(107 days)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of X-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment, and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Siemens intends to market a new software version, SOMARIS/10 syngo CT VB20 for the following SOMATOM Computed Tomography (CT) Scanner Systems:
a) Single Source CT Scanner systems (SOMATOM go. Platform):
- SOMATOM go.Now
- SOMATOM go.Up
- SOMATOM go.All
- SOMATOM go.Top
- SOMATOM go.Sim
- SOMATOM go.Open Pro
In this submission, the above listed CT scanner systems are jointly referred to as subject devices by "SOMATOM go. Platform" CT scanner systems.
b) Dual Source CT Scanner system:
- SOMATOM Pro.Pulse
The above listed subject devices with SOMARIS/10 syngo CT VB20 are Computed Tomography X-ray Systems which feature one (Single Source) or two (Dual Source) continuously rotating tube-detector system and function according to the fan beam principle. The SOMATOM go. Platform and the SOMATOM Pro.Pulse with software SOMARIS/10 syngo CT VB20 produce CT images in DICOM format, which can be used by trained staff for software applications, e.g. post-processing applications, commercially distributed by Siemens Healthcare and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.
The provided FDA 510(k) Clearance Letter for the SOMATOM CT Systems focuses heavily on establishing substantial equivalence to predicate devices through comparisons of technological characteristics, hardware, and software. It generally asserts that the device has met performance criteria through verification and validation testing, but it does not provide a detailed "Acceptance Criteria Table" with specific quantitative metrics and reported device performance. Similarly, it describes the types of studies performed (e.g., bench testing, retrospective blinded rater study), but it lacks the specific details requested regarding sample sizes, data provenance, expert qualifications, and effect sizes that would typically be found in a detailed study report.
Therefore, I will extract and synthesize the information that is available in the document and explicitly state where the requested information is not provided.
Understanding the Device and its Changes
The devices under review are Siemens SOMATOM CT Systems (SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, and SOMATOM Pro.Pulse) with a new software version, SOMARIS/10 syngo CT VB20. This new software version builds upon the previous VB10 version cleared in K233650 and K232206.
The submission focuses on modifications and new features introduced with VB20, including:
- Eco Power Mode: New feature for reduced energy consumption during idle times (not supported on go.Now and go.Up).
- Oncology Exchange: New feature for transferring prescription information from ARIA Oncology Information System.
- myExam Contrast: New feature for exchanging contrast injection parameters.
- FAST 3D Camera/FAST Integrated Workflow: Modifications including retrained algorithms, collision indication, and Centerline/Grid Overlay.
- FAST Planning: Extended to detect additional body regions.
- myExam Companion (myExam Compass/myExam Cockpit): Clinical decision trees now available for child protocols.
- HD FoV 5.0: New extended field of view reconstruction algorithm (for go.Sim and go.Open Pro only).
- CT guided intervention – myAblation Guide interface: New interface.
- Flex 4D Spiral: Modifications regarding dynamic tube current modulation.
- ZeeFree RT: New stack artifact reduced reconstruction for respiratory-related examinations (for go.Open Pro only).
- DirectDensity: Modified to include stopping-power ratio (Kernel St).
- DirectLaser: Patient Marking workflow improvement.
- Respiratory Motion management - Open Online Interface: New interface for respiratory gating.
- DirectSetup Notes: Enabled for certain SOMATOM go. Platform systems.
The core argument for clearance is substantial equivalence to predicate devices. This means that, despite modifications, the device is as safe and effective as a legally marketed device (the predicates).
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not contain a specific table of quantitative acceptance criteria with corresponding reported device performance values. Instead, it describes general acceptance criteria related to verification and validation tests and then provides qualitative statements about the test results demonstrating comparability or improvement over predicate devices.
Here's a summary of the described performance evaluations:
| Feature/Metric | Acceptance Criteria (Qualitative) | Reported Device Performance (Qualitative) |
|---|---|---|
| Overall | Meet acceptance criteria for all software specifications. Enable safe and effective integration. Perform as intended in specified use conditions. | "All software specifications have met the acceptance criteria." "Verification and validation support the claims of substantial equivalence." "Perform(s) as intended in the specified use conditions." "As safe, as effective, and perform as well as or better than the predicate devices." |
| FAST 3D Camera Accuracy (Isocentering, Range, Direction) | Comparable or better accuracy to predicate device for adults; extend support to adolescents. | "Overall, the subject devices with syngo CT VB20 delivers comparable or improved accuracy to the predicate devices with syngo CT VB10 predicate device for adults and extends the support to adolescents." |
| FAST Planning Correctness | High fraction (percentage) of ranges calculated correctly and without needing change. Meets interactive requirements (fast calculation time). | "For more than 90% of the ranges no editing action was necessary to cover standard ranges." "For more than 95%, the speed of the algorithm was sufficient." |
| HD FoV 5.0 Performance (vs. HD FoV 4.0) | As safe and effective as HD FoV 4.0. | "Results obtained with the new HD FoV 5.0 algorithm are compared with its predecessor, the HD FoV 4.0 algorithm, based on physical and anthropomorphic phantoms...This comparison is conducted to demonstrate that the HD FoV 5.0 algorithm is as safe and effective as the HD FoV 4.0 algorithm." (No quantitative metrics provided in this document excerpt regarding this comparison's outcome). |
| Flex 4D Spiral Functionality & Image Quality | Proper function and acceptable image quality. | "The performed bench test report describes the technical background of Flex 4D Spiral and its functionalities with SOMATOM CT scanners, demonstrate the proper function of those, and assess the image quality of Flex 4D Spiral." (No quantitative metrics provided) |
| ZeeFree RT Reconstruction Performance | No relevant errors in CT values and noise in homogeneous phantoms. No relevant errors in CT values in tissue-equivalent phantoms. No relevant geometrical distortions in static phantoms. No relevant deteriorations of position/shape in dynamic phantoms. No relevant new artifacts. Maintain performance with iMAR. Independent of detector width. | "introduces no relevant errors in terms of CT values and noise levels measured in a homogeneous water phantom" "introduces no relevant errors in terms of CT values measured in a phantom with tissue-equivalent inserts, even in the presence of metals and in combination with the iMAR algorithm" "introduces no relevant geometrical distortions in a static torso phantom" "introduces no relevant deteriorations of the position or shape of a dynamic thorax phantom" "does not introduce relevant new artefacts" "can be successfully applied in combination with metal artifact correction (iMAR)" "is independent from the physical detector width" |
| DirectDensity Performance (iBHC variants) | Reduced dependence on tube voltage and filtration for non-water-like tissues. Image values aligned with material properties. | "reduced dependence on tube voltage and filtration compared to the corresponding quantitative kernel (Qr) with iBHC Bone for non-water-like tissues, such as adipose and bone." "generate image value closely aligned with the respective material properties." "has been validated." |
2. Sample Sizes Used for the Test Set and Data Provenance
The document provides very limited, qualitative information:
- FAST 3D Camera: Optimized using "additional data from adults and adolescence patients." No specific number of patients or images mentioned.
- FAST Planning: Evaluated on "patient data." No specific number of patients or images mentioned.
- HD FoV 5.0: Evaluated with "physical and anthropomorphic phantoms."
- Flex 4D Spiral: No specific sample size or data type mentioned for performance assessment.
- ZeeFree RT: Evaluated with "homogeneous water phantom," "phantom with tissue-equivalent inserts," "static torso phantom," and "dynamic thorax phantom." Also, "retrospective blinded rater studies of respiratory 4D CT examinations performed at two institutions." No specific number of phantoms, images per phantom, or patient cases mentioned.
- DirectDensity: Evaluated on "SOMATOM CT scanner models." No specific sample size or data type mentioned.
Data Provenance:
- Country of Origin: Not specified for the patient data used for algorithm optimization/validation.
- Retrospective or Prospective:
- FAST 3D Camera: Implied retrospective as it uses "additional data."
- FAST Planning: Implied retrospective as it uses "patient data."
- HD FoV 5.0: Retrospective for the blinded rater study.
- ZeeFree RT: Retrospective for the blinded rater study of clinical cases. The phantom tests are by nature not retrospective/prospective.
3. Number of Experts and Qualifications for Ground Truth
- HD FoV 5.0: "board-approved radio-oncologists and medical physicists." The number of experts is not specified.
- ZeeFree RT: "board-approved radio-oncologists and medical physicists." The number of experts is not specified.
For other tests, ground truth appears to be established by phantom measurements or internal engineering verification, rather than human expert reads validating clinical ground truth.
4. Adjudication Method for the Test Set
The document mentions "retrospective blinded rater study" for HD FoV 5.0 and ZeeFree RT. However, it does not specify the adjudication method used (e.g., 2+1, 3+1, none) for these studies. It only states they were "blinded."
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned for HD FoV 5.0 and ZeeFree RT. Both were "retrospective blinded rater studies."
- Effect Size: The document does not report specific effect sizes (e.g., how much human readers improve with AI vs. without AI assistance). It only states that the purpose of the comparison was to "demonstrate that the HD FoV 5.0 algorithm is as safe and effective as the HD FoV 4.0 algorithm" and for ZeeFree RT that it "enables the optional reconstruction of stack artefact corrected images, which reduce the strength of misalignment artefacts." This implies an assessment of non-inferiority or improvement in image quality, but specific quantitative results for reader performance are not provided in this excerpt.
6. Standalone (Algorithm Only) Performance
The document describes tests for several algorithms (FAST 3D Camera, FAST Planning, HD FoV 5.0, Flex 4D Spiral, ZeeFree RT, DirectDensity) using phantoms and "patient data." These evaluations seem to be focused on the algorithm's performance in generating images or calculations, independent of human interpretation in some cases (e.g., accuracy of FAST 3D Camera, correctness percentage of FAST Planning).
However, it does not explicitly use the term "standalone performance" to differentiate these from human-in-the-loop assessments. The mention of "retrospective blinded rater studies" for HD FoV 5.0 and ZeeFree RT indicates a human-in-the-loop component for that specific evaluation, but the phantom testing mentioned alongside them would be considered standalone.
7. Type of Ground Truth Used
- Phantom Data: For HD FoV 5.0, Flex 4D Spiral, ZeeFree RT, and DirectDensity, physical and/or anthropomorphic phantoms were used, implying the ground truth is precisely known physical characteristics or pre-defined phantom configurations.
- Expert Consensus/Reads: For HD FoV 5.0 and ZeeFree RT, board-approved radio-oncologists and medical physicists performed retrospective blinded rater studies, implying their interpretations/ratings served as a form of ground truth or evaluation metric. It's not explicitly stated if this was against a clinical gold standard (e.g., pathology) or if it was a comparative assessment of image quality and clinical utility.
- Internal Verification: For FAST 3D Camera, FAST Planning, accuracy was assessed, likely against internal system metrics or pre-defined ideal outcomes.
