K Number
K241220
Date Cleared
2025-01-24

(268 days)

Product Code
Regulation Number
866.5600
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tina-quant Lipoprotein (a) Gen.2 Molarity assay is an in vitro test for the quantitative determination of lipoprotein (a) [Lp(a]] in human serum and plasma on cobas c systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular disease risk, when used in conjunction with clinical evaluation and other lipoprotein tests.

Device Description

The Tina-quant Lipoprotein (a) Gen.2 Molarity assay is an in vitro test for the quantitative determination of lipoprotein (a) [Lp(a)] in human serum and plasma on cobas c systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular disease risk, when used in conjunction with clinical evaluation and other lipoprotein tests.

Tina-quant Lipoprotein (a) Gen.2 Molarity assay quantifies lipoprotein (a) in human serum and plasma and reports the values in nmoVL with calibrator values traceable to the WHO/IFCC SRM2B reference material.

Reagents - working solutions
R1: Glycine buffer: 170 mmol/L. pH 7.0: stabilizers: BSA: rabbit serum 0.1 %, preservative
R3: Latex particles coated with polyclonal anti-human lipoprotein (a) antibodies (rabbit): 0.5 %; glycine buffer: 170 mmol/L, pH 7.3, BSA; preservative

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Tina-quant Lipoprotein (a) Gen.2 Molarity assay, based on the provided document:

This device is an in vitro diagnostic (IVD) assay, not an AI/ML-driven device. Therefore, the concepts of human readers, multi-reader multi-case (MRMC) studies, ground truth establishment by experts for images, training sets, and adjudication methods are not directly applicable in the typical sense as they would be for an AI model that interprets medical images. The "acceptance criteria" here refer to the predefined performance specifications for an analytical assay.

Device Name: Tina-quant Lipoprotein (a) Gen.2 Molarity

1. Table of Acceptance Criteria and Reported Device Performance

The document states, "All acceptance criteria were met" for the non-clinical performance evaluation sections (Precision, Analytical Sensitivity, Linearity, Dilution, High Dose Hook Effect, Endogenous Interferences, Analytical Specificity/Cross-Reactivity, Sample Matrix Comparison, Method Comparison, and Stability). Specific pre-defined acceptance criteria are generally internal to the manufacturer's quality system and not explicitly listed in this 510(k) summary (which focuses on summarizing the results against those criteria). However, the reported device performance is provided, which demonstrates that the device met the manufacturer's internal criteria.

Performance CharacteristicAcceptance Criteria (Implied/General)Reported Device Performance (as stated in document)
Precision (Repeatability)Within pre-defined CV% limitsCV% between 0.4% and 2.1% (for various samples)
Precision (Intermediate Precision)Within pre-defined CV% limitsCV% between 0.7% and 2.6% (for various samples)
Limit of Blank (LoB)LoB should be ≤ 6 nmol/LLoB = 6 nmol/L
Limit of Detection (LoD)LoD should be ≤ 7 nmol/LLoD = 7 nmol/L
Limit of Quantitation (LoQ)LoQ should be ≤ 7 nmol/LLoQ = 7 nmol/L
Linearity/Assay Reportable RangeLinear within 7 – 240 nmol/LConfirmed for the measuring range of 7 – 240 nmol/L
DilutionAccurate dilution of samples > measuring range (1:3 rerun function)Confirmed, supporting 1:3 dilution for samples above measuring range
High Dose Hook EffectNo false results up to a high concentrationConfirmed no false result up to 450 nmol/L
Endogenous InterferencesNo significant interference from tested substances (Icterus, Hemolysis, Lipemia, Rheumatoid factors)All predefined acceptance criteria were met. Claims: Icterus up to I index 60, Hemolysis up to H index 1000, Lipemia up to L index 2000, Rheumatoid factors up to 1200 IU/mL.
Analytical Specificity / Cross-ReactivityNo significant cross-reactivity from tested substances (Plasminogen, Apolipoprotein B)No significant cross-reactivity in tested concentration ranges (Plasminogen up to 150 mg/dL, Apolipoprotein B up to 200 mg/dL).
Exogenous InterferencesNo significant interference from common drug panelsAll predefined acceptance criteria were met for 15 listed drugs at specified concentrations.
Sample Matrix ComparisonAcceptable performance across serum and different plasma typesAll predefined acceptance criteria were met (Slope, Intercept, Correlation Coefficient for comparison of serum vs. serum with gel, Li-Heparin, K2-EDTA, K3-EDTA plasma).
Method Comparison to Lp(a) ELISA reference methodStrong correlation with reference methodSample size (n) = 126. Deming regression: y = 1.023x + 0.692 nmol/L, r = 0.992. Sample concentrations 8.70 - 234 nmol/L.
StabilitySupports manufacturer's claimsStability data supports Roche Diagnostic's claims as reported in the package labeling.

