    K Number
    K112565
    Manufacturer
    Date Cleared
    2011-12-06

    (95 days)

    Product Code
    Regulation Number
    870.4210
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K102532, K072362

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NovaPort® One Vascular Access cannula with its associated Introducer Kit is a Novalung coated percutaneous cannula intended to be used to cannulate blood vessels in patients requiring extracorporeal support such as cardiopulmonary bypass. The application of NovaPort® One cannula is limited to

    Device Description

    The NovaPort® One Vascular Access cannula with Novalung coating are a wire reinforced, thin walled femoral cannula made of polyurethane. A cone without wire reinforcement allows clamping and serves as transition to a connector. The beveled tip is not a separate piece and is part of the whole cannula body. The NovaPort® one cannulas are inserted percutaneously into vessels. Insertion depth marks aid in positioning the cannula. They have a beveled tip that is made as part of the PVC body. The Cannula is intended to be used to cannulate vessels, perfuse vessels or organs and/or connect with accessory extracorporeal equipment. The cannula introducer is intended to facilitate proper insertion and placement of the appropriate sized cannula within the vessel for cardiopulmonary bypass. All surfaces of the NovaPort® One which come into contact with blood are protected with the Novalung Coating. This coating is a bioactive, stable, biocompatible and non-thrombogenic surface. This coating consists of high molecular weight heparin from pig mucosa bound covalently and ionically to immobilized polypeptides. NovaPort® one cannula is available in insertion lengths of 90 mm and with outer diameters of 13 F, 17 F, 19 F and 21 F. All are supplied sterile with Novalung coated, non-pyrogenic, and are single use.

    AI/ML Overview

    The provided 510(k) summary for the Novalung® GmbH NovaPort® One Vascular Access device primarily focuses on demonstrating substantial equivalence to predicate devices rather than establishing specific performance criteria through a detailed clinical or standalone performance study as would be seen for AI/ML devices.

    Here's an analysis based on the information provided, emphasizing what is present and what is absent from the request's specific questions regarding acceptance criteria and study details:


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) for a medical device (a vascular access cannula) and not an AI/ML diagnostic tool, there are no specific diagnostic "performance metrics" like sensitivity, specificity, or AUC. Instead, the acceptance is based on demonstrating equivalence to predicate devices across various attributes. The "acceptance criteria" here implicitly refer to meeting the expectations for functional and safety performance that allow for declaration of substantial equivalence.

    Acceptance Criteria Category (Implied)Reported Device Performance (Summary in 510(k))
    Intended UseIdentical to predicate devices
    Indications for UseIdentical to predicate devices, limited to