(175 days)
The HLS Cannula from MAQUET is intended for use by trained physicians only. The HLS Cannula may be utilized to cannulate suitable vessels to provide circulatory perfusion of organs and vessels by forming a connection with the extracorporeal circulation. Standard surgical or percutaneous insertion techniques can be employed. This product is intended for use up to six hours or less.
The HLS Cannula from MAQUET is a wire-reinforced, thin-wall cannula made of polyurethane. The transparent proximal section has no reinforcement and can be clamped. Each cannula is supplied with a pre-mounted 3/8" connector and an introducer that allows a guide wire up to 0.038" to be inserted. The cannula is available with an optional percutaneous insertion kit for the Seldinger technique and an optional BIOLINE or SOFTLINE Coating. The HLS Cannula comes in a range of sizes and lengths. The HLS Cannula is a sterile and non-pyrogenic device, for single use only and is not to be re-sterilized by the user. The insertion kit from MAQUET comprises various components which permit access to the vessel. One insertion kit is available for the arterial cannula (PIK 100) and one for the venous cannula (PIK 150).
This document describes the premarket notification (510(k)) for the HLS Cannula, which is a medical device used for circulatory perfusion during extracorporeal circulation. The study outlined is a non-clinical performance test comparing the HLS Cannula to predicate devices to establish substantial equivalence.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Flow Rate / Pressure Drop | Substantially equivalent to predicate devices. | The performance data of MAQUET cannulae as compared to the MEDTRONIC predicate cannulae are substantially equivalent for relevant flow of 0-6 liters/minute. |
| Kink Resistance | Wire-reinforced cannula shafts will not kink. | Test performed to verify that the wire re-inforced cannulae shafts will not kink. (Implied acceptance met by satisfactory test.) |
| Tensile Strength (Cannula-Connector) | Bonding stability between cannula and connector is verified. | Test performed to verify the bonding stability between cannulae and connector. (Implied acceptance met by satisfactory test.) |
| Tensile Strength (Introducer Grip) | Bonding of introducer and grip is verified. | Verification of bonding of introducer and grip. (Implied acceptance met by satisfactory test.) |
| Six Hours Pressure Resistance | Durability of all connections for 6 hours at 1.5 bar. | Test performed to verify the durability of all connections of HLS arterial and venous Cannulae for a period of 6 hours at 1.5 bar. (Implied acceptance met by satisfactory test.) |
| Air Tightness | All connections are air-tight at a specified vacuum. | Test performed to verify that all connections of the cannulae are air-tight if put to a specified vacuum. (Implied acceptance met by satisfactory test.) |
| Corrosion | No corrosion. | Test performed to evaluate if cannulae corrode. (Implied acceptance met by satisfactory test.) |
| Printed Depth-Markings Adhesion | Adhesion of printing is sufficient. | Test performed to verify that the adhesion of the printing is sufficient. (Implied acceptance met by satisfactory test.) |
| Biologic Evaluation | Compliance to ISO 10993-1. | Products met these requirements. |
| Intravascular Catheter Standards | Compliance to ISO 10555-1. | Products met these requirements. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not explicitly state the specific number of units or samples used for each test. It generally refers to "HLS Cannulae" in the plural, suggesting multiple units were tested.
- Data Provenance: The studies were non-clinical performance tests conducted by Maquet Cardiopulmonary AG. The data provenance is internal testing performed by the manufacturer to support the 510(k) submission. There is no information regarding country of origin of data beyond the manufacturer's location (Germany) for the submission. The studies are prospective in the sense that they were designed and executed to evaluate the performance of the HLS Cannula prior to market clearance.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable as the described tests are non-clinical performance and material characteristic evaluations (e.g., flow rate, tensile strength, kink resistance, biocompatibility standard compliance). These tests rely on established scientific methods and engineering standards, not on expert consensus or human interpretation of outcomes in a clinical setting to establish "ground truth."