8. Sample Size for the Training Set
The document does not provide any specific information about the sample size used for training the algorithms (e.g., for FAST 3D Camera, FAST Planning, HD FoV 5.0, ZeeFree RT). It only states that FAST 3D Camera was "optimized using additional data" and FAST Planning's algorithm had "product development, validation, and verification on patient data."
9. How the Ground Truth for the Training Set Was Established
The document does not provide any specific information on how the ground truth for the training set was established. It only mentions the data types used for validation/verification (phantoms, patient data from two institutions, expert raters).
FDA 510(k) Clearance Letter - SOMATOM CT Systems
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov
July 3, 2025
Siemens Medical Solutions USA, Inc.
℅ Kenny M Bello
Regulatory Affairs Professional
810 Innovation Drive
KNOXVILLE, TN 37932
Re: K250822
Trade/Device Name: SOMATOM go.Now; SOMATOM go.Up; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM Pro.Pulse
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: May 16, 2025
Received: June 16, 2025
Dear Kenny M Bello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K250822 - Kenny M Bello Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K250822 - Kenny M Bello Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K250822
Device Name: SOMATOM go.Now; SOMATOM go.Up; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM Pro.Pulse
Indications for Use (Describe)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of X-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment, and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1
Page 5
510(k) Summary
SOMATOM go. Platform and SOMATOM Pro.Pulse
with software version SOMARIS/10 syngo CT VB20
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
I. Contact Details
Submitter
Siemens Medical Solutions USA, Inc.
810 Innovation Drive
Knoxville, TN 37932
Establishment Registration Number: 1034973
Importer/Distributor
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355
Establishment Registration Number: 2240869
Location of Manufacturing Site (1)
Siemens Healthineers AG
Siemensstr. 1 -OR- Rittigfeld 1
D-91301 Forchheim, Germany
Establishment Registration Number: 3004977335
Location of Manufacturing Site (2)
SIEMENS SHANGHAI, MEDICAL EQUIPMENT LTD
278 Zhou Zhu Rd
Shanghai, CHINA, 201318
Establishment Registration Number: 3003202425
Note: Descriptions in this submission use the short company name Siemens. It covers both manufacturing locations and names as listed above. Brand name on all products is Siemens Healthineers.
Page 6
Submitter Contact Person:
Kenny M Bello
Regulatory Affairs Professional
Siemens Medical Solutions USA, Inc.
Molecular Imaging
810 Innovation Drive
Knoxville, TN 37932
Phone: (205) 856-6099
monsuru.bello@siemens-healthineers.com
II. Device Name and Classification
Table 1: Subject Device Names
| Product Name | Trade Name |
|---|---|
| SOMATOM go.Now | SOMATOM go.Now |
| SOMATOM go.Up | SOMATOM go.Up |
| SOMATOM go.All | SOMATOM go.All |
| SOMATOM go.Top | SOMATOM go.Top |
| SOMATOM go.Sim | SOMATOM go.Sim |
| SOMATOM go.Open Pro | SOMATOM go.Open Pro |
| SOMATOM Pro.Pulse | SOMATOM Pro.Pulse |
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
III. Predicate Device
Primary predicate device:
Trade Name: SOMATOM go. Platform CT Scanner Systems with SOMARIS/10 syngo CT VB10 software
510(k) Number: K233650
Clearance Date: March 26, 2024
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
Recall Information: All predicate device recalls have been considered in the subject device design.
Page 7
Secondary predicate device:
Trade Name: SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 software
510(k) Number: K232206
Clearance Date: December 6, 2024
Classification Name: Computed Tomography X-ray System
Classification Panel: Radiology
Regulation Number: 21 CFR §892.1750
Device Class: Class II
Product Code: JAK
Recall Information: All predicate device recalls have been considered in the subject device design.
Note:
K233650 was a bundle submission with various Siemens CT Scanner Systems, including the CT scanner systems SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, SOMATOM X.cite, SOMATOM X.ceed.
IV. Device Description
Siemens intends to market a new software version, SOMARIS/10 syngo CT VB20 for the following SOMATOM Computed Tomography (CT) Scanner Systems:
a) Single Source CT Scanner systems (SOMATOM go. Platform):
- SOMATOM go.Now
- SOMATOM go.Up
- SOMATOM go.All
- SOMATOM go.Top
- SOMATOM go.Sim
- SOMATOM go.Open Pro
In this submission, the above listed CT scanner systems are jointly referred to as subject devices by "SOMATOM go. Platform" CT scanner systems.
b) Dual Source CT Scanner system:
- SOMATOM Pro.Pulse
The above listed subject devices with SOMARIS/10 syngo CT VB20 are Computed Tomography X-ray Systems which feature one (Single Source) or two (Dual Source) continuously rotating tube-detector system and function according to the fan beam principle. The SOMATOM go. Platform and the SOMATOM Pro.Pulse with software SOMARIS/10 syngo CT VB20 produce CT images in DICOM format, which can be used by trained staff for software applications, e.g. post-processing applications, commercially distributed by Siemens Healthcare and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.
Page 8
Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.
The software platform provides plugin software interfaces that allow for the use of specific commercially available post processing software algorithms in an unmodified form from the cleared stand-alone post processing version.
New software version syngo CT VB20 (SOMARIS/10 syngo CT VB20) is a modified software version based on syngo CT VB10 (SOMARIS/10 syngo CT VB10) which was cleared for the predicate devices in K233650 (SOMATOM go. Platform) and in K232206 (SOMATOM Pro.Pulse).
Software version SOMARIS/10 syngo CT VB20 will be offered ex-factory and as an optional upgrade for the applicable existing SOMATOM go. Platform and SOMATOM Pro.Pulse CT Scanner Systems.
The bundle approach is feasible for this submission since the subject devices have similar technological characteristics, software operating platform, and supported software characteristics. The supporting data are similar, primarily one review division/group will be involved, and the indications for use is the same between the devices. All subject devices will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use remains unchanged compared to the predicate devices.
V. Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment, and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
VI. Indications for Use Comparison
Subject Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
Page 9
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Primary Predicate Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Secondary Predicate Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Comparison:
The subject device Indications for Use is the exact same as the primary (K233650) and the secondary (K232206) predicate's Indications for Use. As such, the Indications for Use remains unchanged from the predicate devices and is substantially equivalent to the predicate devices.
VII. Comparison of Technological Characteristics with the Predicate Device
Supported by the subject device, SOMARIS/10 syngo CT VB20 software version is a further development of the SOMARIS/10 syngo CT VB10 software version which is cleared in K233650 and K232206.
The SOMATOM go. Platform and SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20 software version provides the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the predicate devices. The software features of the SOMATOM go. Platform and SOMATOM Pro.Pulse have been modified or improved in comparison to the predicate devices to support enhanced device functionality compared to the predicate devices.
Page 10
The new syngo CT VB20 software reuses all unmodified software features of the legacy software syngo CT VB10 cleared in K233650 and K232206. Additionally, no features present in the predicate device are descoped.
Software version SOMARIS/10 syngo CT VB20 is designed to reuse hardware independent extended functionalities and GO technologies provided by Siemens cleared software applications.
The intended use and fundamental scientific technology for the SOMATOM go. Platform and SOMATOM Pro.Pulse remain unchanged from the predicate devices.
At a high level, the subject and predicate devices are based on the same subset of technological elements:
- Scanner Principle – Whole body X-Ray Computed Tomography Scanner
- System Acquisition – Continuously rotating tube detector system
- Iterative Reconstruction – Support of various iterative reconstruction principles
- Workplaces – Support of workplaces that include reconstruction and image evaluation software
- Patient table
- Patient table foot switch for movement
- Stellar Detector technology
- Tin filtration technology
- Athlon/Vectron X-ray Tube
- Power Generator
- Scan&GO
- Mobile workflow (Tablet)
- Optional injector arm
- Optional support of CT guided intervention workflow (myNeedle Companion – myNeedle Guide)
- Optional support of FAST 3D Camera for patient positioning workflow
- Scanner display and control functionality
- Remote Scan Control
- Interfaces for Respiratory Gating Devices
- Standard CARE technologies
- Standard FAST technologies
- Standard GO technologies
- Precision matrix (large image matrices)
- HD FoV
- DirectDensity
- Stack artefact reduced reconstruction
The terms in the following Table 2 are used to indicate various technological characteristics in comparison to the primary and secondary predicate devices information:
Table 2: Overview of term definition.
| Term | Definition |
|---|---|
| new | The feature is newly supported for Siemens CT Scanners and newly introduced within the subject device submission |
Page 11
| Term | Definition |
|---|---|
| modified | This feature is modified from the predicate devices |
| enabled | This feature is currently supported by other cleared Siemens CT systems or cleared Siemens stand-alone software applications. This feature will be supported for the subject device with software version SOMARIS/10 syngo CT VB20 and is substantially equivalent compared to the cleared version of the predicate devices. |
| unmodified | The feature remains unchanged from the predicate device. |
| n.a. | The feature is not supported for the subject/predicate device |
The subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse CT scanner systems with SOMARIS/10 syngo CT VB20 will support hardware and software modifications/further developments in comparison to the predicate devices as listed in Table 3. The column "HW/ SW change" in the table contains reference to a hardware (HW) or/and software (SW) change. The columns under "Predicate devices" of the Table 3 indicate the clearance of the technological properties in the corresponding predicate device.
Page 12
Table 3: Summary of modifications (hardware and software) of the subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse with software version SOMARIS/10 syngo CT VB20 compared to the predicate devices.
| Technological property | HW/SW change | Subject devices | Predicate devices | ||
|---|---|---|---|---|---|
| Single Source | Dual Source | Single Source | Dual Source | ||
| SOMATOM go. Platform with syngo CT VB20 | SOMATOM Pro.Pulse with syngo CT VB20 | SOMATOM go. Platform with syngo CT VB10 (K233650) | SOMATOM Pro.Pulse with syngo CT VB10 (K232206) | ||
| 1. CARE Moodlight | HW/SW | enabled¹ (same as in K233650) | unmodified | cleared | cleared |
| 2. Interfaces for respiratory gating | HW | modified² (same as in K233650, with exception of Open Online Interface which are introduced with syngo CT VB20) | n.a. | cleared | n.a. |
| 3. Eco Power Mode | HW/SW | new³ | new | n.a. | n.a. |
¹ In software version syngo CT VB20, the functionality CARE Moodlight is enabled for the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro. For the subject devices SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, and SOMATOM go.Top, the functionality CARE Moodlight was cleared in K233650 and it remains unchanged in software version syngo CT VB20.
² SOMATOM go.Now does not support respiratory gating interfaces.
³ SOMATOM go.Now and SOMATOM go.Up do not support Eco Power Mode.
