K Number
K203488
Device Name
emField
Date Cleared
2021-02-19

(84 days)

Product Code
Regulation Number
890.5850
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The emField is indicated to be used for:

*Relaxation of muscle spasms;

*Prevention or retardation of disuse atrophy;

*Increasing local blood circulation;

*Muscle re-education;

*Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; and

*Maintaining or increasing range of motion.

Device Description

emField is a non-invasive therapeutic device. The device produces a magnetic field that interacts with the tissues of the human body. By muscle stimulation, the emField is indicated for bulk muscle excitation in the legs or arms for rehabilitative purposes.

The device housing protects the patient from electrical shock and mechanical injuries. The device is mobile standalone equipment with four wheels. Two applicators are available for therapy: a large and a small one. The main body of emField is used to control function of magnetic stimulation. It is operated with parameters such as Frequency, Time and Intensity. These parameters can be controlled by the user on LCD and with the help of a rotary knob at the user control panel.

AI/ML Overview

The provided document is a 510(k) Pre-market Notification from the FDA regarding the Zimmer MedizinSysteme GmbH emField device. It outlines the device's indications for use, technological characteristics, and performance data to demonstrate substantial equivalence to a predicate device.

However, the document does not contain information about acceptance criteria for a study proving device performance in the context of an AI/ML medical device. The emField device is a "Powered Muscle Stimulator" that generates a magnetic field for therapeutic purposes, which falls under conventional medical devices, not AI/ML-driven devices.

Therefore, I cannot provide the requested information about acceptance criteria, study details (sample size, data provenance, expert ground truth, adjudication, MRMC study, standalone performance, training set details) for an AI/ML device, because this document is not about such a device.

The "Performance data" section (starting on page 7) refers to the device's compliance with voluntary standards (like IEC 60601-1 for electrical safety, IEC 60601-1-2 for electromagnetic compatibility, and IEC 60601-2-10 for nerve and muscle stimulators) and preclinical testing for magnetic field output and tissue heating. These are standard performance tests for electromechanical medical devices, not AI/ML performance metrics.

In summary, the provided document does not contain the information needed to answer the prompt, as it concerns a non-AI/ML medical device.

{0}------------------------------------------------

February 19, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zimmer MedizinSysteme GmbH % Scott Blood Principal Regulatory Consultant Quality and Regulatory Services 151 Gleasondale Road Stow, Massachusetts 01775

Re: K203488

Trade/Device Name: emField Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF Dated: November 24, 2020 Received: November 27, 2020

Dear Scott Blood:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Patrick Antkowiak, PhD Assistant Director (acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203488

Device Name emField

Indications for Use (Describe) The emField is indicated to be used for:

*Relaxation of muscle spasms;

*Prevention or retardation of disuse atrophy;

*Increasing local blood circulation;

*Muscle re-education;

*Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; and

*Maintaining or increasing range of motion.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, gray letters. Below that, in smaller gray letters, is the word "MedizinSysteme."

510(k) Summary emField

1. Basic Information-Submitter:

510(k) Owner:Zimmer MedizinSysteme GmbHJunkersstrasse 989231 Neu-UlmGermanyEstablishment Registration: 8010720
Official Contact:Mrs. Ute HaussManager Regulatory AffairsPhone: +49-731-9761-216Fax: +49-731-9761-118E-mail: u.hauss@zimmer.de
Date Summary Prepared:November 24, 2020
Device Name:
Trade Name:emField
Common Name:Powered Muscle Stimulator
Classification Name:Stimulator, Muscle, Powered
Regulation Number:21 CFR 890.5850
Product Code:IPF
Classification:Class II
Predciate Device:HPM-6000 – K160992
Company Name:BTL Industries, Inc.
Reference Device:emFieldPro – K182963
  1. Device Description:

Company Name:

emField is a non-invasive therapeutic device. The device produces a magnetic field that interacts with the tissues of the human body. By muscle stimulation, the emField is indicated for bulk muscle excitation in the legs or arms for rehabilitative purposes.

Zimmer MedizinSysteme GmbH

The device housing protects the patient from electrical shock and mechanical injuries. The device is mobile standalone equipment with four wheels. Two applicators are available for therapy: a large and a small one. The main body of emField is used to control function of magnetic stimulation. It is operated with parameters such as

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, sans-serif font. The logo is in gray color.

