Search Results
Found 1 results
510(k) Data Aggregation
(168 days)
InFix Cannulated Screw System is generally intended for fracture fixation of various bones and bone fragments, such as femoral neck, intercondylar femoral, malleolus, pilon tibial, calcaneus, talus, tibial plateau, tarsal, wrist, metacarpal, carpal, scaphoid and radius fracture. InFix Cannulated Screw System is also intended for fixation arthrodesis, iliosacral dislocations, and hallux valgus corrections.
Accessories implants:
The round washer is used to increase bone contact area for distributing the forces/load and prevent the screw head from sinking into the bone.
The InFix Cannulated Screw System includes eight cannulated screw specifications and associated washers, which are manufactured from stainless steel and titanium alloy. The cannulated screw is a self-tapping and self-drilling screw with a cancellous thread that can be guided into a position via a guided wire. They are used to aid in the alignment and stabilization of fractures to the skeletal system.
Here's an analysis of the provided text regarding the acceptance criteria and study proving device performance:
This document is a 510(k) premarket notification for the InFix Cannulated Screw System, which is a Class II medical device. The primary purpose of this submission is to demonstrate substantial equivalence to existing predicate devices, rather than to prove de novo safety and effectiveness through extensive clinical trials.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Biomechanical Testing (ASTM F543) | Self-tapping performance | Results indicate substantial equivalence to legally marketed devices. |
Torsional strength | Results indicate substantial equivalence to legally marketed devices. | |
Axial pullout strength | Results indicate substantial equivalence to legally marketed devices. | |
Driving torque | Results indicate substantial equivalence to legally marketed devices. | |
Cleaning Validation | AAMI TIR30:2011 guidance requirements | Test result shows that the acceptance criteria is met. |
Sterilization Validation | ANSI/AAMI/ISO 17665-1 requirements | Test result shows that the acceptance criteria is met. |
Details Regarding the Study and Acceptance Criteria:
This submission relies heavily on non-clinical tests to demonstrate substantial equivalence. Clinical effectiveness studies as one might find for a novel AI device are not presented here because this is a hardware medical device for which the FDA determination is about substantial equivalence to already approved devices. Therefore, the "acceptance criteria" here are largely benchmarks against the performance of predicate devices and adherence to recognized standards.
2. Sample size used for the test set and the data provenance:
- Biomechanical Test: The document states "The biomechanical tests ASTM F543 were performed". It does not specify the exact number of screws or test samples used for each performance metric (self-tapping, torsional, axial pullout, driving torque).
- Cleaning Validation & Sterilization Validation: Similarly, the document does not specify the number of samples used for these validation tests.
- Data Provenance: Not explicitly stated, but typically these tests would be conducted in a laboratory setting by the manufacturer or a contracted lab. The tests are non-clinical, so "country of origin of the data" in terms of patient demographics is not applicable. These are prospective tests performed to meet regulatory requirements.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable in this context. For a hardware device demonstrating substantial equivalence through non-clinical testing, ground truth is established by objective engineering measurements against specified physical properties and performance standards (e.g., ASTM F543). There isn't a need for expert consensus on an interpretation of an output, as would be the case for an AI diagnostic device.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. Adjudication methods like 2+1 are used for expert review of images or data to establish a ground truth or resolve discrepancies in diagnostic labeling. This is not relevant for biomechanical or validation testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices or software that assists human interpretation. The InFix Cannulated Screw System is a physical bone fixation device, not a diagnostic tool or AI-assisted product.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a physical implant. The concept of "standalone algorithm performance" does not apply.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the biomechanical tests (ASTM F543), the "ground truth" is established by engineering standards and direct physical measurements. The performance of the InFix Cannulated Screw System is compared against the known performance characteristics of the predicate devices and the requirements of the ASTM standard.
- For cleaning and sterilization validations, the "ground truth" is adherence to the specified industry standards (AAMI TIR30:2011 and ANSI/AAMI/ISO 17665-1).
8. The sample size for the training set:
- Not applicable. This device is hardware and does not involve machine learning or AI, and therefore has no "training set."
9. How the ground truth for the training set was established:
- Not applicable. As there is no training set for this hardware device, there is no ground truth to be established for it.
Ask a specific question about this device
Page 1 of 1