Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161546
    Device Name
    R2P SlenGuide
    Date Cleared
    2016-10-14

    (133 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K142819, K090040

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The R2P (radial to peripheral) SlenGuide is designed for the introduction of interventional and diagnostic devices into the peripheral vasculature of the lower extremities.

    Device Description

    The R2P SlenGuide is a single use, ethylene oxide sterilized device that is designed to perform as a guiding catheter for interventional procedures in the peripheral vasculature. It is packaged with a guiding catheter and an inner guide. The guiding catheter features a three-layer construction with a stainless steel mesh, polyamide elastomer, and polytetrafluoroethyelene. The distal end has a hydrophilic coating and a soft-tip visible under fluoroscopy. The inner guide is an accessory device made of polyester elastomer with a flexible distal portion containing tungsten, visible under fluoroscopy.

    AI/ML Overview

    The provided document describes the R2P SlenGuide, a percutaneous catheter, and its testing to demonstrate substantial equivalence to predicate devices. It does not contain information about a study proving the device meets acceptance criteria in terms of performance metrics like sensitivity, specificity, accuracy, or effect size for AI assistance. Instead, the "acceptance criteria" here refer to meeting standards for safety and performance characteristics through non-clinical testing.

    Here's a breakdown of the available information based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various non-clinical tests performed and states that "Performance testing met the predetermined acceptance criteria." However, it does not explicitly state the numerical acceptance criteria for each test or the specific reported performance results in a detailed, quantitative table. It only confirms that the criteria were met.

    For example, for "Peak tensile force," the test procedure is described (measure peak tensile strength), but the acceptable range (e.g., >X Newtons) and the actual measured value are not provided in this summary. The same applies to other tests.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document mentions that performance tests were performed on "non-aged and accelerated aged samples." However, it does not specify the number of samples used for each test.
    • Data Provenance: The tests are described as non-clinical performance testing conducted by the manufacturer, Terumo Corporation, or its Ashitaka Factory in Japan. The data is retrospective in the sense that it was collected as part of the device development and regulatory submission process.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable as the document describes non-clinical performance and biocompatibility testing of a physical medical device (a catheter), not an AI/software device that would require expert-established ground truth for a test set.

    4. Adjudication Method

    This information is not applicable for the same reason as point 3.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This information is not applicable as the device is a physical catheter, not an AI system. There is no mention of human readers or AI assistance in this context.

    6. Standalone (Algorithm Only) Performance Study

    This information is not applicable as the device is a physical catheter, not an algorithm.

    7. Type of Ground Truth Used

    This information is not applicable as the document describes non-clinical performance and biocompatibility testing of a physical medical device. The "ground truth" for these tests would be the established scientific and engineering principles and the specific requirements outlined in the referenced ISO and ASTM standards, as well as internal standards.

    8. Sample Size for the Training Set

    This information is not applicable as the device is a physical catheter, not an AI/machine learning device that would require a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as point 8.


    Summary of Non-Clinical Tests Performed (as a proxy for "study")

    The study to demonstrate that the device meets performance requirements involved a series of non-clinical, in-vitro tests based on established international standards (ISO and ASTM) and FDA guidance documents. These tests were categorized into performance testing and biocompatibility testing.

    Performance Testing (Guiding Catheter & Inner Guide):

    • Standards Referenced: ISO 10555-1:2013, ASTM F640-12, USP , FDA Guidance "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters."
    • Tests Included:
      • Radio-detectability
      • Surface quality
      • Peak tensile force
      • Freedom from leakage
      • Hub performance (for Guiding Catheter)
      • Distal tip appearance
      • Particulate evaluation
      • Torque Strength (for Guiding Catheter)
      • Coating lubricity (for Guiding Catheter)
      • Evaluation of flexibility and kink resistance (for Guiding Catheter)
      • Distal tip flexibility (for Guiding Catheter)
      • Kink condition (for Guiding Catheter)
      • Flexural rigidity
      • Distal tip strength (for Guiding Catheter)
      • Cleanliness
      • Product dimensions

    Biocompatibility Testing:

    • Standards Referenced: ISO 10993-1, FDA General Program Memorandum #G95-1 (5/1/95), Draft Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993.
    • Classification: Externally Communicating Device, Circulating Blood, Limited Contact (-6.

    In essence, the "study" demonstrating the device meets its "acceptance criteria" here refers to the comprehensive non-clinical testing outlined above, which confirmed adherence to recognized standards for safety, performance, and biocompatibility, thereby supporting the claim of substantial equivalence to predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1