Search Filters

Search Results

Found 32 results

510(k) Data Aggregation

    K Number
    K241220
    Date Cleared
    2025-01-24

    (268 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tina-quant Lipoprotein (a) Gen.2 Molarity assay is an in vitro test for the quantitative determination of lipoprotein (a) [Lp(a]] in human serum and plasma on cobas c systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular disease risk, when used in conjunction with clinical evaluation and other lipoprotein tests.

    Device Description

    The Tina-quant Lipoprotein (a) Gen.2 Molarity assay is an in vitro test for the quantitative determination of lipoprotein (a) [Lp(a)] in human serum and plasma on cobas c systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular disease risk, when used in conjunction with clinical evaluation and other lipoprotein tests.

    Tina-quant Lipoprotein (a) Gen.2 Molarity assay quantifies lipoprotein (a) in human serum and plasma and reports the values in nmoVL with calibrator values traceable to the WHO/IFCC SRM2B reference material.

    Reagents - working solutions
    R1: Glycine buffer: 170 mmol/L. pH 7.0: stabilizers: BSA: rabbit serum 0.1 %, preservative
    R3: Latex particles coated with polyclonal anti-human lipoprotein (a) antibodies (rabbit): 0.5 %; glycine buffer: 170 mmol/L, pH 7.3, BSA; preservative

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Tina-quant Lipoprotein (a) Gen.2 Molarity assay, based on the provided document:

    This device is an in vitro diagnostic (IVD) assay, not an AI/ML-driven device. Therefore, the concepts of human readers, multi-reader multi-case (MRMC) studies, ground truth establishment by experts for images, training sets, and adjudication methods are not directly applicable in the typical sense as they would be for an AI model that interprets medical images. The "acceptance criteria" here refer to the predefined performance specifications for an analytical assay.

    Device Name: Tina-quant Lipoprotein (a) Gen.2 Molarity

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states, "All acceptance criteria were met" for the non-clinical performance evaluation sections (Precision, Analytical Sensitivity, Linearity, Dilution, High Dose Hook Effect, Endogenous Interferences, Analytical Specificity/Cross-Reactivity, Sample Matrix Comparison, Method Comparison, and Stability). Specific pre-defined acceptance criteria are generally internal to the manufacturer's quality system and not explicitly listed in this 510(k) summary (which focuses on summarizing the results against those criteria). However, the reported device performance is provided, which demonstrates that the device met the manufacturer's internal criteria.

    Performance CharacteristicAcceptance Criteria (Implied/General)Reported Device Performance (as stated in document)
    Precision (Repeatability)Within pre-defined CV% limitsCV% between 0.4% and 2.1% (for various samples)
    Precision (Intermediate Precision)Within pre-defined CV% limitsCV% between 0.7% and 2.6% (for various samples)
    Limit of Blank (LoB)LoB should be ≤ 6 nmol/LLoB = 6 nmol/L
    Limit of Detection (LoD)LoD should be ≤ 7 nmol/LLoD = 7 nmol/L
    Limit of Quantitation (LoQ)LoQ should be ≤ 7 nmol/LLoQ = 7 nmol/L
    Linearity/Assay Reportable RangeLinear within 7 – 240 nmol/LConfirmed for the measuring range of 7 – 240 nmol/L
    DilutionAccurate dilution of samples > measuring range (1:3 rerun function)Confirmed, supporting 1:3 dilution for samples above measuring range
    High Dose Hook EffectNo false results up to a high concentrationConfirmed no false result up to 450 nmol/L
    Endogenous InterferencesNo significant interference from tested substances (Icterus, Hemolysis, Lipemia, Rheumatoid factors)All predefined acceptance criteria were met. Claims: Icterus up to I index 60, Hemolysis up to H index 1000, Lipemia up to L index 2000, Rheumatoid factors up to 1200 IU/mL.
    Analytical Specificity / Cross-ReactivityNo significant cross-reactivity from tested substances (Plasminogen, Apolipoprotein B)No significant cross-reactivity in tested concentration ranges (Plasminogen up to 150 mg/dL, Apolipoprotein B up to 200 mg/dL).
    Exogenous InterferencesNo significant interference from common drug panelsAll predefined acceptance criteria were met for 15 listed drugs at specified concentrations.
    Sample Matrix ComparisonAcceptable performance across serum and different plasma typesAll predefined acceptance criteria were met (Slope, Intercept, Correlation Coefficient for comparison of serum vs. serum with gel, Li-Heparin, K2-EDTA, K3-EDTA plasma).
    Method Comparison to Lp(a) ELISA reference methodStrong correlation with reference methodSample size (n) = 126. Deming regression: y = 1.023x + 0.692 nmol/L, r = 0.992. Sample concentrations 8.70 - 234 nmol/L.
    StabilitySupports manufacturer's claimsStability data supports Roche Diagnostic's claims as reported in the package labeling.

    2. Sample Sizes and Data Provenance

    • Precision (Repeatability): n = 84 (number of individual measurements per sample level).
    • Precision (Intermediate Precision): 2 aliquots per run, 2 runs per day, 21 days.
    • LoB: One analyte-free sample measured with three reagent lots, 6 runs, 10-fold determination per run, distributed over >3 days.
    • LoD: 5 serum samples with low analyte concentrations measured with three reagent lots, 2-fold determination per run, 6 runs distributed over 5 days.
    • LoQ: 5 serum samples measured with three reagent lots, 5 runs distributed over 5 days.
    • Linearity: 1 run using 3 reagent lots and 5 replicates per sample. A dilution series prepared from human serum (sample High) and NaCl 0.9% (sample Blank) to obtain 16 levels.
    • Method Comparison (to Lp(a) ELISA): n = 126 native human serum samples.
    • Sample Matrix Comparison: ≥ 50 samples.
    • Reference Range Study:
      • Caucasian/White: n = 425
      • African-American/Black: n = 111
      • Asian: n = 128
      • Hispanic/Latino (among Caucasian/White): n = 110
      • Non-Hispanic/Non-Latino (among Caucasian/White): n = 311
      • Data Provenance for Reference Range Study: Samples from apparently healthy adults in the United States, with equal representation of males and females. This suggests prospective collection for the purpose of establishing reference ranges. Other studies (e.g., method comparison, precision) would typically use laboratory samples which could be retrospective or specifically prepared for the study. The document does not specify "retrospective" or "prospective" for all tests, but the nature of these analytical performance studies often involves controlled, prepared, or collected samples without a specific patient encounter context. The "native human serum samples" for method comparison generally implies biological samples.