2. Sample Sizes and Data Provenance

  • Precision (Repeatability): n = 84 (number of individual measurements per sample level).
  • Precision (Intermediate Precision): 2 aliquots per run, 2 runs per day, 21 days.
  • LoB: One analyte-free sample measured with three reagent lots, 6 runs, 10-fold determination per run, distributed over >3 days.
  • LoD: 5 serum samples with low analyte concentrations measured with three reagent lots, 2-fold determination per run, 6 runs distributed over 5 days.
  • LoQ: 5 serum samples measured with three reagent lots, 5 runs distributed over 5 days.
  • Linearity: 1 run using 3 reagent lots and 5 replicates per sample. A dilution series prepared from human serum (sample High) and NaCl 0.9% (sample Blank) to obtain 16 levels.
  • Method Comparison (to Lp(a) ELISA): n = 126 native human serum samples.
  • Sample Matrix Comparison: ≥ 50 samples.
  • Reference Range Study:
    • Caucasian/White: n = 425
    • African-American/Black: n = 111
    • Asian: n = 128
    • Hispanic/Latino (among Caucasian/White): n = 110
    • Non-Hispanic/Non-Latino (among Caucasian/White): n = 311
    • Data Provenance for Reference Range Study: Samples from apparently healthy adults in the United States, with equal representation of males and females. This suggests prospective collection for the purpose of establishing reference ranges. Other studies (e.g., method comparison, precision) would typically use laboratory samples which could be retrospective or specifically prepared for the study. The document does not specify "retrospective" or "prospective" for all tests, but the nature of these analytical performance studies often involves controlled, prepared, or collected samples without a specific patient encounter context. The "native human serum samples" for method comparison generally implies biological samples.

3. Number of Experts and Qualifications for Ground Truth

For this in vitro diagnostic device, "ground truth" is established by reference methods or validated analytical measurements, not by human expert interpretation in the way it would be for an AI image analysis tool.

  • Not Applicable in the traditional sense for AI/ML image analysis. The "ground truth" for analytical performance tests like precision, linearity, or interference is an established analytical value or the characteristic of the sample itself (e.g., known concentration, presence/absence of interfering substance). For "method comparison," the ground truth is provided by the designated "Lp(a) ELISA reference method."

4. Adjudication Method for the Test Set

  • Not Applicable in the traditional sense for AI/ML image analysis. Adjudication is usually performed by multiple human experts reviewing a case. For an IVD, the "adjudication" is inherent in the rigorous analytical protocols, statistical methods (e.g., CLSI guidelines), and the use of reference methods or certified reference materials for traceability.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

  • Not Applicable. This is an in vitro diagnostic assay, not an AI-assisted diagnostic imaging system that would involve human readers interpreting cases.

6. Standalone (i.e. algorithm only without human-in-the loop performance) Performance

  • Yes, implicitly. The performance studies (Precision, Analytical Sensitivity, Linearity, Interference, Method Comparison) evaluate the device (assay on the cobas c 503 analyzer) as a standalone system. There is no human intervention in the measurement or calculation of results once the sample is loaded. The device generates a quantitative value for Lp(a).