4. Adjudication Method for the Test Set
This information is not applicable for the non-clinical performance and material characteristic tests described. Adjudication methods are typically relevant for studies involving human interpretation or clinical outcomes where discrepancies might arise.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The document describes the premarket notification for a medical device (cannula) and its non-clinical performance testing. It does not involve AI, image interpretation, or human reader performance studies.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. This device is a physical medical instrument, not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the non-clinical tests is derived from engineering specifications, material science standards (e.g., ISO 10993-1, ISO 10555-1), and established predicate device performance. For example, the "ground truth" for flow rate is the measured flow rate and pressure drop of the predicate device. For kink resistance, it's the physical property of the cannula not kinking under test conditions.
8. The sample size for the training set
This information is not applicable. This document describes the testing of a physical medical device, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable as there is no training set mentioned or implied.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date FEB 25 2011, followed by the word "MAQUET" in large, bold letters. Below that, the words "GETINGE GROUP" are printed in smaller letters. Above the word "MAQUET" is the number K102532.
510 (K) Summary [as required by 21 CFR 807.92(c) ]
Maquet Cardiopulmonary AG Submitter: Hechinger Strasse 38 72145 Hirrlingen Germany
Contact Person: Dr. Ingrid Richter Phone: +49 7478 921-337 +49 7478 921-8667 Fax: E-mail: ingrid.richter@maquet-cp.com
August 31, 2010 Date Prepared:
HLS Cannula non-coated, with BIOLINE Coating Device Trade Name: and with SOFTLINE Coating
HLS Femoral Cannula Common/Usual name:
Catheter, Cannulae and Tubing, Vascular, Classification names:
Cardiopulmonary Bypass
Predicate Devices:
Bio-Medicus Femoral Cannula, Medtronic USA, cleared under K924642,
QUADROX-i Adult Microporous Membrane Oxygenator with and without Arterial Filter with SOFTLINE Coating, MAQUET Cardiopulmonary AG, cleared under K082117,
QUADROX-i Adult Microporous Membrane Oxygenator with and without Arterial Filter with BIOLINE Coating, MAQUET Cardiopulmonary AG, cleared under K090511.
{1}------------------------------------------------
Device Description:
The HLS Cannula from MAQUET is a wire-reinforced, thin-wall cannula made of polyurethane. The transparent proximal section has no reinforcement and can be clamped. Each cannula is supplied with a pre-mounted 3/8" connector and an introducer that allows a guide wire up to 0.038" to be inserted. The cannula is available with an optional percutaneous insertion kit for the Seldinger technique and an optional BIOLINE or SOFTLINE Coating. The HLS Cannula comes in a range of sizes and lengths.
The HLS Cannula is a sterile and non-pyrogenic device, for single use only and is not to be re-sterilized by the user.
The insertion kit from MAQUET comprises various components which permit access to the vessel. One insertion kit is available for the arterial cannula (PIK 100) and one for the venous cannula (PIK 150).
Indications for Use:
The HLS Cannula from MAQUET is intended for use by trained physicians only. The HLS Cannula may be utilized to cannulate suitable vessels to provide circulatory perfusion of organs and vessels by forming a connection with the extracorporeal circulation. Standard surgical or percutaneous insertion techniques can be employed. This product is intended for use up to six hours or less.
The insertion kit is used for preparing the cannulation of vessels for extracorporeal circulation. Surgical procedures can be performed under visual control or with percutaneous cannulation using the Seldinger technique.
Statement of Technical Comparison:
HLS Cannula non-coated, with BIOLINE Coating and with SOFTLINE Coating from MAQUET Cardiopulmonary AG, have a comparable intended use as the Bio-Medicus Femoral Cannulae (K924642). Both are intended to be used to cannulate vessels, perfuse vessels or organs up to 6 hours. Standard surgical or percutaneous insertion techniques can be used.
Furtheron the indications for use of the SOFTLINE Coating of the QUADROXi Adult with SOFTLINE Coating is the same as with the HLS Cannula with SOFTLINE Coating and the indications for use of the BIOLINE Coating of the QUADROX-i Adult with BIOLINE Coating is the same as with the HLS Cannula with BIOLINE Coating.