    3. Number of Experts and Qualifications for Ground Truth

    For this in vitro diagnostic device, "ground truth" is established by reference methods or validated analytical measurements, not by human expert interpretation in the way it would be for an AI image analysis tool.

    • Not Applicable in the traditional sense for AI/ML image analysis. The "ground truth" for analytical performance tests like precision, linearity, or interference is an established analytical value or the characteristic of the sample itself (e.g., known concentration, presence/absence of interfering substance). For "method comparison," the ground truth is provided by the designated "Lp(a) ELISA reference method."

    4. Adjudication Method for the Test Set

    • Not Applicable in the traditional sense for AI/ML image analysis. Adjudication is usually performed by multiple human experts reviewing a case. For an IVD, the "adjudication" is inherent in the rigorous analytical protocols, statistical methods (e.g., CLSI guidelines), and the use of reference methods or certified reference materials for traceability.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    • Not Applicable. This is an in vitro diagnostic assay, not an AI-assisted diagnostic imaging system that would involve human readers interpreting cases.

    6. Standalone (i.e. algorithm only without human-in-the loop performance) Performance

    • Yes, implicitly. The performance studies (Precision, Analytical Sensitivity, Linearity, Interference, Method Comparison) evaluate the device (assay on the cobas c 503 analyzer) as a standalone system. There is no human intervention in the measurement or calculation of results once the sample is loaded. The device generates a quantitative value for Lp(a).

    7. The Type of Ground Truth Used

    • Reference Method/Analytical Standards:
      • Precision, Sensitivity, Linearity, Interferences: Ground truth is based on known concentrations in control materials, spiked samples, or by using analyte-free samples, following standardized laboratory practices and CLSI guidelines (e.g., CLSI EP05-A3, EP17-A2, EP06-A-Ed2).
      • Method Comparison: The "ground truth" or reference values were derived from the Lp(a) ELISA reference method.
      • Traceability: The device's calibration values are traceable to the WHO/IFCC SRM2B reference material for nmol/L, indicating a standardized and accepted reference for accuracy.

    8. The Sample Size for the Training Set

    • Not applicable in the AI/ML sense. This is a traditional IVD assay, not an AI model that requires a "training set" to learn features. The assay is based on well-understood principles of immunoturbidimetry and does not involve machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As there is no AI/ML training set, this question is not relevant to this device. The assay's "knowledge" is built into its chemical reagents, optical detection system, and pre-programmed algorithms for calculating concentration from turbidity measurements, based on established analytical principles and calibration to reference materials.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211058
    Device Name
    Lp(a) Ultra
    Manufacturer
    Date Cleared
    2022-12-22

    (622 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lp(a) Ultra assay is intended for in vitro diagnostic use in the immunoturbidimetric quantitation of lipoprotein (a) [Lp(a)] in human serum and plasma using an automated analyzer. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular disease in specific populations, when used in conjunction with clinical evaluation.

    For In Vitro Diagnostic use.

    Device Description

    Lp(a) Ultra assay is composed by 2 ready to use liquid reagents (Reagent 1and Reagent 2) that are supplied in the following configuration: Reagent 1 fill volume 18 mL in a 20 mL wedge and Reagent 2 fill volume 9 mL in a 20 mL wedge, 1 wedge of each/kit.

    The kit contains one plastic (HDPE) vial of Reagent 1 and one plastic (HDPE) vial of Reagent 2, which allows the customer to perform 86 tests (on AU680 automatic analyzer).

    AI/ML Overview

    The provided text describes the performance testing of the Lp(a) Ultra assay, an in vitro diagnostic device, and its acceptance criteria. This is a lab-based assay, not an AI/ML medical device, and therefore the concepts of human readers, AI assistance, ground truth experts, and training/test sets as understood in AI/ML are not directly applicable in the same way. However, I can extract and present the information in a table format that parallels the requested structure for acceptance criteria and performance against those criteria.

