Search Results
Found 2 results
510(k) Data Aggregation
(254 days)
The ABL800 FLEX analyzers are intended for In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cC22+, cClu, cLac, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FHHb and FHbF). In addition the ABL800 FLEX is intended for In vitro testing of samples of expired air for the parameters p02 and for In vitro testing of pleura samples for the pH parameter.
pH: pH is the indispensable measure of acidemia or alkalemia and is therefore an essential part of the pH/blood gas measurement. The normal function of many metabolic processes requires a pH to be within a relatively narrow range.
pO2: The arterial oxygen tension is an indicator of the oxygen uptake in the lungs.
pCO2: pCO2 is a direct reflection of the adequacy of alveolar ventilation in relation to the metabolic rate.
Potassium (cK+): the measurements of the concentration of plasma are used to monitor the electrolyte balance.
Sodium (cNa+); the measurements of the concentration of sodium ions in plasma are used to monitor the electrolyte balance.
Calcium (cCa++): the measurements of the concentration of calcium ions in plasma are used to monitor the electrolyte balance.
Chloride (cCl-): the measurements of the concentration of chloride ions in plasma are used to monitor the electrolyte balance.
Glucose (cGlu): The glucose measure the concentration of glucose in plasma. The glucose measurements are used to screen for, diagnose and monitor diabetes, pre-diabetes and hyper and hypoglycemia.
Lactate (cLac): The lactate measure the concentration of lactate in plasma. Lactate measurements serve as a marker of critical imbalance between tissue oxygen demand and oxygen supply.
Bilirubin (ctBil): The bilirubin measure the total concentration of bilirubin in plasma. ctBil is used to assess the risk of hyperbilirubinemia.
Total Hemoglobin (ctHb): ctHb is a measure of the potential oxygen-carrying capacity of the blood.
Oxygen Saturation (sO2): sO2 is the percentage of oxygenated hemoglobin in relation to the amount of hemoglobin capable of carrying oxygen. sO2 allows evaluation of oxygenation.
Fraction of Oxyhemoglobin (FO2Hb): FO2Hb is a measure of the potential oxygen transport capacity; that is the fraction of oxyhemoglobins present (tHb) including dyshemoglobins.
Fraction of Carboxyhemoglobin (FCOHb is the fraction of carboxyhemoglobin. It is incapable of transporting oxygen.
Fraction of Methemoglobin (FMetHb): FMetHb is the fraction of methemoglobin. It is incapable of transporting oxygen.
Fraction of Deoxyhemoglobin in Total Hemoglobin (FHHb): FHHb is the fraction of deoxyhemoglobin in total hemoglobin. It can bind oxygen then forming oxyhemoglobin.
Fraction of Fetal Hemoglobin (FHbF): Fetal hemoglobin consist of two a-chains and two B-chains, and has a higher oxygen affinity than adult Hb.
Creatinine (cCrea): The creatinine measure the concentration of creatinine in blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.
Pleural pH: The pH measurement of pleural fluid can be a clinically useful tool in the management of patients with parapneumonic effusions. Critical values: pH >7.3 is measured in uncomplicated parapneumonic effusions. All pleural effusions with a pH of
ABL800 FLEX with AQURE connectivity is a stationary, automated system intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCl-, cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions F02Hb, FCOHb, FMetHb, FHHb and FHbF).
The modification consists of integration with the Medical Device Data System (MDDS) called AQURE system. The software enables the initiation of device actions on connected ABL800 series analyzers.
The provided text is a 510(k) summary for the Radiometer ABL800 FLEX with AQURE connectivity. This document focuses on demonstrating substantial equivalence to a predicate device and addresses a software modification (integration with the AQURE system), not a study proving the original device's performance against detailed acceptance criteria for its clinical parameters.
Therefore, the information required to fully answer your request regarding performance criteria and a study proving the device meets those criteria for the measured clinical parameters (pH, pO2, pCO2, etc.) is not present in this document. This document specifically states: "No performance characteristics are affected by the change. The performance data submitted in the original submission (K041874 as modified by K043218, K050869, K051968, K100777 and K110416) still apply."