Page 13
| Technological property | HW/SW change | Subject devices | Predicate devices | ||
|---|---|---|---|---|---|
| Single Source | Dual Source | Single Source | Dual Source | ||
| SOMATOM go. Platform with syngo CT VB20 | SOMATOM Pro.Pulse with syngo CT VB20 | SOMATOM go. Platform with syngo CT VB10 (K233650) | SOMATOM Pro.Pulse with syngo CT VB10 (K232206) | ||
| 4. FAST 3D Camera/ FAST Integrated Workflow | SW | modified (same as in K233650, with exception of retrained FAST 3D Camera algorithms, collision indication, and Centerline/Grid Overlay which are introduced with syngo CT VB20) | modified (same as in K232206, with exception of retrained FAST 3D Camera algorithms, collision indication, and Centerline/Grid Overlay which are introduced with syngo CT VB20) | cleared | cleared |
| 5. Oncology Exchange | SW | new | new | n.a. | n.a. |
| 6. DirectSetup Notes | SW | enabled⁴ (same as in K233650) | enabled⁴ (same as in K233650) | cleared | n.a. |
| 7. FAST Planning | SW | modified (same as in K233650, with exception of the detection of additional body regions which are introduced with syngo CT VB20) | modified (same as in K232206, with exception of the detection of additional body regions which are introduced with syngo CT VB20) | cleared | cleared |
⁴ In software version syngo CT VB20, the functionality DirectSetup Notes is enabled for the subject devices SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, and SOMATOM Pro.Pulse. For the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro, the functionality DirectSetup Notes was cleared in K233650 and it remains unchanged in software version syngo CT VB20.
Page 14
| Technological property | HW/SW change | Subject devices | Predicate devices | ||
|---|---|---|---|---|---|
| Single Source | Dual Source | Single Source | Dual Source | ||
| SOMATOM go. Platform with syngo CT VB20 | SOMATOM Pro.Pulse with syngo CT VB20 | SOMATOM go. Platform with syngo CT VB10 (K233650) | SOMATOM Pro.Pulse with syngo CT VB10 (K232206) | ||
| 8. myExam Companion – myExam Compass/myExam Cockpit | SW | modified (same as in K233650, with exception of the clinical decision trees now also available for child protocols) | modified (same as in K233650, with exception of the clinical decision trees now also available for child protocols) | cleared | cleared |
| 9. HD FoV 5.0 | SW | new⁵ | n.a. | n.a. | n.a. |
| 10. CT guided intervention – myAblation Guide interface | SW | new | new | n.a. | n.a. |
| 11. Flex 4D Spiral | SW | modified⁶ (same as in K233650, with exception of the modifications regarding dynamic tube current modulation which is introduced with syngo CT VB20) | modified (same as in K233650, with exception of the modifications regarding dynamic tube current modulation which is introduced with syngo CT VB20) | cleared⁶ | cleared |
⁵ In software version syngo CT VB20, the functionality HD FoV 5.0 is introduced only for the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro.
⁶ The CT scanner CT systems SOMATOM go.Now and SOMATOM go.Up do not support Flex 4D Spiral.
Page 15
| Technological property | HW/SW change | Subject devices | Predicate devices | ||
|---|---|---|---|---|---|
| Single Source | Dual Source | Single Source | Dual Source | ||
| SOMATOM go. Platform with syngo CT VB20 | SOMATOM Pro.Pulse with syngo CT VB20 | SOMATOM go. Platform with syngo CT VB10 (K233650) | SOMATOM Pro.Pulse with syngo CT VB10 (K232206) | ||
| 12. ZeeFree RT | SW | new⁷ | n.a. | n.a. | n.a. |
| 13. DirectDensity | SW | modified (same as in K233650, with exception of the calculation of stopping-power ratio (Kernel St) which is introduced with syngo CT VB20) | modified (same as in K233650, with exception of the calculation of stopping-power ratio (Kernel St) which is introduced with syngo CT VB20) | cleared | cleared |
| 14. DirectLaser | HW/SW | modified⁸ (same as in K233650, with exception of the Patient Marking workflow improvement introduced with syngo CT VB20) | n.a. | cleared⁸ | n.a. |
⁷ In software version syngo CT VB20, the functionality ZeeFree RT is introduced only for the subject device SOMATOM go.Open Pro.
⁸ DirectLaser is available only for SOMATOM go.Sim and SOMATOM go.Open Pro.
Page 16
| Technological property | HW/SW change | Subject devices | Predicate devices | ||
|---|---|---|---|---|---|
| Single Source | Dual Source | Single Source | Dual Source | ||
| SOMATOM go. Platform with syngo CT VB20 | SOMATOM Pro.Pulse with syngo CT VB20 | SOMATOM go. Platform with syngo CT VB10 (K233650) | SOMATOM Pro.Pulse with syngo CT VB10 (K232206) | ||
| 15. myExam Contrast | SW | new | new | n.a. | n.a. |
Page 17
A tabular summary of the comparable hardware and software properties between the subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse with software version syngo CT VB20 and the predicate devices are listed in Table 4 through Table 8 below (modifications are in gray shaded sections).
Table 4 provides comparison of the hardware properties between the single source subject devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB20 and the predicate devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB10 (K233650).
Table 5 provides comparison of the hardware properties between the single source subject devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB20 and the predicate devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB10 (K233650).
Table 6 provides comparison of the hardware properties between the single source subject devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT VB20 and the predicate devices SOMATOM go.Sim and SOMATOM go.Open.Pro with software version syngo CT VB10 (K233650).
Table 7 provides comparison of the hardware properties between the dual source subject device SOMATOM Pro.Pulse with software version syngo CT VB20 and the predicate devices SOMATOM Pro.Pulse (K233650) and SOMATOM go.Top with software version syngo CT VB10 (233650).
Table 8 compares the software technological characteristics of the subject devices and the predicate devices.
Table 4: Comparison of technical hardware characteristics between the subject devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB20 and the primary predicate devices SOMATOM go.Now and SOMATOM go.Up with software version syngo CT VB10 (K233650).
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Now | SOMATOM go.Up | SOMATOM go.Now | SOMATOM go.Up | ||
| scanner | Whole body Computed Tomography (CT) Scanner System | Whole body Computed Tomography (CT) Scanner System | Same as the primary predicate devices | ||
| System Configuration | Single Source | Single Source | Same as the primary predicate devices |
Page 18
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Now | SOMATOM go.Up | SOMATOM go.Now | SOMATOM go.Up | ||
| Environment of Use | Professional Healthcare Facility | Professional Healthcare Facility | Same as the primary predicate devices | ||
| Generator Max. power (kW) | 32 | 32 | Same as the primary predicate devices | ||
| Detector technology | Stellar based on UFC (Ultra Fast Ceramic) | Stellar based on UFC (Ultra Fast Ceramic) | Same as the primary predicate devices | ||
| Detector volume coverage (mm) | 11.2 | 22.4 | 11.2 | 22.4 | Same as the primary predicate devices |
| Detector physical rows | 16 | 32 | 16 | 32 | Same as the primary predicate devices |
| Detector Slice width (mm) | 0.7 | 0.7 | Same as the primary predicate devices | ||
| Detector DAS channel no. | 768 | 768 | Same as the primary predicate devices | ||
| Detector Image slices | 32 | 64 | 32 | 64 | Same as the primary predicate devices |
| Tube technology | Chronon | Chronon | Same as the primary predicate devices |
Page 19
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Now | SOMATOM go.Up | SOMATOM go.Now | SOMATOM go.Up | ||
| Tube kV steps | 80, 110, 130 | 80, 110, 130 | Same as the primary predicate devices | ||
| Tube Max. current (mA) | 400 | 400 | Same as the primary predicate devices | ||
| Tube focus | • 0.8 x 0.4 / 8° • 0.8 x 0.7 / 8° | • 0.8 x 0.4 / 8° • 0.8 x 0.7 / 8° | Same as the primary predicate devices | ||
| Tube Heat storage capacity (MHU) | 3.5 | 3.5 | Same as the primary predicate devices | ||
| Tube Cooling rate (kHU/ min) | 915 | 915 | Same as the primary predicate devices | ||
| Gantry Bore size | 70 | 70 | Same as the primary predicate devices | ||
| Gantry Scan FoV (cm) | 50 | 50 | Same as the primary predicate devices | ||
| Gantry Extended FoV (cm) | 70 | 70 | Same as the primary predicate devices | ||
| Gantry Rot. time (sec) | 0.5, 0.8, 1.0, 1.5 | 0.5, 0.8, 1.0, 1.5 | Same as the primary predicate devices |
Page 20
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Now | SOMATOM go.Up | SOMATOM go.Now | SOMATOM go.Up | ||
| Gantry Tilt (degree) | n.a. | +/- 30° -30° to +28° with optional 307 kg patient table | n.a. | +/- 30° -30° to +28° with optional 307 kg patient table | Same as the primary predicate devices |
| Patient table Type | Vector (1250 mm), Vario 1 (1600 mm), Vario RT (1600 mm) with table extension | Vario 1 (1600 mm), Vario 2 (2000 mm), Vario RT (1600 mm) with table extension | Vector (1250 mm), Vario 1 (1600 mm), Vario RT (1600 mm) with table extension | Vario 1 (1600 mm), Vario 2 (2000 mm), Vario RT (1600 mm) with table extension | Same as the primary predicate devices |
| Patient table Max. weight capacity (kg) | Vector: 160 Vario 1: 227 Vario RT: 227 | Vario 1: 227 Vario 2: 307 Vario RT: 227 | Vector: 160 Vario 1: 227 Vario RT: 227 | Vario 1: 227 Vario 2: 307 Vario RT: 227 | Same as the primary predicate devices |
| Max. scan length topogram | 1680 mm with table extension | 1680 mm with table extension 2080 mm with table extension | 1680 mm with table extension | 1680 mm with table extension 2080 mm with table extension | Same as the primary predicate devices |
| Max. scan length Image acquisition | Max. 1600 mm with patient table extension | Max. 1600 mm with patient table extension | Max. 1600 mm with patient table extension | Max. 1600 mm with patient table extension | Same as the primary predicate devices |
Page 21
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Now | SOMATOM go.Up | SOMATOM go.Now | SOMATOM go.Up | ||
| Max. 2000 mm with patient table extension | Max. 2000 mm with patient table extension | ||||
| Spectral filtration | Tin Filter | Tin Filter | Same as the primary predicate devices | ||
| High Power 80 | yes | yes | yes | yes | Same as the primary predicate devices |
| FAST 3D Camera | option for patient positioning with 3D Camera: • gantry mounted | option for patient positioning with 3D Camera: • ceiling mounted • gantry mounted | option for patient positioning with 3D Camera: • gantry mounted | option for patient positioning with 3D Camera: • ceiling mounted • gantry mounted | Same as the primary predicate devices |
| x-ray foot switch | Option to trigger hands-free scanning | Option to trigger hands-free scanning | Same as the primary predicate devices | ||
| Table foot switch | Option for table patient movement | Option for table patient movement | Same as the primary predicate devices | ||
| i-joystick | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Same as the primary predicate devices |
Page 22
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Now | SOMATOM go.Up | SOMATOM go.Now | SOMATOM go.Up | ||
| Tablet dock | Option for mounting of the tablet on the patient table. | Option for mounting of the tablet on the patient table. | Same as the primary predicate devices | ||
| CARE Breathe | visualization of breathing instructions to support three messages in respiratory motion management: • free breathing • preparation for breath hold • breath hold itself in combination with the visualization of a countdown visual patient instruction unit is placed on the gantry front and gantry back in the gantry funnel | visualization of breathing instructions to support three messages in respiratory motion management: • free breathing • preparation for breath hold • breath hold itself in combination with the visualization of a countdown visual patient instruction unit is placed on the gantry front and gantry back in the gantry funnel | Same as the primary predicate devices | ||
| CARE 2D camera | one integrated into the gantry front and one mounted on the gantry back to monitor the patient during examination | one integrated into the gantry front and one mounted on the gantry back to monitor the patient during examination | Same as the primary predicate devices | ||
| CARE Moodlight | LED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Scan | LED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Radiation on | Same functionality as the primary predicate devices. In the subject devices with syngo CT VB20, the status "Radiation on" has been renamed to "Scan". |
Page 23
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Now | SOMATOM go.Up | SOMATOM go.Now | SOMATOM go.Up | ||
| Respiratory Motion management - Interfaces for respiratory gating | n.a. | acquisition of respiration-correlated scans supported Interface connectors for: • Anzai respiratory gating system Varian RGSC respiratory gating system • Open interface compatible respiratory gating systems • Open Online interface | n.a. | acquisition of respiration-correlated scans supported Interface connectors for: • Anzai respiratory gating system Varian RGSC respiratory gating system • Open interface compatible respiratory gating systems | Same as the primary predicate device, the subject device SOMATOM go.Up with software version syngo CT VB20 support acquisition of respiration-correlated scans. For measurement of the breathing signal of the patient, interfaces for connecting of 3rd party respiratory gating devices are offered. Introduction of new Open Online Interface. |
Page 24
Table 5: Comparison of technical hardware characteristics between the subject devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB20 and the primary predicate devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB10 (K233650).