    Key takeaway: This document describes a traditional in-vitro diagnostic device, not an AI/ML based device. Therefore, many of the requested fields (e.g., number of experts, adjudication methods, MRMC studies, AI assistance) are not relevant.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Test Performed)Target/Requirement (Implicit)Reported Device Performance
    Precision (Repeatability/Reproducibility)Inter-Assay (Total Imprecision): Good precision across the concentration range.
    Intra-Assay (Within Run): Calculated CV% lower than 5%.Inter-Assay: Total %CV ranged from 2.3% to 5.3% across 5 concentration levels (20 mg/dL to 100 mg/dL).
    Intra-Assay: Total %CV ranged from 0.5% to 4.8% across 5 concentration levels and 3 runs. All calculated CV% were lower than 5%.
    Linearity (Analytical Measuring Range - AMR)Demonstrate linearity up to 100 mg/dL and establish AMR.Linear range found as 8.4 to 105.5 mg/dL (Lot 00507) and 8.4 to 103.8 mg/dL (Lot 10208). The assay is linear up to 100 mg/dL. Claimed AMR: 10 mg/dL to 100 mg/dL (based on LoQ and linearity).
    Analytical Sensitivity/Detection LimitLimit of Blank (LoB): Highest measurement for a blank sample within reasonable limits.
    Limit of Detection (LoD): Lowest analyte concentration reliably distinguished from LoB.
    Limit of Quantitation (LoQ): Lowest amount quantifiable with stated accuracy.LoB: 0.7 mg/dL (for both reagent lots). Supports LoB claim of 0.7 mg/dL.
    LoD: 1.6 mg/dL (Lot 90266), 1.9 mg/dL (Lot 90530). Supports LoD claim of 1.9 mg/dL.
    LoQ: 2.5 mg/dL (Lot 90266), 3.0 mg/dL (Lot 90530). Supports LoQ claim of 3.0 mg/dL.
    Interference (Endogenous Substances)No significant interference from common endogenous substances (e.g., Intralipid, Triglycerides, Bilirubin, Rheumatoid Factor, Hemoglobin, Ascorbic Acid) at specified concentrations.No interference observed for:
    • Intralipid® Sterile Fat Emulsion: up to 1000 mg/dL
    • Conjugated bilirubin: up to 60 mg/dL
    • Unconjugated bilirubin: up to 60 mg/dL
    • Rheumatoid Factor: up to 500 UI/mL
    • Hemoglobin: up to 1000 mg/dL
    • Ascorbic Acid: up to 180 mg/dL
    • Triglycerides: Tested up to 1000 mg/dL; no significant interference from 461 to 715 mg/dL. |
      | Stability (On Board/Calibration) | Demonstrate stated on-board and calibration stability claims. | Calibration Stability: 15 days on AU 680 Analyzer.
      On-board stability: 30 days on AU680 analyzer.
      %bias for levels 1-5 within acceptable ranges (-4.8% to 5.1%). |
      | Prozone Effect | No high-dose hook effect (prozone) observed within the claimed measuring range. | No Prozone effect observed up to 500.0 mg/dL. No prozone effect claimed up to the upper limit of the measuring range. |
      | Method Comparison vs. Predicate Device | Demonstrate substantial equivalence in performance to the predicate device. | Correlation Coefficient (r): 0.995 (Passing & Bablok fit and Linear fit).
      Slope: 0.9850 (0.9706 - 1.0000) for Passing & Bablok; 0.9771 (0.9612-0.9929) for Linear fit. |
      | Matrix Comparison (Serum vs. Plasma) | Demonstrate correlation between serum and various plasma types (Lithium Heparin, Sodium Heparin, Di-Potassium EDTA, Tri-Potassium EDTA). | Lithium Heparin: r=0.997, Slope=0.99 (0.95-1.03).
      Sodium Heparin: r=0.999, Slope=0.99 (0.95-1.01).
      Di-Potassium EDTA: r=0.997, Slope=0.97 (0.95-1.01).
      Tri-Potassium EDTA: r=0.998, Slope=0.99 (0.95-1.03). |

    2. Sample size used for the test set and the data provenance:

    • Precision (Repeatability/Reproducibility):
      • Inter-Assay: 5 samples at different concentrations. 2 replicates per sample, 2 runs per day, for 28 testing days. (Total of 280 measurements per sample level over the entire study for 5 levels? The specific number of measurements is implied but not explicitly stated as 'n=X' for the total dataset.)
      • Intra-Assay: 5 samples at different concentrations. 20 replicates per sample, run on 3 different runs. (Total of 300 measurements over the entire study: 5 samples x 20 replicates x 3 runs).
    • Linearity: 2 different reagent lots. No specific sample count is given, but "dilution series" are implied.
    • Analytical Sensitivity (LoB, LoD, LoQ):
      • LoB: 4 saline samples (zero-analyte) for two reagent lots, tested in 5 replicates in 3 different runs. (Total of 60 measurements per lot).
      • LoD & LoQ: The guidelined CLSI EP17-A2 was followed, but specific sample counts were not explicitly stated for these sections beyond the general method.
    • Interference: "Low: ~30 mg/dL" and "High: ~50 mg/dL" Lp(a) concentrations. Two aliquots of serum pool per concentration. Tested in different replicates for paired difference and then diluted. Specific numbers of individual patient samples tested for interference effects for each substance are not provided.
    • Stability (On Board Calibration): 5 samples at different concentrations (20 mg/dL to 100 mg/dL).
    • Prozone Study: Last calibrator level used to reach a high concentration (approx. 500.0 mg/dL). Sample diluted in saline.
    • Method Comparison vs. Predicate Device: Not explicitly stated but usually involves a significant number of patient samples covering the assay range to demonstrate correlation. The CLSI document EP09c 3rd Edition was followed.
    • Matrix Comparison:
      • Lithium Heparin: 57 serum and Lithium Heparin plasma samples "derived from the same patients", tested in duplicate.
      • Sodium Heparin: 58 serum and Sodium Heparin plasma samples "derived from the same patients", tested in duplicate.
      • Di-Potassium EDTA: 57 serum and Di-Potassium EDTA plasma samples "derived from the same patients", tested in duplicate.
      • Tri-Potassium EDTA: 56 serum and Tri-Potassium EDTA plasma samples "derived from the same patients", tested in duplicate.

    Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective, beyond stating that they used "human serum and plasma" samples. Given it's a 510(k) submission from an Italian company for an in vitro diagnostic device, the samples are typically human biological samples obtained under ethical guidelines, often from healthy volunteers or patient populations relevant to the assay's indication.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable for this type of IVD device. The "ground truth" for quantitative assays is established by reference methods, traceable calibrators, and robust analytical procedures, not through expert consensus like in image interpretation. The performance is assessed by direct measurement and comparison to established analytical performance metrics and standard reference materials.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods are typically used in clinical trials involving human interpretation of medical images or symptoms where there might be inter-reader variability. For an IVD assay, performance is determined by instrumental measurements and statistical analysis against predefined acceptance criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is an in vitro diagnostic device, not an AI/ML-driven analysis tool for human readers. No human interpretation or AI assistance is involved in its direct operation or intended use.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable in the context of an AI algorithm. The device, Lp(a) Ultra, is a standalone reagent (assay) used on an automated analyzer (Beckman Coulter AU680). Its performance is inherently "standalone" in that it measures analyte concentration without human interpretation of raw signals; humans perform the testing and interpret the numerical results from the analyzer.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for this in vitro diagnostic assay is established through:

    • Reference standards and calibrators: Used to define known concentrations of Lp(a).
    • Defined analytical methods: Following recognized guidelines (e.g., CLSI EP05-A3, EP15-A3, EP06, EP17-A2, EP07, EP25-A, EP09c) for precision, linearity, sensitivity, interference, etc.
    • Comparison to a legally marketed predicate device: "Ground truth" for demonstrating substantial equivalence is the performance of the predicate device (Diazyme Lipoprotein (a) Assay).
    • Known concentrations in spiked samples or characterized biological samples: Used for studies like linearity and interference.

    8. The sample size for the training set:

    Not applicable. This is not an AI/ML device, so there is no "training set" in that sense. The device is a chemical reagent kit with defined analytical characteristics.

    9. How the ground truth for the training set was established:

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180074
    Date Cleared
    2018-03-22

    (71 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diazyme Lipoprotein (a) Assay is intended as a latex particle enhanced immunoturbidimetric assay for the in vitro quantitative determination of lipoprotein(a) [Lp(a)] concentration in human serum or Clinical Chemistry Systems. The measurement of Lipoprotein (a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular diseases in specific populations, when used in conjunction with clinical evaluation. For in vitro diagnostic use only.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification acceptance letter from the FDA for a device called "Diazyme Lipoprotein (a) Assay." It is an in vitro diagnostic device, not an AI/ML medical device. Therefore, the information required to answer your specific questions about acceptance criteria, study design (sample size, data provenance, expert adjudication, MRMC, standalone performance), and ground truth establishment, which are typical for AI/ML device submissions, is not present in this document.

    The document focuses on:

    • The FDA's determination of substantial equivalence to a predicate device.
    • Regulatory classifications and general controls (e.g., annual registration, GMP, labeling).
    • Indications for Use for the Diazyme Lipoprotein (a) Assay, which is for the in vitro quantitative determination of lipoprotein(a) [Lp(a)] concentration in human serum.

    To answer your questions, I would need a different type of document, such as a clinical study report or a detailed premarket submission summary (e.g., from the FDA's 510(k) database if it provided more detail than this letter).

    Therefore, I cannot populate the table or answer the specific questions about the study design, ground truth, and expert involvement based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123046
    Date Cleared
    2012-12-20

    (83 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use in the quantitative measurement of lipoprotein(a) (Lp(a)) in human serum or plasma on the ADVIA Chemistry systems. Measurement of Lp(a) may aid in the diagnosis of disorders of lipid (fat) metabolism and assessing persons at risk for cardiovascular diseases when used in conjunction with clinical evaluation and other lipoprotein tests.

    The ADVIA® Chemistry Lipoprotein(a) calibrators is intended for use in the calibration of ADVIA® Chemistry systems for the ADVIA Chemistry Lipoprotein(a) (LPA) assay.

    Device Description

    The Lipoprotein(a) reagents are ready-to-use liquid reagents packaged for use on the automated ADVIA 1650 Chemistry system. They are supplied as a 100 tests/wedge, 2 wedges/kit. ADVIA Chemistry Lipoprotein(a) calibrator is a single analyte, human serum based product containing human lipoprotein (a). The kit consists of 1 vial each of 5 calibrator levels which are lyophilized. The target concentrations of these calibrators are 7.5, 15, 30, 65, and 95 mg/dL. The volume per vial (after reconstitution with deionized water) is 1.0 mL. Deionized water is recommended to be used as a zero calibrator.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the ADVIA® Chemistry Lipoprotein(a) Assay, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state formal "acceptance criteria" in a definitive, quantified table for all performance characteristics. Instead, it describes studies that "gave acceptable results when compared to the predicate device" and lists the numerical outcomes of these studies. Therefore, this table is constructed by inferring the implied acceptance from the reported "acceptable results" and direct comparisons to the predicate device where available.

    Performance CharacteristicImplied Acceptance Criteria (from predicate similarity/acceptable results)Reported Device Performance (ADVIA Chemistry Lipoprotein(a) Assay)
    PrecisionWithin acceptable variability for a diagnostic assay, comparable to predicate. (Precise numerical criteria not explicitly stated but implied by acceptable results)Within Run CV: 0.9% - 1.2% (lower concentrations), 0.8% (higher concentration)
    Total CV: 1.3% - 1.6% (lower concentrations), 1.6% (higher concentration)
    Linearity/Reportable RangeLinear performance across the intended measuring range.Linear/measuring range: 10.00 mg/dL - 85.00 mg/dL
    Limit of Blank (LoB)Below 6.0 mg/dL (claim supported)5.35 mg/dL
    Limit of Detection (LoD)Below 9.0 mg/dL (claim supported)8.90 mg/dL
    Limit of Quantitation (LoQ)Below 10.0 mg/dL (claim supported)9.02 mg/dL
    Method Comparison (Serum)Strong correlation (R > 0.99) and agreement with predicate device. Slope and intercept confidence intervals encompassing 1 and 0 respectively for good agreement.Correlation Coefficient: 0.99
    Linear Regression (y = mx + b): y = 1.01x - 1.02 mg/dL
    Slope 95% CI: 1.00 - 1.02
    Intercept 95% CI: -1.47 - -0.57
    Matrix Comparison (Plasma)Strong correlation (R > 0.99) and agreement with predicate device. Slope and intercept confidence intervals encompassing 1 and 0 respectively for good agreement.Correlation Coefficient: 0.99
    Linear Regression (y = mx + b): y = 1.01x - 0.98 mg/dL
    Slope 95% CI: 0.99 - 1.02
    Intercept 95% CI: -1.49 - -0.47
    Analytical SpecificityNo significant interference (>10% variance) from common interferents (icterus, lipemia, hemolysis) at specified concentrations.No significant interference found at:
    • Unconjugated bilirubin: 0-60 mg/dL
    • Conjugated bilirubin: 0-60 mg/dL
    • Intralipid: 0-1000 mg/dL
    • Hemoglobin: 0-1000 mg/dL |

