Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K143497
    Date Cleared
    2015-11-04

    (330 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K011103, K023752

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Safety Auto-Disable Syringe with Needle (Auto-Lock) is used for aspiration of fluids from vials and ampoules and a variety of fluid injections below the surface of the skin except phlebotomy.

    It has a detachable needle with a dedicated fitting. The Safety Auto-Disable Syringe with Needle (Auto-Lock) contains an inner mechanism used to allow the needle to be retracted inside the plunger rod of the syringe after the operator's thumb force released. After activation the needle is fully contained inside the syringe guarding against accidental needle sticks during normal handling and disposal of the used needle/syringe combination.

    Device Description

    The proposed device is a single use, sterile syringe composed of barrel, plunger, plunger, push-button, spring, spring holder, rubber stopper, needle cap, and autodisable part.

    AI/ML Overview

    The document provided is a 510(k) premarket notification for a "Safety Auto-Disable Syringe with Needle (Auto-Lock)". It describes the device, its intended use, and a comparison to predicate devices, along with summaries of performance testing.

    Here's an analysis of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" against "reported device performance" for each individual test in the format usually expected for a medical device. Instead, it lists various bench tests and indicates that the device conforms to relevant ISO standards. For the simulated clinical study, a specific performance metric related to the failure rate is given.

    Here's a table based on the information provided, focusing on the simulated clinical study's outcome as the most direct "performance" metric stated:

    Acceptance Criteria (Implied)Reported Device Performance
    Failure rate no higher than 0.7% (with 97.5% confidence) OR failure rate no higher than 1.1% (with 99.5% confidence)No failures observed in 512 device trial.

    For other tests, the document generally states "Conforms to [ISO Standard]" or "Comply to [ISO Standard]" as the performance, implying that the standards themselves define the acceptance criteria. Examples include:

    Test NameAcceptance Criteria (Implied from standard)Reported Device Performance
    Sterile testConforms to sterility requirementsMeets standard
    Bacterial Endotoxin testConforms to endotoxin limitsMeets standard
    Residual of EO testConforms to EO residue limitsMeets standard
    Needle point piecing strength testConforms to ISO 7864Meets standard
    Piston in barrel fitness testConforms to ISO 7886-1Meets standard
    Tolerance on graduated capacity (delivery accuracy) testComply to ISO 7886-1Meets standard
    Hub/needle bond strengthConform to ISO 7864Conform to ISO 7864
    Reuse durabilityCan't be re-used according to ISO 7886-4Can't re-used according to ISO 7886-4

    2. Sample size used for the test set and the data provenance

    • Sample Size for Simulated Clinical Study: 512 devices.
    • Data Provenance: The simulated clinical study and a "clinical investigation" were "performed in China per the requirements of China FDA." This indicates the data is prospective, acquired from China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of "experts" to establish ground truth in the context of the simulated clinical study or other performance tests. The tests performed are primarily bench tests and a simulated clinical study, which would rely on objective measurements and established protocols rather than expert consensus on a "ground truth" (e.g., as in image interpretation).

    4. Adjudication method for the test set

    Not applicable. The tests described are objective performance and safety evaluations, not subjective assessments requiring adjudication by multiple readers or experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (syringe), not an AI-powered diagnostic or assistive tool where MRMC studies would be relevant to evaluate AI assistance for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical syringe, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically used in AI/diagnostic studies (e.g., pathology, clinical outcomes) is not directly applicable to the performance testing of this physical medical device. The "ground truth" for the tests mentioned (e.g., sterility, fluid delivery accuracy, needle integrity) would be defined by the specifications and measurable parameters within the referenced ISO standards. For example, for "Sterile test," the ground truth is "sterile" as determined by microbiology methods conforming to standards. For "hub/needle bond strength," the ground truth is a specific force threshold defined in ISO 7864.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1