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.All | SOMATOM go.Top | SOMATOM go.All | SOMATOM go.Top | ||
| scanner | Whole body Computed Tomography (CT) Scanner System | Whole body Computed Tomography (CT) Scanner System | Same as the primary predicate devices | ||
| System Configuration | Single Source | Single Source | Same as the primary predicate devices | ||
| Environment of Use | Professional Healthcare Facility | Professional Healthcare Facility | Same as the primary predicate devices | ||
| Generator Max. power (kW) | 75 | 75 | Same as the primary predicate devices | ||
| Detector technology | Stellar based on UFC (Ultra Fast Ceramic) | Stellar based on UFC (Ultra Fast Ceramic) | Same as the primary predicate devices | ||
| Detector volume coverage (mm) | 22.4 | 38.4 | 22.4 | 38.4 | Same as the primary predicate devices |
| Detector physical rows | 32 | 64 | 32 | 64 | Same as the primary predicate devices |
| Detector Slice width (mm) | 0.7 | 0.6 | 0.7 | 0.6 | Same as the primary predicate devices |
| Detector DAS channel no. | 768 | 840 | 768 | 840 | Same as the primary predicate devices |
Page 25
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.All | SOMATOM go.Top | SOMATOM go.All | SOMATOM go.Top | ||
| Detector Image slices | 64 | 128 | 64 | 128 | Same as the primary predicate devices |
| Tube technology | Athlon | Athlon | Same as the primary predicate devices | ||
| Tube kV steps | 70 to 140 (in 10 kV steps) | 70 to 140 (in 10 kV steps) | Same as the primary predicate devices | ||
| Tube Max. current (mA) | 825 | 825 | Same as the primary predicate devices | ||
| Tube focus | • 1.1 x 1.2 / 8° • 0.8 x 0.8 / 8° | • 1.1 x 1.2 / 8° • 0.8 x 0.8 / 8° | Same as the primary predicate devices | ||
| Tube Heat storage capacity (MHU) | 7 | 7 | Same as the primary predicate devices | ||
| Tube Cooling rate (kHU/ min) | Up to 1700 | Up to 1700 | Same as the primary predicate devices | ||
| Gantry Bore size | 70 | 70 | Same as the primary predicate devices | ||
| Gantry Scan FoV (cm) | 50 | 50 | Same as the primary predicate devices |
Page 26
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.All | SOMATOM go.Top | SOMATOM go.All | SOMATOM go.Top | ||
| Gantry Extended FoV (cm) | 70 | 70 | Same as the primary predicate devices | ||
| Gantry Rot. time (sec) | 0.33, 0.5, 1.0 | 0.33, 0.5, 1.0 | Same as the primary predicate devices | ||
| Gantry Tilt (degree) | +/- 30° -30° to +28° with optional 307 kg patient table | +/- 30° -30° to +28° with optional 307 kg patient table | Same as the primary predicate devices | ||
| Patient table Type | Vario 1 (1600 mm) Vario 2 (2000 mm) Vario RT (1600 mm) with table extension | Vario 1 (1600 mm) Vario 2 (2000 mm) Vario RT (1600 mm) with table extension | Same as the primary predicate devices | ||
| Patient table Max. weight capacity (kg) | Vario 1: 227 Vario 2: 307 Vario RT: 227 | Vario 1: 227 Vario 2: 307 Vario RT: 227 | Same as the primary predicate devices | ||
| Max. scan length topogram | 1680 mm with table extension 2080 mm with table extension | 1680 mm with table extension 2080 mm with table extension | Same as the primary predicate devices | ||
| Max. scan length Image acquisition | Max. 1600 mm with patient table extension Max. 2000 mm with patient table extension | Max. 1600 mm with patient table extension Max. 2000 mm with patient table extension | Same as the primary predicate devices |
Page 27
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.All | SOMATOM go.Top | SOMATOM go.All | SOMATOM go.Top | ||
| Spectral filtration | Tin Filter | Combined Tin Filter/Split filter | Tin Filter | Combined Tin Filter/Split filter | Same as the primary predicate devices |
| High Power 70/ High Power 80 | 825 (@70 kV) 825 (@ 80 kV) | 825 (@70 kV) 825 (@ 80 kV) | Same as the primary predicate devices | ||
| FAST 3D Camera | option for patient positioning with FAST 3D Camera: • ceiling mounted • gantry mounted | option for patient positioning with FAST 3D Camera: • ceiling mounted • gantry mounted | Same as the primary predicate devices | ||
| x-ray foot switch | Option to trigger hands-free scanning | Option to trigger hands-free scanning | Same as the primary predicate devices | ||
| Table foot switch | Option for table patient movement | Option for table patient movement | Same as the primary predicate devices | ||
| i-joystick | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Same as the primary predicate devices | ||
| Tablet dock | Option for mounting of the tablet on the patient table. | Option for mounting of the tablet on the patient table. | Same as the primary predicate devices | ||
| CARE Breathe | visualization of breathing instructions to support three messages in respiratory motion management: | visualization of breathing instructions to support three messages in respiratory motion management: | Same as the primary predicate devices |
Page 28
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.All | SOMATOM go.Top | SOMATOM go.All | SOMATOM go.Top | ||
| • free breathing • preparation for breath hold • breath hold itself in combination with the visualization of a countdown visual patient instruction unit is placed on the gantry front and gantry back in the gantry funnel | • free breathing • preparation for breath hold • breath hold itself in combination with the visualization of a countdown visual patient instruction unit is placed on the gantry front and gantry back in the gantry funnel | ||||
| CARE 2D camera | one integrated into the gantry front and one mounted on the gantry back to monitor the patient during examination | one integrated into the gantry front and one mounted on the gantry back to monitor the patient during examination | Same as the primary predicate devices | ||
| CARE Moodlight | LED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Scan | LED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Radiation on | Same functionality as the primary predicate devices In the subject devices with syngo CT VB20, the status "Radiation on" has been renamed to "Scan". | ||
| Respiratory Motion management - Interfaces for respiratory gating | acquisition of respiration-correlated scans supported Interface connectors for: • Anzai respiratory gating system • Varian RGSC respiratory gating system | acquisition of respiration-correlated scans supported Interface connectors for: • Anzai respiratory gating system • Varian RGSC respiratory gating system | Same as the primary predicate devices, the subject devices SOMATOM go.All and SOMATOM go.Top with software version syngo CT VB20 support acquisition of respiration-correlated scans. |
Page 29
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.All | SOMATOM go.Top | SOMATOM go.All | SOMATOM go.Top | ||
| • Open interface compatible respiratory gating systems • Open Online interface | • Open interface compatible respiratory gating systems | For measurement of the breathing signal of the patient, interfaces for connecting of 3rd party respiratory gating devices are offered. Introduction of new Open Online Interface. |
Page 30
Table 6: Comparison of technical hardware characteristics between the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT VB20 and the primary predicate devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT VB10 (K233650).
