Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K211555
    Date Cleared
    2021-10-18

    (152 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K060211

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hypodermic Safety Needle with Syringe: The Hypodermic Safety Needle with Syringe is intended for use in the aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks.

    Hypodermic Safety Needle: The Hypodermic Safety Needle is intended to be used with a luer lock syringe for aspiration and injection of fluids for medical purpose. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.

    Device Description

    The device is used for subcutaneous injection, intramuscular injection, intravenous injection of blood, and preparation of liquid medicine. The needle tube will be pushed into the safety cap after use. It is designed to aid in the prevention of needlestick injuries.

    The device is a disposable anti-needle stick syringe made of the following components: Needle cap, Needle tube, Safety shield, Needle hub, Barrel, Plunger, Plunger stopper.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Hypodermic Safety Needle and Hypodermic Safety Needle with Syringe. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study that proves the device meets specific acceptance criteria in the way an AI/ML device would.

    However, I can extract the information you've requested based on the regulatory submission's approach to performance and safety, interpreting the "acceptance criteria" as compliance with relevant international standards and the "study" as the non-clinical performance testing.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Compliance with Standards)Reported Device Performance
    ISO 7886-1: Sterile hypodermic syringes for single use - Part 1: Syringes for manual useMeets all performance standards
    ISO 9626: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methodsMeets all performance standards
    ISO 7864: Sterile hypodermic needles for single use - Requirements and test methodsMeets all performance standards
    ISO 594-1: Conical fittings with a 6% (Luer) taper for syringes, needles, and certain other medical equipment - Part 1: General requirementsMeets all performance standards
    ISO 594-2: Conical fittings with a 6% (Luer) taper for syringes, needles, and certain other medical equipment - Part 2: Lock fittingsMeets all performance standards
    ISO 23908: Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters, and needles used for blood samplingMeets all performance standards
    ISO 6009: Hypodermic needles for single use - Color coding for identificationMeets all performance standards
    ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (Biocompatibility)Conforms to ISO 10993-1 (No new testing performed as components unchanged)
    ISO 11135:2014: Sterilization of health care products - Ethylene oxide – Requirements for development, validation and routine control of a sterilization process for medical devicesConforms to ISO 11135:2014 (No new testing performed as sterilization process unchanged)
    AAMI TIR28:2016: Product adoption and process equivalence for ethylene oxide sterilizationConforms to AAMI TIR28:2016 (No new testing performed as sterilization process unchanged)
    Sterility Assurance Level (SAL)10^-6

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the performance testing. The "data provenance" for non-clinical performance testing would typically be from in-house or contracted laboratory testing conducted by the manufacturer (Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. in China). These are prospective tests designed to show compliance with the standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to this type of device submission. The "ground truth" for these tests are the objective technical requirements and measurement methods defined by the international ISO standards. Compliance is determined by whether the device's measured performance falls within the specified limits of these standards, not by expert consensus on clinical interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This type of device performance testing does not involve human adjudication in the way clinical diagnostic studies do. Compliance against standards is an objective measurement.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a conventional medical device (safety needle/syringe), not an AI/ML diagnostic or assistive device. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a conventional physical medical device, not an algorithm or software. No standalone performance testing in this context was performed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this submission are the objective technical requirements and test methods defined by recognized international standards (ISO standards). Device performance is measured against these established, quantitative benchmarks.

    8. The sample size for the training set

    Not applicable. This is a conventional medical device, not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a conventional medical device, not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1