Search Results
Found 4 results
510(k) Data Aggregation
(357 days)
The Su2ura® Approximation Device is indicated in endoscopic surgery for the placement of interrupted or running stitches in soft tissue.
The Anchora Su2ura® Approximation Device is a sterile single use device is indicated in endoscopic surgery for the placement of interrupted or running stitches in soft tissue. The device is a 5mm single-use endoscopic suturing device loaded with threaded anchors. It is designed to deploy up to 19 anchors by a trigger action of the handle of the delivery system. After deployment, each anchor is anchored within the tissue in which it was deployed. After deployment of 2 or more anchors, the suture that passes between the anchors can be tightened to pull the anchors toward each other and approximate the tissue. The delivery system consists of a plastic handle and a 36-cm stainless steel shaft. The handle includes a Trigger, a Knob, and a Toggle Button. A Cap can be added at the distal edge of the Shaft to advance extracorporeal suture knots. The anchors can be deployed in single-stitch or running- stitch configuration. The device deploys the anchors through a needle with an outer diameter of 2.1 mm, available in two lengths: 5 mm and 9 mm protrusion. The Anchor is advanced outside of the needle by a Pushrod. Each anchor, made of stainless steel, is threaded with an off-the-shelf UHMWPE suture manufactured by Teleflex Medical. The device provides the anchors already threaded on this suture. The device is sterile packaged and designed for single use only. The device is manufactured from medical grade materials (mainly plastics and stainless steels).
The Su2ura Approximation Device is a sterile, single-use endoscopic suturing device with the following acceptance criteria and reported performances:
1. Acceptance Criteria and Device Performance
Characteristic | Acceptance Criteria | Reported Device Performance |
---|---|---|
Biocompatibility | Evaluate all patient contact materials in accordance with ISO 10993-1. | Biocompatibility assessments were performed in accordance with ISO 10933-1. Testing included extractable and leachable studies (ISO 10993-18) and a toxicological risk assessment (ISO 10993-17). Implied that the device met the criteria. |
Human Factors | Verify the usability of the device with representative users. Training provided and Instructions for Use (IFU) must be clear and accurate. | A Human Factors Engineering (HFE) Study was performed with 18 representative users, evaluating the usability of all tasks. The study confirmed device usability and demonstrated that the training and IFU were clear and accurate. Implied that the device met the criteria. |
Pre-Clinical Safety & Performance | Assess safety and performance under GLP conditions, in compliance with ISO 10993-6. Incision sites should be successfully closed, and long-term safety should be similar between test and control items. | An animal study (8 pigs) was performed under GLP conditions and in compliance with ISO 10993-6. 5 pigs for the test item and 3 for control. All incisions were successfully closed, and histopathological data indicated long-term safety was similar between Test and Control Items. Implied that the device met the criteria. |
Mechanical Performance | Conform to USP requirements, applicable standards, product specifications (e.g., dimensional verification, component compatibility, tissue anchoring pull-out, corrosion, deployment reliability, suture pull-out strength, tensile strength). | Bench testing included dimensional verification, component compatibility, tissue anchoring pull-out, corrosion testing per ASTM F2129, deployment reliability, suture pull-out strength, and tensile strength per USP . The results of all studies confirmed substantial equivalence between the subject and predicate designs, implying these criteria were met. |
MRI Safety & Compatibility | Verified per ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119. | MRI safety and compatibility was verified per the specified ASTM standards. The device was deemed MR Conditional. Implied that the device met the criteria. |
Packaging Integrity | Confirmed by testing in accordance with ASTM F1929-15, ASTM DI69-16, and ASTM F1886/F1886M-16. | Packaging integrity was confirmed by testing in accordance with the specified ASTM standards. Implied that the device met the criteria. |
2. Sample Size Used for the Test Set and Data Provenance
- Biocompatibility: Not explicitly stated as a test set with a specific sample size of biological samples, but rather a set of material characterization studies (extractables and leachables) and toxicological risk assessments. The input does not provide information on the number of samples for these tests.
- Human Factors: 18 representative users. The provenance of the users (e.g., country of origin, whether they were clinicians or laypersons, etc.) is not specified. The study appears to be prospective.
- Pre-Clinical Performance (Animal Study): 8 pigs (5 for test items, 3 for control items). The provenance of the data is a GLP animal study. It is a prospective study.
