Search Results
Found 2 results
510(k) Data Aggregation
(92 days)
The SureSigns VS3 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSignsVS3 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
Patient Types | |||
---|---|---|---|
Parameter | Adult | Pediatric | Neonatal |
NBP | √ | √ | √ |
SpO2 | √ | √ | √ |
Temperature | √ | √ | √ |
The SureSigns VS4 vital signs monitor is for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The SureSigns VS4 is for monitoring, recording and alarming of multiple physiological parameters in healthcare environments for patient types listed below. Additionally, the monitor may be used in transport situations within a healthcare facility.
Patient Types | |||
---|---|---|---|
Parameter | Adult | Pediatric | Neonatal |
NBP | √ | √ | √ |
SpO2 | √ | √ | √ |
Temperature | √ | √ | √ |
CO2 | √ | √ | √ |
RRa | √ | √ | |
SpHb | √ | √ |
The subject devices are SureSigns VS3 and SureSigns VS4 multi-parameter patient monitors. Modifications include adding connectivity to IntelliVue GuardianSoftware (IGS), adding Temporal Temperature measurement to VS3, and other enhancements such as additional title bar indicators, expanded patient records pane, display patient name or primary ID, expanded QuickCapture entries, default CO2 setting alignment (VS4 only), temperature high and low alarms color change, increased SpO2 alarm delay defaults, option to prevent inadvertently clearing NBP programs, and RFID patient ID entry.
This document describes the Philips SureSigns VS3 and VS4 Vital Signs Monitors, which are multi-parameter patient monitors. The submission is a 510(k) premarket notification for modifications to previously cleared devices.
Here's an analysis of the acceptance criteria and study information provided, focusing on what can be extracted from the text:
1. Table of Acceptance Criteria and Reported Device Performance
The document broadly states that "Pass/Fail criteria were based on the specifications cleared for the subject device." However, it does not explicitly list specific numerical acceptance criteria for performance metrics (e.g., accuracy ranges for NBP, SpO2, Temperature, CO2). Instead, it confirms that the device reportedly meets these implicit criteria.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Safety and Reliability Requirements | Test results demonstrate that the Philips SureSigns VS3 Vital Signs monitor and the Philips SureSigns VS4 Vital Signs monitor meet all safety and reliability requirements. |
Performance Claims (based on cleared specifications) | Test results demonstrate that the Philips SureSigns VS3 Vital Signs monitor and the Philips SureSigns VS4 Vital Signs monitor meet all performance claims. The added Temporal Temperature measurement to VS3 will have the same specification with that from the predicate VS4. |
Functionality | Verification, validation, and testing activities establish the functionality characteristics of the subject device. |
IEC 62304: 2006 compliance (for software) | The modified software for VS3 and VS4 (version A.06) complies with IEC 62304: 2006 - Medical Device Software life-cycle processes. |
ISO 80601-2-56: 2009 + CORR. 2 (2007) compliance (for temperature) | Both VS3 and VS4 (for Temporal Temperature measurement) comply with ISO 80601-2-56: 2009 + CORR. 2 (2007). Medical electrical equipment - Part 2-56: Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement. |
Substantial Equivalence | Test results showed substantial equivalence. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not provide details on the sample size used for the test set.
It also does not specify the data provenance (e.g., country of origin, retrospective or prospective). The testing appears to be internal validation by Philips Medical Systems.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The testing described is verification, validation, and performance testing, which typically involves comparing device outputs to a known reference standard (e.g., calibration equipment, established measurement methods) rather than human expert-established ground truth in a clinical scenario.
4. Adjudication Method for the Test Set
This information is not provided. Given the nature of the device (monitoring physiological parameters), "adjudication" in the sense of resolving discrepancies between human experts for a diagnostic task is unlikely to be relevant here. The testing would involve comparing device readings against a gold standard or reference measurement.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically for diagnostic imaging devices where human readers interpret images with and without AI assistance. The SureSigns VS3 and VS4 are vital signs monitors, not diagnostic imaging devices designed for human interpretation of complex data patterns in the same way. The modifications focus on software connectivity, a new temperature measurement accessory, and feature enhancements.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The document describes "system level tests, performance tests, and safety testing from hazard analysis." This strongly implies that standalone performance testing was done for the device's functions (e.g., accuracy of NBP, SpO2, Temperature measurements). The software verification and validation, along with compliance to specific standards for temperature measurement, suggest testing of the algorithms and hardware in a standalone capacity against reference standards. However, "standalone" in the context of an AI algorithm is typically for diagnostic or predictive AI, which this device, in its described modifications, does not appear to be. It's a measurement device.