7. The Type of Ground Truth Used

  • Reference Method/Analytical Standards:
    • Precision, Sensitivity, Linearity, Interferences: Ground truth is based on known concentrations in control materials, spiked samples, or by using analyte-free samples, following standardized laboratory practices and CLSI guidelines (e.g., CLSI EP05-A3, EP17-A2, EP06-A-Ed2).
    • Method Comparison: The "ground truth" or reference values were derived from the Lp(a) ELISA reference method.
    • Traceability: The device's calibration values are traceable to the WHO/IFCC SRM2B reference material for nmol/L, indicating a standardized and accepted reference for accuracy.

8. The Sample Size for the Training Set

  • Not applicable in the AI/ML sense. This is a traditional IVD assay, not an AI model that requires a "training set" to learn features. The assay is based on well-understood principles of immunoturbidimetry and does not involve machine learning training.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As there is no AI/ML training set, this question is not relevant to this device. The assay's "knowledge" is built into its chemical reagents, optical detection system, and pre-programmed algorithms for calculating concentration from turbidity measurements, based on established analytical principles and calibration to reference materials.

{0}------------------------------------------------

January 24, 2025

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Roche Diagnostics Leslie Patterson Senior Regulatory Affairs Manager 9115 Hague Road Indianapolis, Indiana 46250

Re: K241220

Trade/Device Name: Tina-quant Lipoprotein(a) Gen.2 Molarity Regulation Number: 21 CFR 21 CFR 866.5600 Regulation Name: Low-density Lipoprotein Immunological Test System Regulatory Class: Class II Product Code: DFC Dated: December 18, 2024 Received: December 18, 2024

Dear Leslie Patterson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula V. Caposino -S

Paula Caposino, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241220 Device Name

Tina-quant Lipoprotein (a) Gen.2 Molarity

Indications for Use (Describe)

The Tina-quant Lipoprotein (a) Gen.2 Molarity assay is an in vitro test for the quantitative determination of lipoprotein (a) [Lp(a]] in human serum and plasma on cobas c systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular disease risk, when used in conjunction with clinical evaluation and other lipoprotein tests.

Type of Use (Select one or both, as applicable)

< | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Tina-quant Lipoprotein (a) Gen.2 Molarity K241220 - 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Submitter NameRoche Diagnostics
Address9115 Hague RoadP.O. Box 50416Indianapolis, IN 46250-0457
ContactLeslie PattersonPhone: (317) 225-8563Email: leslie.patterson@roche.com
Date PreparedJanuary 24, 2025
Proprietary NameTina-quant Lipoprotein (a) Gen.2 Molarity
Classification NameLow-density lipoprotein immunological test system.
Product Codes,Regulation NumbersDFC, 21 CFR 866.5600
Predicate DeviceTina-quant Lipoprotein (a) Gen.2 (K122722)
Establishment RegistrationRoche Diagnostics GmbH Mannheim, Germany: 9610126Roche Diagnostics GmbH Penzberg, Germany: 9610529Roche Diagnostics Indianapolis, IN United States: 1823260

{5}------------------------------------------------

DEVICE DESCRIPTION 1.

The Tina-quant Lipoprotein (a) Gen.2 Molarity assay is an in vitro test for the quantitative determination of lipoprotein (a) [Lp(a)] in human serum and plasma on cobas c systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular disease risk, when used in conjunction with clinical evaluation and other lipoprotein tests.

Tina-quant Lipoprotein (a) Gen.2 Molarity assay quantifies lipoprotein (a) in human serum and plasma and reports the values in nmoVL with calibrator values traceable to the WHO/IFCC SRM2B reference material.

Reagents - working solutions 1.1.

R1: Glycine buffer: 170 mmol/L. pH 7.0: stabilizers: BSA: rabbit serum 0.1 %, preservative

R3: Latex particles coated with polyclonal anti-human lipoprotein (a) antibodies (rabbit): 0.5 %; glycine buffer: 170 mmol/L, pH 7.3, BSA; preservative

2. INDICATIONS FOR USE

The Tina-quant Lipoprotein (a) Gen.2 Molarity assay is an in vitro test for the quantitative determination of lipoprotein (a) [Lp(a)] in human serum and plasma on cobas c systems. The measurement of Lp(a) is useful in evaluating lipid metabolism disorders and assessing atherosclerotic cardiovascular disease risk, when used in conjunction with clinical evaluation and other lipoprotein tests.

