Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K131173
    Date Cleared
    2014-09-18

    (511 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K032985

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FIA Blood Glucose Monitoring System (G2) is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The FIA Blood Glucose Monitoring System (G2) is intended to be used by a single patient and should not be shared.

    The FlA Blood Glucose Monitoring System (G2) is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The FIA Blood Glucose Monitoring System (G2) should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The FIA Blood Glucose Test Strips (G2) are for use with the FIA Blood Glucose Meter (G2) to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

    The FIA Glucose Control Solutions are for use with the FIA Blood Glucose Meter (G2) and FIA Blood Glucose Test Strips (G2) to check that the meter and test strips are working together properly and that the test is performing correctly.

    Device Description

    FIA Blood Glucose Monitoring System (G2) consists of:
    (1) Glucose Meter
    (2)Glucose Test Strips
    (3)Two levels of glucose control solutions (L1 and L2) may be purchased separately. Glucose control solutions were previously cleared under K032985.
    (4)Check Strip
    (5)Instruction for use

    FIA Blood Glucose Monitoring System (G2) is an electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor. When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use.

    The FIA Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) assessing the performance of the FIA Blood Glucose monitoring system (G2) and FIA Blood Glucose Test strips (G2). There are two levels of controls (L1 and L2).

    The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover.

    The device is calibrated by implicit coding process. While inserting the test strip into strip slot to perform the blood glucose test, the coding procedure is complete. The meter will apply formula including this parameter of code to calculate the glucose value.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance:

    The document states that the FIA Blood Glucose Monitoring System (G2) is compliant with the standard ISO 15197:2003 In vitro diagnostic test systems-Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. However, the exact numerical acceptance criteria from this standard and the specific reported device performance against those criteria are not explicitly detailed in the provided text.

    The document only makes a general statement: "All the relevant activities were performed by professionals and the results demonstrated that the predetermined acceptance criteria were fully met."

    To properly answer this section, one would typically expect a table like this:

    Performance Metric (from ISO 15197:2003)Acceptance Criteria (from ISO 15197:2003)Reported Device Performance (FIA BGMS G2)
    For glucose concentration
    Ask a Question

    Ask a specific question about this device

    K Number
    K113314
    Date Cleared
    2012-11-09

    (366 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K032985

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary drawn from the fingertips, venous and arterial whole blood samples. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by health care professionals in a clinical setting:

    The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

    The PRECICHEK KP Blood Glucose Test Strip is for use with the PRECICHEK NS-101 POCT Professional Blood Glucose Meter to quantitatively measure glucose in fresh capillary drawn from the fingertips, venous and arterial whole blood samples.

    The PRECICHEK Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals to assess the performance of the PRECICHEK NS-101 POCT Professional Blood Glucose monitoring system and the PRECICHECK KP Blood Glucose Test strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    Device Description

    PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System consists of:

    • (1) Glucose Meter
    • (2) Glucose Test Strips
    • (3) Two levels of glucose control solutions (Level I and Level II) may be purchased separately. Glucose control solutions were previously cleared under K032985.
    • (4) Check Strip
    • (5) Instruction for use

    [Test Principle]
    The PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System is electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor.

    When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use.

    [Control Solution]
    The PRECICHEK Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) by healthcare professionals to assess the performance of the PRECICHEK NS-101 POCT Professional Blood Glucose monitoring system and the PRECICHECK KP Blood Glucose Test strips. There are two levels of controls (Levels 1,2).

    [Check Strip]
    The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover.

    [Device Calibration]
    The device is calibrated by implicit coding process. The code number is the last two digits of the strip lot number. The user should input the strip lot number into the memory of the meter before use.

    While performing the blood glucose test, the strip lot number printed on the vial label is scanned to match the one in the memory. The meter will apply formula including this parameter to calculate the glucose value.

    AI/ML Overview

    The provided 510(k) summary (K113314) describes the PRECICHEK NS-101 POCT Professional Blood Glucose Monitoring System. The summary states that the device's performance was evaluated against the standard ISO 15197:2003.

    Here's the breakdown of the information requested:

    1. A table of acceptance criteria and the reported device performance

    The summary explicitly states that the device is compliant with ISO 15197:2003 for in vitro diagnostic test systems for blood glucose monitoring. The key performance criteria from this standard (and assuming the device met them as stated in the summary) are:

    Acceptance Criteria (from ISO 15197:2003)Reported Device Performance
    Accuracy (System Accuracy):Met Acceptance Criteria
    - For glucose concentrations
    Ask a Question

    Ask a specific question about this device

    K Number
    K120064
    Date Cleared
    2012-09-05

    (240 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K032985

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PRECICHEK Cloudia Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The PRECICHEK Cloudia Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared.

