Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K140571
    Date Cleared
    2014-05-15

    (70 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    FKR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage PureFlow SL module is an optional accessory to the NxStage System One™ that prepares dialysate for use during hemodialysis, as prescribed by the physician.

    Device Description

    The NxStage PureFlow SL module is an optional accessory to the NxStage System One used to prepare water for hemodialysis that meets ANSI/AAMI/ISO 13959:2009 and proportion it with dialysate concentrate to produce dialysate per ANSI/AAMI/ISO 11663:2009. The PureFlow SL consists of the Control Unit (CU), the water Pre-Treatment Unit, the optional OPTA Kit, the Purification Pack (PAK), and the Dialysate Sack (SAK) with Dialysate Concentrate.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the supporting study:

    The provided document is a 510(k) Summary for a device modification, not a comprehensive study report with detailed performance data. Therefore, many of the requested details about specific acceptance criteria, study sizes, ground truth establishment, and MRMC effectiveness are not explicitly stated. The document focuses on demonstrating substantial equivalence to a predicate device based on design control requirements and general performance verification.

    Here's a breakdown of the available information structured as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific quantitative acceptance criteria or detailed numerical performance results. Instead, it makes a general statement about meeting design specifications and being suitable for its intended use.

    However, based on the Technological Characteristics Comparison Table (Table 1) and the description, we can infer some qualitative "acceptance criteria" through comparison with the predicate device and relevant standards:

    Feature/Criteria (Inferred)Reported Device Performance
    Indication for UseSame as predicate device (NxStage PureFlow SL module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician).
    Water purification technologySame as predicate device (Deionization).
    UltrafiltrationSame as predicate device (Redundant ultrafiltration replaced within 12 weeks).
    Water QualityMeets or exceeds requirements per ANSI/AAMI/ISO 26722:2009 (Water treatment equipment for hemodialysis applications).
    Proportioning methodSame as predicate device (Mixed to use Batch via volumetric dosing of purified water with specified amount of electrolyte concentrate).
    Conductivity MeasurementSame as predicate device (System measures final batch conductivity prior to each treatment per ANSI/AAMI/ISO 26722:2009).
    Dialysate ConcentrateSame as predicate device (Standard hemodialysis concentrate per ANSI/AAMI/ISO 13958:2009).
    PackagingSame as predicate device (Packaged in flexible LLDPE bags with standard luer lock connector, ranging in size from 20L to 60L).
    BufferSame as predicate device (Lactate).
    Mix-to-Use TimeSame as predicate device (96-hours).
    BioburdenSame as predicate device (AAMI/ANSI/ISO 11663:
    Ask a Question

    Ask a specific question about this device

    K Number
    K111174
    Date Cleared
    2011-09-19

    (145 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    FKR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage PureFlow SL module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

    Device Description

    The NxStage PFSL module is an optional accessory to the NxStage System One that is used to treat water for hemodialysis per ANSI/AAMI/ISO 13959:2009 and ANSI/AAMI/ISO 26722:2009 and proportion it with dialysate concentrate to produce dialysate per ANSI/AAMI/ISO 11663:2009. The PFSL module consists of the Control Unit (CU), the water Pre-Treatment Unit, the optional OPTA Kit, the Purification Pack (PAK), and the Dialysate Sack (SAK) with Dialysate Concentrate.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the NxStage PureFlow SL Subsystem, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Standard)Reported Device Performance (Result for 96-hour MTU)
    Bioburden:
    Ask a Question

    Ask a specific question about this device

    K Number
    K080919
    Date Cleared
    2008-10-15

    (197 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    FKR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage PureFlow SL module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

    Device Description

    The NxStage PFSL module is an optional accessory to the NxStage System One that is used to treat water for hemodialysis and proportion it with dialysate concentrate to produce dialysate per ANSI/AAMI RD52:2004 and ANSI/AAMI RD62: 2006. The PFSL module consists of the Control Unit (CU), the water Pre-Treatment Unit, the Purification Pack (PAK), and the Dialysate Sack (SAK) with Dialysate Concentrate.

