K Number
K142057

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-01-13

(168 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

InFix Cannulated Screw System is generally intended for fracture fixation of various bones and bone fragments, such as femoral neck, intercondylar femoral, malleolus, pilon tibial, calcaneus, talus, tibial plateau, tarsal, wrist, metacarpal, carpal, scaphoid and radius fracture. InFix Cannulated Screw System is also intended for fixation arthrodesis, iliosacral dislocations, and hallux valgus corrections.

Accessories implants:

The round washer is used to increase bone contact area for distributing the forces/load and prevent the screw head from sinking into the bone.

Device Description

The InFix Cannulated Screw System includes eight cannulated screw specifications and associated washers, which are manufactured from stainless steel and titanium alloy. The cannulated screw is a self-tapping and self-drilling screw with a cancellous thread that can be guided into a position via a guided wire. They are used to aid in the alignment and stabilization of fractures to the skeletal system.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study proving device performance:

This document is a 510(k) premarket notification for the InFix Cannulated Screw System, which is a Class II medical device. The primary purpose of this submission is to demonstrate substantial equivalence to existing predicate devices, rather than to prove de novo safety and effectiveness through extensive clinical trials.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Biomechanical Testing (ASTM F543)Self-tapping performanceResults indicate substantial equivalence to legally marketed devices.
Torsional strengthResults indicate substantial equivalence to legally marketed devices.
Axial pullout strengthResults indicate substantial equivalence to legally marketed devices.
Driving torqueResults indicate substantial equivalence to legally marketed devices.
Cleaning ValidationAAMI TIR30:2011 guidance requirementsTest result shows that the acceptance criteria is met.
Sterilization ValidationANSI/AAMI/ISO 17665-1 requirementsTest result shows that the acceptance criteria is met.

Details Regarding the Study and Acceptance Criteria:

This submission relies heavily on non-clinical tests to demonstrate substantial equivalence. Clinical effectiveness studies as one might find for a novel AI device are not presented here because this is a hardware medical device for which the FDA determination is about substantial equivalence to already approved devices. Therefore, the "acceptance criteria" here are largely benchmarks against the performance of predicate devices and adherence to recognized standards.

2. Sample size used for the test set and the data provenance:

  • Biomechanical Test: The document states "The biomechanical tests ASTM F543 were performed". It does not specify the exact number of screws or test samples used for each performance metric (self-tapping, torsional, axial pullout, driving torque).
  • Cleaning Validation & Sterilization Validation: Similarly, the document does not specify the number of samples used for these validation tests.
  • Data Provenance: Not explicitly stated, but typically these tests would be conducted in a laboratory setting by the manufacturer or a contracted lab. The tests are non-clinical, so "country of origin of the data" in terms of patient demographics is not applicable. These are prospective tests performed to meet regulatory requirements.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable in this context. For a hardware device demonstrating substantial equivalence through non-clinical testing, ground truth is established by objective engineering measurements against specified physical properties and performance standards (e.g., ASTM F543). There isn't a need for expert consensus on an interpretation of an output, as would be the case for an AI diagnostic device.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. Adjudication methods like 2+1 are used for expert review of images or data to establish a ground truth or resolve discrepancies in diagnostic labeling. This is not relevant for biomechanical or validation testing of a physical device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices or software that assists human interpretation. The InFix Cannulated Screw System is a physical bone fixation device, not a diagnostic tool or AI-assisted product.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a physical implant. The concept of "standalone algorithm performance" does not apply.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • For the biomechanical tests (ASTM F543), the "ground truth" is established by engineering standards and direct physical measurements. The performance of the InFix Cannulated Screw System is compared against the known performance characteristics of the predicate devices and the requirements of the ASTM standard.
  • For cleaning and sterilization validations, the "ground truth" is adherence to the specified industry standards (AAMI TIR30:2011 and ANSI/AAMI/ISO 17665-1).

8. The sample size for the training set:

  • Not applicable. This device is hardware and does not involve machine learning or AI, and therefore has no "training set."

9. How the ground truth for the training set was established:

  • Not applicable. As there is no training set for this hardware device, there is no ground truth to be established for it.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 13, 2015

Infix Medical, LLC Mr. James Holley Chief Operating Officer 69 Baybridge Drive, Suite M Gulf Breeze, Florida 32561

Re: K142057

Trade/Device Name: Infix Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: November 21, 2014 Received: November 21, 2014

Dear Mr. Holley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

INFIX

Traditional 510(k) Premarket Notification

Indications for Use

510(k) Number (if known):

Device Name: InFix Cannulated Screw System

Indications for Use:

InFix Cannulated Screw System is generally intended for fracture fixation of various bones and bone fragments, such as femoral neck, intercondylar femoral, malleolus, pilon tibial, calcaneus, talus, tibial plateau, tarsal, wrist, metacarpal, carpal, scaphoid and radius fracture. InFix Cannulated Screw System is also intended for fixation arthrodesis, iliosacral dislocations, and hallux valgus corrections.