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision: Not explicitly stated as a separate "test set" for precision, but studies used:
      • Serum sample pools and serum-based controls.
      • Each sample assayed 2 replicates per run, 2 runs per day, for at least 20 days.
    • Linearity/Assay Reportable Range:
      • Sample Size: Nine diluted samples (prepared from high and low serum pools).
    • Limit of Blank, Limit of Detection, Limit of Quantitation:
      • Sample Size: 160 replicates of "zero" serum pool and several serum pools (number not specified) with Lp(a) concentrations up to 4x LoD level.
    • Method Comparison (Serum):
      • Sample Size: 68 serum samples.
    • Matrix Comparison (Plasma):
      • Sample Size: 44 plasma samples.
    • Analytical Specificity:
      • Specific concentrations of interferents (unconjugated bilirubin 0-60 mg/dL, conjugated bilirubin 0-60 mg/dL, Intralipid 0-1000 mg/dL, hemoglobin 0-1000 mg/dL) were tested with samples at 3 specific Lp(a) levels (e.g., 14, 28, 47 mg/dL for bilirubin). The exact number of individual samples/replicates isn't specified beyond "in 14, 28, and 47 mg/dL lipoprotein (a) samples."
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective). The samples are referred to as "human serum" or "human plasma."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of blood assay does not typically use human experts to establish "ground truth" in the same way an imaging device or AI algorithm would. Instead, the ground truth for an in vitro diagnostic assay is established through:

    • Reference Methods: Comparison against an established, legally marketed (predicate) device.
    • Known Concentrations: Use of accurately prepared and characterized calibrators and controls with known concentrations of the analyte.
    • Spiking Experiments: Addition of known amounts of interferents or analyte to assess impact.

    The document implicitly relies on the predicate device (Randox Lipoprotein (a) assay on the Hitachi 717) as a "ground truth" reference for method and matrix comparison studies.

    4. Adjudication Method for the Test Set

    Not applicable. The "test set" here refers to clinical samples or control materials analyzed by the device, and their values are compared against a predicate device or expected values, not adjudicated by experts in a qualitative sense.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in vitro diagnostic assay for quantitative measurement of a biomarker, not an imaging device or an AI application that assists human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, this entire submission describes the standalone performance of the ADVIA Chemistry Lipoprotein(a) assay. The device measures Lp(a) concentration directly on the ADVIA Chemistry systems without human interpretation or intervention in the measurement process itself, other than preparing samples and operating the instrument.

    7. The Type of Ground Truth Used

    The ground truth used for performance validation is primarily:

    • Predicate Device Measurements: For method and matrix comparisons, the results from the legally marketed Randox Lipoprotein(a) assay on the Hitachi 717 are considered the comparative "ground truth" for demonstrating substantial equivalence.
    • Known Concentrations/Prepared Samples: For linearity, LoB/LoD/LoQ determinations, and analytical specificity, samples with known concentrations (e.g., "zero" serum pool, serum pools with Lipoprotein (a) concentration up to 4 x LOD level, diluted samples with expected values, samples with spiked interferents) serve as the ground truth.

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of machine learning or AI. This device is a classical in vitro diagnostic immunoassay, not an AI algorithm that requires training data. The development and calibration of such assays involve extensive R&D, but the data used in those phases are not typically referred to as a "training set" in this manner.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the AI/machine learning sense for this device. The development of the assay's reagents and methodologies would have involved established chemical and immunological principles, and calibration of the assay (using the ADVIA Chemistry Lipoprotein(a) calibrator) uses products with assigned Lp(a) concentrations.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas c Tina-quant Lipoprotein (a) Gen.2 assay is an in vitro test intended for the quantitative determination of lipoprotein(a) [Lp(a)] in human serum and plasma on the Roche/Hitachi cobas c systems. The measurement of Lp(a) is useful in evaluation of lipid metabolism disorders and assessing atherosclerotic cardiovascular disease in specific populations, when used in conjunction with clinical evaluation and other lipoprotein tests.

    The Preciset Lp(a) calibrator set is intended for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the value sheets.

    The PreciControl Lp(a) Gen.2 control set is intended for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the value sheets.

    Device Description

    The TQ Lp(a) Gen.2 test principle is a particle-enhanced immunoturbidimetric assay. Human lipoprotein (a) agglutinates with latex particles coated with anti-Lp(a) antibodies. The precipitate is determined turbidimetrically. The Preciset Lp(a) Gen.2 calibrator set consists of five lyophilized calibrators based on a stabilized and lyophilized pool of human plasma. The concentrations of the calibrator components have been adjusted to ensure optimal calibration of the appropriate Roche methods on clinical chemistry analyzers. The PreciControl Lp(a) Gen.2 control set contains two lyophilized controls based on a human plasma matrix.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the cobas c Tina-quant Lipoprotein (a) Gen.2 Test System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document primarily focuses on demonstrating substantial equivalence to a predicate device and reports performance characteristics rather than explicit acceptance criteria. However, we can infer some criteria from the comparative data and the general performance evaluation.

