K Number
K223590

Validate with FDA (Live)

Date Cleared
2023-08-23

(265 days)

Product Code
Regulation Number
862.1295
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, lithium heparin plasma, and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia.

Folate levels in serum, lithium heparin plasma, and red blood cells are used to assess folate status. The serum folate levels is an indicator of recent folate intake. A low RBC folate value can indicate a prolonged folate deficiency.

Device Description

The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum and lithium heparin plasma or red blood cells using the Access Immunoassay Systems.

AI/ML Overview

The provided text describes the Beckman Coulter Access Folate Assay, a chemiluminescent immunoassay for the quantitative determination of folic acid levels. The submission is a 510(k) premarket notification for demonstrating substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and supporting studies based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Study TypeAcceptance CriteriaReported Device Performance
Method ComparisonR² ≥ 0.90 and slope 1.00 ± 0.12. Estimated bias at concentration corresponding to reference limits suggestive that values have not changed appreciably.R² = 0.99, Slope = 1.04 (95% CI: 1.01 - 1.07), Intercept = 0.081 (95% CI: -0.074 - 0.19). The study met the acceptance criteria. The estimated bias at concentration corresponding to reference limits defined on the predicate system suggest that such values have not changed appreciably on the DxI 9000 analyzer.
LinearityLinear throughout the analytical measuring interval (2.0 - 24.8 ng/mL).The study met the acceptance criterion, indicating linearity on the DxI 9000 Immunoassay Analyzer throughout the analytical measuring interval (2.0 - 24.8 ng/mL).
Serum ImprecisionNot explicitly stated as acceptance criteria, but based on typical standards for imprecision: Expected low %CV for higher concentrations and low SD for lower concentrations.Within-laboratory (total) % CV: between 2.2% and 4.4% for Folate concentrations > 2.0 ng/mL. Within-laboratory (total) SD: between 0.10 - 0.21 for Folate concentrations ≤ 2.0 ng/mL. Repeatability (within-run) % CV: between 1.6% and 2.7% for Folate concentrations > 2.0 ng/mL. Repeatability (within-run) SD: between 0.08 - 0.09 for Folate concentrations ≤ 2.0 ng/mL.
RBC ImprecisionNot explicitly stated as acceptance criteria, but based on typical standards for imprecision: Expected low %CV.Within-laboratory (total) % CV: ranged from 1.9% to 4.9%.
Limit of Blank (LoB)Claimed LoB of 0.80 ng/mL (1.81 nmol/L).The assay is designed to meet the claimed LoB of 0.80 ng/mL.
Limit of Detection (LoD)Claimed LoD of 1.0 ng/mL (2.27 nmol/L).The assay is designed to meet the claimed LoD of 1.0 ng/mL.
Limit of Quantitation (LoQ)Claimed LoQ of < 2.0 ng/mL (4.53 nmol/L) based on a 20% CV.The LoQ for Access Folate is designed to meet the claimed LoQ of < 2.0 ng/mL (4.53 nmol/L) based on a 20% CV.

2. Sample Size Used for the Test Set and Data Provenance

  • Method Comparison: 123 samples. Data provenance is not specified (e.g., country of origin, retrospective/prospective).
  • Linearity: The number of samples/levels used is not explicitly stated, but it's a verification study.
  • Serum Imprecision: 5 serum samples with varying concentrations. Each sample was assayed in duplicate with two runs per day, over 20-22 days, meeting a minimum requirement of 80 replicates per sample on each instrument and reagent lot combination. This implies a total of (5 samples * 80 replicates/sample) = 400 replicates (for one instrument and reagent lot combination, multiplied by 3 instruments and 3 reagent lots, so 3600 replicates in total across all combinations). Data provenance is not specified.
  • RBC Imprecision: 6 whole blood hemolysate samples with varying concentrations. For each sample, N=80 replicates were collected for imprecision calculations. Data provenance is not specified.
  • LoB/LoD/LoQ: Verification studies were performed but sample sizes are not explicitly stated, though CLSI EP17-A2 is mentioned as the protocol, which specifies methodologies for determining these limits.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This is a laboratory diagnostic assay for measuring analyte concentrations, not an interpretation task typically requiring expert consensus on ground truth like image analysis. The "ground truth" for method comparison and imprecision studies typically refers to the reference method (predicate device) results and the true concentration of controls/samples, respectively.

4. Adjudication Method for the Test Set

Not applicable. As this is a quantitative laboratory assay, there is no expert adjudication process in the traditional sense. The comparisons are against the predicate device or established analytical performance criteria.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This is an in-vitro diagnostic device (IVD) that quantitatively measures a biomarker (folic acid). MRMC studies are typically performed for devices that involve human interpretation, such as imaging devices, to evaluate the impact of AI assistance on reader performance.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

Yes, the studies described (Method Comparison, Linearity, Imprecision, LoB/LoD/LoQ) demonstrate the standalone performance of the Access Folate Assay on the DxI 9000 Access Immunoassay Analyzer. This device is an automated immunoassay system; its output is a quantitative measurement, not an interpretation that involves human-in-the-loop interaction in the same way as, for example, an AI-powered diagnostic imaging tool. The "performance" refers to the analytical accuracy, precision, and range of the automated system itself.

