.0 min, p=0.049). |
Image Registration Error: Comparable to reference device (0.38 ± 0.18 mm). | 0.01 - 0.45 mm: Similar to reference device error. |
Alignment Accuracy Range (Target Error): Within range of reference device (0.1-4mm). | 0.14 - 0.97 mm: Within range of reference device. |
Study Details
- Sample Size and Data Provenance:
- Test Set (Clinical Study): Total of 71 patients. 35 in the Ultrasound-guided (USG) group and 36 in the Robotic Assisted Fluoroscopy (RAFG) group.
- Data Provenance: Japan (randomized, single-blind clinical trial). Prospective in nature.
- Number of Experts and Qualifications for Ground Truth (Test Set):
- The document implies that traditional clinical outcomes (stone-free rate at 3 months, complication rates) were the primary ground truth. While radiologists/clinicians are involved in the procedures and assessments, the "ground truth establishment" isn't described as an expert consensus reading of images purely for validation purposes like in an AI image analysis tool. It's more about objective clinical endpoints.
- The study involved "fellowship-trained surgeons" for the animal study and clinicians/surgeons for the clinical trial.
- Adjudication Method for Test Set:
- Not explicitly described in terms of a multi-reader adjudication process for image interpretation (e.g., 2+1 or 3+1). The primary outcomes are clinical, based on objective measurements (e.g., KUB radiography, CT scans for SFR, Clavien-Dindo classification for complications) rather than subjective image reads requiring adjudication.
- Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, a formal MRMC study as typically understood for diagnostic AI (where human readers interpret images with and without AI assistance to measure improvement in diagnostic accuracy) was not performed.
- The clinical study was a comparative effectiveness study between two different procedural guidance methods (AI-assisted fluoroscopy vs. ultrasound guidance). It showed the effects of the AI-assisted system on procedural outcomes (e.g., fewer punctures, shorter puncture time), indicating how human proceduralists (not image readers) improve when using the ANT-X System compared to ultrasound guidance.
- Effect size of human reader improvement with AI vs. without AI assistance: Not applicable in the context of an MRMC study measuring diagnostic performance. However, for procedural efficiency, the study did show improvement with the ANT-X:
- Mean number of needle punctures: 1.83 (RAFG) vs. 2.51 (USG), a reduction of 0.68 punctures (p=0.025).
- Median needle puncture duration: 5.5 minutes (RAFG) vs. 8.0 minutes (USG), a reduction of 2.5 minutes (p=0.049).
- Standalone Performance (Algorithm Only):
- Yes, the "Bench Testing" section describes tests of the algorithm's performance in isolation or in a controlled environment, disconnected from live patient interaction where a human decision-maker is involved:
- Software Tracking Accuracy (including marker detection)
- Positional Repeatability (of the robot)
- Bullseye Alignment Accuracy
- Puncture Accuracy
- Image Registration Error (0.01-0.45mm) and Alignment Accuracy (0.14-0.97mm) are specific quantitative measures of standalone performance.
- Yes, the "Bench Testing" section describes tests of the algorithm's performance in isolation or in a controlled environment, disconnected from live patient interaction where a human decision-maker is involved:
- Type of Ground Truth Used:
- Bench Testing: Engineering measurements, possibly phantom-based. For example, "Puncture Accuracy" would likely be measured against a known target in a phantom. Image Registration Error and Alignment Accuracy are direct measurements against a reference.
- Clinical Study: Clinical outcomes data (Stone-Free Rate, complication rates, procedural metrics like puncture time and number of punctures), which are objective and generally established by clinical follow-up and imaging.
- Sample Size for Training Set:
- Not specified in the provided text. The document focuses on the validation and testing of the device, not its development or training phase.
- How Ground Truth for Training Set was Established:
- Not specified. This information would typically be proprietary to the manufacturer's development process and is not usually detailed in a 510(k) summary, which focuses on validation data for marketing clearance.
Ask a specific question about this device
(137 days)
The Stealth Autoguide™ System is a positioning and guidance system intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments, based on a pre-operative plan and feedback from an image-guided navigation system with three-dimensional imaging software.
The Stealth Autoguide™ System is a remotely-operated positioning and guidance system, indicated for any neurological condition in which the use of stereotactic surgery may be appropriate (for example, stereotactic EEG, laser tissue ablation, etc.).
