K Number
K112382
Date Cleared
2011-12-14

(118 days)

Product Code
Regulation Number
870.5150
Panel
CV
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Terumo Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

Device Description

The Terumo Aspiration Catheter is a dual lumen rapid exchange catheter. The guidewire lumen is used to facilitate passage of a guide wire which must not exceed 0.014" (0.36 mm) in diameter. The larger extraction lumen allows the removal of thrombus (thrombi) by use of the included aspiration syringe through the extension line. The catheter has a proximal stiff region and a distal flexible region that is coated with hydrophilic polymer which generates lubricity when wet. On the distal tip a radiopaque marker band is incorporated. The proximal end of the catheter is equipped with a standard luer adapter to facilitate the attachment of the included extension line, stopcock and syringes. The provided stylet can be inserted in the catheter to assist in the delivery of the catheter to the vascular lesion. The included flushing tool is used to flush the guide wire lumen in preparation for use. A filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis or any thrombosis.

AI/ML Overview

The Terumo Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. The acceptance criteria and supporting studies are outlined below.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (Bench Testing)Reported Device Performance
ISO 10555-1 Conformance Tests
Surface inspectionConformance to ISO 10555-1 standardsPassed
Corrosion resistanceConformance to ISO 10555-1 standardsPassed
Force at break (shaft, hub)Minimal force at break of 4N (more stringent than 3N requirement of ISO 10555-1)Passed
Freedom from leakageConformance to ISO 10555-1 standardsPassed
Shelf Life Performance Tests (t=0 vs. t=shelf life)
Catheter force at breakNo significant changes in performanceMet all specifications; no significant changes observed
Catheter lubricityNo significant changes in performanceMet all specifications; no significant changes observed
Dimensional verificationNo significant changes in performanceMet all specifications; no significant changes observed
Stylet coil bond tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
Stylet to connector tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
Extension line tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
In-House Comparative Performance Tests
Kink resistanceEquivalent or superior to predicate devices (PRONTO V3, Medtronic Export XT)Equivalent or superior to predicate devices
Aspiration rateEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
Thrombus aspiration capabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
PushabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
TrackabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
Accessory Conformance Tests
StyletConformance to ISO 10555-1 standardsPassed
Stylet connectorConformance to ISO 594-1 & 2 standardsPassed
Aspiration syringeConformance to ISO 7886-1 standardsPassed
Aspiration syringe luerConformance to ISO 594-1 & 2 standardsPassed
Extension LineConformance to ISO 10555-1 standardsPassed
Extension Line luers and stopcockConformance to ISO 594-1 & 2 standardsPassed
Flushing toolConformance to ISO 594-1 & 2 standardsPassed
Biocompatibility Tests
SensitizationMeets the requirements of ISO 10993-10Meets the requirements
HemolysisNon-hemolytic (ASTM F756)Non-hemolytic
Dog thrombo-resistanceNon-thrombogenic (ISO 10993-4)Non-thrombogenic
Complement ActivationMeets the requirements of ISO 10993-4Meets the requirements
Ames AssayMeets the requirements of ISO 10993-3Meets requirements
Lymphoma Forward MutationNon-mutagenic (ISO 10993-3)Non-mutagenic
Bone Marrow MicronucleusNon-clastogenic (ISO 10993-3)Non-clastogenic
Rabbit PyrogenNon-pyrogenic (ISO 10993-11)Non-pyrogenic
CytotoxicityNon-cytotoxic (ISO 10993-5)Non-cytotoxic
Intracutaneous reactivityMeets the requirements of ISO 10993-10Meets requirements
Acute Systemic toxicityNon-toxic (ISO 10993-11)Non-toxic
Physicochemical profile (aged, sterile)Meets the requirements of USP 661Meets the requirements
Cytotoxicity (aged, sterile)Non-cytotoxic (ISO 10993-5)Non-cytotoxic
Hemolysis (aged, sterile)Non-hemolytic (ASTM F756)Non-hemolytic
Sterilization and Residuals
Sterility Assurance Level (SAL)10⁻⁶ (validated per ANSI/AAMI/ISO 11135)Achieved 10⁻⁶ SAL
Ethylene Oxide (EO) residuals≤ 4 mg per device (per ANSI/AAMI/ISO 10993-7)Will meet requirements (prior to use)
Ethylene Chlorohydrin (ECH) residuals≤ 9 mg per device (per ANSI/AAMI/ISO 10993-7)Will meet requirements (prior to use)
Shelf Life36-months (based on performance testing and package integrity validation)36-months

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes bench testing and biocompatibility testing, not a clinical study with a "test set" in the context of patient data.

  • Bench Testing: The sample sizes for each specific bench test (e.g., force at break, aspiration rate) are not explicitly stated. However, the tests were conducted using both real-time aged and accelerated aged samples to assess performance over the specified shelf life.
  • Biocompatibility Testing: The number of samples for each biocompatibility test (e.g., Sensitization, Hemolysis, Pyrogen) is not explicitly stated. The tests were performed on non-aged (sterile) and aged (sterile) samples.
  • Data Provenance: The studies are described as "in-house bench testing" and "biocompatibility testing." This implies the data were generated in a laboratory setting, likely at the manufacturer's (Terumo Medical Corporation) or a contracted testing facility's site, and are not derived from human patients or a specific country of origin in the clinical sense. These are pre-market studies to demonstrate safety and performance characteristics.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. The reported studies are bench and biocompatibility tests, which do not involve establishing ground truth by human experts in the context of diagnostic interpretation or clinical outcomes. The "ground truth" for these tests refers to established scientific and engineering standards (e.g., ISO, ASTM).

4. Adjudication Method for the Test Set

Not applicable. As these are bench and biocompatibility tests, there is no human adjudication process involved in the data collection or interpretation in the way there would be for a clinical trial or image analysis study. The results are compared against predefined pass/fail criteria from international and industry standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The provided document details pre-market bench and biocompatibility studies for a physical medical device (aspiration catheter). It does not describe an AI/software as a medical device or diagnostic imaging study that would typically involve an MRMC comparative effectiveness study to assess human reader performance with and without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. The Terumo Aspiration Catheter is a physical medical device, not an algorithm or software. Therefore, there is no "standalone" algorithm performance study described.

7. Type of Ground Truth Used

The "ground truth" for the various tests were defined by:

  • International Standards: ISO 10555-1, ISO 594-1 & 2, ISO 7886-1, ISO 10993 (various parts), ISO 11135, ISO 11607-1, ISO 11607-2.
  • National Standards: ASTM F756, ANSI/AAMI/ISO 11135, ANSI/AAMI/ISO 10993-7, USP 661.
  • Manufacturer Specifications: In-house bench testing compared the device's performance to predicate devices and likely internal design specifications that reflect expected performance characteristics (e.g., "equivalent or superior to predicate devices" for kink resistance, aspiration rate, etc.).

8. Sample Size for the Training Set

Not applicable. The studies described are for physical device testing and do not involve machine learning algorithms that require a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, for the same reason as above.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).