7. Type of Ground Truth Used

  • Method Comparison: The predicate device's results (Access Folate assay on Access 2 Instrument) served as the comparator for showing substantial equivalence.
  • Linearity, Imprecision, LoB/LoD/LoQ: The "true" concentrations of control materials and samples, as determined by highly controlled experimental conditions and accepted statistical methods (e.g., CLSI guidelines), serve as the basis for evaluating performance against pre-defined analytical criteria.

8. Sample Size for the Training Set

The document does not provide information about a "training set." This type of IVD, an immunoassay, is typically developed and characterized through analytical verification and validation studies rather than machine learning training sets. While there is an "algorithm" (the immunoassay process), it's based on chemical reactions and detection, not a learned model from a large dataset. Therefore, the concept of a training set as used in AI/ML is not directly applicable here.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as a training set, in the machine learning sense, is not described for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 23, 2023

Beckman Coulter, Inc. Kuljeet Kaur Regulatory Affairs Manager 1000 Lake Hazeline Drive Chaska, Minnesota 55318

Re: K223590

Trade/Device Name: Access Folate Assay Regulation Number: 21 CFR 862.1295 Regulation Name: Folic Acid Test System Regulatory Class: Class II Product Code: CGN Dated: November 30, 2022 Received: July 27, 2023

Dear Kuljeet Kaur:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-torres -S

Marianela Perez-Torres, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223590

Device Name Access Folate Assay

Indications for Use (Describe)

The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, lithium heparin plasma, and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia.

Folate levels in serum, lithium heparin plasma, and red blood cells are used to assess folate status. The serum folate levels is an indicator of recent folate intake. A low RBC folate value can indicate a prolonged folate deficiency.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or ribbon. To the right of the circle, the words "BECKMAN" are stacked on top of the word "COULTER" in a bold, sans-serif font. The text is black, providing a strong contrast against the white background.

Access Folate Assay 510(k) Summary

510(k) Number: K223590

Date Prepared: Aug 21, 2023

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Submitted By:

Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318

Kuljeet Kaur Regulatory Affairs Manager Phone: (952)-465-1914 Email: kkaur@beckman.com

Alternate Contact:

Madhuri Boppana Senior Regulatory Affairs Analyst Email: mboppana@beckman.com

Common Name: Folate test system Trade Name: Access Folate Assay Classification Name: Folic acid test system Classification Requlation: 21 CFR 862.1295 Classification Product Code: CGN

Predicate Device: Device Name: Access Folate Assay 510(k) Numbers: Primary predicate: K060774, Secondary predicate: K111952

Device Description:

The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum and lithium heparin plasma or red blood cells using the Access Immunoassay Systems.

{4}------------------------------------------------

Intended Use:

The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, lithium heparin plasma, and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia.

Folate levels in serum, lithium heparin plasma, and red blood cells are used to assess folate status. The serum folate levels is an indicator of recent folate intake. A low RBC folate value can indicate a prolonged folate deficiency.

SystemAttribute/CharacteristicPredicate AccessImmunoassay Systems(K060774)Access Folate on Dxl9000 AccessImmunoassayAnalyzer
Intended Use/Indicationsfor UseThe Access Folate assay is aparamagnetic particle,chemiluminescent immunoassavfor the quantitative determinationof folic acid levels in human serumand plasma (heparin) or red bloodcells using the AccessImmunoassay Systems.Same
Analyte MeasuredAccess FolateSame
TechnologyCompetitive binding ImmunoassaySystemSame
FormatChemiluminescentSame
MethodAutomatedSame
CalibrationUtilizes a stored calibration curveSame
Sample TypeSerum, plasma, or red blood cellsSame
StabilityStable at 2 to 10°C for 14 daysafter initial useSame
InstrumentAccess 2 Immunoassay SystemDxI 9000 AccessImmunoassay Analyzer
SubstrateAccess SubstrateLumi-Phos PRO
Measuring Range1.0 - 24.8 ng/mL2.0 - 24.8 ng/mL

Comparison of Technological Characteristics to the Predicate

Comparison of Technological Characteristics to the Predicate

SystemAttribute/CharacteristicPredicate AccessImmunoassay Systems(K111952)Access Folate on Dxl9000 AccessImmunoassay Analyzer
StandardizationWorld Health Organization(WHO) International StandardsSame

{5}------------------------------------------------

Summary of Studies:

Method Comparison: The results of the within range method comparison study met the acceptance criteria of R2 ≥ 0.90 and slope 1.00 ± 0.12 and supports the equivalence of the Access Folate assay on Dxl 9000 to the predicate device, the Access Folate assay on Access 2 Instrument. The estimated bias at concentration corresponding to reference limits defined on the predicate system suggest that such values have not changed appreciably on the Dxl 9000 analyzer.

