Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    DEN150040
    Manufacturer
    Date Cleared
    2016-06-03

    (280 days)

    Product Code
    Regulation Number
    876.5011
    Type
    Direct
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K083627, K061231, K112543, K000281

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WallFlex Biliary RX Fully Covered Stent System RMV is indicated for indwell up to 12 months in the treatment of benign biliary strictures secondary to chronic pancreatitis.

    Device Description

    The WallFlex Biliary RX Fully Covered Stent System RMV consists of 2 components: the implantable fully covered self-expanding metallic biliary stent (FC-SEMBS) and the Rapid Exchange (RX) delivery device. The FC-SEMBS is made out of Nitinol monofilament wire with a radiopaque platinum core, braided in a tubular mesh configuration. The stent has proximal and distal flares at each end to aid in preventing migration. The proximal flare also contains the biliary stent retrieval loop. The biliary stent is covered with a Permalume™ silicone covering along its entire length except for 2 mm on the retrieval loop end. The RX delivery device is a coaxial tube design. The exterior tube is used to constrain the biliary stent before deployment and re-constrain the biliary stent, if biliary stent repositioning is necessary, after partial deployment. The exterior tube has a clear section so that the constrained stent is visible. A yellow transition zone on the inner tube of the delivery system is visible between the stent and the blue outer sheath. There are four radiopaque (RO) markers to aid in the deployment of the stent while using fluoroscopy. There are two RO markers on the inner tube of the delivery system identifying the ends of the constrained biliary stent. Between these RO markers is an additional RO marker that indicates at what point re-constrainment is no longer possible. The fourth RO marker at the leading end of the exterior tube indicates how far the biliary stent has been deployed. There is one visual marker on the interior tube between the handles to aid in the deployment of the biliary stent. The visual marker indicates the point at which re-constrainment is no longer possible. The interior tube has a single central lumen to accommodate a 0.035 inch (0.86 mm) guidewire.

    AI/ML Overview

    The provided text describes the WallFlex Biliary RX Fully Covered Stent System RMV, its design, nonclinical/bench studies, and a clinical study conducted to assess its safety and effectiveness.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria & Reported Device Performance

    The acceptance criteria are primarily derived from the "Special Controls" section and the "Summary of Nonclinical Studies" (Table 2). The reported device performance is extracted from the "Summary of Nonclinical/Bench Studies" section and the "Results" section of the clinical study.