    The PRECICHEK Cloudia Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The PRECICHEK Cloudia Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

    The PRECICHEK ACH Blood Glucose Test Strips are for use with the PRECICHEK Cloudia Blood Glucose Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

    The PRECICHEK Glucose Control Solutions are for use with the PRECICHEK Cloudia Blood Glucose Monitoring System and PRECICHEK ACH Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    Device Description

    PRECICHEK Cloudia Blood Glucose Monitoring System consists of:

    • (1)Glucose Meter
    • (2)Glucose Test Strips
    • (3) Two levels of glucose control solutions (Level I and Level II) may be purchased separately. Glucose control solutions were previously cleared under K032985.
    • (4)Check Strip
    • (5)Instruction for use

    [Test Principle]

    PRECICHEK Cloudia Blood Glucose Monitoring System is an electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor. When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use.

    [Control Solution]

    The PRECICHEK Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) assessing the performance of the PRECICHEK Cloudia Blood Glucose monitoring system and PRECICHEK ACH Blood Glucose Test strips. There are two levels of controls (Levels 1,2).

    [Check Strip]

    The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover.

    [Device Calibration]

    The device is calibrated by implicit coding process. While inserting the test strip into strip slot to perform the blood glucose test, the coding procedure is complete. The meter will apply formula including this parameter of code to calculate the glucose value.

    AI/ML Overview

    The PRECICHEK Cloudia Blood Glucose Monitoring System acceptance criteria and study information are as follows:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document explicitly states that the device is compliant with ISO 15197:2003. This standard specifies requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. Therefore, the acceptance criteria are those defined by ISO 15197:2003. While the specific numerical performance metrics are not detailed in the provided text, the overall statement of compliance implies that the device met these criteria.

    Acceptance Criteria (Based on ISO 15197:2003)Reported Device Performance
    Met the requirements of ISO 15197:2003 for in vitro diagnostic test systems - Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus."All the relevant activities were performed by professionals and the results demonstrated that the predetermined acceptance criteria were fully met."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample size used for the test set. It also does not provide details on the country of origin or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    The document does not specify the number of experts used to establish the ground truth or their qualifications. It only states, "All the relevant activities were performed by professionals."

    4. Adjudication Method for the Test Set:

    The document does not describe the adjudication method used for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No mention of an MRMC comparative effectiveness study or any effect size regarding human readers improving with AI vs without AI assistance. This device is a blood glucose monitoring system and does not involve AI assistance for human readers in diagnostic interpretation.

    6. Standalone Performance Study:

    Since this is a blood glucose monitoring system, the performance described relates to the accuracy and precision of the device itself (algorithm only without human-in-the-loop diagnostic interpretation). The document indicates that the device's performance was evaluated against the ISO 15197:2003 standard, which assesses the standalone performance of the system for measuring glucose.

    7. Type of Ground Truth Used:

    The ground truth for evaluating the PRECICHEK Cloudia Blood Glucose Monitoring System would typically be established by a reference laboratory method (e.g., a laboratory glucose analyzer) that is traceable to an accepted standard, as per the ISO 15197:2003 guidelines for blood glucose meters. While not explicitly stated, this is the standard practice for such devices.

    8. Sample Size for the Training Set:

    The document does not specify a training set sample size. For a medical device like a blood glucose meter, the "training set" would likely refer to the data used during the development and calibration phases, rather than a distinct 'training set' in the machine learning sense. This detail is not provided.

    9. How the Ground Truth for the Training Set was Established:

    The document does not provide details on how the ground truth for any potential "training set" was established, beyond the general adherence to ISO 15197:2003 for performance evaluation. Calibration of blood glucose meters typically involves comparing the device's readings against highly accurate laboratory reference methods across a range of glucose concentrations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113307
    Date Cleared
    2012-07-27

    (262 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K032985

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For single use-
    GoodLife AC-300 Blood Glucose Monitoring System
    The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
    The GoodLife AC-300 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-300 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
    The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-300 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-300 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-301 Blood Glucose Monitoring System
    The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
    The GoodLife AC-301 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-301 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
    The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-301 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-301 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-302 Blood Glucose Monitoring System
    The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
    The GoodLife AC-302 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-302 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
    The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-302 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-302 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-303 Blood Glucose Monitoring System
    The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
    The GoodLife AC-303 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-303 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
    The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-303 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-303 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-304 Blood Glucose Monitoring System
    The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
    The GoodLife AC-304 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-304 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
    The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-304 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-304 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-305 Blood Glucose Monitoring System
    The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.
    The GoodLife AC-305 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-305 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.
    The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-305 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-305 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    For multiple use-
    GoodLife AC-300 Professional Blood Glucose Monitoring System
    The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
    The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-300 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-300 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-300 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-301 Professional Blood Glucose Monitoring System
    The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
    The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-301 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-301 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-301 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-302 Professional Blood Glucose Monitoring System
    The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
    The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-302 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-302 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-302 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-303 Professional Blood Glucose Monitoring System
    The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
    The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-303 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-303 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-303 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-304 Professional Blood Glucose Monitoring System
    The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
    The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-304 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-304 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-304 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    GoodLife AC-305 Professional Blood Glucose Monitoring System
    The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.
    The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-305 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.
    The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-305 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.
    The GoodLife Glucose Control Solutions are for use with the GoodLife AC-305 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