NConcentration Range* (ng/mL)SlopeSlope 95% CIInterceptIntercept 95% CICorrelation Coefficient R
1231.4 - 251.041.01 - 1.070.081-0.074 - 0.190.99

*Range is Access 2 values

Linearity: A verification study was performed to evaluate the linearity of the Access Folate assay on the Dxl 9000 Access Immunoassay Analyzer based on CLSI EP06-Ed2. The results of this study met the acceptance criterion, indicating that the Access Folate assay is linear on the Dxl 9000 Immunoassay Analyzer throughout the analytical measuring interval (2.0 - 24.8 ng/mL).

Serum Imprecision: Verification studies were performed to determine the imprecision of the Access Folate assay for serum samples on the Dxl 9000 Access Immunoassay Analyzer using a protocol based on CLSI EP-05-A3. The study was run on three Dxl 9000 Immunoassay analyzers, three reagent lots, and three calibrator lots. Five (5) serum samples, with varying Folate concentrations, were assayed in duplicate with two runs per day, over 20-22 days. The study met the minimum requirement of 80 replicates per sample on each instrument and reagent lot combination. The within-laboratory (total) % CV was between 2.2% and 4.4% for Folate concentrations > 2.0 ng/mL. The within-laboratory (total) SD was between 0.10 - 0.21 for Folate concentrations ≤ 2.0 nq/mL.

The repeatability (within-run) % CV was between 1.6% and 2.7% for Folate concentrations > 2.0 ng/mL. The repeatability (within-run) SD was between 0.08 -0.09 for Folate concentrations ≤ 2.0 ng/mL.

Serum Folate(ng/mL)Repeatability(Within-Run)Between-RunBetween-DayWithin-Laboratory
SampleNMeanSD%CVSD%CVSD%CVSD%CV
Sample 1881.70.09N/A0.10N/A0.16N/A0.21N/A
Sample 2884.70.102.10.071.40.173.60.214.4
Sample 3889.20.252.70.000.00.272.90.363.9
Sample 488160.32.00.21.20.42.70.63.6
Sample 588210.42.10.31.30.62.70.83.7

{6}------------------------------------------------

RBC Imprecision: Verification studies were performed to determine the imprecision of the Access Folate assay for hemolysate samples on the Dxl 9000 Access Immunoassay Analyzer using a protocol based on CLSI EP-05-A3. The study was run on one Dxl 9000 Access Immunoassay Analyzer, one reagent lot, and one calibrator lot. Six whole blood hemolysate samples, with varying Folate concentrations, were tested in the study. The within-laboratory (total) % CV ranged from 1.9% to 4.9%.

Hemolysate FolateConcentration (ng/mL)Repeatability(Within-run)Between-runBetween-dayWithin-Laboratory
SampleNMeanSD%CVSD%CVSD%CVSD%CV
Sample 1805.00.214.30.000.0040.000.0010.214.3
Sample 280110.32.80.32.80.00.020.54.0
Sample 380110.54.40.00.020.22.20.54.9
Sample 480130.21.90.10.50.00.0020.31.9
Sample 580180.42.20.31.80.31.40.63.1
Sample 680220.41.80.31.20.00.0030.52.1

LoB/LoD: Verification studies were performed to determine the Limit of Blank (LoB) and Limit of Detection (LoD) for the Access Folate assay on the Dxl 9000 Access Immunoassay Analyzer using a protocol based on CLSI EP17-A2. The assay is designed to meet the claimed LoB of 0.80 ng/mL (1.81 nmol/L), LoD of 1.0 ng/mL (2.27 nmol/L).

LoQ: Verification studies were performed to determine the Limit of Quantitation (LoQ) for the Access Folate assay on the Dxl 9000 Access Immunoassay Analyzer using a protocol based on CLSI EP17-A2. The LoQ for Access Folate is designed to meet the claimed LoQ of < 2.0 ng/mL (4.53 nmol/L) based on a 20% CV.

Substantial Equivalence Comparison Conclusion

Beckman Coulter's Access Folate Assay on the Dxl 9000 Access Immunoassay Analyzer is substantially equivalent to the Access Folate Assay on the Access 2 Immunoassay System as demonstrated through the information and data provided in this submission. The performance testing presented in this submission provides evidence that the device is safe and effective in its intended use.

§ 862.1295 Folic acid test system.

(a)
Identification. A folic acid test system is a device intended to measure the vitamin folic acid in plasma and serum. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia, which is characterized by the presence of megaloblasts (an abnormal red blood cell series) in the bone marrow.(b)
Classification. Class II.