Search Results
Found 2 results
510(k) Data Aggregation
(163 days)
The SleepNet Mojo and Veraseal masks are intended to be used with positive airway pressure devices, such as CPAP or bi-level, operating at or above 3 cm H2O. The masks are to be used on adult patients (>30kg) for whom positive airway pressure therapy has been prescribed. Veraseal - Disposable single use (hospital/institutional) Single patient, multi-use up to 7 days (hospital/institutional) Mojo - Single patient, multi-use (home or hospital/institutional)
The SleepNet Mojo and Veraseal Full Face mask are modifications of the SleepNet Mojo Full Face mask cleared under K060273. The modifications to the cleared Mojo Full Face mask are: - different durations of use - o single use, disposable (Hospital/Institutional) Veraseal - o single patient, multi-use up to 7 days (Hospital/Institutional) Veraseal - o single patient, multi-use (Home or Hospital/Institutional) MoJo - a change in materials . - updated design of the exhaust elbow used with the mask As a result of these modifications we have established two (2) product tradenames to help distinguish them. Veraseal model has 2 durations of use - single use, disposable or single patient, multi-use up to 7 days in the hospital. Mojo model is single patient, multi-use (typically more than 7 days) for use in the home or hospital.
Here's a breakdown of the acceptance criteria and study information for the SleepNet Mojo and Veraseal Full Face Mask, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The submission primarily focuses on demonstrating substantial equivalence to predicate devices using comparative performance testing, rather than explicit acceptance criteria with numerical targets for each attribute. However, the "Comparative Performance" section and the "Table of Comparison to Predicates" implicitly serve this purpose by showing that the modified devices perform similarly to or better than the predicates.
Test / Performance Metric | Acceptance Criteria (Implicit - based on predicate performance) | Reported Device Performance (Modified Mojo and Veraseal) | Predicate Device Performance (Mojo K060273) | Predicate Device Performance (ResMed Mirage Quattro K063122) |
---|---|---|---|---|
CO2 washout (per ISO 17510-2) | Results demonstrating substantial equivalence to predicates. | Demonstrated substantial equivalence to predicates. | Performed. | Performed. |
ETCO2% at mask (increase) | Similar to or lower than predicates. | |||
- 3 cm H₂O | ≤ 5.9 (15%) | Mojo: 6.1 (17%) | ||
Veraseal: N/A (not explicitly listed for Veraseal at this pressure) | 5.9 (15%) | Implicitly not tested for this predicate | ||
- 5 cm H₂O | ≤ 5.8 (14%) | Mojo: 5.9 (14%) | ||
Veraseal: N/A (not explicitly listed for Veraseal at this pressure) | 5.8 (14%) | Implicitly not tested for this predicate | ||
- 10 cm H₂O | ≤ 5.6 (9%) | Mojo: 5.8 (12%) | ||
Veraseal: N/A (not explicitly listed for Veraseal at this pressure) | 5.6 (9%) | Implicitly not tested for this predicate | ||
- Occluded (Mojo) | ≤ 6.0 (17%) | 6.3 (21%) | 6.0 (17%) | Implicitly not tested for this predicate |
Deadspace | Similar to or lower than predicates. | |||
- Small (Mojo) | ≤ 200 ml | 160 ml | 200 ml | N/A (varies by cushion size) |
- Medium (Mojo) | ≤ 245 ml | 185 ml | 245 ml | N/A (varies by cushion size) |
- Large (Mojo) | ≤ 295 ml or similar to ResMed (242ml) | 205 ml | 295 ml | 242 ml |
Exhaust - pressure / flow | Similar flow characteristics to predicates for given pressures. | |||
- 3 cm H₂O | Mojo: ~32.00 lpm | |||
Veraseal: N/A | Mojo: 21.20 lpm | |||
Veraseal: 21.10 lpm | 32.00 lpm | N/A | ||
- 10 cm H₂O | Mojo: ~43.00 lpm | |||
Veraseal: N/A | Mojo: 37.20 lpm | |||
Veraseal: 36.40 lpm | 43.00 lpm | 8 cm H₂O: 32 lpm | ||
- 20 cm H₂O | Mojo: ~55.00 lpm | |||
Veraseal: N/A | Mojo: 53.10 lpm | |||
Veraseal: 53.00 lpm | 55.00 lpm | 20 cm H₂O: 54 lpm | ||
Pressure AAV (Anti-asphyxia valve) | Opening/closing pressures similar to predicate (Mojo K060273). | Mojo: Opening - 1.10 cm H₂O, Closing - 1.60 cm H₂O | ||
Veraseal: Opening - 1.29 cm H₂O, Closing - 1.45 cm H₂O | Opening - 0.65 cm H₂O, Closing - 1.38 cm H₂O | Not disclosed | ||
Resistance to Flow | Pass / Fail criteria is |
Ask a specific question about this device
(57 days)
The Mirage Micro channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.
The Mirage Micro is:
- to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed.
- intended for single patient re-use in the home environment and multi-patient re-use in the hospital/institutional environment.
The Mirage Micro provides seal such that airflow from a positive pressure source is directed to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face.
Mirage Micro is safe when used under the conditions and purposes intended as indicated in the labeling provided with the product.