{6}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS 3.

The following table compares the Tina-quant Lipoprotein (a) Gen.2 Molarity assay with the predicate device, Tina-quant Lipoprotein (a) Gen.2 (K122722).

Candidate Device:Tina-quant Lipoprotein (a)Gen.2 MolarityPredicate Device:Tina-quant Lipoprotein (a)Gen.2 (K122722)
Intended Use /Indications forUseThe Tina-quant Lipoprotein (a) Gen.2Molarity assay is an in vitro test for thequantitative determination of lipoprotein (a)[Lp(a)] in human serum and plasma oncobas c systems. The measurement ofLp(a) is useful in evaluating lipidmetabolism disorders and assessingatherosclerotic cardiovascular disease risk,when used in conjunction with clinicalevaluation and other lipoprotein tests.In vitro test for the quantitativedetermination of lipoprotein (a) in humanserum and plasma on Roche/Hitachicobas c systems. The measurement ofLp(a) is useful in evaluating lipidmetabolism disorders and assessingatherosclerotic cardiovascular disease inspecific populations, when used inconjunction with clinical evaluation andother lipoprotein tests.
Assay MethodSameParticle enhanced immunoturbidimetricassay
InstrumentPlatformSamecobas c systems
SampleType/MatrixSameHuman serum and Li-heparin, K2-EDTA,K3-EDTA plasma
CalibratorSamePreciset Lp(a) Gen.2
Traceability/StandardizationThis method has been standardizedagainst the IFCC reference materialSRM2B for nmol/L.This method has been standardizedagainst an in-house reference material.
CalibrationMethodSame6-point, spline
CalibrationIntervalAutomatic full calibration- after reagent lot changeFull calibration- as required following quality controlproceduresFull calibration- after reagent lot change- as required following quality controlprocedures
ControlsSamePreciControl Lp(a) Gen.2
Measuring Range7-240 nmol/L6.0-80 mg/dL
Lower Limits ofMeasurementLoB (Limit of Blank): 6 nmol/LLoD (Limit of Detection): 7 nmol/LLoQ (Limit of Quantitation): 7 nmol/LLoB (Limit of Blank): 3 mg/dLLoD (Limit of Detection): 4 mg/dLLoQ (Limit of Quantitation): 6 mg/dL
Candidate Device:Tina-quant Lipoprotein (a)Gen.2 MolarityPredicate Device:Tina-quant Lipoprotein (a)Gen.2 (K122722)
MethodComparisonMethod comparison between the candidate method, Tina-quant Lipoprotein (a) Gen.2Molarity assay and the Lp(a) ELISA reference method:Sample size (n) = 126
Deming regression
$y = 1.023x + 0.692$ nmol/L$r = 0.992$
The sample concentrations were between 8.70 and 234 nmol/L.

Technical Characteristics Comparison Table between Tina-quant Lipoprotein Table 1: (a) Gen.2 Molarity and Tina-quant Lipoprotein (a) Gen.2

{7}------------------------------------------------

4. NON-CLINICAL PERFORMANCE EVALUATION

Performance characteristics were evaluated with Tina-quant Lipoprotein (a) Gen.2 Molarity on the cobas c 503 analyzer and are briefly summarized below.

All acceptance criteria were met.

Precision 4.1.

4.1.1. Repeatability and Intermediate Precision

Precision was determined in accordance with the CLSI EP05-A3 requirements with repeatability (n = 84) and intermediate precision (2 aliquots per run, 2 runs per day, 21 days). Results for repeatability and intermediate precision were obtained on the cobas c 503 analyzer. The results are summarized below. All acceptance criteria were met.

Repeatability
SampleMean (nmol/L)SD (nmol/L)CV (%)
Lp(a) Control Level Low35.20.4121.2
Lp(a) Control Level High97.70.4350.4
Human Serum 119.70.4142.1
Human Serum 249.30.5161.0
Human Serum 380.50.4340.5
Human Serum 41200.5460.5
Human Serum 52030.7940.4

{8}------------------------------------------------

Intermediate Precision
SampleMean (nmol/L)SD (nmol/L)CV (%)
Lp(a) Control Level Low35.10.5621.6
Lp(a) Control Level High99.01.141.2
Human Serum 119.70.5082.6
Human Serum 249.30.6761.4
Human Serum 380.50.8311.0
Human Serum 41191.141.0
Human Serum 52031.330.7

Analytical Sensitivity 4.2.