- Mechanical Performance, MRI Safety, Packaging Integrity: These involved various bench tests. The sample sizes for each specific test (e.g., number of devices tested for tensile strength or deployment reliability) are not explicitly stated in the provided text. The data provenance is from laboratory/bench testing.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Biocompatibility: No explicit external experts are mentioned for establishing ground truth. Compliance was based on ISO standards.
- Human Factors: The "representative users" assessed usability. It's not stated whether these were "experts" in the sense of clinicians or if their feedback constituted a "ground truth" to be adjudicated. Their feedback was used to confirm usability.
- Pre-Clinical Performance (Animal Study): Not explicitly stated how many experts were involved in evaluating the histopathological data, but it would typically involve veterinary pathologists. Their specific qualifications are not detailed.
- Mechanical Performance, MRI Safety, Packaging Integrity: Ground truth for these studies relies on recognized ASTM and USP standards and internal engineering measurements. No external "experts" were mentioned for ground truth establishment beyond standard scientific and engineering practices.
4. Adjudication Method for the Test Set
The document does not specify any formal adjudication method (like 2+1 or 3+1) for the various studies conducted as part of the regulatory submission.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is a surgical tool, and its effectiveness is not assessed through human reader interpretation of images, but rather through direct use in surgery, performance in bench tests, and pre-clinical animal studies. Clinical testing was expressly stated as "not required to demonstrate substantial equivalence."
6. Standalone Performance Study
The studies described (bench testing, animal study, human factors) are essentially "standalone" in the sense that they evaluate the device itself. However, the term "standalone performance" often refers to the algorithm-only performance of AI devices without human-in-the-loop. This device is a mechanical surgical tool, not an AI or imaging diagnostic device, so this concept does not directly apply.
7. Type of Ground Truth Used
- Biocompatibility: Compliance with ISO 10993 standards (e.g., acceptable levels of extractables/leachables, lack of toxicological risk).
- Human Factors: User feedback and observed usability were used to confirm clarity of training and IFU, and the overall usability of the device. This is a form of self-reported or observed performance against predetermined usability goals.
- Pre-Clinical Performance (Animal Study): Histopathological data (from tissue analysis) and clinical monitoring of animals provided the ground truth for safety and performance (e.g., successful closure, similar long-term safety profile to controls).
- Mechanical Performance, MRI Safety, Packaging Integrity: These rely on objective measurements and adherence to established industry standards (ASTM, USP) and predefined product specifications.
8. Sample Size for the Training Set
This question is not applicable. The Su2ura® Approximation Device is a mechanical surgical device, not a machine learning or AI model, so there is no "training set" in the context of algorithm development.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(29 days)
Force Fiber® Non-absorbable Surgical Sutures are indicated for use in approximation and/or ligation of soft tissues, including use in cardiovascular surgeries and the use of allograft tissue for orthopedic surgeries.
Force Fiber Suture - Black is a non-absorbable, sterile, surgical suture composed of Ultra High Molecular Weight Polyethylene (UHMWPE) dyed black using D&C Black #4 not to exceed 1.0% by weight. It is an uncoated braid offered in a variety of cut lengths, with or without needles, and provided sterile for single use only. Force Fiber Suture - Black is available in sizes 0, 1 and 2, and meets all surgical suture requirements established by the USP for non-absorbable surgical sutures except for oversize diameter.
This document is a 510(k) summary for a medical device called "Force Fiber® Suture - Black." It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than presenting a study where specific performance criteria are measured against acceptance criteria. Therefore, much of the requested information regarding study design, sample sizes, expert involvement, and ground truth establishment is not present in this document.
However, I can extract the general acceptance criteria and what is reported about the device's performance based on the provided text.
Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Details of Acceptance Criteria | Reported Device Performance |
---|---|---|
Surgical Suture Requirements (USP) | Meets all surgical suture requirements established by the USP for non-absorbable surgical sutures, specifically for: |
- Suture diameter
- Tensile strength
- Needle attachment | "Force Fiber Suture - Black is available in sizes 0, 1 and 2, and meets all surgical suture requirements established by the USP for non-absorbable surgical sutures except for oversize diameter."
