K Number
K200392

Validate with FDA (Live)

Device Name
REXSIN
Manufacturer
Date Cleared
2020-07-07

(140 days)

Product Code
Regulation Number
878.4493
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.

Device Description

Synthetic Absorbable PGA Suture With or Without Needle (REXSIN) are produced and provided by SM ENG Co., Ltd. REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue. Suture of REXSIN is Trisorb manufactured by SAMY ANG Biopharmaceuticals Corporation. SM Eng is receiving bulk Trisorb from Samyang, and then go through cutting, adhesion of need and thread, winding and sterilization process according to SM Eng procedure, which becomes REXSIN suture. REXSIN is a sterilized medical device composed with absorbable, braided, coated suture with needle, stainless steel SUS 304. REXSIN Suture is a synthetic absorbable, braided, coated suture composed of a Polyglycolic acid (PGA). The suture is available undyed (natural) or dyed (D&C Violet No.2). Approximately over 65% of tensile strength remain after 2week. Complete absorption in tissues takes around 90 days. REXSIN Sutures meets requirements established by the United States Pharmacopoeia (USP) for Absorbable Surgical Sutures. REXSIN Sutures is available in 8-0, 7-0, 6-0, 5-0, 2-0, 0, 1 and 2, which are the sizes identified in the currently recognized United States.

AI/ML Overview

The provided text is related to a 510(k) Pre-market Notification for a medical device (surgical suture). It describes the device, its intended use, and non-clinical tests conducted to demonstrate substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria for an AI/ML powered device, a study proving device performance against such criteria, or details regarding ground truth, expert consensus, or specific effect sizes for human readers with AI assistance.

Therefore, I cannot fulfill your request for:

  • A table of acceptance criteria and reported device performance.
  • Sample size used for the test set and data provenance.
  • Number of experts used to establish ground truth and their qualifications.
  • Adjudication method.
  • MRMC comparative effectiveness study results or effect sizes.
  • Standalone algorithm performance.
  • Type of ground truth used.
  • Sample size for the training set.
  • How ground truth for the training set was established.

This document describes a traditional medical device (suture) and its non-clinical testing for substantial equivalence, not an AI/ML-powered diagnostic or decision-support system.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

July 7, 2020

Sm Eng Co., Ltd % Sanglok Lee Manager Wise Company Inc. #303, 142, Gasan digital 1-ro Geumcheon-gu, 08507 Kr

Re: K200392

Trade/Device Name: Rexsin Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: June 4, 2020 Received: June 11, 2020

Dear Sanglok Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cindy Chowdhurv, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200392

Device Name REXSIN

Indications for Use (Describe)

REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for SM Eng, which is a company that specializes in engineering. The logo is purple and white, and it features the company's name and website address. The website address is www.sm-eng.net. The logo is simple and modern, and it is likely used on the company's website and marketing materials.

46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea TEL.82-51-305-8016 FAX.82-51-305-8021

510(k) Summary

The assigned 510(k) Number: K200392

Date of Submission: February 12, 2020

Applicant

Company name: SM ENG CO., LTD Address: 46, Nakdong-daero 1302beon-gil, Sasang-gu, Busan, Korea TEL: +82 51 3058016 FAX: +82513058021 Email: leesg@sm-eng.net

Submission Correspondent

Sanglok, Lee Wise COMPANY Inc. #303, 142, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org

Subject Device Identification

Proprietary Name: REXSIN Common Name: Synthetic Absorbable PGA Suture With or Without Needle Device Class: Class II Regulation Number:21 C.F.R. 878.4493 Product Code: GAM Classification Name: Suture, Absorbable, Synthetic, Polyglycolic Acid

Indication for use

REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue

Predicate devices

510(k) Number: K073614 Device Name: WG-Surgical Sutures with Needle Manufacturer: FOOSIN MEDICAL SUPPLIES INC.LTD

De vice Description

Synthetic Absorbable PGA Suture With or Without Needle (REXSIN) are produced and provided by SM ENG Co., Ltd. REXSINis indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue

{4}------------------------------------------------

SM Eng
www.sm-eng.net

46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea

TEL.82-51-305-8016 FAX.82-51-305-8021

Suture of REXSIN is Trisorb manufactured by SAMY ANG Biopharmaceuticals Corporation. SM Eng is receiving bulk Trisorb from Samyang, and then go through cutting, adhesion of need and thread, winding and sterilization process according to SM Eng procedure, which becomes REXSIN suture.

REXSIN is a sterilized medical device composed with absorbable, braided, coated suture with needle, stainless steel SUS 304

REXSIN Suture is a synthetic absorbable, braided, coated suture composed of a Polyglycolic acid (PGA). The suture is available undyed (natural) or dyed (D&C Violet No.2). Approximately over 65% of tensile strength remain after 2week. Complete absorption in tissues takes around 90 days.

REXSIN Sutures meets requirements established by the United States Pharmacopoeia (USP) for Absorbable Surgical Sutures. REXSIN Sutures is available in 8-0, 7-0, 6-0, 5-0, 2-0, 0, 1 and 2, which are the sizes identified in the currently recognized United States.