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Sim | SOMATOM go.Open Pro | SOMATOM go.Sim | SOMATOM go.Open Pro | ||
| scanner | Whole body Computed Tomography (CT) Scanner System | Whole body Computed Tomography (CT) Scanner System | Same as the primary predicate devices | ||
| System Configuration | Single Source | Single Source | Same as the primary predicate devices | ||
| Environment of Use | Professional Healthcare Facility | Professional Healthcare Facility | Same as the primary predicate devices | ||
| Generator Max. power (kW) | 75 | 75 | Same as the primary predicate devices | ||
| Detector technology | Stellar based on UFC (Ultra Fast Ceramic) | Stellar based on UFC (Ultra Fast Ceramic) | Same as the primary predicate devices | ||
| Detector volume coverage (mm) | 19.2 | 38.4 | 19.2 | 38.4 | Same as the primary predicate devices |
| Detector physical rows | 32 | 64 | 32 | 64 | Same as the primary predicate devices |
| Detector Slice width (mm) | 0.6 | 0.6 | 0.6 | 0.6 | Same as the primary predicate devices |
Page 31
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Sim | SOMATOM go.Open Pro | SOMATOM go.Sim | SOMATOM go.Open Pro | ||
| Detector DAS channel no. | 920 | 920 | 920 | 920 | Same as the primary predicate devices |
| Detector Image slices | 64 | 128 | 64 | 128 | Same as the primary predicate devices |
| Tube technology | Athlon | Athlon | Same as the primary predicate devices | ||
| Tube kV steps | 70 to 140 (in 10 kV steps) | 70 to 140 (in 10 kV steps) | Same as the primary predicate devices | ||
| Tube Max. current (mA) | 825 | 825 | Same as the primary predicate devices | ||
| Tube focus | • 1.1 x 1.2 / 7° • 0.8 x 0.8 / 7° | • 1.1 x 1.2 / 7° • 0.8 x 0.8 / 7° | Same as the primary predicate devices | ||
| Tube Heat storage capacity (MHU) | 7 | 7 | Same as the primary predicate devices | ||
| Tube Cooling rate (kHU/ min) | Up to 1700 | Up to 1700 | Same as the primary predicate devices |
Page 32
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Sim | SOMATOM go.Open Pro | SOMATOM go.Sim | SOMATOM go.Open Pro | ||
| Gantry Bore size | 85 | 85 | Same as the primary predicate devices | ||
| Gantry Scan FoV (cm) | 60 | 60 | Same as the primary predicate devices | ||
| Gantry Extended FoV (cm) | 85 | 85 | Same as the primary predicate devices | ||
| Gantry Rot. time (sec) | 0.35, 0.5, 1.0 | 0.35, 0.5, 1.0 | Same as the primary predicate devices | ||
| Gantry Tilt (degree) | -30° to +25° | -30° to +25° | Same as the primary predicate devices | ||
| Patient table Type | Vario 2 (2000 mm) Vario RT (1600 mm) with table extension | Vario 2 (2000 mm) Vario RT (1600 mm) with table extension | Same as the primary predicate devices | ||
| Patient table Max. weight capacity (kg) | Vario 2: 307 Vario RT: 227 | Vario 2: 307 Vario RT: 227 | Same as the primary predicate devices | ||
| Max. scan length topogram | 1680 mm with table extension 2080 mm with table extension | 1680 mm with table extension 2080 mm with table extension | Same as the primary predicate devices |
Page 33
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Sim | SOMATOM go.Open Pro | SOMATOM go.Sim | SOMATOM go.Open Pro | ||
| Max. scan length Image acquisition | Max. 1600 mm with patient table extension Max. 2000 mm with patient table extension | Max. 1600 mm with patient table extension Max. 2000 mm with patient table extension | Same as the primary predicate devices | ||
| Spectral filtration | Tin Filter | Combined Tin Filter/Split filter | Tin Filter | Combined Tin Filter/Split filter | Same as the primary predicate devices |
| High Power 70/ High Power 80 | 825 (@70 kV) 825 (@ 80 kV) | 825 (@70 kV) 825 (@ 80 kV) | Same as the primary predicate devices | ||
| FAST 3D Camera | option for patient positioning with FAST 3D Camera: • ceiling mounted | option for patient positioning with FAST 3D Camera: • ceiling mounted | Same as the primary predicate devices | ||
| CARE Moodlight | LED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Scan | LED lights integrated on the gantry front cover indicating⁹: • System startup • Scan countdown • Radiation on | CARE Moodlight is introduced for SOMATOM go.Sim and SOMATOM go.Open Pro In the subject devices with syngo CT VB20, the status "Radiation on" has been renamed to "Scan". |
⁹ Added based on non-filing documentation (document-to-file) after the clearance of K233650. The modification has been considered as minor modification to the system and do not change the indications for use nor do significantly affect the safety and effectiveness of the device, hence non-filing documentation.
Page 34
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Sim | SOMATOM go.Open Pro | SOMATOM go.Sim | SOMATOM go.Open Pro | ||
| x-ray foot switch | Option to trigger hands-free scanning | Option to trigger hands-free scanning | Same as the primary predicate devices | ||
| Table foot switch | Option for table patient movement | Option for table patient movement | Same as the primary predicate devices | ||
| i-joystick | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Same as the primary predicate devices | ||
| Tablet dock | Option for mounting of the tablet on the patient table. | Option for mounting of the tablet on the patient table. | Same as the primary predicate devices | ||
| CARE Breathe | n.a. | n.a. | Same as the primary predicate devices | ||
| CARE 2D camera | Patient observation camera integrated on gantry front | Patient observation camera integrated on gantry front | Same as the primary predicate devices | ||
| Respiratory Motion management - Interfaces for respiratory gating | acquisition of respiration-correlated scans supported Interface connectors for: | acquisition of respiration-correlated scans supported Interface connectors for: | Same as the primary predicate devices, the subject devices SOMATOM go.Sim and SOMATOM go.Open Pro with software version syngo CT |
Page 35
| Hardware property | Subject device SOMATOM go. Platform with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go. Platform with SOMARIS/10 syngo CT VB10 (K233650) | Assessment of the substantial equivalency (SE) | ||
|---|---|---|---|---|---|
| SOMATOM go.Sim | SOMATOM go.Open Pro | SOMATOM go.Sim | SOMATOM go.Open Pro | ||
| • Anzai respiratory gating system • Varian RGSC respiratory gating system • Open interface compatible respiratory gating systems • Open Online interface | • Anzai respiratory gating system • Varian RGSC respiratory gating system • Open interface compatible respiratory gating systems | VB20 support acquisition of respiration-correlated scans. For measurement of the breathing signal of the patient, interfaces for connecting of 3rd party respiratory gating devices are offered. Introduction of new Open Online Interface. |
Page 36
Table 7: Comparison of technical hardware characteristics between the subject device SOMATOM Pro.Pulse with software version syngo CT VB20 and the predicate devices SOMATOM Pro.Pulse (K232206) and SOMATOM go.Top with software version syngo CT VB10 (233650).
| Hardware property | Subject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) |
|---|---|---|---|---|
| scanner | Whole body Computed Tomography (CT) Scanner System | Whole body Computed Tomography (CT) Scanner System | Whole body Computed Tomography (CT) Scanner System | Same as the primary and the secondary predicate device |
| System Configuration | Dual Source | Single Source | Dual Source | Same as the secondary predicate device |
| Environment of Use | Professional Healthcare Facility | Professional Healthcare Facility | Professional Healthcare Facility | Same as the primary and the secondary predicate device |
| Generator Max. power (kW) | 150 (2 x 75) | 75 | 150 (2 x 75) | Same as the secondary predicate device |
| Detector technology | Stellar detector based on Multislice UFC (Ultra Fast Ceramic) | Stellar detector based on Multislice UFC (Ultra Fast Ceramic) | Stellar detector based on Multislice UFC (Ultra Fast Ceramic) | Same as the primary and the secondary predicate device |
| Detector volume coverage | 38.4 mm | 38.4 mm | 38.4 mm | Same as the primary and secondary predicate device |
| Detector physical rows | 2 x 64 | 64 | 2 x 64 | Same as the primary and the secondary predicate device |
¹⁰ The primary predicate device of this eSTAR submission is the SOMATOM go. Platform with syngo CT VB10 (K233650). K233650 is a bundle 510(k) which the SOMATOM go.Top is part of.
Page 37
| Hardware property | Subject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) |
|---|---|---|---|---|
| Detector Slice width | 0.6 mm | 0.6 mm | 0.6 mm | Same as the primary and secondary predicate device |
| Detector DAS channel no. | 840 (detector A) 600 (detector B) | 840 | 840 (detector A) 600 (detector B) | Same as the secondary predicate device |
| Detector Image slices | 2 x 128 (with Interleaved Volume Reconstruction, IVR) 2 x 384 (reconstructed slices) | 128 (with Interleaved Volume Reconstruction, IVR) 384 (reconstructed slices) | 2 x 128 (with Interleaved Volume Reconstruction, IVR) 2 x 384 (reconstructed slices) | Same as the secondary predicate device |
| Tube technology | 2 x Athlon DS X-ray tube | Athlon X-ray tube | 2 x Athlon DS X-ray tube | Same as the secondary predicate device. |
| Tube collimator | Tube collimator DS4-A, DS4-B • Equivalent to 0.5 mm Al in the isocenter • 1 mm Al with cardio wedge | Tube collimator • Equivalent to 0.5 mm Al in the isocenter • 1 mm Al with cardio wedge | Tube collimator DS4-A, DS4-B • Equivalent to 0.5 mm Al in the isocenter • 1 mm Al with cardio wedge | Same as the secondary predicate device. |
| Tube kV steps | 70–140 kV in 10 kV steps | 70–140 kV in 10 kV steps | 70–140 kV in 10 kV steps | Same as the primary and the secondary predicate device |
| Tube Max. current | Single Source: 13-825 mA Dual Source: 26-1650 mA | Standard range: 13 – 825 mA | Single Source: 13-825 mA Dual Source: 26-1650 mA | Same as the secondary predicate device. |
Page 38
| Hardware property | Subject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) |
|---|---|---|---|---|
| Tube focus | • 0.8 × 0.8 / 7° • 1.0 × 1.2 / 7° | • 0.8 × 0.8 / 7° • 1.0 × 1.2 / 7° | • 0.8 × 0.8 / 7° • 1.0 × 1.2 / 7° | Same as the primary and the secondary predicate device. |
| Tube Heat storage capacity | 2 x 7.0 MHU | 7.0 MHU | 2 x 7.0 MHU | Same as the primary and the secondary predicate device. |
| Tube Cooling rate (MHU/ min) | up to 1.7 | up to 1.7 | up to 1.7 | Same as the primary and the secondary predicate device. |
| Gantry Bore size | 70 | 70 | 70 | Same as the primary and the secondary predicate device |
| Gantry Scan FoV (cm) | 50 | 50 | 50 | Same as the primary and the secondary predicate device |
| Gantry Extended FoV (cm) | 70 | 70 | 70 | Same as the primary and the secondary predicate device |
| Gantry Rot. time (sec) | 0.33 (optional), 0.5, 1.0 | 0.33 (optional), 0.5, 1.0 | 0.33 (optional), 0.5, 1.0 | Same as the primary and the secondary predicate device |
| Gantry Tilt (degree) | n.a. | ± 30° | n.a. | Same as the secondary predicate device. |
| Maximum temporal resolution in ECG gated or triggered examination (ms) | 86 ms (mono-segment) 43 ms (bi-segment) | 165 ms (mono-segment) 82.5 ms (bi-segment) | 86 ms (mono-segment) 43 ms (bi-segment) | Same as the secondary predicate device |
Page 39
| Hardware property | Subject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) |
|---|---|---|---|---|
| Maximum scan speed at pitch (mm/s at pitch x) | 372 mm/s at pitch 3.2 | 175 mm/s at pitch 1.5 | 372 mm/s at pitch 3.2 | Same as the secondary predicate device |
| Patient table Type | Vario 2 (2000 mm) Vario 2.D (2000 mm) | Vario1 (1600 mm) Vario RT (1600 mm) Vario 2 (2000mm) | Vario 2 (2000 mm) Vario 2.D (2000 mm) | Same as the secondary predicate device. |
| Patient table Max. weight capacity (kg) | 307 | 227 307 | 307 | Same as the secondary predicate device. |
| Max. scan length topogram | 2080 mm with table extension | 1680 mm with table extension 2080 mm with table extension | 2080 mm with table extension | Same as the secondary predicate device. |
| Max. scan length Image acquisition | Max. 2000 mm with patient table extension | Max. 1600 mm with patient table extension Max. 2000 mm with patient table extension | Max. 2000 mm with patient table extension | Same as the secondary predicate device. |
Page 40
| Hardware property | Subject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) |
|---|---|---|---|---|
| Spectral filtration | Tin Filter | Combined Tin Filter / Split Filter | Tin Filter | Same as the secondary predicate device |
| FAST 3D Camera | option for patient positioning with FAST 3D Camera: • Ceiling mounted • Gantry mounted | option for patient positioning with FAST 3D Camera: • ceiling mounted • gantry mounted | option for patient positioning with FAST 3D Camera: • Ceiling mounted • Gantry mounted | Same as the primary and the secondary predicate device |
| High Power 70/ High Power 80 | up to 825 mA (@70 kV)/ up to 825 mA (@80 kV) | up to 825 mA (@70 kV)/ up to 825 mA (@80 kV) | up to 825 mA (@70 kV)/ up to 825 mA (@80 kV) | Same as the primary and the secondary predicate device |
| x-ray foot switch | Option to trigger hands-free scanning | Option to trigger hands-free scanning | Option to trigger hands-free scanning | Same as primary and the secondary predicate device |
| Table foot switch | Option for table patient movement | Option for table patient movement | Option for table patient movement | Same as the primary and the secondary predicate device |
| i-joystick | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Option for patient table movements, including an electrical connection for the tablet dock which allows charging the tablet when mounted. | Same as the primary and the secondary predicate device |
| Tablet dock | Option for mounting of the tablet on the patient table. | Option for mounting of the tablet on the patient table. | Option for mounting of the tablet on the patient table. | Same as the primary and the secondary predicate device |
Page 41
| Hardware property | Subject device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB20 | Primary predicate device SOMATOM go.Top¹⁰ with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device SOMATOM Pro.Pulse with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) |
|---|---|---|---|---|
| CARE 2D camera | one integrated into the gantry front and one mounted on the gantry back to monitor the patient during examination | one integrated into the gantry front and one mounted on the gantry back to monitor the patient during examination | one integrated into the gantry front and one mounted on the gantry back to monitor the patient during examination | Same as the primary and the secondary predicate device |
| CARE Moodlight | LED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Scan | LED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Radiation on | LED lights integrated on the gantry front cover indicating: • System startup • Scan countdown • Radiation on | Same as the primary and the secondary predicate device. In the subject device with syngo CT VB20, the status "Radiation on" has been renamed to "Scan". |
| Moodlight | • Gantry ring Moodlight: ambient light during examination. • CARE Funnel Light: ambient light which illuminates the gantry funnel | • Gantry ring Moodlight: ambient light during examination. | • Gantry ring Moodlight: ambient light during examination. • CARE Funnel Light: ambient light which illuminates the gantry funnel | Same as the secondary predicate device |
Page 42
Table 8: Comparison of software characteristics between the subject devices with software version syngo CT VB20 and the predicate devices with software version syngo CT VB10 (K233650 and K232206).