"Force Fiber Suture - Black is tested in accordance with USP - non-absorbable surgical sutures for suture diameter, tensile strength and needle attachment, and meet the requirements of the Class II Special Controls Guidance: Surgical Sutures; Guidance for Industry and FDA; June 3, 2003." |
| Material Biocompatibility (ISO 10993-1:2018) | Evaluation through biological qualification safety tests as outlined in ISO 10993-1: 2018 --Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing. This implies meeting the safety requirements for medical devices in contact with the body. | "All materials used in the fabrication of the Force Fiber Suture - Black were evaluated through biological qualification safety tests as outlined in ISO 10993-1: 2018 --Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing." |
| MRI Safety | "MR Safe" designation, meaning it poses no known hazards in MR environments. | "Force Fiber Suture - Black is tested to demonstrate it is "MR Safe" and poses no known hazards in MR environments." |
| Substantial Equivalence | No differences in technological characteristics and performance characteristics compared to predicate devices, and the difference (addition of black dye) does not raise new questions of safety or efficacy, making it as safe and effective as currently marketed predicate devices. This is the primary "acceptance criterion" for a 510(k) pathway. | "Force Fiber Suture – Black is substantially equivalent to its predicate Force Fiber Suture devices because there are no differences in technological characteristics and performance characteristics between the proposed and predicate devices. The proposed sutures have the same fundamental design and intended use as the predicate devices. The difference between Force Fiber Suture - Black and its predicate Force Fiber Suture devices is the addition of black dye... This difference does not raise new questions of safety or efficacy. Therefore, Force Fiber Suture - Black is as safe and effective as its currently marketed predicate devices." |
Information Not Available in the Document:
Because this document is a 510(k) summary demonstrating substantial equivalence for a surgical suture, not a clinical study report for an AI/imaging device, the following information is not provided:
- Sample size used for the test set and the data provenance: Not applicable in this context. The testing mentioned refers to material properties and biocompatibility, not performance against a specific "test set" of patient data.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically understood for AI algorithms (e.g., diagnostic accuracy) is not relevant here.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a surgical suture, not an AI-assisted diagnostic tool.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For material properties, the "ground truth" would be established by validated testing methods and specifications (e.g., USP standards), not expert consensus on medical images or outcomes.
- The sample size for the training set: Not applicable. This is not an AI/ML device requiring a training set.
- How the ground truth for the training set was established: Not applicable.
In summary, the provided document describes a medical device (surgical suture) and demonstrates its compliance with established regulatory standards (USP, ISO) and substantial equivalence to existing predicate devices. It does not involve AI or imaging diagnostics, and therefore, many of the requested data points related to those types of studies are not relevant or present.
Ask a specific question about this device
(131 days)
The ExpressBraid Graft Manipulation is intended for use in soft tissue approximation and/or ligation. The suture may be provided individually or be incorporated as a component, into surgeries where constructs including those with allograft or autograft tissue are used for repair.
The ExpressBraid Graft Manipulation device consists of a stainless steel, straight needle attached to a strand of MaxBraid suture. The two ends of the strand of suture are crimped into the needle. The suture may be all white or blue/white MaxBraid suture. The MaxBraid sutures are braided, non-absorbable sutures available in a white configuration consisting of 100% Ultra High Molecular Weight Polyethylene (UHMWPE) fibers and a blue and white co-braid configuration consisting of Deklene II, Polypropylene suture braided into 100% UHMWPE fibers. The suture is provided uncoated. The colorant used in the blue and white co-braid consists of [phthalocyaninato (-2)] cooper
This document describes the ExpressBraid Graft Manipulation device, a non-absorbable surgical suture. The information provided focuses on demonstrating substantial equivalence to a predicate device rather than a study proving the device meets specific acceptance criteria in a clinical context with AI.
However, based on the provided text, we can extract details related to the performance testing conducted to support substantial equivalence. It's important to note that this is not a study assessing AI performance or human reader improvement.
Here's an analysis of the provided text in relation to your request, with the understanding that the "device" in this context is the ExpressBraid Graft Manipulation suture:
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly present a table of acceptance criteria with corresponding performance data in the typical sense for an AI device. Instead, it states that the device was tested against USP (United States Pharmacopeia) standards for non-absorbable surgical sutures and also compared to a predicate device.
Acceptance Criteria Category | Specific Criteria (Implicitly from USP) | Reported Device Performance |
---|---|---|
Tensile Strength | Meets USP requirements | Meets USP requirements |
Diameter | Meets USP requirements | Meets USP requirements |
Needle Attachment | Meets USP requirements | Meets USP requirements |
Equivalence to Predicate | Performs equivalently to Arthrex #2 Suture | Testing concluded equivalent performance |
Biocompatibility | Meets ISO 10993-1 standard requirements | Meets ISO 10993-1 requirements |
2. Sample size used for the test set and the data provenance
The document does not specify the sample size for the performance testing (tensile strength, diameter, needle attachment, or comparative testing).