7. The Type of Ground Truth Used
The ground truth used for these types of physiological parameter monitors would typically involve:
- Reference standard instruments: Highly accurate and calibrated devices used to establish the true value of parameters like NBP, SpO2, and Temperature.
- Known physical inputs: For example, precise temperature probes, blood pressure simulators, or SpO2 simulators with defined oxygen saturation levels.
- Compliance with industry standards: The mention of ISO 80601-2-56 for temperature indicates that the ground truth for temperature measurement accuracy would be established by following the specific protocols and reference measurements outlined in that standard.
The document does not explicitly state "expert consensus" or "pathology" as ground truth, which are more relevant for diagnostic AI applications.
8. The Sample Size for the Training Set
The document does not describe any "training set." This is because the device, based on the provided information, is a vital signs monitor and not an AI/ML-based diagnostic or predictive system that requires a training set for model development. The focus is on measurement accuracy and software functionality.
9. How the Ground Truth for the Training Set Was Established
Since there is no mention of a training set, this information is not applicable/provided.
Ask a specific question about this device
(244 days)
The IntelliVue CL Respiration Pod is indicated for use by health care professionals whenever there is a need for intermittent or spot-check acquisition and monitoring of physiological patient parameters respiration rate and pulse rate wirelessly in specific hospital areas. The IntelliVue CL Respiration Pod is mainly indicated for use in general medical and surgery wards and in waiting areas of emergency rooms.
It is not indicated for use in hospital areas in which continuous patient monitoring is needed, such as intensive care units or operating rooms.
The intended use of the IntelliVue CL Respiration Pod when used together with a patient monitor is for intermittent or spot-check monitoring and recording of, and to generate alarms for, respiration rate and pulse rate of adult patients.
The IntelliVue CL Respiration Pod is also intended for acquisition of respiration rate and pulse rate data of adult patients for a clinical information management system.
The IntelliVue CL Respiration Pod is intended for use by health care professionals. It is not intended for home use. It is not a therapeutic device.
The IntelliVue CL Respiration Pod is not intended for use on patients with extremely high values for respiration rate (above 60 rpm).
The IntelliVue CL Respiration Pod is not intended for use on acutely ill cardiac patients with the potential to develop life threatening arrhythmias, e.g. very fast atrial fibrillation or ventricular tachycardia (rapid irregular pulse rate). For monitoring of these patients, a device for continuous ECG monitoring is necessary. The IntelliVue CL Respiration Pod is not a substitute for an ECG monitor.
Warning: Do not use the CL Respiration Pod on patients with rapid, irregular heart rates greater than 110 bpm. Use under these conditions has not been clinically validated.
Rx only: US Federal Law restricts these devices to sale by or on the order of a physician.
The IntelliVue CL Respiration Pod is a small body-worn, cableless, battery powered, respiration rate and pulse rate measuring and monitoring device. It uses specifically designed adhesive attachment (Mobile CL Resp Attachment, cleared with K122223), which holds the CL Respiration Pod at the patient chest.
The CL Respiration Pod provides and communicates measurements values and other information (e.g. battery state) wirelessly via Short Range Radio (SRR) to an assigned compatible IntelliVue Patient Monitor or IntelliVue GuardianSoftware. It can also provide information on basic posture (like supine, prone, upright etc.) and activity of the patient. The CL Respiration Pod can be controlled from the assigned IntelliVue Patient Monitor or IntelliVue GuardianSoftware.
The CL Respiration Pod is a small, battery powered, wireless, and body worn respiration and pulse rate measuring and monitoring device. It is attached to the patient's left costal arch using a specialized accessory, the adhesive Mobile CL Resp Attachment. The CL Respiration Pod has a multicolor LED and a single key to display states and allow basic operation locally. The CL Respiration Pod is connected during its use wirelessly via SRR a compatible IntelliVue Patient Monitor or IntelliVue to GuardianSoftware.