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| Operating System | Windows based SOMARIS/10 syngo CT VB20 Note: the short version syngo CT VB20 is also used as labeling information | Windows based SOMARIS/10 syngo CT VB20 Note: the short version syngo CT VB20 is also used as labeling information | Windows based SOMARIS/10 syngo CT VB10 Note: the short version syngo CT VB10 is also used as labeling information | Windows based SOMARIS/10 syngo CT VB10 Note: the short version syngo CT VB10 is also used as labeling information | Software version upgraded due to new functionalities. |
| Workplace | syngo Acquisition Workplace (ICS) named as "myExam Console" Image Reconstruction System (IRS) | syngo Acquisition Workplace (ICS) named as "myExam Console" Image Reconstruction System (IRS) | syngo Acquisition Workplace (ICS) named as "myExam Console" Image Reconstruction System (IRS) | syngo Acquisition Workplace (ICS) named as "myExam Console" Image Reconstruction System (IRS) | Same as the primary and secondary predicate devices |
Page 43
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| 2nd workplace option named as "myExam Satellite" with Remote Recon function | 2nd workplace option named as "myExam Satellite" with Remote Recon function | 2nd workplace option named as "myExam Satellite" with Remote Recon function | 2nd workplace option named as "myExam Satellite" with Remote Recon function | ||
| Standard system software | • syngo Examination • syngo Viewing • syngo Filming • syngo Archiving & Network | • syngo Examination • syngo Viewing • syngo Filming • syngo Archiving & Network | • syngo Examination • syngo Viewing • syngo Filming • syngo Archiving & Network | • syngo Examination • syngo Viewing • syngo Filming • syngo Archiving & Network | Same as the primary and secondary predicate devices |
| Detector firmware | Stellar detector firmware supported | Stellar detector firmware supported | Stellar detector firmware supported | Stellar detector firmware supported | Same as the primary and secondary predicate devices |
Page 44
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| Teamplay | Support of Teamplay protocols | Support of Teamplay protocols | Support of Teamplay protocols | Support of Teamplay protocols | Same as the primary and secondary predicate devices |
| Protocols | Support of: • Protocols for Radiation Therapy Planning support patient marking • Protocols that allow scanning with support of an external respiratory gating system | Support of: • Protocols for Radiation Therapy Planning support patient marking • Protocols supporting contrast bolus-triggered data acquisition | Support of: • Protocols for Radiation Therapy Planning support patient marking • Protocols that allow scanning with support of an external respiratory gating system | Support of: • Protocols for Radiation Therapy Planning support patient marking • Protocols supporting contrast bolus-triggered data acquisition | SOMATOM go. Platform: Same as the primary predicate devices SOMATOM Pro.Pulse: |
Page 45
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| • Protocol supporting contrast bolus-triggered data acquisition • Contrast media protocols (including coronary CTA) • Pediatric Protocols • Flex Dose Profile • Turbo Flash Spiral • Dual Energy acquisition (TwinBeam DE and TwinSpiral DE) • Dynamic imaging (Flex 4D Spiral) | • Contrast media protocols (including coronary CTA) • Pediatric Protocols • Flex Dose Profile • Dynamic imaging (Flex 4D Spiral) • interventional scan protocols (scan modes: i-sequence, i-spiral, i-Fluoro) • Protocols supporting Cardiac Scanning, Spectral imaging for child examination, Spectral imaging with high resolution | • Protocol supporting contrast bolus-triggered data acquisition • Contrast media protocols (including coronary CTA) • Pediatric Protocols • Flex Dose Profile • Turbo Flash Spiral • Dual Energy acquisition (TwinBeam DE and TwinSpiral DE) • Dynamic imaging (Flex 4D Spiral) | • Contrast media protocols (including coronary CTA) • Pediatric Protocols • Flex Dose Profile • Dynamic imaging (Flex 4D Spiral) • interventional scan protocols (scan modes: i-sequence, i-spiral, i-Fluoro) • Protocols supporting Cardiac Scanning, Spectral imaging for child examination, Spectral imaging with high resolution | Same as the secondary predicate device |
Page 46
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| • Protocols supporting CT Intervention (scan modes: 2D i-sequence, 3D i-sequence, i-spiral, i-Fluoro) • Protocols supporting Cardiac Scanning • Protocols for DirectBreathhold • Turbo Flash Spiral • various i-spiral and i-sequence scan protocols (applying different scanning parameters such as different slice thicknesses, kV settings or reconstructions kernels to support different clinical scenarios) and using features like iMAR, CARE Dose4D and CARE kV | • Protocols supporting CT Intervention (scan modes: i-sequence, i-spiral, i-Fluoro) • Protocols supporting Cardiac Scanning • Protocols for DirectBreathhold • Turbo Flash Spiral • various i-spiral and i-sequence scan protocols (applying different scanning parameters such as different slice thicknesses, kV settings or reconstructions kernels to support different clinical scenarios) and using features like iMAR, CARE Dose4D and CARE kV |
Page 47
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| • TwinkV (Dual Source Dual Energy acquisition) | • TwinkV (Dual Source Dual Energy acquisition) | ||||
| Advanced Reconstruction | Recon&GO: - Spectral Recon (Dual Energy Reconstruction including Virtual Unenhanced, Monoenergetic plus) - DE SPP (Spectral Post-Processing) - Inline Anatomical ranges (Parallel/Radial) incl. Virtual | Recon&GO: - Spectral Recon (Dual Energy Reconstruction including None, low (L3D) and high (H3D) energy images) - DE SPP (Spectral Post-Processing) - Inline Anatomical ranges (Parallel/Radial) incl. Virtual | Recon&GO: - Spectral Recon (Dual Energy Reconstruction including Virtual Unenhanced, Monoenergetic plus) - DE SPP (Spectral Post-Processing) - Inline Anatomical ranges (Parallel/Radial) incl. Virtual | Recon&GO: - Spectral Recon (Dual Energy Reconstruction including None, low (L3D) and high (H3D) energy images) - DE SPP (Spectral Post-Processing) - Inline Anatomical ranges (Parallel/Radial) incl. Virtual | SOMATOM go. Platform: Same as the primary predicate devices SOMATOM Pro.Pulse: Same as the secondary predicate device |
Page 48
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| Unenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removal | Unenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removal | Unenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removal | Unenhanced, Monoenergetic plus - Inline Spine and Rib Ranges - Inline table and bone removal | ||
| Image viewing | CT View&GO offers: - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI) | CT View&GO offers: - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI) | CT View&GO offers: - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI) | CT View&GO offers: - 2D and 3D (MPR, VRT, MIP and minIP) - Evaluation tools, Filming, Printing - Interactive Spectral Imaging (ISI) | Same as the primary and the secondary predicate devices |
Page 49
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| - Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average | - Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average | - Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average | - Basic visualization tools: Endo View - Basic manipulation tools: DE ROI, ROI HU, Average | ||
| Software interface | • Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical devices in their own right. | • Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical devices in their own right. | • Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical devices in their own right. | • Recon&GO Inline Results Software interface to post-processing algorithms which are unmodified when loaded onto the CT scanners and 510(k) cleared as medical devices in their own right. | In software version syngo CT VB20, the subject devices introduce plug-in interface for the stand alone software application myAblation Guide. |
Page 50
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| • software interfaces for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. Note: The clearance of standalone Advanced Visualization Application software is mandatory precondition. These advanced visualization tools are designed to support the technician & physician in | • software interfaces for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. Note: The clearance of standalone Advanced Visualization Application software is mandatory precondition. These advanced visualization tools are designed to support the technician & physician in | • software interfaces for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. Note: The clearance of standalone Advanced Visualization Application software is mandatory precondition. These advanced visualization tools are designed to support the technician & physician in | • software interfaces for post-processing functionalities to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. Note: The clearance of standalone Advanced Visualization Application software is mandatory precondition. These advanced visualization tools are designed to support the technician & physician in |
Page 51
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners. • Plug-in interface for myAblation Guide | the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners. • Plug-in interface for myAblation Guide | the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners. | the qualitative and quantitative measurement & analysis of clinical data acquired and reconstructed by Computed Tomography scanners. | ||
| Cybersecurity | IT Hardening | IT Hardening | IT Hardening | IT Hardening | Same as the primary and the secondary predicate devices |
| HD FoV | HD FoV 4.0 | HD FoV 4.0 | HD FoV 4.0 | HD FoV 4.0 | The subject devices |
Page 52
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| HD FoV 5.0¹¹ | SOMATOM go. Sim and SOMATOM go.Open Pro with syngo CT VB20 support a new generation of extended field of view reconstruction algorithm: HD FoV 5.0. |
¹¹ HD FoV 5.0: In software version syngo CT VB20, this version of HD FoV is available for SOMATOM go.Sim and SOMATOM go.Open Pro only.
