Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K102187
    Date Cleared
    2010-12-10

    (129 days)

    Product Code
    Regulation Number
    880.2910
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

    Device Description

    The Radiant Innovation Inc., Infrared Ear Thermometer, Model TH709L(E) is electronic thermometer using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces of the patient.

    To measure ear temperature, the ear thermometer is inserted into a patient's outer ear canal. A start button is pressed to start the measurement through the radiation exchanges. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The measured temperature then appears on a LCD display. The total operation takes a few seconds.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting study for the Radiant Innovation Inc. Infrared Ear Thermometer TH709L(E), based on the provided document:

    This document is a 510(k) summary for a Special 510(k) submission, meaning the device TH709L(E) is being asserted as substantially equivalent to a previously cleared predicate device (TH8 series) due to minor modifications. The focus of such submissions is typically demonstrating that the modifications do not raise new questions of safety or effectiveness, often through compliance with recognized standards and comparison to the predicate. Therefore, the "study" described is primarily about demonstrating compliance with accuracy standards rather than a comparative effectiveness study or a complex standalone algorithm validation.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from ASTM E1965-98 and EN12470-5:2003, as implied by compliance)Reported Device Performance (TH709L(E))
    Accuracy: ±0.2°C (±0.4°F) for 35.5°C to 42°C (95.9°F to 107.6°F)±0.2°C (0.4°F) for 35.542°C (95.9107.6°F)
    Accuracy: ±0.3°C (±0.5°F) for other temperature ranges±0.3°C (0.5°F) for other temperature ranges

    Note: The document explicitly states "Compliance to applicable voluntary standards includes ASTM E1965-98 and EN12470-5;2003". These standards define the accuracy requirements for clinical infrared thermometers. The reported performance in the comparison table mirrors these standard accuracy criteria, implying the device met these thresholds.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state a specific "test set" sample size or data provenance in terms of human subjects for a clinical study. The evaluation appears to be based on bench testing ("bench testing contained in this submission supplied demonstrate that the modification of TH709L(E) do not raise any new questions of safety or effectiveness.") and compliance with voluntary standards (ASTM E1965-98 and EN12470-5;2003). These standards typically involve specified test procedures and equipment to verify accuracy under controlled conditions, rather than a clinical trial with a specific number of patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the evaluation focuses on compliance with technical standards through bench testing, not on interpreting medical images or data that would require expert adjudication. The "ground truth" for thermometer accuracy is established by highly accurate reference thermometers and controlled temperature environments as part of the standard testing protocols specified in ASTM E1965-98 and EN12470-5:2003.

    4. Adjudication method for the test set

    This information is not applicable for the same reasons as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is an infrared ear thermometer, not an AI-powered diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable in the context of an "algorithm only" performance as would be relevant for an AI device. The device performance itself (measuring temperature) is essentially a standalone function, and its accuracy is validated against established standards.

    7. The type of ground truth used

    The ground truth used for establishing the device's accuracy is derived from measurements taken by highly accurate reference instruments under controlled environmental conditions, as dictated by the specified voluntary standards (ASTM E1965-98 and EN12470-5:2003).

    8. The sample size for the training set

    This information is not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101253
    Date Cleared
    2010-06-29

    (56 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

    Device Description

    The Radiant Innovation Inc., Infrared Ear Thermometer, Model TH60N is electronic thermometer using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces of the patient.

    To measure ear temperature, the ear thermometer is inserted into a patient's outer ear canal. A start button is pressed to start the measurement through the radiation exchanges. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The measured temperature then appears on a LCD display. The total operation takes a few seconds.

    Radiant Infrared Ear Thermometer TH60N is an ear thermometer without probe cover with some sub-functions such as: temperature unit switchable, recall memory, scan function, last reading display, ambient temperature, and backlight.

    AI/ML Overview

    The provided text is a 510(k) summary for the Radiant Innovation Inc. Infrared Ear Thermometer TH60N. It does not contain detailed acceptance criteria and a study proving the device meets those criteria in the way typically found for AI/ML medical devices. Instead, it focuses on substantial equivalence to a predicate device based on regulatory compliance and non-clinical testing.

    Therefore, I cannot extract the specific information requested in your bullet points because the document does not contain that level of detail for a performance study.

    Here's what I can extract and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    • Acceptance Criteria/Performance Standards: The document states compliance with "applicable voluntary standards includes ASTM E1965-98 and EN12470-5:2003, as well as EN 60601-1 (IEC 60601-1) and EN 60601-1-2 (IEC 60601-1-2) requirements."
      • However, it does not provide the specific quantitative acceptance criteria within those standards (e.g., maximum allowable error in temperature readings) nor does it report the device's specific performance against these criteria (e.g., "Our device had an accuracy of X degrees Celsius, which is within the Y degrees Celsius specified by ASTM E1965-98"). The document only states that "bench testing contained in this submission supplied demonstrate that the modification of TH60N do not raise any new questions of safety or effectiveness." This implies the device met the standards, but the specific data is not in this summary.

