510(k) Data Aggregation

    K Number
    K170262
    Date Cleared
    2017-09-28

    (244 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital Thermometer is intended to measure oral temperature of human body. It can be used in home environment or at a doctor's clinic. It is indicated for use by people of all ages.

    Device Description

    The Digital Thermometer consists two parts. The thermometer probe part is a compact, small, light-weight, and battery-powered unit with thermometer sensor for oral temperature measurements. The thermometer probe must be connected to a compatible mobile phone device through its headphone jack for displaying measured temperatures.

    The mobile application could be conducted for operation and display temperature measurement when the Digital Thermometer connects with mobile device. The mobile application can be conducted on iOS operating system and Android operating system of mobile device.

    The mobile application could perform the functionality on mobile device screen including measurement activity, historical data activity, trend activity, basic setting and initial setting activity.

    AI/ML Overview

    The provided text describes a digital thermometer, Model 2180, and its substantial equivalence to a predicate device, the Basal Digital Thermometer (K141505). The information primarily focuses on the device's characteristics, intended use, and compliance with various standards.

    Here's an analysis of the acceptance criteria and the study information as requested, based only on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document lists several standards and requirements that the Digital Thermometer complies with, implying these act as acceptance criteria. The reported performance is that the device "complies" or "met the endpoints."

    Acceptance Criteria / StandardReported Device Performance
    Measurement Accuracy
    ±0.1°C (35.5°C to 42.0°C)Digital Thermometer: ±0.1°C (35.5°C to 42.0°C)
    ±0.2°C (other range)Digital Thermometer: ±0.2°C (other range)
    Display Resolution
    0.1°C (°F)Digital Thermometer: 0.1°C (°F)
    Displayed Temperature Range
    (Implicit from ASTM E1112-00)Complies with ASTM E1112-00 displayed temperature range requirements
    Environmental (Operating)
    10°C~40°C (50 ~ 104°F)Digital Thermometer: 10°C~40°C (50 ~ 104°F)
    15%-95% RHDigital Thermometer: 15%-95% RH
    Environmental (Storage)
    -20°C ~50°C (-4.0 ~ 122°F)Digital Thermometer: -20°C ~50°C (-4.0 ~ 122°F)
    15%-95% RHDigital Thermometer: 15%-95% RH
    Biocompatibility (Cytotoxicity)
    No Cytotoxicity (ISO 10993-5)Met endpoints for cytotoxicity
    Biocompatibility (Sensitization & Irritation)
    No Sensitization & Irritation-negligible (ISO 10993-10)Met endpoints for sensitization & irritation-negligible
    Electrical Safety (AAMI / ANSI ES60601-1)Complies with general requirements, protection against electrical/mechanical hazards, excessive temperatures, etc.
    Electromagnetic Compatibility (AAMI / ANSI/ IEC 60601-1-2)Complies with radiated emission, electrostatic discharge immunity, RF electromagnetic field immunity, etc.
    Human Interface and Mechanical Test (ASTM E1112-00)Complies with ASTM E1112-00 requirements
    Software Guidance (Premarket Submissions for Software)Complies with guidance requirements including level of concern, software description, V&V documentation.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample size used for any specific test set, nor does it mention data provenance (e.g., country of origin, retrospective/prospective). It generally states that the device was "verified and validated" or "complies" with standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The document refers to compliance with established standards (e.g., ASTM E1112-00 for accuracy), which inherently have their own defined test methodologies, but details about specific expert involvement in this device's testing ground truth establishment are absent.

    4. Adjudication method for the test set

    The document does not describe any adjudication method (e.g., 2+1, 3+1) for establishing ground truth for a test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the device is a digital thermometer, not an AI-assisted diagnostic tool involving human readers. No MRMC study was mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a standalone digital thermometer that provides a temperature reading. Its performance criteria (e.g., accuracy) were assessed in this context. The "software validation" mentioned indicates evaluation of the algorithm within the device and mobile application, which can be seen as a standalone performance assessment for its accuracy and functionality.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the performance of the thermometer, the "ground truth" would be established by reference standards or calibrated equipment used in accordance with the specified testing methodologies of standards like ASTM E1112-00. The text does not detail the specifics of such "ground truth" generation for its internal testing beyond stating compliance with the standard.

    8. The sample size for the training set

    This information is not applicable. The device is a digital thermometer, not a machine learning model that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no "training set" for this type of device.