K Number
K131173

Validate with FDA (Live)

Date Cleared
2014-09-18

(511 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FIA Blood Glucose Monitoring System (G2) is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The FIA Blood Glucose Monitoring System (G2) is intended to be used by a single patient and should not be shared.

The FlA Blood Glucose Monitoring System (G2) is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The FIA Blood Glucose Monitoring System (G2) should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The FIA Blood Glucose Test Strips (G2) are for use with the FIA Blood Glucose Meter (G2) to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The FIA Glucose Control Solutions are for use with the FIA Blood Glucose Meter (G2) and FIA Blood Glucose Test Strips (G2) to check that the meter and test strips are working together properly and that the test is performing correctly.

Device Description

FIA Blood Glucose Monitoring System (G2) consists of:
(1) Glucose Meter
(2)Glucose Test Strips
(3)Two levels of glucose control solutions (L1 and L2) may be purchased separately. Glucose control solutions were previously cleared under K032985.
(4)Check Strip
(5)Instruction for use

FIA Blood Glucose Monitoring System (G2) is an electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor. When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use.

The FIA Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) assessing the performance of the FIA Blood Glucose monitoring system (G2) and FIA Blood Glucose Test strips (G2). There are two levels of controls (L1 and L2).

The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover.

The device is calibrated by implicit coding process. While inserting the test strip into strip slot to perform the blood glucose test, the coding procedure is complete. The meter will apply formula including this parameter of code to calculate the glucose value.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Acceptance Criteria and Reported Device Performance:

The document states that the FIA Blood Glucose Monitoring System (G2) is compliant with the standard ISO 15197:2003 In vitro diagnostic test systems-Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. However, the exact numerical acceptance criteria from this standard and the specific reported device performance against those criteria are not explicitly detailed in the provided text.

The document only makes a general statement: "All the relevant activities were performed by professionals and the results demonstrated that the predetermined acceptance criteria were fully met."

To properly answer this section, one would typically expect a table like this:

Performance Metric (from ISO 15197:2003)Acceptance Criteria (from ISO 15197:2003)Reported Device Performance (FIA BGMS G2)
For glucose concentration < 75 mg/dLx% of results within ± y mg/dL of referenceData not provided
For glucose concentration ≥ 75 mg/dLx% of results within ± y% of referenceData not provided
MARD (Mean Absolute Relative Difference)Specific MARD value requirementData not provided
etc.etc.Data not provided

Without the actual numerical performance data and specific ISO 15197:2003 criteria, a detailed table cannot be populated.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for the Test Set: Not explicitly stated in the provided text. The document mentions "clinical tests" were performed, but does not specify the number of blood samples or participants involved.
  • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It only states that the clinical tests were performed by "professionals."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • The document does not provide any information about the number of experts used or their qualifications to establish ground truth. For blood glucose monitoring systems, ground truth is typically established by comparing results to a laboratory reference method (e.g., YSI analyzer), rather than expert consensus on images or interpretations.

4. Adjudication Method for the Test Set:

  • The document does not mention any adjudication method for the test set. Given that this is a blood glucose monitoring system, the "ground truth" would usually be established by a laboratory reference method, implying a direct comparison rather than a separate adjudication process by experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images. The FIA Blood Glucose Monitoring System (G2) is an in vitro diagnostic device for self-testing blood glucose, where the output is a numerical value directly from the device, not an interpretation by a human reader. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.

6. Standalone (i.e., algorithm only without human-in-the-loop performance) Performance:

  • Yes, in spirit, the primary performance evaluation described is standalone. The device itself (meter and test strip) is designed to provide a direct quantitative measurement of glucose. The "clinical tests" performed against ISO 15197:2003 evaluate the accuracy of the device's measurements relative to a reference method, which is essentially its standalone performance. There isn't a "human-in-the-loop" component in the sense of a human interpreting the device's output and then having that interpretation evaluated.

7. Type of Ground Truth Used:

  • The document implies that the ground truth for the clinical tests was based on a reference method to which the device's measurements were compared, aligning with the requirements of ISO 15197:2003. For blood glucose monitors, this is typically a laboratory reference method (such as a YSI glucose analyzer) that provides highly accurate glucose concentration values. The text doesn't explicitly name the reference method, but states compliance with a standard designed for such comparisons.

8. Sample Size for the Training Set:

  • The document does not provide any information about a training set size. For a blood glucose monitoring system, the "training set" would refer to data used for the initial development and calibration of the device's algorithm, which is typically an internal process and not usually detailed in a 510(k) summary in this manner. The focus here is on the verification and validation of the final product.

9. How the Ground Truth for the Training Set Was Established:

  • The document does not provide any information on how the ground truth for a training set (if one was formally defined for algorithm development) was established. As mentioned above, for this type of device, it would generally involve comparison to a laboratory reference method.

{0}------------------------------------------------

510(k) Summary

    1. Submitter Information
Company name:HMD BioMedical Inc.
Contact person:Axel Lin
Address:No. 181, Minsheng St., Xinpu Township, HsinchuCounty, Taiwan
Phone:886-3-6017000
FAX:886-3-5885500
E-mail:axellin@hmdbio.com
Date Prepared:Sep 17, 2014
    1. Name of Device
      Trade/Proprietary Name:

FIA Blood Glucose Monitoring System (Model no.: G2)

Common name:Blood Glucose Test System
Classification name:Glucose Test System
Classification Panel:Clinical Chemistry (75)
Regulation no.:862.1345 (Class II), 862.1660 (Class I)
Product code:NBW, LFR, JJX, JQP
    1. Predicate Device
Trade/Proprietary name:PRECICHEK Cloudia Blood GlucoseMonitoring System
Common name:Blood Glucose Test System
Submitter:HMD BioMedical Inc.
510(k) no.:K120064
Product code:NBW. LFR

{1}------------------------------------------------

HMD BioMedical Inc.