Page 53
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| Standard technologies | • FAST technologies • CARE technologies • GO technologies | • FAST technologies • CARE technologies • GO technologies | • FAST technologies • CARE technologies • GO technologies | • FAST technologies • CARE technologies • GO technologies | Same as the primary and the secondary predicate devices |
| Iterative Reconstruction Methods | ADMIRE iMAR SAFIRE | ADMIRE iMAR | ADMIRE iMAR SAFIRE | ADMIRE iMAR | SOMATOM go. Platform: Same as the primary predicate devices SOMATOM Pro.Pulse: |
Page 54
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| Same as the secondary | |||||
| Matrix sizes | 256 x 256 pixels 512 x 512 pixels 768 x 768 pixels 1024 x 1024 pixels (Precision Matrix) | 256 x 256 pixels 512 x 512 pixels 768 x 768 pixels 1024 x 1024 pixels (Precision Matrix) | 256 x 256 pixels 512 x 512 pixels 768 x 768 pixels 1024 x 1024 pixels (Precision Matrix) | 256 x 256 pixels 512 x 512 pixels 768 x 768 pixels 1024 x 1024 pixels (Precision Matrix) | Same as the primary and the secondary predicate device |
| DirectDensity | including relative electron density, relative mass density, and stopping power ratio | including relative electron density, relative mass density, and stopping power ratio | including relative electron density, relative mass density | including relative electron density and relative mass density | Compared to the predicate devices, the subject devices with syngo CT VB20 support a |
Page 55
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| modified DirectDensity functionality. Same as the secondary predicate | |||||
| Matrix sizes | 256 x 256 pixels | 256 x 256 pixels | 256 x 256 pixels | 256 x 256 pixels | Same as the primary and the secondary predicate device |
| 512 x 512 pixels | 512 x 512 pixels | 512 x 512 pixels | 512 x 512 pixels | ||
| 768 x 768 pixels | 768 x 768 pixels | 768 x 768 pixels | 768 x 768 pixels | ||
| 1024 x 1024 pixels (Precision Matrix) | 1024 x 1024 pixels (Precision Matrix) | 1024 x 1024 pixels (Precision Matrix) | 1024 x 1024 pixels (Precision Matrix) | ||
| DirectDensity | including relative electron density, relative mass density, and stopping power ratio | including relative electron density, relative mass density, and stopping power ratio | including relative electron density, relative mass density | including relative electron density and relative mass density | Compared to the predicate devices, the subject devices with syngo CT VB20 support a |
Page 56
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| Stack artifact reduced reconstruction | Provides images with reduced stack artifacts after the reconstruction: ZeeFree: StackRecon types in cardiac-related examinations ZeeFree RT¹²: StackRecon types in respiratory-related examinations | Provides images with reduced stack artifacts after the reconstruction: ZeeFree: StackRecon types in cardiac-related examinations | Provides images with reduced stack artifacts after the reconstruction: ZeeFree: StackRecon types in cardiac-related examinations | Provides images with reduced stack artifacts after the reconstruction: ZeeFree: StackRecon types in cardiac-related examinations | ZeeFree RT is a new functionality introduced with syngo CT VB20 for the subject device SOMATOM go.Open Pro. |
¹² ZeeFree RT: In software version syngo CT VB20, ZeeFree RT is available for SOMATOM go.Open Pro only.
Page 57
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| Oncology exchange | transfer of prescription information from the ARIA Oncology Information System (OIS) to the CT scanner | transfer of prescription information from the ARIA Oncology Information System (OIS) to the CT scanner | n.a. | n.a. | New functionality introduced with syngo CT VB20. |
| DirectSetup Notes | patient and accessory setup documentation in the Scan&GO workflow | patient and accessory setup documentation in the Scan&GO workflow | n.a. | n.a. | The subject devices SOMATOM go. Platform CT scanner systems are enabled to support DirectSetup Notes. |
Page 58
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| DirectSetup Notes was cleared in K233650 for SOMATOM go.Sim and SOMATOM go.Open Pro. | |||||
| FAST Integrated Workflow (FAST 3D Camera) | FAST integrated workflow: • FAST Range • FAST Isocentering • FAST Direction • Collision Indication | FAST integrated workflow: • FAST Range • FAST Isocentering • FAST Direction • Collision Indication | FAST integrated workflow: • FAST Range • FAST Isocentering • FAST Direction | FAST integrated workflow: • FAST Range • FAST Isocentering • FAST Direction | Compared to the predicate devices, the FAST 3D Camera of the subject devices supports in syngo CT VB20 a Collision indication and |
Page 59
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| • Centerline/ Grid Overlay | • Centerline/ Grid Overlay | Centerline/Grid Overlay function. DirectSetup Notes was cleared in K233650 for SOMATOM go.Sim and SOMATOM go.Open Pro. | |||
| FAST Planning | detects globally used scan and recon ranges from the topogram image additional body regions: - Head/Neck (Vascular), Lateral/APPA - Vascular – Runoff Knee - Neck and chest - Neck and Lung | detects globally used scan and recon ranges from the topogram image additional body regions: - Head/Neck (Vascular), Lateral/APPA - Vascular – Runoff Knee - Neck and chest - Neck and Lung | detects globally used scan and recon ranges from the topogram image | detects globally used scan and recon ranges from the topogram image | As the primary and secondary devices, the subject devices with syngo VB20 support FAST Planning functionality with wide range of body regions. In syngo CT VB20, FAST Planning has been extended |
Page 60
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| - Extremities – Both knees | - Extremities – Both knees | with additional body regions. | |||
| Eco Power Mode | reduce the energy consumption during idle times.¹³ | reduce the energy consumption during idle times. | n.a. | n.a. | New feature introduced with syngo CT VB20 of the subject devices. |
| myExam Contrast | exchange of contrast injection parameters, scan and patient parameters | exchange of contrast injection parameters, scan and patient parameters | n.a. | n.a. | New feature introduced with syngo CT VB20. |
¹³ SOMATOM go.Now and SOMATOM go.Up do not support Eco Power Mode.
Page 61
| Software property | Subject device with SOMARIS/10 syngo CT VB20 | Primary predicate device with SOMARIS/10 syngo CT VB10 (K233650) | Secondary predicate device with SOMARIS/10 syngo CT VB10 (K232206) | Assessment of the substantial equivalency (SE) | |
|---|---|---|---|---|---|
| SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | SOMATOM go.Now SOMATOM go.Up SOMATOM go All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro | SOMATOM Pro.Pulse | ||
| Breathhold technique | Respiratory Motion Management support breath hold triggered spiral scans with manual breath hold triggered examinations. DirectBreathold: externally triggered workflow to start a spiral scan by receiving a trigger signal from an external gating device. | n.a. | Respiratory Motion Management support breath hold triggered spiral scans with manual breath hold triggered examinations. DirectBreathold: externally triggered workflow to start a spiral scan by receiving a trigger signal from an external gating device. | n.a. | SOMATOM go. Platform: Same as the primary predicate devices SOMATOM Pro.Pulse does not support DirectBreathhold |
Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation is completed. Test results show that the subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse with syngo CT VB20 are comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the primary predicate devices.
Page 62
VIII. Performance Data
Non-Clinical Testing
Non-clinical testing, (integration and functional) including phantom tests were conducted for the SOMATOM go. Platform and SOMATOM Pro.Pulse CT scanner systems with syngo CT VB20 during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.
The general purpose of each test is to verify and validate the functionality of the subject device modifications.
Testing will cover all related subsystems that contribute to the device modifications. Test levels are defined. For each test level several test activities are performed. The test specification and acceptance criteria are related to the corresponding requirements. Various test activities are performed to specific modifications on different test levels to ensure safe and effective integration in the system. Three test levels are defined:
System Validation test:
- Acceptance test (workflow and user manual test)
- Legal and Regulatory test
System Verification test:
- System Integration Test (functional)
- Functionality verification
- Image Quality (IQ) Evaluation
Tests are conducted for all software components developed in product development and for the complete product itself. Several activities are considered for this process, including creation of test specifications that relate to software/hardware requirements including tests to address risk mitigations that are identified, documented, and traced by hazard keys.
Bench testing:
Additional evaluation tests are performed as bench tests to support the new device or device modification on Non-Clinical Performance Testing as listed in Table 9 below.
Table 9: Non-clinical performance testing (bench testing).