However, I can extract information related to the software modification and its acceptance:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (for AQURE integration) | Reported Device Performance (Adherence to criteria) |
---|---|
Mitigation of all identified hazards to As Low As Reasonably Practicable (ALARP) as per ISO 14971. | FMEA risk analysis conducted, and all identified hazards mitigated to ALARP. |
Verification of software mitigations by using test protocols. | Software mitigations verified by using test protocols. |
Results met predefined acceptance criteria. | Results met predefined acceptance criteria. (Specific criteria for software functionality are not detailed in this summary document.) |
Remote control functionality to other devices than ABL90 FLEX not available in US. | The functionality is not available to the customer, and any information regarding this functionality is either removed from the manual or indicated as unavailable. Device actions for other analyzers have a note added stating "This feature is not available in the USA". |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not explicitly stated for the software verification. The document mentions "test protocols" were used.
- Data Provenance: Not specified, but given it's a product from Radiometer Medical ApS in Denmark, it can be inferred that the testing likely occurred in a controlled lab or manufacturing environment. The study is retrospective in the sense that it relies on previously established performance data for the core ABL800 FLEX device and focuses on the impact of the new software integration.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- The document does not mention the use of experts or ground truth establishment for the software verification. The assessment appears to be based on engineering and risk management principles (FMEA, test protocols).
4. Adjudication method for the test set
- Not applicable as this is a software modification verification, not a clinical study involving human judgment.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. The device is for in-vitro diagnostic testing (blood gas, electrolytes, metabolites, oximetry), not an imaging or interpretive AI device where MRMC studies are typically performed. The software modification is for data management connectivity (MDDS), not AI-assisted interpretation.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- The device itself (ABL800 FLEX) is a standalone automated diagnostic system. The AQURE connectivity is a software integration to manage data and device actions. The validation of the software integration focuses on its functionality and safety, not standalone diagnostic performance, as the diagnostic performance relies on the already established ABL800 FLEX analyzer.
7. The type of ground truth used
- For the software modification, the "ground truth" would be the expected functional behavior and safety requirements defined during the design and risk analysis phases. For example, a "ground truth" might be that a specific command sent via AQURE results in the correct action on the ABL800 FLEX analyzer, or that data transfer is accurate. However, this is not a biological or clinical ground truth.
8. The sample size for the training set
- Not applicable. This is a medical device connectivity software update and risk assessment, not a machine learning or AI algorithm requiring a training set in the conventional sense.
9. How the ground truth for the training set was established
- Not applicable for the same reason as above.
In summary: This 510(k) summary explicitly states that the software modification (AQURE connectivity) does not affect the performance characteristics of the ABL800 FLEX analyzer. Therefore, performance data for the clinical parameters refers back to the original submissions (K041874 and subsequent modifications), which are not detailed in this document. The provided text only describes the verification and validation activities conducted for the software change itself, primarily focusing on risk management and functional testing rather than clinical performance studies.
Ask a specific question about this device
(119 days)
The ABL 80 FLEX is a portable, automated system that measures pH, blood gases, electrolytes, glucose, and hematocrit in whole blood. The ABL80 FLEX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of- care setting.
These tests are only performed under a physicians order:
pH: pH is the indispensable measure of acidemia or alkalemia and is therefore an essential part of the pH/Blood gas measurement.
pO2: The arterial oxygen tension is an indicator of the oxygen uptake in the lungs.
pCO2: pCO2is a direct reflection of the adequacy of alveclar ventilation in relation to the metabolic rate.
Potassium (cK*): The potassium level is the predominant intracellular cation. It is fundamental for correct neuromuscular activity
Sodium (cNa"): The sodium ion is the most abundant cation in plasma. It is the foremost agent involved in maintenance of osmolality and body fluid volumes.
Calcium (cCa2): The calcium ion is the most abundant mineral element in the human body and is involved in numerous enzymatic processes, blood coagulation, cell growth, and membrane transport mechanisms as well as plays an important role in nervous impulse conduction, neuromuscular transmission, and muscular contraction and relaxation.