    Performance CharacteristicAcceptance Criteria (Inferred from Predicate/Good Practice)Reported Device Performance (TQ Lp(a) Gen.2 assay)
    Measuring RangeComparable to predicate (2.0 – 80.0 mg/dL)6.0 – 80.0 mg/dL
    Lower Limits of MeasureLDL: 2.0 mg/dL (predicate)LoB: 3 mg/dL, LoD: 4 mg/dL, LoQ: 6 mg/dL
    Hook EffectNo significant effect within expected rangeNo hook effect up to 190 mg/dL
    Precision (Within-run)Comparable to predicate or betterControl L: 1.4% CV, Control H: 1.0% CV
    Pool 1: 5.4% CV, Pool 2: 6.2% CV
    Pool 3: 2.4% CV, Pool 4: 0.9% CV
    Precision (Intermediate/Total)Comparable to predicate or betterControl L: 1.6% CV, Control H: 1.1% CV
    Pool 1: 7.6% CV, Pool 2: 6.4% CV
    Pool 3: 2.9% CV, Pool 4: 1.1% CV
    Icterus InterferenceNo significant interferenceNo significant interference up to an I index of 60 (approx. 60 mg/dL)
    Hemolysis InterferenceNo significant interferenceNo significant interference up to an H index of 1000 (approx. 1000 mg/dL)
    Lipemia InterferenceNo significant interferenceNo significant interference up to an L index of 2000
    Plasminogen Cross-ReactivityNo significant cross-reactivityNo significant cross-reactivity up to 150 mg/dL
    Apolipoprotein B Cross-ReactivityNo significant cross-reactivityNo significant cross-reactivity up to 200 mg/dL
    Rheumatoid Factor InterferenceNot specified for predicateNo significant interference up to 1200 IU/mL
    Drugs InterferenceNot specified for predicateNo interference at therapeutic concentrations using common drug panels
    Reagent On-board StabilityStable for a reasonable period6 weeks
    Reagent Unopened StabilityStable until expiration date2-8°C until expiration date
    Calibration FrequencyAfter reagent lot change and as requiredSame

    Study Proving Acceptance Criteria:

    The document describes the submission as a 510(k) premarket notification, indicating a substantial equivalence study. The study's purpose is to demonstrate that the cobas c Tina-quant Lipoprotein (a) Gen.2 Test System (candidate device) is as safe and effective as a legally marketed predicate device. This is primarily achieved through direct comparison of features and performance characteristics, as summarized in the tables provided in the document. The document lists "Evaluations summary" in Section 9, indicating that performance characteristics were evaluated.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific test (e.g., precision, interference studies, linearity, method comparison). It mentions evaluating "several performance characteristics."

    • Data Provenance: The document does not specify the country of origin for the data or whether it was retrospective or prospective. It is implied to be laboratory-generated data for performance evaluations.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to the evaluation of an in vitro diagnostic (IVD) device like the cobas c Tina-quant Lipoprotein (a) Gen.2 Test System. The "ground truth" for these tests is established by reference methods or accepted analytical principles, not by expert consensus in the same way it would be for, for example, image interpretation. The device measures a specific analyte concentration.

    4. Adjudication Method for the Test Set

    Not applicable for this type of IVD device. Analytical performance is typically evaluated against defined statistical metrics and analytical specifications, not through expert adjudication of results.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This is an in vitro diagnostic device for quantitative measurement of a biomarker, not an AI-assisted diagnostic imaging or interpretation system involving human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is an IVD assay, so its performance is inherently "standalone" in terms of measurement. The "algorithm" here refers to the immunoassay chemistry and detection system. Its performance is evaluated independently of human interpretation of raw signals, although human operators perform the testing and interpret the final quantitative results.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this device would be established by:

    • Reference materials/calibrators: The Preciset Lp(a) Gen.2 calibrator set is mentioned as being based on a "stabilized and lyophilized pool of human plasma," with concentrations adjusted for optimal calibration. This implies traceability to a higher-order reference method or material for determining Lp(a) concentrations.
    • Method comparison against predicate device: The substantial equivalence comparison implies that the predicate device serves as a benchmark for acceptable performance.
    • Physiological samples with known characteristics: For interference studies, samples spiked with known interferents would be used.

    8. The Sample Size for the Training Set

    No information is provided about a "training set" in the context of machine learning. For an IVD assay, method development involves extensive experimentation for optimization of reagents, reaction conditions, and calibration models. However, these are not typically referred to as "training sets" in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for a chemical immunoassay, as there is no machine learning "training set" in the conventional sense. The "ground truth" for assay development and validation is established through analytical chemistry principles, use of reference materials, and comparison to established methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082488
    Manufacturer
    Date Cleared
    2009-01-13

    (138 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diazyme Lp(a) is intended as a latex particle enhanced immunoturbidimetric assay for the in vitro quantitative determination of lipoprotein(a) [Lp(a)] concentration in human serum er plasma (EDTA) on Clinical Chemistry Systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular diseases in specific populations, when used in conjunction with clinical evaluation.

    Diazyme Lp(a) Control is intended for use in monitoring the quality control of results obtained with the Diazyme Lp(a) reagents by turbidimetry.

    Diazyme Lp(a) standard is intended for use in establishing the calibration curve for the Diazyme Lp(a) reagents by turbidimetry.

    Device Description

    The Diazyme Lipoprotein (a) Assay is based on a latex enhanced immunoturbidimetric assay. Lp(a) in the sample binds to specific anti-Lp(a) antibody, which is coated on latex particles, and causes agglutination. The degree of the turbidity caused by agglutination can be measured optically and is proportional to the amount of Lp(a) in the sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Diazyme Lp(a) Assay, based on the provided text:

    Acceptance Criteria and Device Performance

    The provided document describes the Diazyme Lp(a) Assay by comparing its performance to a predicate device (Denka Lp(a) Assay) and reporting specific performance characteristics such as precision and accuracy. The "acceptance criteria" are implied by the performance of the legally marketed predicate device and the new device demonstrating "good agreement" and "substantially similar" results.