Common name:Glucose Control Solution
Submitter:HMD BioMedical, Inc.
510(k) no.:K032985
Product code:JJX
    1. Device Description
      FIA Blood Glucose Monitoring System (G2) consists of:
  • (1) Glucose Meter

  • (2)Glucose Test Strips

  • (3)Two levels of glucose control solutions (L1 and L2) may be purchased separately. Glucose control solutions were previously cleared under K032985.

  • (4)Check Strip

  • (5)Instruction for use

[Test Principle]

FIA Blood Glucose Monitoring System (G2) is an electrochemical biosensor system that measures the amount of electric current produced then displays the result as a blood glucose level on the LCD monitor. When the blood is drawn into the blood reaction zone of the test strip, the glucose in the blood sample mixes with a special chemical in the test strip, which produces a small electric current. The reaction current is proportional to the amount of glucose in the blood. The result is displayed on the LCD monitor and automatically stored in the meter for future use.

[Control Solution]

The FIA Glucose control solution is intended for in vitro diagnostic use (i.e. for external use only) assessing the performance of the FIA Blood Glucose monitoring system (G2) and FIA Blood Glucose Test strips (G2). There are two levels of controls (L1 and L2).

{2}------------------------------------------------

[Check Strip]

The Check Strip can be used to check that the meter is operating properly. It is composed of PCB, resistor, top cover and bottom cover.

[Device Calibration]

The device is calibrated by implicit coding process. While inserting the test strip into strip slot to perform the blood glucose test, the coding procedure is complete. The meter will apply formula including this parameter of code to calculate the glucose value.

5. Intended Use

The FIA Blood Glucose Monitoring System (G2) is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The FIA Blood Glucose Monitoring System (G2) is intended to be used by a single patient and should not be shared.

The FIA Blood Glucose Monitoring System (G2) is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The FIA Blood Glucose Monitoring System (G2) should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The FIA Blood Glucose Test Strips (G2) are for use with the FIA Blood Glucose Meter (G2) to quantitatively measure qlucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The FIA Glucose Control Solutions are for use with the FIA Blood Glucose Meter (G2) and FIA Blood Glucose Test Strips (G2) to check that the meter and test strips are working together properly and that the test is performing correctly.

    1. Comparison of Subject Devices and predicate device
      Technological Characteristics Comparison Table of FIA Blood Glucose Monitoring System (G2) and PRECICHEK Cloudia Blood Glucose Monitoring System (K120064)

{3}------------------------------------------------

HMD BioMedical Inc.

Similarities
ItemSubject deviceFIA BGMS (G2)Predicate devicePRECICHEK Cloudia BGMS (K120064)
Intended useIt is designed to quantitatively measure the concentration of glucose in fresh capillary whole blood.Same
Detection methodAmperometry: Current produced by chemical reactionSame
Test range20~600mg/dLSame
Operating conditions50104°F (1040°C), 20~80%Same
AutocodingYesSame
EnzymeGlucose Dehydrogenase (FAD) (Aspergillus oryzae)Same
Capillary testing siteFingertips onlySame
Sample volume0.5ulSame
Memory999Same
Test time5 secSame
Time & date settingReload the batteries then press the buttons to set the time and date.Same

{4}------------------------------------------------

HMD BioMedical Inc.

Differences
ItemSubject deviceFIA BGMS (G2)Predicate devicePRECICHEK Cloudia BGMS(K120064)
DatatransmissionGSMN/A
AverageN/A7, 14, 21, 28 days

7. Discussion of Clinical Tests Performed

FIA Blood Glucose Monitoring System (Subject Device) is compliant to the standard of ISO 15197:2003 In vitro diagnostic test systems-Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. All the relevant activities were performed by professionals and the results demonstrated that the predetermined acceptance criteria were fully met.

8. Conclusion

The subject device was tested and fulfilled the requirements from those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered horizontally and vertically in the image. The background is white.

Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design resembling a bird or a human figure with outstretched arms. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the design.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 2, 2014

HMD BIOMEDICAL, INC. AXEL LIN MANAGER OF QUALITY ASSURANCE #181 MINSHENG STREET, XINPU TOWNSHIP, HSINCHU COUNTY, TW 305 TAIWAN

Re: K131173 Trade/Device Name: FIA Blood Glucose Monitoring System (G2) Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, LFR, JJX Dated: September 5, 2014 Received: September 8, 2014

Dear Ms. Lin:

This letter corrects our substantially equivalent letter of September 25, 2014. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2 - Ms. Lin

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known) K131173

Device Name FIA Blood Glucose Monitoring System (G2)

Indications for Use (Describe)

The FIA Blood Glucose Monitoring System (G2) is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The FIA Blood Glucose Monitoring System (G2) is intended to be used by a single patient and should not be shared.

The FlA Blood Glucose Monitoring System (G2) is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The FIA Blood Glucose Monitoring System (G2) should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The FIA Blood Glucose Test Strips (G2) are for use with the FIA Blood Glucose Meter (G2) to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The FIA Glucose Control Solutions are for use with the FIA Blood Glucose Meter (G2) and FIA Blood Glucose Test Strips (G2) to check that the meter and test strips are working together properly and that the test is performing correctly.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.