    Device Description

    GoodLife AC-300305 Blood Glucose Monitoring System and GoodLife AC-300305 Professional Blood Glucose Monitoring System consist of:
    (1)Glucose Meter
    (2)Glucose Test Strips
    (3) Two levels of glucose control solutions (Level I and Level II) may be purchased separately. Glucose control solutions were previously cleared under K032985.
    (4)Check Strip
    (5)Instruction for use
    These models have the same intended use and technological characteristics except the housing of the meter. Although the different shapes of housing, the materials are the same.
    [Test Principle]
    GoodLife AC-300305 Blood Glucose Monitoring System and GoodLife AC - 300305 Professional Blood Glucose Monitoring System are electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor. When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use.
    [Control Solution]
    The GoodLife Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) to assess the performance of the GoodLife AC-300305 with GoodLife AC Blood Glucose Test strips or AC-300305 Professional Blood Glucose monitoring system with AC Professional Blood Glucose Test strips. There are two levels of controls (Levels 1,2).
    [Check Strip]
    The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover.
    [Device Calibration]
    The device is calibrated by implicit coding process. While inserting the test strip into strip slot to perform the blood glucose test, the coding procedure is complete. The meter will apply formula including this parameter of code to calculate the glucose value.

    AI/ML Overview

    The document describes several GoodLife AC-300~305 blood glucose monitoring systems and their professional counterparts. The acceptance criteria and the study proving the device meets them are outlined, primarily referring to consistency with ISO 15197:2003.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device claims compliance with ISO 15197:2003. This international standard sets the requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. While the document does not explicitly list the detailed clauses of ISO 15197:2003 as acceptance criteria, it states that the device is "compliant to the standard" and that "the results demonstrated that the predetermined acceptance criteria were fully met."

    Therefore, the acceptance criteria are implicitly those defined by ISO 15197:2003. The document does not provide specific numerical performance metrics from the study to directly compare against these criteria.

    Acceptance Criteria (from ISO 15197:2003 - implied)Reported Device Performance
    Accuracy Requirements (e.g., within ±15% or ±20% of a reference method for specified glucose concentrations)"fully met" the predetermined acceptance criteria.
    Precision Requirements (e.g., coefficient of variation)Study conducted according to CLSI/NCCLS EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods
    Linearity (range)Study conducted according to CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement. Tested range: 20~600 mg/dL
    Interference effectsStudy conducted according to CLSI EP07-A2: Interference Testing in Clinical Chemistry
    Method comparisonStudy conducted according to CLSI/NCCLS EP09-A2: Method Comparison and Bias Estimation Using Patient Samples
    Other requirements of ISO 15197:2003"fully met" the predetermined acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states, "All the relevant activities were performed by professionals and the results demonstrated that the predetermined acceptance criteria were fully met." However, it does not specify the sample size for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth. The studies cited (CLSI/NCCLS EP05-A2, EP06-A, EP07-A2, EP09-A2) are guidelines for evaluating the performance of quantitative measurement methods, implying that an established reference method (often laboratory-based) would serve as the ground truth.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method. For quantitative measurements like blood glucose, adjudication among multiple human readers is typically not applicable in the same way it would be for qualitative assessments (e.g., image interpretation). Instead, the device's readings are compared against a reference method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is more relevant for diagnostic imaging or subjective assessments where multiple human readers interpret cases. Blood glucose monitoring systems are quantitative measurement devices not typically evaluated with MRMC studies in this context. Therefore, there is no effect size of human readers improving with AI vs. without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The performance described for the blood glucose monitoring system is inherently a standalone performance. The device provides a quantitative measurement of glucose. A "human-in-the-loop" would involve interpretation or decision-making aided by the device's output, but the device's core function is the measurement itself. The studies performed (e.g., accuracy, precision, linearity) evaluate this standalone measurement capability. Therefore, yes, a standalone performance evaluation was done.

    7. The Type of Ground Truth Used

    Based on the nature of blood glucose monitoring and the CLSI standards referenced, the ground truth would typically be established by a highly accurate laboratory reference method for glucose measurement. The document states a "reference method" is used, which is standard practice in these types of evaluations. The CLSI/NCCLS EP09-A2 standard explicitly deals with "Method Comparison and Bias Estimation Using Patient Samples," which involves comparing the device's results to a more established, accurate method.

    8. The Sample Size for the Training Set

    The document does not specify the sample size used for the training set. Blood glucose meters are calibrated implicitly during the manufacturing process and through quality control rather than through an "AI training set" in the conventional sense of machine learning algorithms. The calibration involves applying a formula that includes a code parameter specific to the test strip.

    9. How the Ground Truth for the Training Set Was Established

    The device is calibrated by an "implicit coding process" that occurs when a test strip is inserted. This suggests a pre-programmed calibration based on manufacturing specifications and quality control data rather than a dynamic ground truth established for a "training set" in the machine learning context. The document does not detail how the ground truth for this initial calibration was established, but it would typically involve rigorous laboratory testing against reference methods during the device and strip development and manufacturing stages.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1