| Feature/Non-clinical supportive testing | Bench Testing performed |
|---|---|
| FAST 3D Camera/ FAST Integrated Workflow | The FAST 3D camera sub-features FAST Isocentering, FAST Range, and FAST Direction have been optimized using additional data from adults and adolescence patients. The bench test evaluates and compares the accuracy of the three sub-features in software version syngo CT VB20 to the accuracy of the predicate devices with syngo CT VB10. The objectives of the bench tests are to demonstrate that the FAST 3D camera achieves comparable or better results for both, adults and adolescents, as the predicate device for adults. |
Page 63
| Feature/Non-clinical supportive testing | Bench Testing performed |
|---|---|
| Overall, the subject devices with syngo CT VB20 delivers comparable or improved accuracy to the predicate devices with syngo CT VB10 predicate device for adults and extends the support to adolescents. | |
| FAST Planning | The purpose of the test is to provide a clear reporting on the applied algorithm, its product development, validation, and verification on patient data, which enable the claims. Objective of the test is to assess the fraction (percentage) of ranges calculated by the FAST Planning algorithm that are correct and can be applied without change. Additionally, calculation time was measured to check whether it meets interactive requirements. The test results show that the editing actions for the scanner technician can be reduced to a minimum and that the calculation time is fast enough for interactive speed during scanning. For more than 90% of the ranges no editing action was necessary to cover standard ranges. For more than 95%, the speed of the algorithm was sufficient. |
| HD FoV 5.0 | The bench test contains a detailed description and evaluation of the new HD FoV 5. 0 algorithm for extended field of view reconstruction. Results obtained with the new HD FoV 5. 0 algorithm are compared with its predecessor, the HD FoV 4. 0 algorithm, based on physical and anthropomorphic phantoms. In addition to the bench test, the performance of the algorithm was evaluated by board-approved radio-oncologists and medical physicists by means of retrospective blinded rater study. This comparison is conducted to demonstrate that the HD FoV 5. 0 algorithm is as safe and effective as the HD FoV 4. 0 algorithm. |
| Flex 4D Spiral | The performed bench test report describes the technical background of Flex 4D Spiral and its functionalities with SOMATOM CT scanners, demonstrate the proper function of those, and assess the image quality of Flex 4D Spiral. |
| ZeeFree RT | The bench test evaluates the performance of the ZeeFree RT reconstruction. The objectives of the tests are to demonstrate that compared to the Standard reconstruction, ZeeFree RT • introduces no relevant errors in terms of CT values and noise levels measured in a homogeneous water phantom, and • introduces no relevant errors in terms of CT values measured in a phantom with tissue-equivalent inserts, even in the presence of metals and in combination with the iMAR algorithm, and • introduces no relevant geometrical distortions in a static torso phantom, and |
Page 64
| Feature/Non-clinical supportive testing | Bench Testing performed |
|---|---|
| • introduces no relevant deteriorations of the position or shape of a dynamic thorax phantom when moving a spherical shape according to regular, irregular, and patient breathing motion. In addition to the bench tests, the performance of the algorithm was evaluated by board-approved radio-oncologists and medical physicists by means of retrospective blinded rater studies of respiratory 4D CT examinations performed at two institutions. The bench test results show that the ZeeFree RT reconstruction • can successfully be applied to 4D respiratory-gated sequence images (Direct i4D), and • enables the optional reconstruction of stack artefact corrected images, which reduce the strength of misalignment artefacts, if such stack alignment artefacts are identified in non-corrected standard images, and • does not introduce relevant new artefacts, which were previously not present in the non-corrected standard reconstruction, and • does not affect CT values and noise levels in a homogenous water phantom outside of stack-transition areas compared to the non-corrected standard reconstruction, and • can be successfully applied in combination with metal artifact correction (iMAR) and does not introduce new artifacts, which were previously not present in the non-corrected standard reconstruction, even in presence of metals, and • can be successfully applied to phantom data if derived from a suitable motion phantom demonstrating its correct technical function on the tested device, and • is independent from the physical detector width of the acquired data. By design, input images for the correction part of the algorithm are independent of kernel, slice thickness and increment since those parameters are fixed. | |
| DirectDensity | The test results show for the iBHC variants Artificial120, eDDensity, mDDensity and StoppingPowerRatio a reduced dependence on tube voltage and filtration compared to the corresponding quantitative kernel (Qr) with iBHC Bone for non-water-like tissues, such as adipose and bone. Furthermore, the iBHC variants Artificial120, eDDensity, mDDensity and StoppingPowerRatio generate image value closely aligned with the respective material properties. In conclusion, the feature DirectDensity at any kV has been validated for the release Som/X VB20 on all supported SOMATOM CT scanner models |
Page 65
A list of recognized consensus standards and general use consensus standards considered for the subject devices is provided as Table 10 and Table 11 below.
Table 10: Recognized Consensus Standards.
| Date of Entry | Recognition Number | Standard Developing Organization | Standard Designation Number and Date | Title of Standard |
|---|---|---|---|---|
| 12/18/2023 | 12-352 | NEMA | PS 3.1 - 3.20 2023e | Digital Imaging and Communications in Medicine (DICOM) Set |
| 07/06/2020 | 12-325 | NEMA | XR 25-2019 | Computed Tomography Dose Check |
| 07/06/2020 | 12-330 | NEMA | XR 28-2018 | Supplemental Requirements for User Information and System Function Related to Dose in CT |
| 12/23/2019 | 12-328 | IEC | 61223-3-5 Edition 2.0 2019-09 | Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests and constancy tests - Imaging performance of computed tomography X-ray equipment [Including: Technical Corrigendum 1 (2006)] |
| 03/14/2011 | 12-226 | IEC | 61223-2-6 Second Edition 2006-11 | Evaluation and routine testing in medical imaging departments - Part 2-6: Constancy tests - Imaging performance of computed tomography X-ray equipment |
| 06/07/2021 | 12-336 | IEC | 60601-1-3 Edition 2.2 2021-01 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment |
| 06/27/2016 | 12-302 | IEC | 60601-2-44 Edition 3.2: 2016 | Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and |
Page 66
| Date of Entry | Recognition Number | Standard Developing Organization | Standard Designation Number and Date | Title of Standard |
|---|---|---|---|---|
| essential performance of x-ray equipment for computed tomography | ||||
| 12/23/2019 | 5-125 | ANSI AAMI ISO | 14971: 2019 | Medical devices - Applications of risk management to medical devices |
| ISO 14971 Third Edition 2019-12 | Medical devices - Application of risk management to medical devices | |||
| 01/14/2019 | 13-79 | ANSI AAMI IEC | 62304:2006/A1:2016 | Medical device software - Software life cycle processes [Including Amendment 1 (2016)] |
| IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION | Medical device software - Software life cycle processes | |||
| 05/30/2022 | 19-46 | ANSI AAMI | ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)] |
| 12/21/2020 | 19-36 | ANSI AAMI IEC | 60601-1-2:2014 [Including AMD 1:2021] | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
Page 67
| Date of Entry | Recognition Number | Standard Developing Organization | Standard Designation Number and Date | Title of Standard |
|---|---|---|---|---|
| 07/06/2020 | 5-129 | ANSI AAMI IEC | 62366-1:2015+AMD1:2020 (Consolidated Text) | Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1 |
| IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION | Medical devices - Part 1: Application of usability engineering to medical devices | |||
| 07/09/2014 | 12-273 | IEC | 60825-1 Edition 2.0 2007-03 | Safety of laser products - Part 1: Equipment classification, and requirements |
| 12/21/2020 | 5-132 | IEC | 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
| 12/23/2019 | 12-309 | IEC | 60601-2-28 Edition 3.0 2017-06 | Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis |
| 12/20/2021 | 12-341 | IEC | 62563-1 Edition 1.2 2021-07 CONSOLIDATED VERSION | Medical electrical equipment - Medical image display systems - Part 1: Evaluation methods |
Table 11: General Use Consensus Standards.
| Standard Developing Organization | Standard Designation Number and Date | Title of Standard | How was Standard Used |
|---|---|---|---|
| IEC | 60601-1:2005+A1:2012+A2:2020 | Medical electrical equipment - part 1: general requirements for basic safety and essential performance | ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & |
Page 68
| Standard Developing Organization | Standard Designation Number and Date | Title of Standard | How was Standard Used |
|---|---|---|---|
| A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | |||
| IEC/ISO | 17050-1 | Conformity Assessment – Supplier's declaration of conformity – Part 1: General requirements | Declaration of conformance to FDA recognized consensus standards. |
| IEC/ISO | 17050-2 | Conformity assessment – Supplier's declaration of conformity – Part 2: Supporting documentation. | General consensus standards not currently recognized by FDA. |
A list of applicable guidance documents considered for this submission is provided as Table 12 below.
Table 12: FDA Guidance Document and Effective Date.
| FDA Guidance Document | Issue date |
|---|---|
| User Fees and Refunds for Premarket Notification Submissions (510(k)s): Guidance for Industry and Food and Drug Administration Staff | 10/05/2022 |
| Refuse to Accept Policy for 510(k)s: Guidance for Industry and Food and Drug Administration Staff | 04/21/2022 |
| Electronic Submission Template for Medical Device 510(k) Submissions | 10/2/2023 |
| Deciding When to Submit a 510(k) for a Change to an Existing Device | 10/25/2017 |
| The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] | 07/28/2014 |
| Content of Premarket Submissions for Software Contained in Medical Devices | 06/14/2023 |
| Off-The-Shelf Software Use in Medical Devices | 09/27/2019 |
| Applying Human Factors and Usability Engineering to Medical Devices | 02/03/2016 |
| Pediatric Information for X-ray Imaging Device Premarket Notifications | 11/28/2017 |
| Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions | 09/27/2023 |
| Electromagnetic Compatibility (EMC) of Medical Devices | 06/06/2022 |
| Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices | 09/06/2017 |
Page 69
| FDA Guidance Document | Issue date |
|---|---|
| Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices | 09/14/2018 |
Verification and Validation
The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claims of substantial equivalence.
Cybersecurity
Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.
Wireless Coexistence Testing
Additionally, Siemens conforms to the requirements for Radio Frequency Wireless Technology as defined in FDA guidance document "Radio Frequency Wireless Technology in Medical Devices, Guidance for Industry and Food and Drug Administration Staff" issued on August 14, 2013 by adhering to the EMC and risk based verification and validation requirements in design, testing, and labeling of the wireless remote control components of the subject devices.
The Radio Frequency Wireless Technology of the optional Remote Scan Control and supporting Control Device iPad for Scan&GO complies to 47 CFR part 15 subpart c – Intentional Radiators. All Radio device labels will show an FCC ID code to show compliance. Shielding requirement applicable to the SOMATOM go. Platform and SOMATOM Pro.Pulse and respective Scatter Radiation diagrams for typical room installations are provided in the User Documentation and Planning Guide of the intended Scanners in accordance with IEC60601-2-44.
Siemens has considered several measures to address wireless coexistence by design to ensure the safe operation of the wireless components in combination with the applicable system supported functionality. Wireless technology in the system setup to perform a task in a given shared environment where other systems have an ability to perform their tasks and may or may not be using the same set of rules has been considered. According to FDA guidance 'Radio Frequency Wireless Technology in Medical Devices" Siemens has addressed the safety, effectiveness, and high likelihood of coexistence with other devices of this technology in our product design by our Risk Management Process, Failure Mode and Effects Analysis (FMEA) Process, and Requirement Engineering Process. As part of the risk management process, hazardous situations associated with the Scan&GO and its connection to the host system via Wi-Fi were addressed as part of the Risk Management process.
Testing for co-existence considered for following scenarios:
- Co-Channel Testing
- Adjacent Channel Testing
Page 70
- RF Interference Testing
- Separation Distance/Location Testing
Scan&GO is designed to allow dynamic frequency selection and transmission power control by default in accordance with IEEE 802.11h. Adjacent channel testing is addressed by the fact that Scan&GO does not support shared medium access to Siemens Wi-Fi network. RF interference was tested by successfully ensuring that wireless communications were actively transmitting in situations where possible interference may exist. Recommended distance and router locations requirements are documented in the user documentation.
Summary
The features described in this premarket notification are supported with verification and validation testing, dosimetry and imaging performance, and analysis of phantom images to assess device and feature performance during product development. The risk analysis was completed, and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria. Verification and validation testing of the device was found acceptable to support the claim of substantial equivalence.
General Safety and Effectiveness Concerns
The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a system related risk analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing according to the risk management process. In order to minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
IX. Conclusions
Verification and validation and phantom testing were performed. The non-clinical data supports the safety of the device and the hardware and software verification and validation demonstrates that the subject devices SOMATOM go. Platform and SOMATOM Pro.Pulse CT scanner systems with software version syngo CT VB20 perform as intended in the specified use conditions. The data included in this submission demonstrates that the SOMATOM go. Platform and SOMATOM Pro.Pulse CT scanner systems with software version syngo CT VB20 with the described modifications performs comparably to the predicate devices currently marketed for the same intended use. The conclusion drown from the non-clinical tests demonstrates that the subject devices are as safe, as effective, and perform as well as or better than the predicate devices.
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.