    Acceptance Criterion (Implied by Predicate/Good Agreement)Reported Diazyme Lp(a) Assay Performance
    Reportable Range (Serum)5.44 mg/dL to 100.0 mg/dL (Predicate: 0.5mg/dL to 80.0 mg/dL)
    Reportable Range (Plasma)5.44 mg/dL to 100.0 mg/dL (Predicate: 0.5mg/dL to 80.0 mg/dL)
    Precision/Serum (Within Run)1.1% - 2.6% (Predicate: 1.26% - 2.00%)
    Precision/Serum (Total)2.4% - 3.6% (Predicate: 0.99% - 2.22%)
    Accuracy/Serum (Correlation Coefficient)0.998 (Predicate: 0.989)
    Accuracy/Serum (Slope/Intercept)y = 0.9895x + 0.0279 (Predicate: y = 1.108x - 1.44)
    Accuracy/Plasma (Correlation Coefficient)0.9803 (Predicate: 0.990)
    Accuracy/Plasma (Slope/Intercept)y = 1.044x + 0.0407 (Predicate: y = 1.079x - 0.16)
    Interference (Triglycerides)Less than 10% interference at 1000 mg/dL
    Interference (Ascorbic acid)Less than 10% interference at 10 mmol/L
    Interference (Bilirubin)Less than 10% interference at 40 mg/dL
    Interference (Bilirubin Conjugated)Less than 10% interference at 40 mg/dL
    Interference (Hemoglobin)Less than 10% interference at 1000 mg/dL

    Note: The acceptance criteria are based on demonstrating "substantial equivalence" to the predicate device. This implies that the new device's performance should be comparable and within acceptable analytical limits, which are not explicitly defined as numerical thresholds beyond the reported performance metrics. The statement "good agreement with legally marketed assay" and "no significant deviation" act as the overarching acceptance criteria for accuracy.

    Study Details

    1. Sample Size used for the test set and the data provenance:

      • Test Set Sample Size:
        • Accuracy (Correlation studies): 76 serum samples.
        • Precision: Three levels of serum specimens. While not explicitly stating the number of unique patient samples, the study design involved testing these three levels with 2 runs per day, with duplicates, over 20 working days.
        • Interference: Human serum samples were used, but the exact number is not specified.
      • Data Provenance: Not explicitly stated, but the samples are "human serum" and "human plasma." There is no mention of country of origin or whether the data was retrospective or prospective.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This device is an in vitro diagnostic assay that quantifies a biomarker (Lp(a)). The "ground truth" for such assays is typically established by comparative methods, such as a legally marketed predicate device or a reference method. It does not involve human expert interpretation of images or clinical findings in the same way as, for example, a radiology AI device. Therefore, no information on human experts establishing ground truth for the test set or their qualifications is provided or relevant in this context.
    3. Adjudication method for the test set:

      • Not applicable. As explained above, this is a quantitative diagnostic assay where the "ground truth" is typically the result from a reference method or predicate device, not subject to human expert adjudication. The comparison method is direct measurement against the predicate device.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is an in vitro diagnostic assay, not an AI-powered diagnostic imaging device that assists human readers. Therefore, an MRMC study and AI-assisted performance improvement are not relevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, this is primarily a standalone device performance study. The Diazyme Lp(a) Assay is an automated quantitative assay. The reported performance metrics (precision, accuracy, interference) are for the device itself operating without human interpretation of its primary output beyond standard laboratory procedures for running an assay. The comparison against the predicate device reflects the standalone performance.
    6. The type of ground truth used:

      • The "ground truth" for the accuracy study was established by comparison with an existing commercial Lp(a) assay method (the predicate device, Denka Lp(a) Assay, K013359). This is a common method for establishing accuracy and substantial equivalence for in vitro diagnostic devices.
    7. The sample size for the training set:

      • No information about a "training set" is provided. This is typical for in vitro diagnostic assays based on chemical/immunological reactions, as they are not machine-learning algorithms that require a training phase with labeled data in the same way an AI device does. The device's performance characteristics are determined through analytical validation studies (precision, accuracy, interference, linearity, etc.).
    8. How the ground truth for the training set was established:

      • Not applicable, as no training set for an AI/machine learning algorithm is mentioned or relevant for this type of in vitro diagnostic device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K063838
    Manufacturer
    Date Cleared
    2007-09-27

    (275 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the in vitro quantitative determination of the Lipoprotein(a) in human serum, Liheparin plasma and EDTA plasma on T60 instruments.

    Measurement of low-density lipoprotein in serum may aid in the diagnosis of disorders of lipid (fat) metabolism and help to identify specific populations at risk from cardiovascular diseases

    Lipoprotein(a) Calibrator is used as a calibrator for quantification of Lipoprotein(a) in serum and plasma by immunoturbidimetry with T60 instruments using methods defined by Thermo Electron Oy

    For in vitro diagnostic use on T60 instrument. Lipoprotein(a) Control is intended to be used as an assayed control serum to monitor precision of Lipoprotein(a) test defined by Thermo Electron Oy

    For in vitro diagnostic use on T60 instrument. Lipoprotein(a) Control High is intended to be used as an assayed control serum to monitor precision of Lipoprotein(a) test defined by Thermo Electron Oy.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification decision letter from the FDA to Thermo Electron Oy for their Lipoprotein(a) test system. This document grants market clearance for the device but does not contain the detailed study information or acceptance criteria requested.

    Therefore, I cannot extract the specific information required to answer your questions about acceptance criteria, device performance, study details (sample sizes, provenance, expert qualifications, adjudication, MRMC, standalone), or ground truth establishment. This type of information is typically found within the actual 510(k) submission document, which is not provided here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050487
    Device Name
    QUANTIA LA(A)
    Manufacturer
    Date Cleared
    2005-04-26

    (60 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quantia Lp(a) is intended as a latex particle enhanced immunoturbidimetric assay for the in vitro quantitative determination of lipoprotein(a) [Lp(a)] concentration in human serum or plasma (EDTA, Heparin, Citrate) on the Clinical Chemistry Systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular diseases in specific populations, when used in conjunction with clinical evaluation.