The Midas Rex™ Legend™ depth stop attachment and tools are incision, cutting, removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.
Stealth Autoguide™ System: The Stealth Autoguide System is a robotic positioning and guidance system intended to interpret navigation tracker coordinates and surgical plan coordinates from the StealthStation to robotically position and orient instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments to pre-defined plans.
Midas Rex™ Legend™ Depth Stop System: The Midas Rex™ Legend™ Depth Stop System consists of a Depth Stop Attachment and specific surgical dissecting tools that will be used in conjunction with the Stealth Autoguide System to create cranial access holes for neurosurgical procedures.
The provided text describes the Medtronic Stealth Autoguide System and Midas Rex Legend Depth Stop System. It includes information on performance testing for the Stealth Autoguide System, but lacks specific details on acceptance criteria and a study to prove the device meets all acceptance criteria in a comprehensive format. It also doesn't contain the requested information about training sets, expert ground truth development, MRMC studies, or standalone performance.
However, based on the provided text, I can extract the following information concerning the performance testing for the Stealth Autoguide System's accuracy:
Acceptance Criteria and Reported Device Performance for Stealth Autoguide™ System
Acceptance Criterion | Reported Device Performance (Mean) | Standard Deviation | 99% CI* Upper |
---|---|---|---|
3D Positional Accuracy: Mean error ≤ 2.0 mm | |||
Biopsy Needle Accuracy Validation - StealthStation S7 | 0.92 mm | 0.47 mm | 3.03 mm |
Biopsy Needle Accuracy Validation - StealthStation S8 | 0.97 mm | 0.26 mm | 1.70 mm |
sEEG bolts/Visualase Accuracy Validation - StealthStation S7 | 1.50 mm | 0.68 mm | 3.08 mm |
sEEG bolts/Visualase Accuracy Validation - StealthStation S8 | 1.48 mm | 0.48 mm | 2.60 mm |
Trajectory Angle Accuracy: Mean error ≤ 2.0 degrees | |||
Biopsy Needle Accuracy Validation - StealthStation S7 | 1.22 degrees | 0.51 degrees | 2.41 degrees |
Biopsy Needle Accuracy Validation - StealthStation S8 | 0.59 degrees | 0.23 degrees | 1.11 degrees |
sEEG bolts/Visualase Accuracy Validation - StealthStation S7 | 1.04 degrees | 0.76 degrees | 2.81 degrees |
sEEG bolts/Visualase Accuracy Validation - StealthStation S8 | 0.42 degrees | 0.17 degrees | 0.82 degrees |
Details of the Accuracy Study:
-
Sample size used for the test set and the data provenance: The document states that performance was determined using "overall end-to-end worst-case system level accuracy testing which incorporated clinically relevant anatomical phantoms." Further specifics about the sample size (e.g., number of phantoms, number of measurement points per phantom) and data provenance (e.g., country of origin, retrospective or prospective) are not provided in this document.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided in the document. The accuracy testing seems to be based on direct physical measurements against defined targets on phantoms rather than expert interpretation of images.
-
Adjudication method for the test set: This information is not provided. Given the nature of the accuracy testing (physical measurements), traditional adjudication methods for image interpretation would likely not apply.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This information is not provided. The assessment described is a technical accuracy validation of the device's navigation and positioning capabilities, not a study involving human readers or AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The "Stealth Autoguide™ System" is described as a "robotic positioning and guidance system" and the accuracy validation focuses on its "performance in 3D positional accuracy" and "trajectory angle accuracy." This implies standalone technical performance testing of the system's ability to achieve planned trajectories, before a human surgeon uses it to guide instruments. The system is designed to "robotically position and orient instrument holders or tool guides," suggesting its core function is algorithm-driven positioning. However, the evaluation here focuses on the accuracy of the guidance provided, which would then be utilized by a surgeon.
-
The type of ground truth used: The ground truth for the accuracy study was established by defining "clinically relevant anatomical phantoms" and measuring the device's "performance in 3D positional accuracy" and "trajectory angle accuracy" against the known positions and trajectories on these phantoms. This is a phantom-based measurement ground truth.
-
The sample size for the training set: This information is not provided. The document describes an accuracy validation study, not the development or training of an AI algorithm.
-
How the ground truth for the training set was established: This information is not provided, as details about a training set are not included in the document.
Ask a specific question about this device
Page 1 of 1