Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K012216
    Date Cleared
    2001-08-17

    (32 days)

    Product Code
    Regulation Number
    876.5010
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K003929

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NIROYAL® Biliary Premounted Stent System is indicated for use in the treatment of biliary strictures produced by malignant neoplasms.

    Device Description

    The NIROYAL® Biliary Premounted Stent System (NIROYAL® Biliary System) is a stent system consisting of a 0.018" balloon delivery catheter and a gold plated stainless steel stent. The modified NIROYAL® Biliary System is the same as the NIROYAL® Biliary System cleared under K003929, with the exception of the delivery balloon size for the 7 mm x 14 mm stent. The 7 mm x 14 mm stent will be mounted on a 20 mm balloon, instead of a 15 mm balloon as described in K003929.

    AI/ML Overview

    Acceptance Criteria and Device Performance for NIROYAL® Biliary Premounted Stent System

    This document outlines the acceptance criteria and the summary of the study performed to demonstrate that the NIROYAL® Biliary Premounted Stent System (with a modified balloon size) meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The following table details the functional testing performed on the modified NIROYAL® Biliary System and the corresponding acceptance criteria. The document states that "Functional testing was conducted to verify the integrity of the modified balloon length of 20 mm for the 7 mm x 14 mm NIROYAL® Biliary System." and implies that the device met these criteria, as substantial equivalence was determined.

    TestAcceptance CriteriaReported Device Performance
    Tracking Force1.5 lbf maximumMet Acceptance Criteria
    Stent Deployment Pressure103 psi maximumMet Acceptance Criteria
    Balloon Burst Pressure Within the Stent176 psi minimumMet Acceptance Criteria
    Stent Profile0.075" maximumMet Acceptance Criteria
    Stent Expansion Uniformity at Nominal and Rated Pressure5% maximumMet Acceptance Criteria
    Stent Foreshortening at Nominal and Rated Pressure20% maximumMet Acceptance Criteria
    Balloon Burst Pressure Out of the Stent176 psi minimumMet Acceptance Criteria
    Stent Securement0.30 lbf minimumMet Acceptance Criteria

    2. Sample Size and Data Provenance

    The document does not specify the exact sample size used for each functional test. The study appears to be a non-clinical bench test and therefore does not involve human subjects or data from a specific country of origin in the context of clinical trials. The data provenance is internal to Boston Scientific Scimed, Inc. and is generated during product development and modification. This is a retrospective analysis of the modified features against pre-defined engineering specifications.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. This study involves functional, non-clinical bench testing of a medical device, not a diagnostic or AI-driven system that requires expert consensus for ground truth.

    4. Adjudication Method

    Not applicable. Adjudication methods are typically used in clinical studies or for establishing ground truth in diagnostic accuracy studies. This document describes non-clinical functional testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a submission for a modification to an existing medical device, involving non-clinical functional testing, not an MRMC comparative effectiveness study of a diagnostic AI system.

    6. Standalone Performance Study

    Yes, a standalone performance study was conducted. The "Non-Clinical Test Summary" section describes various functional tests performed on the device itself, independent of human interaction beyond the execution of the tests. This is a standalone (algorithm/device only) performance evaluation of the physical aspects of the stent system.

    7. Type of Ground Truth Used

    The "ground truth" for this study is based on engineering specifications and pre-defined performance limits established for the device's functional integrity and safety. These limits (e.g., maximum force, minimum pressure, maximum profile) serve as the objective standards against which the device's performance is measured.

    8. Sample Size for the Training Set

    Not applicable. This is a medical device modification submission based on non-clinical functional testing, not an AI/ML model that requires a training set.

    9. How Ground Truth for the Training Set Was Established

    Not applicable. As noted above, this is not an AI/ML model; therefore, there is no training set or ground truth established for a training set in this context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1