    2. Sample size used for the test set and the data provenance

    • Not explicitly stated for clinical performance. The document mentions "bench testing" but doesn't specify sample sizes for any clinical validation or the origin of data beyond "bench testing." For a thermometer, a "test set" would typically refer to a clinical study measuring human body temperatures, which is not described.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/Not stated. This type of information is typically for diagnostic devices where expert consensus (e.g., radiologists, pathologists) is needed to establish ground truth. For a thermometer, "ground truth" would be established by a reference thermometer or a controlled environment, not expert interpretation. This document does not describe how ground truth was established, only that it complies with standards.

    4. Adjudication method for the test set

    • Not applicable/Not stated. Similar to point 3, adjudication methods (like 2+1) are used for interpretive tasks in diagnostic imaging or similar fields. They are not relevant for a simple temperature measurement device like an ear thermometer.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done

    • No. This type of study is for evaluating human performance (e.g., radiologists) with and without AI assistance in diagnostic tasks. It is not relevant for an infrared ear thermometer.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, in essence. A thermometer is inherently a standalone device; the "algorithm" (its internal processing) produces a direct measurement. The "bench testing" mentioned would be evaluating this standalone performance against reference standards. However, specific details of this testing (e.g., number of measurements, temperature ranges tested, results) are not provided in this summary.

    7. The type of ground truth used

    • Implied by compliance to standards: For a thermometer, ground truth is typically established using highly accurate reference thermometers in controlled environments (e.g., blackbody cavities for infrared calibration, or direct contact thermometry for clinical validation). The document does not explicitly state the specific ground truth used but relies on adherence to standards like ASTM E1965-98 and EN12470-5:2003 which would specify such methods.

    8. The sample size for the training set

    • Not applicable/Not stated. This device is a traditional electronic thermometer, not an AI/ML device that requires a "training set."

    9. How the ground truth for the training set was established

    • Not applicable/Not stated. As it's not an AI/ML device, there's no training set or associated ground truth for training.

    In summary, this 510(k) submission focuses on meeting established regulatory standards and demonstrating substantial equivalence to a predicate device through "bench testing." It is not a clinical study report for an AI/ML device, and therefore the detailed performance metric breakdown you're looking for is absent from this specific document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070976
    Date Cleared
    2007-04-20

    (14 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K063185, K051269, K040377, K030324, K020504, K011059

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

    Device Description

    The Radiant Innovation Inc., Infrared Ear Thermometer, Model TH01BN are electronic thermometers using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces of the patient.

    To measure ear temperature, the ear thermometer is inserted into a patient's outer ear canal. A start button is pressed to start the measurement through the radiation exchanges. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The measured temperature then appears on a LCD display. The total operation takes a few seconds.

    AI/ML Overview

    The provided text is a 510(k) summary for the Radiant Innovation Inc. Infrared Ear Thermometer TH01BN. It primarily focuses on demonstrating substantial equivalence to a predicate device based on compliance with voluntary standards and bench testing, rather than a clinical study with detailed acceptance criteria and performance data as typically seen in AI/ML medical devices.

    Therefore, much of the requested information (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set details) is not present in this document because it describes a traditional medical device (thermometer) and not an AI or machine learning-based device. The "study" referenced is primarily conformance to relevant standards for clinical thermometers.

    However, I can extract information related to the acceptance criteria and reported device performance based on the standards mentioned.

    1. A table of acceptance criteria and the reported device performance

    The document states that the device demonstrated compliance to applicable voluntary standards ASTM E1965-98 and EN12470-5:2003 for clinical electronic thermometers. These standards define the acceptance criteria for accuracy.

    Acceptance Criteria (from Standards)Reported Device Performance (Implied by Compliance)
    ASTM E1965-98:
    Accuracy within specified limits for ear thermometersCompliant with ASTM E1965-98
    Temperature range coveredTested and compliant with the standard's requirements for ear thermometers
    Response timeCompliant with ASTM E1965-98
    EN12470-5:2003:
    General requirements for ear thermometersCompliant with EN12470-5:2003
    Accuracy standardsCompliant with EN12470-5:2003
    Temperature measurement uncertaintyMeets the specified uncertainty limits in the standard

    Note: The specific numerical values for accuracy and other parameters from these standards are not detailed in the provided summary, only that the device "compli(ed) to applicable voluntary standards." The device performance is "reported" as being compliant with these standards.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The "tests performed" refer to bench testing for compliance with standards, not detailed clinical trial data with patient samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The ground truth for a physical measurement device like a thermometer is established through traceable calibration standards, not expert consensus in the way an AI model's ground truth for medical images would be.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. Adjudication methods are typically used in studies involving human interpretation or annotation, which is not the case for a thermometer's performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not provided. An MRMC study is relevant for AI-assisted diagnostic tools involving human readers, which this device is not.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable in the context of an infrared ear thermometer. The device itself is a "standalone" instrument for measuring temperature. There is no "algorithm only" performance separate from the device's function.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For a thermometer, the "ground truth" for its accuracy is typically established by comparing its readings against a calibrated reference thermometer in a controlled environment, traceable to national/international standards (e.g., against a black body radiator or a liquid-in-glass thermometer of known accuracy). This is part of the bench testing implied by compliance with ASTM and EN standards.

    8. The sample size for the training set

    This information is not applicable and not provided. This device is an infrared ear thermometer, a hardware device, not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1