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    SterilizationThe sterility assurance level (SAL) shall be 10-6.Passed: Sterilization was re-assessed in 2015 to ensure compliance to most recent version of ANSI/AAMI/ISO 10993-7. Testing was conducted on the 10x120mm fully covered stent, considered worst-case.
    Ethylene Oxide ResidualsEtOH residual amounts are below the maximum amount allowed: For the stent, the average daily dose of EO to patient shall not exceed 0.1 mg/day, max EO dose of 20 mg in first 24 hr, 60 mg in first 30 days, 2.5 g in a lifetime. For the delivery catheter, the average daily dose shall not exceed 4 mg.Passed: EtOH residual amounts are below the maximum amount allowed.
    BiocompatibilityCytotoxicity: Grade 0 (non-cytotoxic); Irritation: Non-irritant (based on histopathologic evaluations); Sensitization: No signs of sensitization from polar and non-polar extracts; Acute Systemic Toxicity: No evidence of mortality or systemic toxicity; Genotoxicity: Non-mutagenic; Implantation: Non-irritant (based on histopathologic evaluations); Subchronic Toxicity: No signs of systemic toxicity due to leachable components; Nickel leaching: Acceptable resistance to corrosion and acceptable nickel leach rates; Toxicological Risk: Not likely to lead to systemic toxicity.Passed: Cytotoxicity Grade 0. Irritation non-irritant. No signs of sensitization. No evidence of mortality or systemic toxicity. Non-mutagenic. Non-irritant after implantation. No signs of systemic toxicity. Acceptable nickel leaching demonstrated. Exposure to the stent is not likely to lead to systemic toxicity.
    Deployment TestingThe delivery catheter must safely and accurately deliver the stent to the intended anatomic location. No damage to the stent or simulated stricture model should occur.Passed: Stent sizes tested represent the smallest and largest sizes available and can be used as surrogates. The stent was successfully placed in 100% (127/127) of patients in the clinical study.
    Expansion/Compression Force TestingRadial expansion forces must be at least 85% of specified N values for various stent diameters and compressions. Radial compression forces must be at least 85% of specified N values for various stent diameters and compressions.Passed: The device met the specified radial expansion and compression forces.
    Dimensional TestingAll dimensions should be within design tolerance ranges.Passed: Dimensions of each stent size and delivery catheter were measured and verified to meet the acceptance criteria and labeled dimensions.
    Tensile Strength Testing (Bond Integrity)All bonded components of the delivery catheter should withstand forces that exceed those encountered during clinical use.Passed: Bonded joints of delivery catheters were tested to failure and withstood forces exceeding those encountered clinically.
    Stent IntegrityNo crevice or pitting corrosion; No more than 2 weld breaks; No significant weight loss; Meet specification for radial compression and expansion forces; Wire must meet specification for tensile properties; Stent cover integrity must meet requirements for holes and delamination.Passed: Stents tested represent worst-case scenario. No degradation observed.
    MRI CompatibilityMRI compatibility labeling must be supported by testing. Displacement forces should not damage tissue, and localized temperature increases should not damage tissues when patients are scanned as outlined in the labeling.Evaluated: Mean deflection angle of 3°. Worst case projected temperature rise of 5.5°C for whole body average SAR of 4 W/kg. Artifacts appeared as localized signal voids extending ~10 mm from stent wall perimeter and ~2mm beyond each end. This data supports the MRI Conditional labeling.
    Shelf Life (Package Integrity)Packaging must not be compromised below a minimum pressure threshold. No channels, pinholes, or other barrier breaches allowed. All samples must not have channels visible at 10X magnification.Passed: Package integrity testing demonstrated maintenance of sterility.
    Shelf Life (Functional Performance)Deployment Testing: Delivery catheter should safely and reliably deliver the stent to the intended location without damage to the stent or patient after 25 months real time aging. Expansion/Compression Force Testing: Forces must not be impacted by aging. Flexural Rigidity: Stents must not become rigid after aging. Covering Material Integrity: Maximum of 6 diamonds containing compromised coating integrity. Weld Integrity: No more than two broken wire weld joints observed. Tensile Strength: Bonded components withstand forces after aging.Passed: Deployment testing passed with largest stent size after aging. Expansion/compression forces passed after aging. Flexural rigidity passed after aging. Covering material integrity passed (maximum 6 diamonds with compromised coating). Weld integrity passed (no more than 2 broken welds). Tensile strength testing passed after aging.
    Clinical Performance (Removal)The ability to safely place and subsequently remove the stent after the maximum labeled indwell period.Stent Removability: 72.4% (92/127) ITT, 78.0% (92/118) PP. Removal Success: 84.3% (107/127) ITT, 90.7% (107/118) PP. Stent indwell up to 613 days (~20 months).
    Clinical Performance (Functionality)The stent resolves strictures during the maximum labeled indwell period.Stent Functionality During Stent Indwell: 77.2% (98/127) ITT, 83.1% (98/118) PP (lack of required reintervention). Stricture Resolution: 74.0% (94/127) ITT, 79.7% (94/118) PP (lack of stricture-related re-intervention). Biliary obstructive symptoms reduced from 51.2% (baseline) to 6.3% (Indwell month 6). Bilirubin and alkaline phosphatase levels decreased.
    Clinical Performance (Recurrence)Stricture resolution is maintained post-stent removal.No Stricture Recurrence: 85.1% (80/94) ITT/PP patients who initially achieved stricture resolution did not experience recurrence over median 19.0 months follow-up.
    Clinical Performance (Adverse Events)All adverse event data including bile duct obstruction and trauma to the bile duct should be reported.SAEs (patients with at least one SAE): 36.2% (46/127) ITT, 36.4% (43/118) PP. Common SAEs: Pancreatitis (16%), Cholangitis (9%). No stent or stent removal related deaths.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Study (Test Set): 127 patients for the Intent-to-Treat (ITT) cohort, and 118 patients for the Per-Protocol (PP) cohort.
    • Data Provenance:
      • Country of Origin: The clinical study was a "large prospective multinational study utilizing 13 centers in 11 countries outside of the US (OUS)." Specific countries are not listed, but it indicates a broad international representation.
      • Retrospective/Prospective: The clinical study was explicitly prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    The document does not specify the number or qualifications of experts establishing ground truth for the clinical study directly. The "ground truth" for clinical outcomes (e.g., stricture resolution, stent removability, adverse events) would have been established by the clinical investigators at the 13 centers based on patient follow-up, medical records, imaging, and endoscopic findings. These would be qualified medical professionals (e.g., gastroenterologists, interventional radiologists) involved in the care of the patients and the conduct of the trial. However, no specific details on their number or specific qualifications (e.g., "radiologist with 10 years of experience") are provided within this document.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1) for establishing clinical outcomes. Clinical outcomes would likely have been determined based on predefined criteria in the study protocol and assessed by the attending clinicians at each site. Serious adverse events (SAEs) were reported and summarized, indicating a system for tracking and categorizing these events, but not explicitly an independent adjudication panel.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study involving human readers improving with AI vs. without AI assistance was not done. This device is a medical stent, not an AI diagnostic tool, so such a study would not be applicable. The clinical study assessed the device's performance directly in patients.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone (algorithm only) study was not done. This device is a physical medical device (stent system) that requires human intervention (endoscopic placement and removal). Its performance is inherently linked to human use in a clinical setting.

