Search Results
Found 1 results
510(k) Data Aggregation
(86 days)
ArTT Augments and Buttresses are indicated to be used, in combination with EMPOWR Acetabular Cup System, as an alternative to structural allograft in skeletally mature patients in cases of segmental acetabular deficiencies.
ArTT Augments and Buttresses are indicated for cementless use to the bone interface; and are affixed to the mating acetabular cup using bone cement.
This 510(k) submission aims at introducing new device components, the ArTT Augments and Buttresses and Bone Screws (Revision Bone Screws (Dia. 6.5mm), Locking Bone Screws (Dia. 6.5mm), Locking Bone Screws (Dia. 4mm), Bone Screws (Dia. 4mm)) to be used in combination with the EMPOWR Acetabular Cup System (K190057) manufactured by Encore Medical, L.P.
ArTT Augments and Buttresses are manufactured from Ti6Al4V 3D printed (ISO 5832-3) and are indicated for cementless use to the bone interface; they are affixed to the mating acetabular cup using bone cement. ArTT Augments are available in 9 diameters (Dia. 50mm to 66mm) and 3 eccentricities (Ecc. 10mm, Ecc. 15mm, Ecc. 20mm), while ArTT Buttresses are available in 3 diameters (Dia.50mm, Dia.56mm, Dia.62mm), 3 heights (H. 15mm, H. 30mm, H. 60mm) and 3 configurations (Neutral, Left, Right).
Bone Screws are manufactured from Ti6Al4V (ISO 5832-3 - ASTM F1472); they are intended for cancellous or cortical bone and are available in several lengths.
This FDA 510(k) clearance letter pertains to
"ArTT Augments and Buttresses and Bone Screws" for orthopedic use. It's important to note that this document is a clearance letter, not a detailed scientific study publication. As such, it provides summary information about the device and the types of testing performed to demonstrate substantial equivalence, rather than a deep dive into specific study methodologies or acceptance criteria with detailed performance metrics.
Based on the provided text, here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state specific quantitative acceptance criteria (e.g., minimum tensile strength, maximum wear rate, specific pull-out force thresholds) or precise numerical performance results for the ArTT Augments and Buttresses and Bone Screws.
Instead, it states:
- "Mechanical tests demonstrated that the device performance fulfilled the intended use, and that the device is substantially equivalent to the predicate device."
- "Other performance requirements were fulfilled through rationales and comparisons with previously cleared components (K210717, latest 510(k) approval)."
This implies that the acceptance criteria were based on:
Acceptance Criteria | Reported Device Performance |
---|---|
Performance fulfills intended use. | Demonstrated by mechanical testing and comparison with predicate devices. |
Substantially equivalent to predicate device (Zimmer Trabecular Metal Acetabular Augments - K061067 and DePuy Gription TF 5.5mm Sterile Locking Screws - K123924). | Confirmed through mechanical testing on "worst case components or constructs" and rationales. Specific tests mentioned include Fretting fatigue, Torsional properties, driving torque, and pull-out load. |
Mechanical properties (fretting fatigue, torsional properties, driving torque, pull-out load) meet established standards/protocols. | "Fretting fatigue testing of the ArTT Augments and Buttresses with Bone screws (Internal protocol)" and "Torsional properties, driving torque and pull-out load of Bone Screws (ASTM F543)". The results fulfilled these requirements. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for Test Set: The document states that mechanical testing was performed on "worst case components or constructs." However, it does not specify the exact number of samples or specimens used for each test (e.g., how many augments, how many screws of each type).
- Data Provenance: The studies were "Non-clinical testing" conducted by the manufacturer, Limacorporate S.p.A. The location of the testing laboratories or specific origin of data (e.g., retrospective/prospective in a clinical sense) is not described, as these are non-clinical (mechanical) tests.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:
This section is not applicable as the studies described are non-clinical (mechanical) tests. The "ground truth" for mechanical testing is typically established by engineering standards (e.g., ASTM F543) and internal protocols, not by expert consensus from medical professionals.
4. Adjudication Method for the Test Set:
This section is not applicable for non-clinical mechanical testing, where adjudication methods like 2+1 or 3+1 by human experts are not used. Performance is measured against engineering specifications and standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:
No, an MRMC comparative effectiveness study was not done. The clearance is for a physical orthopedic implant system, not an AI software/diagnostic device that would involve human readers interpreting cases.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
No, a standalone (algorithm only) performance study was not done. This device is a physical implant system, not an algorithm or AI.
7. The Type of Ground Truth Used:
For the mechanical tests conducted, the "ground truth" implicitly refers to:
- Engineering Standards: Specifically mentioned is ASTM F543 for torsional properties, driving torque, and pull-out load of bone screws.
- Internal Protocols: Used for fretting fatigue testing.
- Performance of Predicate Devices: The measured performance of the ArTT Augments and Buttresses was deemed "substantially equivalent" to that of the predicate devices. This means the performance of the predicate likely served as a benchmark or reference point for establishing acceptable performance.
8. The Sample Size for the Training Set:
This section is not applicable as this is a physical medical device, not an AI/machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
This section is not applicable for the same reason as above (physical medical device, no training set).
Ask a specific question about this device
Page 1 of 1