    AI/ML Overview

    The provided text describes a Special 510(k) submission for a device modification, which primarily focuses on demonstrating substantial equivalence to a predicate device. This type of submission relies on design control requirements and performance testing to ensure the modified device meets design specifications and remains as safe and effective as the original. As such, the information you've requested regarding detailed acceptance criteria, specific study designs (like MRMC or standalone performance), sample sizes for test and training sets, and expert qualifications for ground truth establishment is not typically present in a 510(k) summary for such a modification.

    The 510(k) summary focuses on demonstrating that the NxStage PureFlow SL (PFSL) module with modifications is substantially equivalent to the legally marketed predicate device (cleared through K060296).

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with detailed reported performance metrics. Instead, it states:

    "Performance, verification and validation testing was conducted to characterize performance of the proposed PFSL to provide a basis of comparison to the predicate device as all features are not identical. Results of this testing have documented that the proposed PFSL module is substantially equivalent to the predicate device and is suitable for the labeled indications for use."

    This indicates that internal performance specifications were met to ensure substantial equivalence, but the specific numerical acceptance criteria and results are not disclosed in this summary. The general acceptance criterion is that the modified device performs comparably to the predicate device and meets its design specifications, preparing dialysate per ANSI/AAMI RD52:2004 and ANSI/AAMI RD62: 2006.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size used for any test set or the data provenance. The testing mentioned is "Performance, verification and validation testing," which typically refers to bench testing rather than clinical study data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The study described is a non-clinical, bench-testing comparison for substantial equivalence, not a clinical study requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. The study described is a non-clinical, bench-testing comparison for substantial equivalence.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is an accessory for hemodialysis, and the evaluation is non-clinical, focusing on technological characteristics and performance rather than human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A "standalone" performance evaluation in the context of an algorithm or AI would not apply here. However, the non-clinical "Performance, verification and validation testing" can be considered a standalone evaluation of the device's functional performance against its design specifications and the predicate device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device modification is its design specifications and compliance with relevant industry standards (ANSI/AAMI RD52:2004 and ANSI/AAMI RD62: 2006) for preparing dialysate. The performance was characterized to demonstrate substantial equivalence to the predicate device.

    8. The sample size for the training set

    This information is not applicable and not provided. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided. This is not an AI/ML device that requires a training set.

    In summary:

    This 510(k) focuses on a "Special 510(k) Device Modification" which primarily relies on demonstrating compliance with 21 CFR §820.30 Design Control Requirements and conducting performance, verification, and validation (bench) testing to confirm that the modified device remains substantially equivalent to its predicate device. This type of submission does not typically involve the detailed clinical study metrics, expert consensus, or AI-specific evaluation criteria often associated with novel medical devices or AI-powered diagnostics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060296
    Date Cleared
    2006-03-31

    (53 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    FKR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage Dialysate Preparation Module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

    Device Description

    The NxStage Dialysate Preparation Module is an optional accessory to the NxStage System One that pre-treats and purifies water that mixes with concentrate to produce dialysate meeting the requirements of the ANSI/AAMI Standards, RD52:2004 and The dialysate is to be used with the NxStage System One during RD62:2001. hemodialysis treatments.

    AI/ML Overview

    The provided text describes a Special 510(k) Device Modification for the NxStage Dialysate Preparation Module, focusing on its substantial equivalence to a predicate device. It does not contain information about acceptance criteria, device performance, or a study that uses a test set, ground truth, or expert review in the way a typical AI/ML medical device submission would.

    Therefore, the following information is based on the provided text, and many sections will indicate that the information is "Not applicable" or "Not provided" for this type of device modification.


    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Substantial Equivalence to Predicate DeviceThe modified NxStage Dialysate Preparation Module was compared to the baseline device (K043436) and found to be substantially equivalent based on device indications for use, comparison of descriptive and technological characteristics, and design control certification.
    Meeting Design SpecificationsDesign validation testing was performed to ensure that the modified device meets design specifications.
    Compliance with ANSI/AAMI StandardsThe device produces dialysate meeting the requirements of ANSI/AAMI Standards, RD52:2004 and RD62:2001.
    Compliance with Design Control RequirementsCertification of compliance to 21 CFR 820.30 Design Control requirements was provided.
    Intended UseThe device meets the minimum requirements considered acceptable for its intended use (preparing dialysate for hemodialysis with the NxStage System One).