The HLS Cannulae from MAQUET have similar technical characteristics, performance specification and materials as the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Maquet Getinge Group. The word "MAQUET" is in large, bold, sans-serif font on the top line. Below that, "GETINGE GROUP" is in a smaller, sans-serif font.
Both cannulae are made of polyurethane and are protected from kinking - to a greater or lesser extent - by wire reinforcement. The standard adult sizes are equipped with 3/8" connectors; all feature a designated clamping section.
Non-clinical Testing:
Performance testing has resulted in data that demonstrate that the HLS Cannulae performs according to the applied performance standards. The following performance characteristics of the HLS Cannulae were compared with the predicate device to determine substantial equivalence:
Evaluation of flow rate
For a relevant flow of 0 - 6 liters /minute the pressure drop of arterial and venous HLS Cannulae was measured in comparison to the predicate devices Bio-Medicus Femoral Cannulae from Medtronic.
The performance data of MAQUET cannulae as compared to the MEDTRONIC predicate cannulae are substantially equivalent.
To evaluate the safety and effectiveness of the HLS Cannula additional testing was performed:
Kink resistance
This test is to verify that the wire re-inforced cannulae shafts will not kink.
Tensile test of connection of cannula-connector
This test is to verify the bonding stability between cannulae and connector.
Tensile test of connection of introducer grip
This test is a verification of bonding of introducer and grip.
Verification of six hours pressure resistance
This test is to verify the durability of all connections of HLS arterial and venous Cannulae for a period of 6 hours at 1.5 bar.
Air tightness
This test is to verify that all connections of the cannulae are air-tight if put to a specified vacuum.
Evaluation of Corrosion
This test is to evaluate if cannulae corrode.
Evaluation on printed depth-markings of cannulae
The test is to verify that the adhesion of the printing is sufficient.
{3}------------------------------------------------
MAQUET
The HLS Cannulae non coated, with BIOLINE Coating and with SOFFENEROUP Coating have been tested or evaluated for compliance to ISO 10993-1 Biologic Evaluation of Medical Devices as well as to ISO 10555-1 "Sterile, sincle-use intravascular catheters - Part 1 general requirements". The products met these requirements.
Determination of Substantial Equivalence
MAQUET Cardiopulmonary AG has considered indications for use, design, materials, performance characteristics, and effectiveness of the proposed HLS Cannulae non-coated, with BIOLINE Coating and with SOFTLINE Coating as compared to the predicate devices.
Based on the review and testing provided in this submission MAQUET Cardiopulmonary AG believes that the HLS Cannula non-coated, with BIOLINE Coating and with SOFTLINE Coating is substantially equivalent to the named predicate devices that are FDA cleared.
Conclusion
The HLS Cannulae non coated, with BIOLINE Coating and with SOFTLINE Coating are substantially equivalent to the named predicate devices which currently hold market clearance.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Maquet Cardiopulmonary AG Whitney Torning Director of Regulatory Affairs 45 Barbour Pond Drive Wayne, NJ 07470
FEB 2 5 2011
Re: K102532
Trade Name: HLS Cannula Non-coated, with BIOLINE Coating, and with SOFTLINE Coating Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II Product Code: DWF Dated: January 28, 2011 Received: February 2, 2011
Dear Ms. Torning:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{5}------------------------------------------------
Page 2 - Ms. Whitney Torning
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely Fours,
Sincerely yours,
Bram D. Zuckerman, M.D.
m D. Directof Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known): ≤102532
Device Name: HLS Cannula
Indications for Use:
The HLS Cannula from MAQUET is intended for use by trained physicians only. The HLS Cannula may be utilized to cannulate suitable vessels to provide circulatory perfusion of organs and vessels by forming a connection with the extracorporeal circulation. Standard surgical or percutaneous insertion techniques can be employed. This product is intended for use up to six hours or less.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Page 1 of 1
(Division Sigh-Off) of Cardiovas r Devices
§ 870.4210 Cardiopulmonary bypass vascular catheter, cannula, or tubing.
(a)
Identification. A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.(b)
Classification. Class II (performance standards).