Accessories implants:

The round washer is used to increase bone contact area for distributing the forces/load and prevent the screw head from sinking into the bone.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Orthopedic Devices

510(k) Number:

{3}------------------------------------------------

510(k) Summary

Submitter's Name: Infix Medical, LLC Address: 69 Baybridge Dr. Gulf Breeze, FL 32561 Tel: 850-324-7675 Fax: 855-228-1339

Contact Name: James C Holley

Preparation date: January, 5, 2015

Registration Number: 3010759323

Device Name: InFix Cannulated Screw System

Common Name: Bone Fixation Screws

Classification Name: Class II, as per Title 21 of the Code of Federal Regulations, Sec. 888.3040 Smooth or threaded metallic bone fixation fastener

Product Code: HWC

Predicate Device Information:

SYNTHES 7.0/7.3 MM CANNULATED SCREWS (K962011) SYNTHES 6.5 MM CANNULATED SCREWS (K021932) SYNTHES STERILE 4.5 MM CANNULATED SCREWS (K963172) SYNTHES STERILE 3.5 MM AND 4.0 MM CANNULATED SCREWS (K963192) SYNTHES STERILE 3.0 MM CANNULATED SCREW AND THREADED WASHER (K962823) SYNTHES 2.4 MM CANNULATED SCREW (K012945) ORTHO SOLUTIONS EXTREMITY FIXATION IMPLANTS FOR OSTEOSYSNTHESIS (K111678)

Device Description:

The InFix Cannulated Screw System includes eight cannulated screw specifications and associated washers, which are manufactured from stainless steel and titanium alloy. The cannulated screw is a self-tapping and self-drilling screw with a cancellous thread that can

{4}------------------------------------------------

be guided into a position via a guided wire. They are used to aid in the alignment and stabilization of fractures to the skeletal system.

Indication for use:

InFix Cannulated Screw System is generally intended for fracture fixation of various bones and bone fragments, such as femoral neck, intercondylar femoral, malleolus, pilon tibial, calcaneus, talus, tibial plateau, tarsal, metatarsal, wrist, metacarpal, carpal, scaphoid and radius fracture. InFix Cannulated Screw System is also intended for fixation arthrodesis, iliosacral dislocations, and hallux valgus corrections.

Accessories implants:

The round washer is used to increase bone contact area for distributing the forces/load and prevent the screw head from sinking into the bone.

Technological Characteristics

The technological characteristic of the InFix Cannulated Screw System are similar to the predicate devices including design, dimensions, and materials. The InFix Cannulated Screw System screws and washers are fabricated from stainless steel 316L per ASTM F138 and Ti-6Al-4V alloy per ASTM F136. The stainless steel 316L and Ti-6Al-4V are commonly used material in orthopedic implants.

Summary of Substantial Equivalence:

The InFix Cannulated Screw System is substantially equivalent to the predicated devices. No new issues of safety or efficacy have been raised.

Summary of Performance Data (Nonclinical and/or Clinical) Non-Clinical Tests

  • Biomechanical Test
    The biomechanical tests ASTM F543 were performed to determine substantial equivalence of the InFix Cannulated Screw System including self-tapping, torsional, axial pullout and driving torque. Results indicate that the subject screws are substantially equivalent to legally marketed devices offering a reasonable assurance of safety and effectiveness.

  • Cleaning Validation The cleaning validation was performed in according with the AAMI TIR30:2011 guidance document. The test result shows that the acceptance criteria is met.

  • Sterilization Validation

{5}------------------------------------------------

The sterilization validation was performed in according with the ANSI/AAMI/ISO 17665-1. The test result shows that the acceptance criteria is met.

  • The InFix Cannulated Screw System has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration, or image artifact in the MR environment. The safety of the InFix Cannulated Screw System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury.

Clinical Tests

Clinical data and conclusions were not needed for these devices to show substantial equivalence.

Conclusions

The InFix Cannulated Screw System has been shown to be substantially equivalent the predicate devices. Result of preclinical tests/engineering justification and the similarities with legal marketed predicated devices indicate the device will perform within the intended use and no new issues of safety or effectiveness have been raised

N/A