The CL Respiration Pod uses a measurement principle that is based the thoracic movements. Inclination changes of the on incorporated accelerometer sensor, caused by chest and abdomen movements during breathing and heart contraction, produce a voltage signal, from which respiration and pulse rate signals are derived. The modified Philips IntelliVue CL Respiration Pod uses the same measurement principle (respiratory effort and heart contraction/beating) to derive the respiration and pulse signal as the predicate Philips IntelliVue CL Respiration Pod.
With the means of the built-in accelerometer sensor, the modified CL Respiration Pod also detects activity and basic posture of the patient. This is the same technology as used by the legally marketed Philips IntelliVue CL Respiration Pod (K122223).
The Philips IntelliVue CL Respiration Pod aims to measure respiration rate and pulse rate. The study conducted for this 510(k) submission (K132320) is primarily focused on demonstrating substantial equivalence to a predicate device (IntelliVue CL Respiration Pod, K122223) rather than establishing new acceptance criteria or a comparative effectiveness of AI with human readers.
Here's an analysis of the provided information:
1. Table of Acceptance Criteria and Reported Device Performance
The submission describes verification and validation activities, but it doesn't explicitly state quantitative acceptance criteria in a table format that the device needed to meet for de novo clearance. Instead, the study aims to show substantive equivalence to a predicate device.
The reported device performance is described qualitatively as meeting defined specifications and demonstrating substantial equivalence. The document states:
- "The test results have shown that the respiration rate and pulse rate accuracy, resolution, and signal quality met in the whole measurement range the defined specifications."
- "The results of the statistical comparison demonstrate the substantial equivalency of of the modified IntelliVue CL Respiration Pod with the predicate IntelliVue CL Respiration Pod (K122223)."
Without specific numerical criteria for accuracy, resolution, and signal quality, a direct table of acceptance criteria vs. performance cannot be fully constructed from the provided text. The "acceptance criteria" appear to be implicit in the demonstration of substantial equivalence and meeting internal specifications.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The additional clinical data set used for validating algorithm changes consisted of 26 patients (21 male, 5 female).
- Total Duration of Recording: 621 hours and 10 minutes.
- Data Points Used for Comparison: 2,276,855 ECG beats and 255,307 breaths.
- Data Provenance: The document does not explicitly state the country of origin. It mentions "clinical data," which implies real patient data. It is a prospective or, more accurately, retrospective analysis of a prospectively collected dataset, as it explicitly states, "An additional not yet used clinical data set with patient signals was used to validate the algorithm changes." This implies data were pre-existing before being used for this specific validation. It also notes that an "identical data set of K122223 has been reused for the development of the modifications," which would be retrospective for this new device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not mention the use of experts to establish ground truth for the test set. The comparison is statistical, between the modified device and the predicate device.
4. Adjudication Method for the Test Set
No adjudication method is mentioned or implied, as the comparison is statistical against a predicate device's measurements, not against a consensus-based ground truth established by experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. The device is a physiological parameter monitor, and the study focuses on its measurement accuracy and equivalence to another device, not on how it impacts human reader performance or diagnostic tasks. The device itself is not an AI algorithm that assists human readers in interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the evaluations described are primarily standalone performance assessments of the device itself (or its underlying algorithms for respiration and pulse rate derivation) compared to a predicate device. The clinical performance tests involve a "Bland-Altman statistical comparison between respiration rate and pulse rate performance measured by the modified IntelliVue CL Respiration Pod and the predicate IntelliVue CL Respiration Pod." This is an assessment of the device's output.
7. The Type of Ground Truth Used
The "ground truth" in this context is the measurements obtained from the legally marketed predicate device (IntelliVue CL Respiration Pod, K122223). The study aimed to show that the modified device's measurements are substantially equivalent to those of the predicate device. This is a form of reference standard comparison rather than an independent gold standard like pathology or expert consensus. For bench testing, "simulated values" were used as the reference.
8. The Sample Size for the Training Set
The document states, "Two separate clinical data sets have been used. The identical data set of K122223 has been reused for the development of the modifications." The sample size for this reused dataset (presumably for training or development of the modifications) is not specified in the provided text.
9. How the Ground Truth for the Training Set Was Established
The ground truth for the "training set" (the data from K122223 used for development) is implicitly linked to how the original predicate device (K122223) established its performance and potentially its own reference measurements. The current submission does not detail how the ground truth for this development dataset was established but indicates it was previously used for K122223.
Ask a specific question about this device
Page 1 of 1