Non-Clinical Test Conclusion

Bench tests were conducted to verify that the subject device (REXSIN) met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:

USP <861> SUTURES - DIAMETER USP <871> SUTURES - NEEDLE ATTACHMENT

USP <881> TENSILE STRENGTH USP MONOGRAPH OF ABSORBABLE SURGICAL SUTURE

Sterile barrier system testing

ISO 11607-1, Packaging For Terminally Sterilized Medical Devices - Part 1: Requirements For Materials, Sterile Barrier Systems And Packaging Systems ISO 11607-2, Packaging For Terminally Sterilized Medical Devices - Part 2: Validation Requirements For Forming, Sealing And Assembly Processes ASTM F 88, Standard Test Method for Seal Strength of Flexible Barrier Materials ASTM F 1929, Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration

Accelerated aging stability testing

  • ASTM F 1980, Standard Guide For Accelerated Aging Of Sterile Barrier Systems For Medical Devices. (Sterility)

Biocompatibility testing

    1. Suture Needle
  • ISO 10993-12: Sample Preparation
  • ISO 10993-5: Test for Cytotoxicity
  • ISO 10993-10: Test for Irritation and Sensitization
  1. Suture
  • ISO 10993-1: Selection of Tests

{5}------------------------------------------------

46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea

TEL.82-51-305-8016 FAX.82-51-305-8021

  • ISO 10993-2: Animal Welfare
  • ISO 10993-12: Sample Preparation
  • ISO 10993-5: Test for Cytotoxicity
  • ISO 10993-10: Test for Irritation and Sensitization
  • ISO 10993-11: Testfor Systemic Toxicity
  • ISO 10993-3: Tests for Genotoxicity
  • ISO 10993-6: Test for Local Effects after Implantation
  • ISO 10993-4: Selection of Tests for Interaction with Blood

Real-time and accelerated aging stability testing was performed to support shelf life of REXS IN. Additionally, the residual strength and absorption rate studies were evaluated in accordance with the requirements outlined in FDA's Class II Special Controls Guidance Document: Surgical

Sutures.

Substantially Equivalent Conclusion

The following table compares the subject device (REXSIN) to the predicate devices with respect to intended use, technological characteristics and principles of operation, etc.

PropertySubject device:REXSINPredicate device:WG-Surgical Sutures withNeedle
510(k)NumberK200392K073614
Product CodeGAMSame
Indicationsfor usestatementsREXSIN is indicated for use ingeneral soft tissue approximationand/or ligation, including use inophthalmic procedures, but notfor use in cardiovascular orneural tissue.Same
ClassllSame
SterileYesSame
Single UseYesSame
ConfigurationPGA Suture and NeedleSame
Suture
MaterialPolyglycolic acidSame
Coating materialPolycaprolactone and CalciumstearateSame
ColorUndyed (natural) and dyed (D&CViolet No.2)Same
Absorbable/NonabsorbableAbsorbableSame
Braided/MonofilamentBraidedSame
Barbed/Not BarbedNot BarbedSame
Suture SizeThe subject device is available in8-0, 7-0, 6-0, 5-0, 4-0,3-0,2-0, 0,1 and 2., which are the sizesidentified in the currentlyrecognized United StatesPharmacopoeia.Same. The subject device isavailable in 10-0 through 3 or4, which are the sizesidentified in the currentlyrecognized United StatesPharmacopoeia.
Length of Suture15, 30, 45, 50, 60, 75, 90, 100,120, 125, 140, 250cmUnknown

Table. Comparison of Technology Characteristics

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for SM Eng. The logo is in a light gray color and has a stylized design. The website address, www.sm-eng.net, is written in a smaller font below the logo. The logo is simple and modern.

46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea TEL.82-51-305-8016 FAX.82-51-305-8021

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for SM Eng. The logo is in a serif font and is a dark gray color. The website address, www.sm-eng.net, is written in a smaller font below the logo. A purple line underlines the logo and website address. There is a gray swirl to the left of the logo.

46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea

TEL.82-51-305-8016 FAX.82-51-305-8021

Diameter of SutureThe tensile strengths of subjectdevice comply with the tensilerequirement listed in USP <861>Same
Tensile strengthThe tensile strengths of subjectdevice comply with the tensilerequirement listed in USP <881>Same
Needle AttachmentThe bond between suture andneedle of the applicant devicemeet the requirements defined inUSP <871>.Same
Needle
MaterialStainless SteelSame
Needle typeTaper point, Reverse Cutting,Conventional cutting, Tapercutting, Spatula, Blunt pointUnknown
BiocompatibilityComply with ISO 10993-5,10993-10Same

The subject devices, REXSIN are determined to be Substantially Equivalent (SE) to the predicate device, and as safe and effective as the predicate device.

§ 878.4493 Absorbable poly(glycolide/l-lactide) surgical suture.

(a)
Identification. An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.