Frequency, Time and Intensity. These parameters can be controlled by the user on LCD and with the help of a rotary knob at the user control panel.

Indications for Use Statement:

emField is indicated to be used for:

  • Relaxation of muscle spasms; ●
  • . Prevention or retardation of disuse atrophy;
  • Increasing local blood circulation;
  • Muscle re-education;
  • Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; . and
  • Maintaining or increasing range of motion.
ATTRIBUTESUBJECT DEVICEZimmerMedizinSysteme GmbHemFieldThis SubmissionPREDICATE DEVICEBTL Industries, Inc.HPM-6000K160992REFERENCE DEVICEZimmerMedizinSysteme GmbHemFieldProK182963
Intended UsePhysical Medicine21 CFR 890.5850IPF - Stimulator, Muscle, PoweredThe emField is indicated to be used for:• Relaxation of muscle spasms;• Prevention or retardation of disuse atrophy;• Increasing local blood circulation;• Muscle re-education;• Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; and• Maintaining or increasing range of motion.Physical Medicine21 CFR 890.5850IPF - Stimulator, Muscle, PoweredThe HPM-6000 is indicated to be used for:• Relaxation of muscle spasms;• Prevention or retardation of disuse atrophy;• Increasing local blood circulation;• Muscle re-education;• Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis; and• Maintaining or increasing range of motion.Physical Medicine21 CFR 890.5850NGX- Stimulator, Muscle, Powered, For Muscle ConditioningThe emFieldPro is indicated to be used for:• Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.• Strengthening, Toning and Firming of buttocks and thighs.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo is in black and white. The word "zimmer" is in a bold, sans-serif font, and the words "MedizinSysteme" are in a smaller, sans-serif font below the word "zimmer."

The Indications for Use statement for emField is identical to the predicate device. The Indications for Use is not the same for the reference device, however, the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for conditioning and stimulation of muscles.

    1. Technological Characteristics:
      The emField device has the equivalent technology and principles of operation as its predicate device. The emField device and its predicate are comprised of a system console and applicators. The system console consists of the electromagnetic field generators, computer, and the touch-screen control panel. The emField is equipment that generates a magnetic field by applying a strong current to an applicator. The technical characteristics of the emField and its predicate device, are equivalent.
TechnologicalCharacteristicSUBJECT DEVICEZimmerMedizinSysteme GmbHemFieldThis SubmissionPREDICATE DEVCIEBTL Industries, Inc.HPM-6000K160992REFERENCE DEVICEZimmerMedizinSysteme GmbHemFieldProK182963
PrimaryFunctionMuscle stimulationMuscle stimulationMuscle stimulation
Principle ofActionInitiating action potentialof nerves results inmuscle contractionInitiating action potentialof nerves results inmuscle contractionInitiating action potentialof nerves results inmuscle contraction
ElectricalProtectionClass I, BFClass II, BFClass I, BF
User InterfaceTouch screenTouch screenTouch screen
Touch screensize7"8.4"7"
Type of EnergyMagnetic fieldMagnetic fieldMagnetic field
Number ofApplicators222
Number ofMagnetic Coilsin theApplicator111
Magnetic FieldIntensityLarge applicator:0.5 – 1.5 T +/-20%Small applicator:299-1 applicator:0.5 – 1.8 T299-2 applicator:Large applicator:0.5 – 1.5 T +/-20%Small applicator:
0.5 – 2.0 T +/-20%0.7 – 2.5 T0.5 – 2.0 T +/-20%
Total InducedCurrent inTissue (mA)251unknown251
Type ofOperationContinuousContinuousContinuous

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, gray letters. Underneath the word "zimmer" is the phrase "MedizinSysteme" in a smaller, lighter gray font.