Mirage Micro is a prescription device supplied nonsterile.
The provided document is a 510(k) summary for the ResMed Mirage Micro Traditional, a vented nasal mask, seeking substantial equivalence to predicate devices. It is not a study that proves the device meets acceptance criteria in the way typically seen for AI/ML-based medical devices or diagnostics. Instead, it demonstrates substantial equivalence through comparisons of technological characteristics and performance data via bench testing.
Therefore, many of the requested fields (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable to this type of submission.
Here's a breakdown of the available information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly state formal "acceptance criteria" with numerical thresholds in the sense of a diagnostic device. Instead, it relies on demonstrating substantial equivalence to predicate devices through comparative performance and technological characteristics. The "reported device performance" is essentially a statement of equivalence.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
CO2 performance substantially equivalent to predicate device. | "The CO2 performance of the new device and the predicate device are substantially equivalent." |
Pressure-flow characteristics and flow impedance substantially equivalent to predicate device. | "The pressure-flow characteristics and flow impedance of both the new device and the predicate device are substantially equivalent." |
Operation on the same standard flow generator setting as predicate device. | "Both the new device and the predicate are designed to operate on the same standard flow generator setting." |
Provision of seal via dual wall silicone interface. | "The new device and the predicate mask, provide seal via dual wall silicone interface." |
Incorporation of vent holes for continuous air leak. | "Both the masks incorporate vent holes to provide continuous air leak to flush out the dead space within the mask and minimize the amount of CO2 rebreathed by the patient." |
Connection to conventional air delivery hose via standard conical connectors. | "Both the masks connect to conventional air delivery hose between the mask and the positive airway-pressure source via standard conical connectors (ref: ISO 5356-1:2004)." |
Provisions for connecting oxygen and pressure sensing tubing. | "Both the masks have provisions for connecting oxygen and pressure sensing tubing via luer ports." |
Construction using molded plastic components and fabric headgear with safe materials. | "Both the masks are constructed using molded plastic components and fabric headgear. All the components of both the masks are fabricated using materials deemed safe. (ref: ISO 10993-1)." |
Intended use matching predicate devices. | "it has the same intended use;" |
Technological characteristics similar to predicate devices. | "it has similar technological characteristics to both predicates:" |
Does not raise new questions of safety and effectiveness. | "it does not raise new questions of safety and effectiveness:" |
At least as safe and effective as predicate devices. | "it is at least as safe and effective as the predicate devices Mirage Micro and Mirage Quattro." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample size for the test set: Not applicable. The submission refers to "bench testing" and "comparison with predicate" rather than a clinical trial or a test set of data points in the context of an AI/ML device.
- Data provenance: Not applicable. Bench testing is typically performed in a laboratory setting. No country of origin for clinical data is mentioned as none was used.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of experts: Not applicable. Ground truth establishment by experts is not described as part of this bench testing for a mechanical device.
- Qualifications of experts: Not applicable.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication method: Not applicable. This concept is relevant for expert disagreement resolution in diagnostic studies, not for mechanical device bench testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC study: No. This is a medical device, not an AI/ML diagnostic.
- Effect size: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone performance: Not applicable. This is a medical device, not an AI/ML diagnostic.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Type of ground truth: Not explicitly stated as "ground truth" in the AI/ML sense. The performance of the new device was compared against the performance of the predicate devices (Mirage Micro K071808 and Mirage Quattro K063122) through bench testing parameters like CO2 performance, pressure-flow characteristics, and flow impedance. The "truth" is based on the established safety and efficacy of the predicate devices.
8. The sample size for the training set
- Sample size for training set: Not applicable. This is a medical device, not an AI/ML system.
9. How the ground truth for the training set was established
- Ground truth for training set establishment: Not applicable. This is a medical device, not an AI/ML system.
Summary of the Study (as described in the 510(k) Summary):
The submission for the Mirage Micro Traditional nasal mask relies on a bench testing study to demonstrate substantial equivalence to its predicate devices (Mirage Micro Mask K071808 and Mirage Quattro Mask K063122).
The core of the "study" involved:
- Comparison of Technological Characteristics: Detail how the new device's design, materials, and features are similar to the predicates. This includes aspects like the dual-wall silicone interface, vent hole design, standard connectors, luer ports, and biocompatible materials.
- Performance Data Comparison (Bench Testing): Direct comparison of the new device's performance against the predicate Mirage Micro regarding:
- CO2 performance: Stated as "substantially equivalent."
- Pressure-flow characteristics: Stated as "substantially equivalent."
- Flow impedance: Stated as "substantially equivalent."
- Operation on standard flow generator settings: Stated as designed to operate on the same settings.
The conclusion drawn from this comparison and bench testing is that the new Mirage Micro is substantially equivalent because it shares the same intended use, similar technological characteristics, does not raise new questions of safety and effectiveness, and is at least as safe and effective as the predicate devices. Clinical data was deemed unnecessary as "Use of vented nasal masks with CPAP or Bilevel therapy is proven technology and is well accepted by the medical community. Bench testing is sufficient to demonstrate safety and efficacy..."
Ask a specific question about this device
Page 1 of 1