    Study Details

    1. Sample size used for the test set and the data provenance: Not applicable. This was a Special 510(k) Device Modification focusing on design validation and substantial equivalence, not a clinical study on a test set in the context of an AI/ML device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for AI/ML performance evaluation (e.g., expert consensus on medical images) was not established for this type of device modification.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a Dialysate Preparation Module, not an AI-assisted diagnostic or interpretative device that would involve human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a medical device, not an AI algorithm. Its performance is evaluated based on its physical and functional characteristics to prepare dialysate.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's performance would be the physical and chemical properties of the dialysate it produces, validated against established standards (ANSI/AAMI RD52:2004 and RD62:2001) and its design specifications. It is based on objective measurements and engineering principles, not expert consensus on medical images or pathology.
    7. The sample size for the training set: Not applicable. This device does not use machine learning, so there is no "training set."
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K043436
    Date Cleared
    2005-03-17

    (94 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    FKR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage Dialysate Preparation Module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

    Device Description

    Device Booonproniment and optional accessory to the NxStage System The NAGtage Dialyone in that mixes with concentrate to produce One that pro treats crequirements of the ANSI/AAMI Standards, RD52:2004 and The dialysate is to be used with the NxStage System One during RD62:2001. hemodialysis therapies.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the NxStage Dialysate Preparation Module, not a study evaluating acceptance criteria or device performance in the typical sense of a clinical or algorithmic study.

    Therefore, most of the requested information cannot be extracted from this document, as it pertains to a regulatory submission demonstrating substantial equivalence to predicate devices, rather than a detailed performance study with quantifiable acceptance criteria and results.

    However, based on the document, I can provide the following:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria. Instead, it states that the device's performance was characterized to enable comparison to predicate devices, focusing on demonstrating substantial equivalence.

    Acceptance Criteria (Implied)Reported Device Performance
    Adherence to ANSI/AAMI Standards (RD52:2004 and RD62:2001)The device "pro treats requirements of the ANSI/AAMI Standards, RD52:2004 and RD62:2001."
    Produces AAMI dialysate from purified AAMI water and AAMI concentrateThe device "uses purified AAMI water that mixes with AAMI concentrate to produce AAMI dialysate."
    Incorporates proper safety features (conductivity measurement, automatic safety clamp, mix to use timer, ultrafilters and I.V. sterilization filters to insure dialysate purity)The device "incorporates the proper safety features, such as conductivity measurement, automatic safety clamp, mix to use timer, and ultrafilters and I.V. sterilization filters to insure dialysate purity."
    Safe for intended use as an optional accessory to the NxStage System One during hemodialysis"Verification and Validation testing, which includes simulated use testing, sufficiently demonstrates that the Prep Module can be used safely as intended."
    Substantially equivalent to predicate devices"Results of the testing presented in this 510(k) have demonstrated that the NxStage Prep Module is substantially equivalent to the predicate devices and is suitable for the labeled indication for use."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The submission references "Verification and Validation testing" and "simulated use testing," but does not specify sample sizes, specific test sets, or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided as the submission does not detail ground truth establishment by experts. The "ground truth" here is regulatory compliance and functional performance against standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a medical accessory (Dialysate Preparation Module), not an AI or imaging device that would typically involve human readers or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable. The device is a hardware accessory; there is no "algorithm only" performance concept in this context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" indirectly used here is adherence to industry performance standards (ANSI/AAMI RD52:2004 and RD62:2001) for dialysate quality and the successful functioning of safety features, as well as comparison to legally marketed predicate devices. It is based on engineering and functional testing rather than clinical outcomes or expert consensus on diagnostic interpretations.

    8. The sample size for the training set

    This information is not provided. This is not an AI/machine learning device, so the concept of a "training set" is not applicable.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1