Limit of Blank (LoB) 4.2.1.

For determination of LoB, one analyte-free sample was measured with three reagent lots in 6 runs, each run with 10-fold determination, distributed over >3 days, on the cobas c 503 analyzer. The LoB was determined according to CLSI EP17-A2. The LoB claim in the labeling will be set to LoB = 6 nmol/L.

Limit of Detection (LoD) 4.2.2.

For determination of LoD, 5 serum samples with low analyte concentrations were measured each three reagent lots with 2-fold determination per run on one cobas c 503 analyzer. Six runs were distributed over 5 days. The LoD was determined according to CLSI EP17-A2. The LoD claim in the labeling will be set to LoD = 7 nmol/L

4.2.3. Limit of Quantitation (LoQ)

For determination of LoQ, 5 serum samples were measured with three reagent lots on one cobas c 503. Five runs were distributed over 5 days. The Limit of Quantitation (LoQ) was determined according to CLSI EP17-A2. The LoQ claim in the labeling will be set to 7 nmol/L.

4.3. Linearity/Assay Reportable Range

The linearity of the Tina-quant Lipoprotein (a) Gen.2 Molarity assay was assessed according to CLSI EP06-A-Ed2.

{9}------------------------------------------------

Linearity of the Tina-quant Lipoprotein (a) Gen.2 Molarity assay was assessed on one cobas c 503 analyzer in 1 run using 3 reagent lots and 5 replicates per sample. A dilution series was prepared from human serum (sample High) and NaCl 0.9% (sample Blank). The dilution series spanning the measuring range were prepared to obtain 16 levels.

Linearity was confirmed for the measuring range of 7 – 240 nmol/L.

4.4. Dilution

Post Dilution Check experiments were performed for samples above the measuring range and verify dilution of samples via the rerun function is a 1:3 dilution. Samples with concentrations above the measuring range were measured with the Tina-quant Lipoprotein (a) Gen.2 Molarity assay on the cobas c 503 via the automatic rerun function. The target values were determined with the ELISA reference method.

High Dose Hook Effect 4.5.

High-dose hook effect study was performed using the Tina-quant Lipoprotein (a) Gen.2 Molarity assay and confirmed that no false result occurs up to a lipoprotein (a) concentration of 450 nmol/L.

Endogenous Interferences 4.6.

Endogenous substances were evaluated for potential interference with the Tina-quant Lipoprotein (a) Gen.2 Molarity assay on the cobas c 503. All predefined acceptance criteria were met.

EndogenousSubstanceClaimNo significant interference up to
IcterusI index of 60 for conjugated and unconjugated bilirubin(approximate conjugated and unconjugated bilirubin concentration: 1026µmol/L or 60 mg/dL).
HemolysisH index of 1000(approximate hemoglobin concentration: 621 µmol/L or 1000 mg/dL).
Lipemia (Intralipid)L index of 2000There is poor correlation between the L index (corresponds to turbidity) andtriglycerides concentration.
Rheumatoid factors1200 IU/mL

{10}------------------------------------------------

Analytical Specificity / Cross-Reactivity 4.7.

Plasminogen and Apolipoprotein B were evaluated for cross-reactivity and the labeling claims for each substance can be found below:

Plasminogen: No significant cross-reactivity in the tested concentration range (up to 150 mg/dL).

Apolipoprotein B: No significant cross-reactivity in the tested concentration range (up to 200 mg/dL).

4.8. Exogenous Interferences

Common drug panels were evaluated for potential interference with the Tina-quant Lipoprotein (a) Gen.2 Molarity assay on the cobas c 503 analyzer. All predefined acceptance criteria were met.