    7. Type of Ground Truth Used

    The ground truth for the clinical study was based on clinical outcomes data from patients, including:

    • Endoscopic findings (stent placement, removability, removal success)
    • Patient symptoms (biliary obstructive symptoms)
    • Laboratory test results (bilirubin, alkaline phosphatase levels)
    • Medical records (re-interventions, recurrence, adverse events, surgical outcomes)

    For the non-clinical/bench studies, the ground truth was established through engineering and laboratory test standards, specifications, and physical measurements (e.g., ISO, ASTM standards, pre-defined tolerance ranges, visual inspections, mechanical testing).

    8. Sample Size for the Training Set

    The document describes a single prospective clinical study acting as the primary evidence for the device's clinical performance. It does not mention a separate "training set" in the context of machine learning or AI models. This is a traditional medical device submission, not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" in the context of an AI/ML model for this traditional medical device, this question is not applicable. The clinical study data was used to demonstrate the device's safety and effectiveness directly, not to train an algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061877
    Date Cleared
    2006-09-15

    (74 days)

    Product Code
    Regulation Number
    878.3610
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K042065, K060323, K000281

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wallflex™ Enteral Colonic Stent with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstruction prior to colectomy in patients with malignant strictures.

    Device Description

    The proposed Wallflex Enteral Stent with Anchor Lock Delivery System consists of 2 components: an implantable metal stent and a delivery system. The proposed stent is manufactured of Nitinol and offered in two diameters, a 25mm body with a 30mm flare, and a 22mm body with a 27mm flare. Each diameter is offered in three lengths, 6cm, 9cm, and 12cm. The proposed Anchor Lock delivery system consists of a coaxial tubing assembly that constrains the stent on the delivery catheter shaft until the stent is released by retracting the exterior tube.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Wallflex™ Enteral Colonic Stent with Anchor Lock Delivery System. It states that this is a request for an expanded indication for a device that already has existing market clearance and introduces "no new materials or design changes."