Chloride (cCl): The chloride ion is the main extracellular anion and plays an important role in maintaining electrical neutrality.
Hematocrit (Hct): The hematocrit measurement is the ratio of the volume of red blood cells in whole blood in comparison to the total volume.
Glucose (cGlu): The glucose measurements are used to screen for, diagnose and monitor glycemic levels in potential pre-diabetic, diabetic, hypoglycemic patients.
The ABL80 FLEX System consists of a modular analyzer incorporating a user interface module with a large color touch screen interfacing to analyzer electronic and fluidic modules. The user interface module contains the analyzer CPU and all of the required electronic interfaces for external communication and data storage. The system also includes a reagent cartridge for the calibration and automatic quality control of the analyzer and sensor system. The calibration and quality control reagents are packaged in sealed foil pouches, similar to the existing ABL 77 cal pack. The analyzer and consumables incorporate "smart chip" technology for unique identification and lot specific calibration data.
The ABL80 FLEX System is a portable, automated system that measures pH, blood gases, electrolytes, glucose, and hematocrit in whole blood.
The following information describes the acceptance criteria and the study that proves the device meets the acceptance criteria:
- Table of acceptance criteria and reported device performance:
Analyte | Acceptance Criteria (Comparison to ABL™700 or other standards) | Reported Device Performance |
---|---|---|
pH | Method comparison against 2 ABL725 analyzers | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
Sodium (cNa+) | Method comparison against 2 ABL725 analyzers | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
Potassium (cK+) | Method comparison against 2 ABL725 analyzers | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
Calcium (cCa2+) | Method comparison against 2 ABL725 analyzers | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
Chloride (cCl-) | Method comparison against 2 ABL725 analyzers | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
pCO2 | Method comparison against target tonometry levels | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
pO2 | Method comparison against target tonometry levels | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
Glucose | Method comparison against serum standard per NCCLS Publication RS1-A | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
Hematocrit | Method comparison against micro-hematocrit method (centrifugation per CLSI standard H7-A3) | Not explicitly stated, but assumed to meet criteria based on substantial equivalence |
-
Sample size used for the test set and data provenance:
The document does not specify the exact sample size for the test set. The provenance of the data (e.g., country of origin, retrospective/prospective) is also not specified. -
Number of experts used to establish the ground truth for the test set and qualifications of those experts:
This information is not provided in the document. The ground truth was established by comparison to other analytical methods/devices rather than expert consensus. -
Adjudication method for the test set:
Not applicable, as the performance testing involved method comparisons against established standards or predicate devices, rather than expert adjudication of results. -
If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
No, an MRMC comparative effectiveness study was not done. The device is an automated system for measuring analytes, not an imaging or diagnostic AI system requiring human reader performance studies. -
If a standalone study (algorithm only without human-in-the-loop performance) was done:
Yes, the performance data described is a standalone evaluation of the ABL80 FLEX System's analytical performance (algorithm only) compared to reference methods and predicate devices. There is no mention of human-in-the-loop performance in the context of this device's validation. -
The type of ground truth used:
The ground truth used was based on established analytical methods and reference devices:- For pH, sodium, potassium, calcium, and chloride: The measurement standard was the ABL™700, and method comparison was conducted against 2 ABL725 analyzers.
- For blood gases (pCO2 and pO2): Target tonometry levels, with values calculated from certified tonometry gas mixtures traceable to NIST.
- For glucose: Serum standard per NCCLS Publication RS1-A.
- For hematocrit: Micro-hematocrit method using centrifugation per CLSI standard H7-A3.
-
The sample size for the training set:
This information is not specified in the document. The ABL80 FLEX System is a medical device for analyte measurement, and the concept of a "training set" in the context of machine learning is not directly applicable to its analytical validation as described. The system's design and performance are based on established electrochemical and spectrophotometric principles, likely calibrated during manufacturing. -
How the ground truth for the training set was established:
Not applicable, as the document does not describe a training set in the context of machine learning. The ground truth for the performance evaluation was established through comparison with established analytical methods and reference devices, as detailed in point 7.
Ask a specific question about this device
Page 1 of 1