TechnologicalCharacteristicSUBJECT DEVICEZimmerMedizinSysteme GmbHemFieldThis SubmissionPREDICATE DEVCIEBTL Industries, Inc.HPM-6000K160992REFERENCE DEVICEZimmerMedizinSysteme GmbHemFieldProK182963
PulseRepetition Rate1 - 150 Hz1 - 150 Hz1 - 150 Hz
Pulse DurationLarge applicator:400 µs +/- 20%Small applicator:250 µs +/- 20%280 +/- 20% µsLarge applicator:400 µs +/- 20%Small applicator:250 µs +/- 20%
PulseAmplitude0 - 100 %0 - 100 %0 - 100 %
Selection ofparameters(Intensity.Time)YesunknownYes
Therapy TimeUp to 60 minUp to 60 minUp to 60 min
Shape ofStimulationPulseSymmetrical BiphasicSine WaveSine, biphasicSymmetrical BiphasicSine Wave
Energy Source100-240 V AC, 50-60Hz, max 12.5 A100-240 V AC, 50-60Hz, max. 14 A100-240 V AC, 50-60Hz, max 12.5 A
SystemDimensions(WxHxD)542 x 501 x 993 mm500 × 970 × 580 mm(20×38×23 in)542 x 501 x 993 mm
OperatingAmbientTemperature10° C to 30° C10° C to 30° C10° C to 30° C
OperatingAmbientHumidity30-85%30-75%30-85%
WeightApprox. 60 kg33 kgApprox. 60 kg
EnvironmentalSpecificationsFor indoor use onlyFor indoor use onlyFor indoor use only

There are no technological differences between the emField device and the reference device, as they are the same device. There are few differences between the technological differences of the subject device and those of the predicate device. Those differences have been discussed and do not affect device safety or performance. The subject device has all features of the predicate device. emField does not raise any new types of safety or effectiveness questions.

The Zimmer MedizinSysteme GmbH emField has the same technological characteristics as the predicate device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, sans-serif font. Below that, in a smaller font, are the words "MedizinSysteme".

6. Performance data

The emField device has been investigated and tested against and complies with the following voluntary standards:

StandardsStandardsOrganizationStandards Title
60601-1IECMedical electrical equipment – Part 1: Generalrequirements for basic safety and essential performance
60601-1-2IECMedical electrical equipment – Part 1-2: Generalrequirements for basic safety and essential performance –Collateral standard: Electromagnetic disturbances –Requirements and tests
60601-1-6IECMedical electrical equipment – Part 1-6: Generalrequirements for basic safety and essential performance -Collateral standard: Usability
60601-2-10IECMedical electrical equipment – Part 2-10: Particularrequirements for the Basic Safety and EssentialPerformance of Nerve and Muscle Stimulators
62366-1IECMedical devices – Application of usability engineering tomedical devices
62304IECMedical devices software –software life cycle processes
14971ISOMedical devices – Application of risk management tomedical devices

The following table shows a comparison of the performance testing in comparison to the predicate devices:

StandardsSUBJECT DEVICEPREDCIATE DEVICEREFERENCE DEVICE
ZimmerMedizinSytemeGmbHBTL Industries,Inc.ZimmerMedizinSyteme GmbH
emFieldThis SubmissionHPM-6000K160992emFieldProK182963
IEC 60601-1XXX
IEC 60601-1-2XXX
IEC 60601-2-10XXX

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z" and "i". Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, sans-serif font. The logo is simple and modern, and the use of a bold font makes it easily recognizable.

SUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE
StandardsZimmerMedizinSytemeGmbHemFieldThis SubmissionBTL Industries,Inc.HPM-6000K160992ZimmerMedizinSyteme GmbHemFieldProK182963
ISO 10993-1ISO 10993-5ISO 10993-10Not applicable(no patient-contactingmaterials)XNot applicable(no patient-contactingmaterials)

According to this comparison table all required performance tests were conducted and show substantial equivalence with the predicate devices.

Preclinical Testing Results

The following tests were performed on the subject device in addition to the testing listed above:

  • Magnetic Field testing
  • Tissue Heating study
  • Usability testing

The testing above confirmed that the large and small applicator each operate within the magnetic field intensity specifications for each applicator and that the tissue being treated by the device does not present an appreciable rise in temperature at maximum intensity to cause a risk to the patient.

Testing has been performed and all components, subassemblies and/or full devices and systems have met the required specifications for the completed tests.

  1. 510(k) Summary:

Zimmer MedizinSysteme GmbH has demonstrated that the emField device is substantially equivalent to the predicate device and reference device.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).