The data provenance is from laboratory testing performed by the manufacturer, Biomet Manufacturing Corp., as part of their 510(k) submission. No information on country of origin of the data or whether it's retrospective or prospective is given, as it's mechanical and biocompatibility testing, not clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to the type of device and testing described. The "ground truth" for mechanical and biocompatibility testing is established by adherence to recognized standards (USP, ISO 10993-1) and direct measurement, not by expert consensus on clinical findings.
4. Adjudication method for the test set
This question is not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus on clinical diagnoses or image interpretations, not for direct mechanical or chemical property testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader, multi-case comparative effectiveness study was not done. This type of study is relevant to AI-assisted diagnostic tools, not to a surgical suture. The document describes comparative performance between two sutures (the subject device and a predicate) in terms of mechanical properties.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. This device is a physical surgical suture, not a software algorithm.
7. The type of ground truth used
The "ground truth" used for this device's performance assessment relies on:
- Established standards: United States Pharmacopeia (USP) requirements for surgical sutures and ISO 10993-1 for biocompatibility.
- Direct measurement and comparison: Mechanical and chemical properties were measured and compared to the predicate device.
8. The sample size for the training set
This question is not applicable. There is no "training set" as this is not an AI/machine learning device.
9. How the ground truth for the training set was established
This question is not applicable. As there is no training set for an AI model, there is no ground truth established for one.
In summary: The provided FDA 510(k) summary focuses on demonstrating the substantial equivalence of a surgical suture based on its mechanical properties and biocompatibility as compared to predicate devices and recognized standards. It does not contain information related to AI performance, clinical trials with human readers, or expert-based ground truth establishment as would be relevant for an AI/ML medical device.
Ask a specific question about this device
(283 days)
The MedShape, Inc., Porous Morphix® Suture Anchor with Force Fibere is intended for fixation of suture to bone in the shoulder, footlankle, knee, hand/wrist and elbow in the following procedures:
- Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Shoulder: Biceps Tenodesis. Acromio-Clavicular Separation Repair, Deltoid Repair. Capsular Shift or Capsulolabral Reconstruction
- Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Foot/Ankle: Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament Repair
- Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Knee: Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Ulnar Collateral Ligament Scapholunate Ligament Reconstruction, Hand/Wrist: Reconstruction, Radial Collateral Ligament Reconstruction
- Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Elbow: collateral Ligament Reconstruction
The proposed Porous Morphix® Suture Anchor with Force Fiber® Suture is a line extension to our Morphix® product family and is substantially equivalent to the Morphix® Suture Anchor with Opti-Fiber™ Suture, cleared under K091202, in which the features and intended uses are the same. In addition, the technological characteristics of the Morphix® and the Porous Morphix® Suture Anchor are equivalent. Differences between the Porous Morphix® Suture Anchor and the MedShape predicate Morphix® Suture Anchor include an extruded porous layer on the top surface of the device and a change in the preferred suture supplier (suture supplier CP Medical, previously cleared under K041894, was switched to suture supplier Teleflex Medical, previously cleared under K033654). This porous surface is derived directly from the implant body and is not a sintered coating.
The provided text describes a 510(k) submission for a medical device, the Porous Morphix® Suture Anchor with Force Fiber® Suture. However, it does not contain information about acceptance criteria or a study proving the device meets specific acceptance criteria in the way typically associated with AI/ML device performance.
Instead, the document focuses on demonstrating substantial equivalence to a predicate device (MedShape, Inc., WedgeLoc™ Suture Anchor with Opti-Fiber™ Suture - now referred to as Morphix®, K091202) based on:
- Same Intended Uses: The new device has the same indications for use as the predicate.
- Equivalent Technological Characteristics: Analysis of non-clinical test results (monotonic and post cyclic suture fixation strength in Sawbone® bone analogue [pull-to-failure], dimensional verification, and material safety data [both bio and MRI compatibility]) supports this equivalence.
- Minor Differences: The differences are an extruded porous layer on the top surface of the device and a change in suture supplier. These differences were determined not to raise questions concerning safety and effectiveness.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and reported device performance.
- Sample size used for the test set and data provenance.
- Number of experts and their qualifications for ground truth.
- Adjudication method.
- MRMC comparative effectiveness study.
- Standalone performance.
- Type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
These points are relevant to AI/ML device validation, which is a different regulatory pathway and type of study than what is presented in this 510(k) summary for a physical medical device (suture anchor).
Ask a specific question about this device
Page 1 of 1