    Quantia Lp(a) Control is intended for use in monitoring the quality control of results obtained with the Quantia Lp(a) reagents by turbidimetry.

    Quantia Lp(a) Standard is intended for use in establishing the calibration curve for the Quantia Lp(a) reagents by turbidimetry.

    Device Description

    The Quantia Lp(a) is intended as a latex particle enhanced immunoturbidimetric assay for the in vitro quantitative determination of lipoprotein(a) [Lp(a)] concentration in human serum or plasma (EDTA, Heparin, Citrate) on the Clinical Chemistry Systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular diseases in specific populations, when used in conjunction with clinical evaluation.

    Quantia Lp(a) Control is intended for use in monitoring the quality control of results obtained with the Quantia Lp(a) reagents by turbidimetry.

    Quantia Lp(a) Standard is intended for use in establishing the calibration curve for the Quantia Lp(a) reagents by turbidimetry.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Quantia Lp(a) device, structured to address your specific points:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state pre-defined acceptance criteria for the Quantia Lp(a) device. Instead, it presents performance data and compares it to a predicate device to demonstrate substantial equivalence.

    Performance MetricReported Device Performance (Quantia Lp(a) vs Predicate)Implicit Acceptance Criteria (based on predicate equivalence)
    Method Comparison
    Slope1.121Close to 1 (indicating proportional agreement)
    Correlation Coeff. (r)0.9754High (indicating strong linear relationship)
    Within-run Precision (CV)
    Control Level 1 (mean 16.1 mg/dL)2.3%Low CVs (indicating good reproducibility)
    Control Level 2 (mean 57.9 mg/dL)0.9%Low CVs (indicating good reproducibility)
    Control Level 3 (mean 38.1 mg/dL)1.5%Low CVs (indicating good reproducibility)

    Study to Prove Device Meets Acceptance Criteria:

    The study proving the device meets the implicit acceptance criteria is a method comparison study and within-run precision assessment, detailed in the "Summary of Performance Data" section.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: 104 samples for the method comparison study.
    • Data Provenance: Not explicitly stated, but originating from Biokit S.A. in Spain. The study compares the Quantia Lp(a) device to a predicate device, which implies the 104 samples were run on both systems. The document doesn't specify if the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not applicable. The "ground truth" for the method comparison study was established by the predicate device (K013128 N-latex Lp(a) from Dade Behring).
    • Qualifications of Experts: N/A, as expert consensus was not used to establish the ground truth.

    4. Adjudication method for the test set:

    • Adjudication Method: Not applicable. The comparison is directly to a predicate device, not against an adjudicated ground truth from multiple human readers.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, an MRMC study was not done. This device is an in-vitro diagnostic (IVD) assay, not an AI-powered image analysis tool or decision support system that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Standalone Performance: Yes, the performance data presented (method comparison and precision) represents the standalone performance of the Quantia Lp(a) assay system, as it's an automated immunoturbidimetric assay. There is no "human-in-the-loop" component in the direct measurement process.

    7. The type of ground truth used:

    • Type of Ground Truth: The "ground truth" for comparative effectiveness in this context is the results obtained from the predicate device, K013218 N-latex Lp(a) (Dade Behring).

    8. The sample size for the training set:

    • Training Set Sample Size: Not applicable. This document describes the validation of a laboratory assay, not a machine learning algorithm that requires a distinct training set. The assay's "training" or calibration would involve its own internal standards (Quantia Lp(a) Standard), not a separate dataset in the machine learning sense.

    9. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not applicable in the machine learning sense. The Quantia Lp(a) Standard is used for establishing the calibration curve. The "ground truth" for these standards would have been established by the manufacturer through rigorous analytical methods (e.g., using reference materials and established metrological traceability) to determine their assigned values. This process is distinct from establishing ground truth for a machine learning training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K023853
    Date Cleared
    2002-12-13

    (24 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The H-ASSAY® Lp(a) Control Set is intended for use as an assayed quality control material for monitoring the performance of Lp(a) immunoturbidimetric assays.

    Device Description

    Not Found

    AI/ML Overview

    While the provided text describes the FDA's decision regarding the K-ASSAY® Lp(a) Controls device, it does not contain any information about acceptance criteria, study details, or performance data. The document is a 510(k) clearance letter (K023853), which states that the device is substantially equivalent to a legally marketed predicate device.

    Therefore, I cannot extract the requested information. The text focuses on regulatory approval rather than the technical details of the device's performance study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021660
    Date Cleared
    2002-07-25

    (66 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Product Code :

    DFC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kamiya K-ASSAY® Lp(a) Assay is an in vitro diagnostic reagent for the quantitative determination of human Lp(a) in human serum by immunoturbidimetric assay. The test may provide in conjunction with other lipoprotein tests, the risk assessment of coronary artery disease for specific populations.

    The K-ASSAY ® Lp(a) Calibrator Set is an in vitro diagnostic reagent for calibration of the K-ASSAY® Lp(a) Assay.

    For in vitro diagnostic use.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for an in vitro diagnostic device, not a study report. Therefore, it does not contain the detailed information requested regarding acceptance criteria, device performance, study design, or ground truth establishment.

    The letter simply states that the FDA has reviewed the premarket notification and determined the device is "substantially equivalent" to legally marketed predicate devices. This means it has met the regulatory requirements for clearance, not that a specific performance study with acceptance criteria has been detailed in this document.

    To provide the requested information, a separate study report or regulatory submission document detailing the performance evaluation of the K-Assay® Lipoprotein (a) Immunoturbidimetric Assay would be necessary.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4