    Intended Use

    The MAGNETOM systems described above are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.

    Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.

    The MAGNETOM systems described above may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.

    Device Description

    The subject device, software syngo MR E11A for MAGNETOM Aera and MAGNETOM Skyra offers two new applications, LiverLab (an application of non-invasive liver evaluation) and MyoMaps (an application designed to provide a means to generate pixel maps for myocardial MR relaxation times). In addition, software syngo MR E11A makes the Dot Cockpit available for the user to modify and create Siemens Dot Engine workflows in a very intuitive way which supplements some of the support of an application specialist. The software syngo MR E11A also includes new and modified sequences as well as minor modifications of already existing features. In additional coils are offered and some hardware components have been modified.

    Siemens Medical Solutions, USA Inc., intends to market MAGNETOM Aera and MAGNETOM Skyra with new software, syngo MR E11A. While syngo MR E11A offers additional capabilities with respect to the predicate device, the MAGNETOM Aera and MAGNETOM Skyra have the same technological characteristics as the predicate device (K121434; Cleared November, 5, 2012).

    Furthermore, Siemens Medical Solutions, USA Inc., intends to market a new configuration of the MAGNETOM Skyra with 24 receive channels with software syngo MR E11A.

    The MAGNETOM Aera and MAGNETOM Skyra will be offered ex-factory (new production) as well as in-field upgrades for the currently installed MAGNETOM Aera and MAGNETOM Skyra systems. The new MAGNETOM Skyra configuration with 24 receive channels will be offered as an ex-factory option (new production).

    AI/ML Overview

    This FDA 510(k) summary describes a new software version (syngo MR E11A) for existing Siemens MAGNETOM Aera and MAGNETOM Skyra MRI systems. The primary focus of the document is to demonstrate substantial equivalence to previous versions and other cleared devices, rather than establishing new acceptance criteria for a novel device.

    Therefore, the requested information regarding acceptance criteria, device performance, and specific study details (like sample size for test sets, expert qualifications, and adjudication methods) is largely not present for the overall system or its new features as this is an equivalence submission. The closest equivalent to "acceptance criteria" are the results of performance tests demonstrating the device performs as intended and is "substantially equivalent."

    However, I can extract the available information regarding testing for the new features:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Implied)Reported Device Performance
    New CoilsPerform as intended with appropriate SNR, image uniformity, and heating levels.Tested for SNR, image uniformity, and heating. The results demonstrated the device performs as intended.
    New/Modified Sequences & AlgorithmsPerform as intended; maintain or improve image quality compared to predicate features.Dedicated phantom testing conducted. Image quality assessments completed. In some cases, image quality was compared. The results demonstrated the device performs as intended.
    Quiet SequencesMeet specified acoustic noise levels.Acoustic noise measurements performed. The results demonstrated the device performs as intended.
    LiverLab (non-invasive liver evaluation)Functional and accurate, potentially comparable to existing methods.Validated with volunteer and phantom scans, and synthetic raw data. The results demonstrated the device performs as intended.
    MyoMaps (myocardial MR relaxation times)Functional and accurate.Tested on volunteers after ECGs were applied. Compared to Thalassaemia Tools in a 100-person study. The results demonstrated the device performs as intended.
    All other software featuresVerified and validated as functional and safe.Verified and validated. The results demonstrated the device performs as intended.
    Overall System ComplianceConform to relevant standards (IEC 62304:2006, IEC, NEMA, ISO 14971:2007).Conforms to IEC 62304:2006, IEC, NEMA standards. Risk management in compliance with ISO 14971:2007.

    2. Sample size used for the test set and the data provenance

    • MyoMaps Comparison Study:
      • Sample Size: 100 persons.
      • Data Provenance: Not specified, but involved "volunteers."
    • LiverLab Validation:
      • Sample Size: Not specified beyond "volunteer" and "phantom scans."
      • Data Provenance: Not specified, beyond "volunteer" and "phantom scans" and "synthetic raw data."
    • New Coils, Sequences, Algorithms, Acoustic Noise: Sample sizes not specified; phantom testing mentioned.
    • Clinical tests (overall device): No clinical tests were conducted to support the substantial equivalence argument beyond the provision of clinical images to support new coils and software features.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided in the document. The document states that the images and spectra, "when interpreted by a trained physician, yield information that may assist in diagnosis." However, for validation studies, the specifics of expert involvement or ground truth establishment are not detailed.

    4. Adjudication method for the test set

    • This information is not provided in the document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • A MRMC study comparing human readers with and without AI assistance was not mentioned or performed as part of this submission. The "MyoMaps" feature was compared to "Thalassaemia Tools," which is a comparison of two tools, not a human-AI comparison.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • The document describes "LiverLab" validation using "volunteer as well as phantom scans, and synthetic raw data" and "MyoMaps" being "tested on volunteers." This suggests standalone performance evaluation for these specific features. However, detailed metrics of standalone performance are not provided, only that the "results... demonstrate that the device performs as intended."

    7. The type of ground truth used

    • MyoMaps: The comparison was against "Thalassaemia Tools." For the "volunteers" testing, the method of establishing ground truth for myocardial MR relaxation times beyond direct measurement is not specified.
    • LiverLab: Validation involved "volunteer as well as phantom scans, and synthetic raw data." The ground truth for phantom scans would be known parameters. For volunteer scans, the ground truth source for liver evaluation is not explicitly stated (e.g., biopsy results, clinical diagnosis).
    • For other features (coils, sequences), the ground truth generally relies on physical measurements and expected image properties.

    8. The sample size for the training set

    • This information is not provided in the document. The submission focuses on verification and validation of implemented features rather than detailing the development or training of algorithms.

    9. How the ground truth for the training set was established

    • This information is not provided in the document. Given that details on a training set are absent, the method for establishing its ground truth is also not mentioned.