DrugConcentration Tested [mg/L]
N-Acetylcysteine150
Acetylsalicylic acid30
Ampicillin-Na75
Ascorbic acid52.5
Cefoxitin6600
Doxycyclin18
Heparin3300 IU/L
Levodopa7.5
Methyldopa22.5
Metronidazole123
Rifampicin48
Acetaminophen156
Cyclosporine1.8
Ibuprofen219
Theophylline60
Phenylbutazone321

{11}------------------------------------------------

Sample Matrix Comparison 4.9.

The effect on quantitation of lipoprotein (a) values in the presence of anticoagulants with the Tina-quant Lipoprotein (a) Gen.2 Molarity assay was determined on the cobas c 503 analyzer by comparing values obtained from samples collected in serum, Li-Heparin, K2-EDTA and K3-EDTA plasma tubes. The study was performed using ≥ 50 samples measured on the cobas c 503 analyzer. All predefined acceptance criteria were met, supporting the labeling claim that serum, Li-Heparin, K2-EDTA and K3-EDTA plasma are acceptable sample types.

AnticoagulantSlopeIntercept(nmol/L)Correlation Coefficient(Pearson r)Concentration of Samples(nmol/L, serum)
Serum vs. Serum with gel separation1.00001.0007.51 - 239
Serum vs. Li-Heparin plasma0.9810.2860.9987.51 - 239
Serum vs. K2-EDTA plasma0.9720.1341.0007.51 - 233
Serum vs. K3-EDTA plasma0.9440.3891.0007.51 - 233

4.10. Method Comparison to Lp(a) ELISA reference method

A method comparison of the Tina-quant Lipoprotein (a) Gen.2 Molarity assay on the cobas c 503 analyzer compared to the Lp(a) ELISA reference method was completed. One hundred twenty-six native human serum samples were tested in singlicate using the cobas c 503 analyzer. The sample concentrations were between 8.70 and 234 nmoVL. The results can be found below:

Deming regression

y = 1.023x + 0.692 nmol/L

r = 0.992

4.11. Stability

The stability data supports Roche Diagnostic's claims as reported in the package labeling.

{12}------------------------------------------------

CLINICAL PERFORMANCE EVALUATION 5.

Reference Range Study 5.1.

Roche conducted a reference range study and measured Lp(a) levels among various subpopulations. The expected ranges were determined using samples from apparently healthy adults in the United States, with an equal representation of males and females.

Lp(a) by Race

nMedian2.5th percentile(90 %-CI)97.5th percentile(90 %-CI)Unit
Caucasian/White42517.8< 7 (n.a.)259 (229; 265)nmol/L
African-American/Black11166.2< 7 (n.a.)267 (207; 313)nmol/L
Asian12819.9< 7 (n.a.)210 (166; 330)nmol/L

Lp(a) by Ethnicity (among Caucasian/White)

nMedian2.5th percentile(90 %-CI)97.5th percentile(90 %-CI)Unit
Hispanic/Latino11014.5< 7 (n.a.)260 (162; 267)nmol/L
Non-Hispanic/Non-Latino31118.1< 7 (n.a.)259 (229; 287)nmol/L

CLINICAL LITERATURE 6.

Associations between Lp(a) and ASCVD risk were demonstrated in multiple published clinical studies in and outside of the US, including the Dallas Heart Study and the UK Biobank Study (PMID: 27831500, PMID: 33115266). The UK Biobank Study was also used to derive Lp(a) thresholds for ASCVD risk (PMID: 36036785, PMID: 38565461). Information was provided to bridge the results from the published studies to the candidate device.

CONCLUSIONS 7.

The analytical performance data for Tina-quant Lipoprotein (a) Gen.2 Molarity assay met the acceptance criteria. The analytical performance data and clinical performance evaluation support the substantial equivalence of Tina-quant Lipoprotein (a) Gen.2 Molarity assay on the cobas c 503 analyzer to the predicate.

§ 866.5600 Low-density lipoprotein immunological test system.

(a)
Identification. A low-density lipoprotein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the low-density lipoprotein in serum and other body fluids. Measurement of low-density lipoprotein in serum may aid in the diagnosis of disorders of lipid (fat) metabolism and help to identify young persons at risk from cardiovascular diseases.(b)
Classification. Class II (performance standards).