    Therefore, the performance data presented in the original K042065 submission was not repeated for this K061877 submission. This means that the current document does not contain the acceptance criteria or the study that proves the device meets those criteria for the expanded indication. Instead, it relies on the predicate device's prior approval.

    As such, I cannot fill in the requested table and information based solely on the provided text. The document explicitly states:

    "As this is a request for an expanded indication and introduces no new materials or design changes the performance testing presented in K042065 was not repeated."

    To answer the prompt fully, one would need to refer to the original K042065 submission, which is not provided here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060323
    Date Cleared
    2006-05-12

    (92 days)

    Product Code
    Regulation Number
    878.3610
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K030769, K000281

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultraflex™ Precision™ Colonic Stent is indicated for palliative treatment of colonic strictures in the descending colon, sigmoid colon and rectum caused by malignant neoplasms and to relieve large bowel obstruction prior to colectomy in patients with malignant strictures.

    Device Description

    The proposed Ultraflex™ Precision™ Colonic Stent System consists of a self-expanding metal stent and a delivery catheter. The proposed stent consists of Nitinol wires wound together to form a cylinder with a flared distal section. The proposed stent is mounted on a delivery catheter. The stent is compressed onto the tapered delivery catheter shaft and bound with braided nylon suture using a crochet stitch. The stent is deployed by holding the handle stationary and pulling a ring to unravel the crochet stitch.

    AI/ML Overview

    This document describes a 510(k) submission for the Ultraflex™ Precision™ Colonic Stent System. It is an application for an expanded indication for a device that is already on the market. Therefore, the information provided focuses on the substantial equivalence to predicate devices and clinical data for the new indication, rather than detailed performance studies for the device itself.

    Here's an analysis based on the provided text, addressing the requested information points where possible:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific quantitative acceptance criteria or reported device performance metrics in a tabular format. The submission relies on demonstrating substantial equivalence to existing predicate devices for its technological characteristics and presenting clinical data to support the new indication.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not specified in the provided text. The document mentions "Clinical data in support of the proposed indication has been collected and is presented within this submission." However, the exact number of patients or cases is not detailed here.
    • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • This information is not provided in the given text.

    4. Adjudication Method for the Test Set

    • This information is not provided in the given text.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study is not mentioned or implied. The submission focuses on device performance and clinical effectiveness for a palliative indication, not on the improvement of human reader performance with AI assistance.

    6. If a Standalone (Algorithm Only) Performance Study Was Done

    • No, a standalone (algorithm only) performance study was not done. This device is a physical medical device (stent system), not an AI algorithm. The performance data refers to the device's functional characteristics and clinical outcomes.

    7. Type of Ground Truth Used

    • Given the context of a colonic stent for malignant strictures, the "ground truth" for the clinical data would likely be based on clinical observation, patient outcomes, medical diagnoses (e.g., confirmed malignant neoplasms), and potentially imaging or endoscopic confirmation of stricture palliation. The text itself does not explicitly define the "type of ground truth," but these are standard for such medical devices.

    8. Sample Size for the Training Set

    • This information is not applicable as this is a physical medical device, not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable as this is a physical medical device, not an AI algorithm requiring a training set.

    Summary based on the document:

    The submission for the Ultraflex™ Precision™ Colonic Stent System asserts substantial equivalence based on:

    • Technological Characteristics: The proposed device has identical materials, construction, and manufacturing processes as the currently marketed Ultraflex Precision Colonic Stent (K030769).
    • Performance Data: For the new indication, "Clinical data in support of the proposed indication has been collected and is presented within this submission." The document explicitly states that performance testing from the prior K030769 submission was not repeated because there are no new materials or design changes.

    The focus is on demonstrating that the new indication is safe and effective and that the device is substantially equivalent to legally marketed predicate devices (K030769 and K000281). Detailed breakdowns of clinical study methodology (sample sizes, expert qualifications, adjudication, specific metrics) are not provided in this summary document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1